Summarizes Status of Part 21 Potential Safety Hazard Review Re Rosemount Model 1152 Pressure Transmitters.Tests on Limited Sample Show Normal Output for over-range & under-range Conditions
ML19323E386
Person / Time
Site: Summer South Carolina Electric & Gas Company icon.png
Issue date: 04/29/1980
From: Crews E
SOUTH CAROLINA ELECTRIC & GAS CO.
To: James O'Reilly
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION II)
References
REF-PT21-80-228-004 NUDOCS 8005230475
Download: ML19323E386 (1)


Text

. .

south CARollNA ELECTRIC a gas COMPANY

.,on.e ... ...

CoLumsiA, south CAmotinA tute t.H. Ca r w s, J a. '

v.c.. .. .. ... c .. c..com. April 29, 1980

c. .... c.. -

Mr. James F. O'Reilly, Director U. S. Nuclear Regulatory Commission Region II, Suite 3100 101 Marietta Street, N. W.

Atlanta, Georgia 30303

Subject:

Virgil C Summer Nuclear Station Potential Substantial Safety Hazard Rosemount Model 1152 Pressure Transmitters Nuclear Engineering File 3.1051

Dear Mr. O'Reilly:

On March 25, 1980, Mr. Mike Heffley of NRC Region II was notified orally of a potential substantial safety hazard concerning Rosemount model 1152 pressure transmitters.

Rosemount informed us of the potential problem for our evaluation and possible reporting. In summary, Rosemount has found in tests on a limited sample of transmitters that a normal output between 4 and 20 mA can occur for an over-range condition with input pressures greater than 140% of the upper range limit or for an under-range condition exceeding 100% of the span.

We are continuing to review each of our applications of this model transmitter to determine if erroneous outputs might occur during normal or accident system conditions nd, if so whether or not such outputs constitute a substantial safety hazar as defined by 10CFR21. A final report providing all information required 10rFR21 will be issued when this review is com-pleted. If you have any uest .ons, please contact us.

Ve pryly yours, A- . ,_ _

. H. rews, Jr.

SMC:EHC:rh cc: H. T. Babb ,

]$O/7 E. H. Crews, Jr. -

g/

O. S. Bradham D. A. Nauman /[{}

H. Radin J. Skolds V..C'. Summer 0005 R Sg yp5 NPCF/Dixon l File UfrICIAL COPY l 3 i