Notice of Nonconformance from Insp on 870406-09. Nonconformance Noted:Licensee Issued Engineering Change Notice NJ-17436 to Delete Requirement Re Glyptal Coating of Metal Tubing W/O Technical JustificationML20236G605 |
Person / Time |
---|
Issue date: |
06/12/1987 |
---|
From: |
Office of Nuclear Reactor Regulation |
---|
To: |
|
---|
Shared Package |
---|
ML20236G568 |
List: |
---|
References |
---|
REF-QA-99900403 NUDOCS 8708040321 |
Download: ML20236G605 (1) |
|
|
---|
Category:NOTICE OF DEVIATION/NONCONFORMANCE FROM A REGULATORY
MONTHYEARML20059H8811993-10-28028 October 1993 Notice of Nonconformance from Insp on 930907-10. Nonconformance Noted:Changes Made to Computer Codes Redya, Odyna & Safer Not Being Documented & Verified in Formal Manner ML20057C0521993-09-16016 September 1993 Notice of Nonconformance from Insp on 930712-15. Nonconformance Noted:Licensee Failed to Maintain User Manual for SEISMIC4 Finite Element Analysis Program Under Formal Document Change Control ML20056F2301993-07-16016 July 1993 Notice of Nonconformance from Insp on 930504-05 Nonconformance Noted:Qa Manual,Rev 10,dtd Dec 1992 Did Not Contain Provisions for Dedicating Items Purchased by Vendor as Commercial Grade ML20062F7831990-11-0909 November 1990 Notice of Nonconformance from Insp on 900625-27. Nonconformance Noted:Licensee Sold Listed Items as Nuclear safety-related W/O Completely Evaluating Suitability for Use in Such Applications ML20062G3091990-11-0808 November 1990 Notice of Nonconformance from Insp on 901015-19. Nonconformance/Deviation Noted:Licensee Failed to Identify Excessive Raised Metal in Form of Burr Adjacent to One Grid Pad Bore Hole on Lower End Fitting ML20058F0131990-11-0202 November 1990 Notice of Nonconformance from Insp on 900423-27. Nonconformance Noted:Failure to Effectively Control Design Calculation Verification Process for Certain Piping & Pipe Support Stress Calculations Re Bulletin 79-14 ML20058D5821990-10-30030 October 1990 Notice of Nonconformance from Insp on 900716-18. Nonconformance Noted:Failure to Implement Program to Control Insp & Process Monitoring of Mfg of safety-related Structural Steel Tubing Sections ML20059L4661990-09-17017 September 1990 Notice of Nonconformance from Insp on 890515-18. Nonconformance Noted:Purchased Commercial Grade 5/16-18 Inches Hex Heavy Nuts from Uchimoto Mfg Co,Neither Approved Nor Qualified Vendor Licensee ML20059L4441990-09-13013 September 1990 Notice of Nonconformance from Insp on 880509-12. Nonconformance Noted:Licensee Could Not Produce Adequate Documentation to Substantiate Qualification of Insp & Auditing Staff ML20058M2681990-08-0707 August 1990 Notice of Nonconformance from Insp on 900312-15. Nonconformance Noted:Five of Six Employees That Perform Activities Affecting nuclear-related Sys & Components Not Indoctrinated in QA Program ML20056A5261990-08-0303 August 1990 Notice of Nonconformance from Insp on 900423-26. Nonconformance Noted:Licensee Failed to Establish Adequate Measures to Assure That Conditions Adverse to Quality Promptly Corrected ML20059P0411990-07-31031 July 1990 Notice of Nonconformance from Insp on 900730-31. Nonconformance Noted:Audit Repts for Abb,Seitz & Poppe & Potthoff Not Completed at Time of Insp & Audits Conducted 7 Months Prior to Insp Date ML20055F7311990-07-11011 July 1990 Notice of Nonconformance from Insp on 900205-08. Nonconformance Noted:Failure to Clearly Establish & Delineate in Writing Authority & Responsibility of Persons Affecting safety-related Function of Field Engineering ML20055D8121990-07-0202 July 1990 Notice of Nonconformance from Insp on 900510 & 11. Nonconformances Noted:Licensee Program to Assure Design Control & Equivalency to Original Part Supplied for safety- Related Spare & Replacement Part Orders,Inadequate ML20247F2931989-07-21021 July 1989 Notice of Nonconformance from Insp on 881102-04 & 890201-03. Nonconformance Noted:Licensee Certified & Supplied Stock Matl W/O Testing Each Piece of Stock Matl & All Nuclear Sales Order Folders Not Color Coded ML20246F0761989-07-10010 July 1989 Notice of Nonconformance from Insp on 890508-11. Nonconformances Noted:Failure to Test Latest Rev of Replacement Circuit Boards at Battery Charger Final Acceptance Test to Verify That Charger Output within Specs ML20245D2171989-04-25025 April 1989 Notice of Nonconformance from Insp on 890206-10. Nonconformance Noted:Nuclear Safety Subcommittee Not Convened to Evaluate Problems Identified in Repair for safety-related Type Pressure Signal Limiter Circuit Boards ML20248K4551989-04-0707 April 1989 Notice of Nonconformance from Insp on 890123-27. Nonconformances Noted:Company Shipped Ductwork w/3/16 Inch by 2 Inch Welds Attaching Stiffeners to Ducts Which Were Spaced More than 8 Inch Apart ML20247J8781989-03-31031 March 1989 Notice of Nonconformance from Insp on 890123-27. Nonconformance Noted:Vendor Initiated 35 Purchase Orders for safety-related Electrical Matl W/Gismo Co W/O Supplier Listed on Ebasco Waterford 3 Approved Supplier List ML20155E4151988-10-0707 October 1988 Notice of Nonconformance from Insp on 880801-05. Nonconformance Noted:Nuclear Raw Matl Stored in Hold Area W/Govt Level 1 Matl & Lack of Evidence That Spherical & Cylindrical Stds Used to Calibr Optical Compactor ML20154H0351988-09-14014 September 1988 Notice of Nonconformance from Insp on 880627-0701. Nonconformance Noted:Qa Dept Director Does Not Rept to Mgt Level Which Provides Adequate Independence from Cost & Schedule When Opposed to Safety Considerations ML20153H4561988-09-0202 September 1988 Notice of Nonconformance from Insp on 880711-15. Nonconformance Noted:Licensee Did Not Include Chemical Product Analysis Performed by Schevers Labs on CMTRs ML20151U9401988-08-12012 August 1988 Notice of Nonconformance from Insp on 880627-29. Nonconformance Noted:Requirement for Vendor to Have Approved QA Program Not Stated on Purchase Orders & Lack of Documentation Re Certification of Two Current Inspectors ML20151S3301988-08-0404 August 1988 Notice of Nonconformance from Insp on 880411-13. Nonconformance Noted:Code Sponsor for Gt Strudl Computer Program Failed to Perform Actions on Error Notices Sent by Control Data Corp from Jan 1985 to Present ML20207G9081988-08-0404 August 1988 Notice of Nonconformance from Insp on 880201-10. Nonconformance Noted:Ge Responsible Engineer Procured 100 Replacement Metric Phillips Flat Head Screws W/O Purchase Documents Being Processed & Verification of Matl Quality ML20151R5991988-08-0202 August 1988 Notice of Nonconformance from Insp on 880524-26. Nonconformance Noted:Eight Radioactive Matls Packages Did Not Contain Foam & One Package Contained Rubber Instead of Foam & Records of Fabrication & Use Not Kept ML20151S2191988-08-0101 August 1988 Notice of Nonconformance from Insp on 880502-06. Nonconformance Noted:Measures to Ensure That Shaft Keys & Terminal Blocks Procured from Commercial Subvendors Per Applicable Design Requirements Not Established ML20151Q3331988-07-26026 July 1988 Notice of Nonconformance from Insp on 880418-22. Nonconformance Noted.Major Areas Inspected:Evaluation & Records Re Several Recent Events at Westinghouse PWR-type Nucelar Power Plants ML20155D2981988-06-0303 June 1988 Notice of Nonconformance from Insp on 880125-28. Nonconformances Noted:Failure to Clearly Establish Duties & Authorities of Engineering Personnel to Ensure safety- Related