LICENSEE EVENT llEPOllT e

CON THot uloCK: l i

l l l l l lh e,

(PLEASE PRI?JT OR TYPE ALL REoutRED INFORMATloN)

.6lW

f. # 9 l I l LICEN5EE K l NlCODC P l 1 l@l 0 l 0l -l 0 l LICENSE 14 lb 0 l ONUMUEH l 0 l 0 l- l 0l 026l@l4 26 l lLICEN50 l l 1JOl@lb7 CATl 68l@

l 1 l TYPE CON'T l o l ti  ;;',%' I L j@l 0 l 5 l0 l0 l 0 l 3l 0 l 5 ]@l 014 il l 0 l 8 l 2 l@l75 0l5 l1 l0 l 81 P80l@

4 3 # 60 61 DOCKE T fAUMRE H 68 69 EVENT DATg 14 R E T'OR T D AT E f VENT DELCHIPTlON AND PHOHAULE CONSEQUENCES h

[o 121 l With the unit at hot-shutdown prior to a scheduled refuelinq outane (normal operatinq l j o lai l pressure and temperature), both source range nuclear instruments were inoperable. l

[ola1[ This is an LCO per T.S. Table 3.5-2; at least one of two source range channels shall I l.o I s l I be operable, or achieve and maintain hot shutdown. This is renortable per T.S. 1

[ O I r, i l 6.9.2.b.(2). Since hot shutdown was maintained, there was no effect on public safety I lo I7I I or olant operation. I I o Ih l I I

'#$'o'E C '^"dE CO s'oC55E COMPONENT CODE SUBC D'E SU E lol91 i e l I l A l@ LEJ@ l E l@ l I l Nl Sl Tl Rl Ul@ W@ W @

9 10 1I 12 13 IS 19 20

_ SE QUE N TI AL OCCURRENCE REPORT REVISION L E HtRd EVf NT YE AR REPOHT NO. CODE TYPE NO.

@ ,ay;oa', t,

_ 21 l 8l 2l 22 l-l 23 1010191 24 2G M

27 I 01 3 l 28 23 lL l 30 l-l 31 W

32 i Kt A il f O PL T 1 I HOURS 22 S B 4 0 FO t BUS. SU PLIE MAN F ACTUF Ef t

(_C.jgl 33 J4 X l@ l35 Zl@ W@

36 l0l0l0l0l 31 40 y@

41 l Nlg W@

42 43 l Wl I l Z l 04;l@

44 CAUSE DESCHIPTIOfJ AfJD CORRECTIVE ACTIONS h lil0ll The source range detector failures are attributed to an accumulation of water in I

[TTr] I the instrument cannisters. A spare detector was installed until replacement of the I g,j7gl normal channel detectors was made. Investigation is continuing to determine the l gl necessary corrective action. l li l41 I l

$1A S  % POWER OTHER STATUS dis HY DISCOVERY DESCHIPTION I 1151 (_Gj@ l 0 l 0 l 0 l@l NA l l A l@l Operator Observaticn l

' * ' ' ' " "

  • 8 e.CTiviTY CO0 TENT HI L E ASE O OF HELEASE AYOUNT OF ACTIVITY LOCATION OF RELEASE i o I I NA L7J @ L2J@l NA " "

l PEnsONdL ExPOssEs NUYdfR TYPE DESCRIPTtoN li l 71 [0 l o l cl@[lj@l NA l

,E nsONNd'iH;u','iEs "

  • NUveEn oEsCniPfiON@

lilall0lol0l@l f 0 0 11 NA 12 80 l

e oen = ,= '^c>u'v e 820 s 18 0 (ol7 l1l'alLZ_J@!NA 8 9 10 l

7 80 NRC UsE ONLY ,

2 O issujf7@""o'EsCniPTiOu

[m l NA @ l llllllllllll15

/ 8 9 10 68 69 83 ;;

NAME OF PREPARER C*S* OI PHONE: (414) 388-2560 Ext. 340