RO-77-79:on 770708,1 & 4 Vol % Hydrogen Balance Nitrogen Failed to Meet Tech Spec Acceptance Criteria During Gas Chromatograph Calibr.Caused by Separation Column Resolution Loss.Replacement Column Will Be Maintained
ML19317G221
Person / Time
Site: Crystal River Duke Energy icon.png
Issue date: 08/03/1977
From: Stewart W
FLORIDA POWER CORP.
To:
Shared Package
ML19317G217 List:
References
NUDOCS 8002280855
Download: ML19317G221 (2)


Text

l i--- LICEhiEIEE EVENT REPORT i{

  • CONTROL* OLOCK:l l l l l l l [pLEASE PRIfJT ALL PEQUtntD INFCRT.1AT;O7;;

l . 1 S NAME LCENSE NUMBER

  • TYPE lFl} lCl Rl Pl 3l l 0 l 0-l-l 0 l 0 l 0 l-0 l 0 l-l O l 0 l l 4l 1l 1l ll 1l l0l3l 7 09 14 15 25 25 30 31 32 CATECORY TYP O 00CMET NUMOER EVENT CATE REPORT oATE D 1 CONT l - l - l lL l lL l l 0 l 5 l 0 l-l 0 l 3 l 0 l 2 l l 0 l 7 l 0 l8 l7 l7 l l0 l8 l0 l 3 l7 l7 l 7 8 57 50 59 60 61 68 69 74 75 00 EVENT OESCRIPTION Ql During calibration of the Fisher gas chromatograph, the 1 and 4 volume percent hydrogen l 7 89 e

Q[ balance nitrogen did' not meet the acceptance criteria required by Tech Spec 4.6.4.1.

7 89 . go O l Redundancy was maintained as the H2 analyzer was available and operable. This was the 7 89 l 00 05 l first occurrence of this event. A replacement separation colu=n was ordered and the l 7 U9 gl chromatograph was restored operable.

(LER 77-79) 7 89 60 E 00E COMPONENT COCE A MAP UA R VCLAICN lPlBl O

7 89 10 l,,,,Ej lXlXlXlXlXlXl lN l l Zl 9 [ 9 l 9 l lN [

11 12 17 43 44 47 48 .

CAUSE DESCRIPTION *.*s ldTEl l This occurrence was caused by the loss of resolution of the chromatograph's separation y 7 89 OO l column. A replacement column will be maintained at all times to ensure timely 7 7 l 80 corrective action.

FAC:UTY

]

METHoo oF STATUS 4 POWER oTHER STATUS DISCOVERY DISCoVERf CESCRPTON 70 ss 9

{ l 1l 0l 0l l 10 12 13 NA 44 l l Cl 45 l

46 NA l c0 ACTMTV CONTENT RELEASEo AMOUNT CF ACTMTY . LoCATCN OF RELEASE 7 0 lZ l 0

OF lZjRELEASE l 10 11 NA l 44 l

45 NA l PERSONNEL EXPOSURES 80 '

NUM0ER TYPE CESCRIPTCN III3l l0 l 0l 0l [ Zj l NA l

7 09 11 12 13 60 PERSONNEL INJUPIES NUM0E R DESCR.PTON 3 l0l0l0l11 12l 7 8D NA l

60 OFFSITE CONSEQUENCES 7 00 00 '

LOSS OR OAMAGE TO FACILITY l T Yl'E CESCRPTCN .

$dl 7 8D 10 NA l

LU PUBUCITY

$l NA l

. .g DITIONAL FACTORS 00 See attached Suppletanntarv Information 8 002 2H .j tiu 8I 7 oo 1

nu U. P. St?# art NAME:

PHONE: -(813) 866-4159

_; an.u,., \

. - SUPPLEMENTARY INFORMATION

1. Repdre No.: 50-302/77-79
2. Facility: Crystal River Unit #3
3. Report Date: 3 August 1977
4. Occurrence Date: 8 J917 1977
5. Identification of Occurrence:

Inoperability of gas chromatograph contrary to Technical Specification 4.6.4.1.

6. Conditions Prior to Occurrence:

Mode 1 power operation.

7. Description of Occurrence:

At 2150 the Fisher gas chromatograph failed to meet the calibration acceptance i criteria as required by Technical Specifications. The 1 and 4 volume percent hydrogen, ...

balsace nitrogen calibration would not meet the requirements due to apparent separation

column degradation resulting in a loss of resolution. A replacement separation column was ordered and the gas chromatograph has been restored operable.

B ,, Designation of Apparent Cause:

The cause of this event was the failure of the colum to perform its intended function.

9. Analysis of Occurrence:

The safety and health of the plant and public were not affected as a redundant hydrogen analyzer was available and operable.

10. Corrective Action:

A spare separation column will be maintained at all times in order that corrective action can be accomplished in a timely manner.

11. Failure Data:

This was the first occurrence of this event. It was the second occurrence of the inoperability of this device. Refer to LER 77-72.