Design & Engineering Activities Performed ML20151V6071988-04-21021 April 1988 Notice of Nonconformance from Insp on 880216-19. Nonconformance Noted:Trent Tube Facility Did Not Include Chemical Product Analyses Performed by Steel Mfgs &/Or Lab on Listed Certified Matl Test Repts for Nuclear Orders ML20151F8731988-04-15015 April 1988 Notice of Nonconformance from Insp on 880307-11. Nonconformance Noted:Matl Being Upgraded for Westinghouse Purchase Orders Lacked Required Number of Tensile Tests Needed to Meet ASME Boiler & Pressure Vessel Code ML20148H3051988-03-22022 March 1988 Notice of Nonconformance from Insp on 871207-11. Nonconformance Noted:Failure to Establish Adequate QA Program to Control safety-related Activities Re Fire Damper Design & QA Monitoring Design Areas ML20149M2211988-02-22022 February 1988 Notice of Nonconformance from Insp on 880111-14. Nonconformance Noted:Matl Not Purchased to Astm/Asme Spec, Chemical Composition Limits Incomplete & Actual Mechanical Property Limits Missing ML20196D2991988-02-11011 February 1988 Notice of Nonconformance from Insp on 871207-11 ML20148M6691988-01-25025 January 1988 Notice of Nonconformance from Insp on 871116-20. Nonconformance Noted:Inadequate Translation of App B Into QA Program & Inadequate Translation of Design Requirements Into Testing Parameter Requirements ML20237B3501987-12-15015 December 1987 Notice of Nonconformance from Insp on 871005-09. Nonconformances Noted:No Documented Evidence That Survey or Audit Undertaken on Eight Vendors Who Performed Svcs for Energy Svcs Group ML20236N6211987-11-12012 November 1987 Notice of Nonconformance from Insp on 870928-1002. Nonconformances Noted:Matl Test Repts Not Supplied for Listed Purcahse Orders,Thermometer SN-2319 Not Calibr Since 830804 & Forms Not Approved or Ref in QCP-42,Rev 12 ML20236M1081987-11-10010 November 1987 Notice of Nonconformance from Insp on 870721-24. Nonconformance Noted:Facility QA Deficiency Repts Not Properly Closed Out & Signed Off.Matl Release,Verification & Matl Cutting Activities Not Documented & Signed Off ML20236G7511987-10-28028 October 1987 Notice of Nonconformance from Insp on 870824-28. Nonconformance Noted:End Cap Protection Not Installed on Machined End of Guide Tubes Although Sequence Signed Off as Complete on Traveller Es 120-B ML20236H6971987-10-28028 October 1987 Notice of Nonconformance from Insp on 870805-07. Nonconformance Noted:Adjustment Screw 102/103 P Inspected in Accordance W/Drawing 8305-034,Rev 1,dtd 840622 Instead of Rev 2,dtd 861113 ML20150C1471987-10-20020 October 1987 Notice of Nonconformace from Insp on 870727-0806. Nonconformance Noted:Documentation Reviewed Could Not Support GE Performance of Seismic Analysis ML20235K2481987-10-0101 October 1987 Notice of Nonconformance from Insp on 870810-14. Nonconformances Noted:Seminar Re QA Manual & Std Practices Not Conducted Since 870103 & Calibr Label for Thermocouple Serial Number TC-11 Did Not Identify Date of Last Calibr ML20238E3711987-09-11011 September 1987 Notice of Nonconformance from Insp on 870525-29. Nonconformance Noted:Furnace Charts for Shop Orders 53332, 53318 & 53317 Did Not Indicate Time Load Charged & Removed ML20236G3701987-07-28028 July 1987 Notice of Nonconformance from Insp on 870512-16 & 27-28. Nonconformance Noted:Audits of Mfg Process Not Performed ML20236J5121987-07-0909 July 1987 Notice of Nonconformance from Insp on 870420-23. Nonconformances Noted:Failure to Examine Welds Prepared to Qualify Wps CP202,Rev 1 by Radiographic or Ultrasonic Testing & Failure to Perform Internal Audit Since 1983 ML20215G8201987-06-18018 June 1987 Notice of Nonconformance from Insp on 870316-20 ML20236G6051987-06-12012 June 1987 Notice of Nonconformance from Insp on 870406-09. Nonconformance Noted:Licensee Issued Engineering Change Notice NJ-17436 to Delete Requirement Re Glyptal Coating of Metal Tubing W/O Technical Justification ML20213G1651987-05-12012 May 1987 Notice of Nonconformance from Insp on 870323-27 1993-09-16
[Table view]Some use of "" in your query was not closed by a matching "". Category:TEXT-INSPECTION & AUDIT & I&E CIRCULARS
MONTHYEARML20149H4691994-11-15015 November 1994 Notice of Violation from Insp on 930201-0308 & Investigation Conducted Between Feb 1990 & Nov 1993.Violation Noted: Licensee Quality Implementation Procedure Not Adequately Established ML20059H3401994-01-11011 January 1994 Insp Rept 99900786/93-01 on 930826-27 & 30-31.No Violations Noted.Major Areas inspected:RMS-9 Overcurrent Trip Devices, Thmthjk & Tjj molded-case Circuit breakers,CR124 Overload Relays & Certain Other Issues ML20059K8641993-12-27027 December 1993 Exam Rept 50-002/OL-93-04 on 931207-08.Exam Results: One RO & Two Sroi Were Administered Exam & Operating Test. All Candidates Passed ML20059K8901993-12-23023 December 1993 Exam Rept 50-097/OL-93-02 & 50-157/OL-93-02 on 931004-06. Exam Results:Three SRO Passed Exams IR 05000317/19930261993-12-17017 December 1993 Insp Repts 50-317/93-26 & 50-318/93-26 on 931026-28.No Overall Exercise Strengths or Weaknesses Identified ML20059G3171993-12-0909 December 1993 Exam Rept 50-128/OL-93-03 on Stated Date.Exam Results: Candidate Successfully Passed SRO Upgrade Exam ML20058K0871993-11-30030 November 1993 Exam Rept 50-284/OL-93-01 on 931025-27.Exam Results:Reactor Operator Operating Test Retake Applicant Failed Exam.All Other Applicants Passed Exams ML20058A8161993-11-18018 November 1993 Notice of Nonconformance from Insp on 930809-13. Nonconformance Noted:Tracg Input Decks Used to Model Gist Facility Not Independently Verified to Be Correct ML20058D6691993-11-15015 November 1993 Exam Rept 50-020/OL-93-01 on 930913-15.Exam Results:All Applicants Passed Exams & Issued Appropriate License ML20059K9471993-11-10010 November 1993 Notice of Violation from Insp on 931006-08.Violation Noted: Licensee Alcohol Testing Procedures Did Not Address Those QCs Used to Ensure Test Results Were Valid ML20058A8251993-11-0404 November 1993 Insp Rept 99900403/93-01 on 930809-13.Nonconformances Noted. Major Areas Inspected:Review Input Modeling of Tragg Computer Code for Gist Facility ML20058A5411993-11-0404 November 1993 Insp Rept 99901268/93-01 on 931018-21.No Violations Noted. Major Areas Inspected: QA Program Implementations in Areas of Matl & Svc & Controls for Subsuppliers,Matl Control Insps & ASME Code Certification ML20059H8811993-10-28028 October 1993 Notice of Nonconformance from Insp on 930907-10. Nonconformance Noted:Changes Made to Computer Codes Redya, Odyna & Safer Not Being Documented & Verified in Formal Manner ML20059H8961993-10-26026 October 1993 Insp Rept 99900403/93-02 on 930907-10.Nonconformance Noted.Major Areas Inspected:Quality Activities of Design of Advanced BWR IR 05000327/19932011993-10-13013 October 1993 Insp Repts 50-327/93-201 & 50-328/93-201 on 930823-930902. Several Deficiencies Noted.Major Areas Inspected:Plant Operations,Maint,Engineering & Quality Programs ML20059G1471993-10-0808 October 1993 Initial Exam Rept 50-134/OL-93-03 Administered During Wk of 930913.Candidates Administered Retake Exam Passed Applicable Portion of Exam ML20057G2431993-09-28028 September 1993 Exam Rept 50-113/OL-93-01 on 930823-24.Exam Results:Both SROs Passed Exam ML20057D8761993-09-24024 September 1993 Exam Rept 50-027/OL-93-03 on 930816-17.Exam Results:One SRO Administered Written & Operating Requalification Exam,Passed Exam ML20057F0861993-09-24024 September 1993 Exam Rept 50-027/OL-93-02 on 930816-17.Exam Results:Initial Written Exam & Operating Tests Administered to One RO Applicant.Operating Tests Also Adminsitered to One SRO Applicant.All Applicants Passed Exam & Issued Licenses ML20057E6741993-09-23023 September 1993 Insp Rept 99900404/93-01 on 930816-17.No Violations or Deviations Noted.Major Areas Inspected:Control of Activities Re re-manufacture of X-ray & Other Components Installed in DB & Dhp Type 480 Volts Breakers ML20057D1921993-09-23023 September 1993 Exam Rept 50-166/OL-93-02 on 930909.Exam Results:Ro Passed Retake of Section C of Written Exam ML20057C0501993-09-16016 September 1993 Notice of Violation from Insp on 930712-15.Violations Noted: Licensee Failed to Appropriate Procedures to Implement Provisions of 10CFR21 ML20057C0521993-09-16016 September 1993 Notice of Nonconformance from Insp on 930712-15. Nonconformance Noted:Licensee Failed to Maintain User Manual for SEISMIC4 Finite Element Analysis Program Under Formal Document Change Control ML20057C0531993-09-0202 September 1993 Insp Rept 99900105/93-01 on 930712-15.Violations & Noncompliance Noted.Major Areas Inspected:Qa program,10CFR21 Program & follow-up on Concern Re Licensee Valve Assembly Seismic Stress Calculations ML20056G0581993-08-12012 August 1993 Exam Rept 50-186/OL-93-01 on 930726-27.Exam Results: Candidate Passed Written & Operating Portions of Exam ML20056D4851993-08-0404 August 1993 Exam Rept 50-150/OL-93-01 on 930628.Exam Results:Two Operator Requalification Retake Exams Administered.Both Operators Passed Exams ML20056E4431993-07-26026 July 1993 Exam Rept 50-184/OL-93-01 on 930713.Exam Results:Applicant Passed Both Portions of Licensing Exam ML20056F2181993-07-16016 July 1993 Notice of Violation from Insp Rept on 930504-05.Violation Noted:Qa Manual,Rev 10,dtd Dec 1992 for Implementing 10CFR21 Inadequate Due to Listed Reasons ML20056F2301993-07-16016 July 1993 Notice of Nonconformance from Insp on 930504-05 Nonconformance Noted:Qa Manual,Rev 10,dtd Dec 1992 Did Not Contain Provisions for Dedicating Items Purchased by Vendor as Commercial Grade ML20056D4061993-07-0808 July 1993 Exam Rept 50-116/OL-93-01 on 930623.Exam Results:All Three Candidates Passed Their Respective Portion of Exams & Issued Appropriate License ML20056F2361993-07-0101 July 1993 Insp Rept 99901265/93-01 on 930504-05 & 0604.Violation & Nonconformances Noted.Insp Conducted to Determine Whether Tests Conducted by Vendor in Accordance W/Approved App B QA Program & Applicable Purchase Order Requirements ML20056D8781993-06-28028 June 1993 Notice of Violation from Insp on 930510-14.Violation Noted:Failed to Complete Evaluation & Revise Procedure ML20056D9661993-06-21021 June 1993 Insp Rept 99901266/93-01 on 930602-03.No Violations or Deviations/Noncompliance Noted.Major Areas Inspected: Licensee Purchase Orders,Related Records & Associated Activities of Huron Industries,Inc ML20056C4571993-06-0707 June 1993 Exam Rept 50-128/OL-93-02 on 930518.Exam Results:Applicant Passed Both Written Exam & Operating Evaluation ML20056D8901993-05-27027 May 1993 Insp Rept 99900100/93-01 on 930510-14.Violations Noted: Failed to Complete Evaluation & Revise Procedures.Major Areas Inspected:Mfg & Engineering Programs for safety-related Items ML20056C3601993-05-0505 May 1993 Exam Rept 50-188/OL-93-01 on 930428-29.Exam Results:One SRO Passed Both Written & Operating Requalification Exams ML20058C6381993-03-19019 March 1993 Exam Rept 50-288/OL-93-01 on 930125-27.Exam Results:One Reactor Operator (RO) Applicant Passed Written Exam Retake of Section C,Four RO Applicants Passed Initial Written Exam & One Senior RO Applicant Passed Operating Tests IR 05000282/19922011992-12-31031 December 1992 Insp Repts 50-282/92-201 & 50-306/92-201 on 920921-1120.No Violations Noted.Major Areas Inspected:Working Relationships & Communication Channels Between Operations & Mgt Involvement Oversight ML20127B6111992-12-22022 December 1992 Insp Rept 99901258/92-01 on 921117-20.Violations & Nonconformance Noted.Major Areas Inspected:Qa Program & Implementation in Areas,Including Control of Work & Special Processes,Purchased Matl & Svcs & Insp & Tests ML20126L8821992-12-17017 December 1992 Insp Rept 99900404/92-02 on 921116-19.Noncompliance Noted. Major Areas Inspected:Dedication of Commercial Grade Components for Nuclear safety-related Applications & Previous Insp Concerns & 10CFR Part 21 Notifications IR 05000331/19922011992-12-0202 December 1992 Insp Rept 50-331/92-201 on 920511-15.Violations Noted.Major Areas Inspected:Review of Procedures & Representatives Records Including Procurement Packages for Mechanical & Electrical Items ML20127M1321992-11-23023 November 1992 Notice of Violation from Insp on 920914-17.Violation Noted: as of 920914,licensee Had Not Revised Procedures to Address Substantive Revs to 10CFR21 That Became Effective on 911029 1994-11-15
[Table view]Some use of "" in your query was not closed by a matching "". |
Text
_ _ _ _ _ _ - _ _
.,o
-c APPENDIX A General Electric Company Nuclear Energy Business Operations Docket No. 99900403/87-01 NOTICE OF NONCONFORMANCE During an inspection conducted April 6-9, 1987, the implementation of GE's Quality Assurance (QA) program at the San Jose, California facility was reviewed.
In addition, documentation supporting GE's response to items of nonconformance and unresolved items in relation to NRC Inspection Report No. 99900403/86-01 was also reviewed. This review was in response to allegations submitted to the NRC by the Government Accountability Project, based upon a review of work records supplied them by a former GE employee, Mr. Sam A. Milam III. These concerns allege potential deficiencies in design control activities within GE's QA program during the period March 1978 to April 1982. The QA program requirements are discussed in GE Topical Report HED0-11209-04A, Revision 6, dated June 30,1986, which describes the QA program GE applies to those design, procurement, and fabrication activities involving safety-related structurer, systems, and components of nuclear power plants within the GE '
scope of work. Section 5 of NEDO-11209-04A, Revision 6, requires, in part,
.that activities affecting quality be accomplished in accordance with documented instructions, procedures, or drawings.
Based on the results of this inspection, it appears that certain activities at the San Jose facility were not conducted in accordance with commitments niade to the NRC.
E0P 55-2.00, Section.4.1.2.(a), " Engineering Change Control," issued December 19, 1978 requires, in part, that the responsible engineer evaluate and technical impact of a proposed change and develop appropriate change documentation.
Contrary to this procedure, GE issued Engineering Change Notice (ECN) NJ-17436, dated June 2,1980, without a technical justification to: (1) delete the requirement for glyptal coating of GE Electrical Metallic Tubing (EMT) because of unavailability; and (2) delete the notation " Approved Vendor: GE Company, Requires Engineering Approval." (67-01-01). This item of nonconformance was previously identified in KRC Inspection Report No. 99900403/86-01 as unresolved j item (86-01-10).
l 8708040321 870612 PDR GA999 ENVGENE 99900403 PDR 1
- _ - _ _ _ - _ _ _ _ - _ - _ _ _ _ _ - _ - _ - _ _ _ - _ _ _ _ _ - _ .-