ML16293A141

From kanterella
Jump to navigation Jump to search
Radiological Survey and Dose Assessment Report for the Western New York Nuclear Service Center and Off-Site Areas in Follow Up to Aerial Gamma Radiation Survey Conducted in 2014, Rev. 0, Reference 5
ML16293A141
Person / Time
Site: West Valley Demonstration Project, P00M-032
Issue date: 08/22/2016
From:
MJW Technical Services
To:
Office of Nuclear Material Safety and Safeguards, State of NY, Energy Research & Development Authority, Office of Nuclear Reactor Regulation
Shared Package
ML16293A155 List: ... further results
References
Download: ML16293A141 (68)


Text

Reference 5-Evaluation of Radiochemical Data Usability (DOE.1997)

Reference 5 - Evaluation ofRacliochemical Data Usability (DOE, 1997)

Evaluation of Radiochemical Data*

Usability J.G. Paar University o~ Tennessee, Knoxville/Oak Ridge National Laboratory D.R. Porterfield Cheinical Science and Technology Division Los Alamos National Laboratory Environmental Re8toration Division P.O. Box 2003 Oak Ridge, 1N 37831-7294 Date Issued -

April,' 1997 Prepared for the U.S. Department of Energy Office of Environmental Management under budget and reporting code EW 20 Environmental Management Activities at the EAST TENNES.SEE TECHNOLOGLY PARK Oak Ridge, Tennessee 37831 managedby

  • LOCKHEED MARTIN ENERGY SYSTEMS, INC.

for the U.S. DEPARTMENT OF ENERGY under contract DE-AC-05-840R21400 es/er/ms-5 ADMIN RECORD sw-A-006155

PREFACE Environmental remedial decisions require the acquisition of data of known and sufficient quality for intended use. Data verification arid validation (V & V) are tWo of the tools that ensure a level of quality assurance in data usability. When applied to radiochemistry, no national standard currently cov:ers V&V concepts adequately, and the need for a document of this type was recognized by most*

of the Department of Energy Complex. This document Wa8 'developed through intersite cooperation and provides a reasonable approach for the evaluation of radioanalytical data for purposes of environmental remediation but can also be applied to data intended for non-remedial purposes.

This document is intended to provide a :frainework onto which implementing procedures can be written. It is a stand-alone document for the puiposes of data evaluation; however, sufficient laboratory deliverables must exist to enable the y&V tools to be used. It is recommended that if V &V procedures are written based on this document, that corresponding specifications for analytical laboratory deliverables are also written.

r-=* :;;;;:**"'==~----. --*

I I

ACKNOWLEDGEMENTS Many persons were of great help in-developing thiS document and are listed below. As this document is not a "consensus standard," consensus is not implied in listing these names. The intention here is to thank all those who helped in developing this document.

D_. Bottrell

  • D. Carden J. Chambers B. Crandall A.Daniel J. Fisk J. Griggs L. Hook J. Karnofsky '

T.Killeen C.KLiu

. D.McCurdy M. Miller D. Porterfield M. Ridley M. Schwab J. Warren R. Wells L. Withrow C. Wrotriiak DOE-EM263 DOE-ORO FERM CO WSRC

MMES LANL EPA-NAREL
MMES WSRC DOE-RL Yankee Atomic EQM LANL LLNL BHC EPNQAMS INEL Quanterra Dames & Moore
  • '.... ~'
  • .~. ".**

l PURPOSE........................................................ *................ 1 II.

  • REFERENCES.............................. ~............ *...................... 1 Ill DEFINITIONS................ *...... ;.. *...,......... :............................ 3 N.

GENERAL INFORMATION....................................................... 6 A

Relation to Laboratory Statement of Work...................................... 6 B.

Reporting Recommendations.........,........................................ 6 C.

Qualification of Data... *.................................................... 7 D.

Problem Identification....................................................... 7 E.

Data Verification.......................................................... 7

1.

Verification Report.....*............................................. 8

  • p.

Analytical Data Validation................................. _.................. 8

1.

Validation Report................... *................................ 8 V.

PROCEDURE....................... :........................................... 8 A

Custody of Samples and Sample Documentation................................. 9 B.

Holding Time and Tum-around Time...................... *.................... 9 C.

Sanlple Preservation................. :................................ *...... 10 D.

Instrument Calibration. *..................................................... 11

1.

Scope.............................................. *............... 11

2.

Mechanism for Initial Calibration Evaluation............................ 12

3.

General Technical Specifications far Initial Calibration.................... 12

a.
  • Check Source Statistics ~........... :.......................... 12
b.

Radioactive Sources......................................... l3

c.

Control Criteria................ :............................. 13

d.

Establishment of control points................................ 13

  • e.

Control ofBackgtound....................................... 14

f.

Geometry.................................,................ 14

g.

Background subtraction...................................... 14

h.

Recalibration................................ *.................. 14

4.

Specific Technical SpeCifications for Calibration Verification............... : 15

5.

Initial Calibration for Gross Alpha and Gross Beta by Gas Proportional Systems 1 S

6.

Calibration Verification for Gross Alpha and Gross Beta by Gas Proportional Systems.................................... ~..................... 17

7.

Initial Calibration for Liquid Scintillation............................... 18

8.

Calibration Verification for Liquid Scintillation................... ;...... 19

9.

Initial Calibration for Alpha Spectrometry............................... 21

10.

Calibration Verification for Alpha Spectrometry.......................... 22

11.

Initial Calibration for Gamma Spectrometry............................. 24

12.

Calibration Verification for Gamma Spectrometry........................ 25 E.

Quality-indicator samples....................................... *........... 27

1.

Total Propagated Uncertainty (TPU)....................................' 28

2..

Standai"d Traceability.............................................. *. 29

3.

Laboratory Control Sample (LCS).. :.................................. 29

4.

Laboratory Duplicate................................ :.............. 31

5.

Matrix Spike (MS)................... ;...... *........................ 32

6.

Method Blank...... *............................................... 34 I

~* '*.. ;...,..,..

F.

Chemical Yield-Tracers and Carriers......................................... 36 G.

Required Detection Limits (RDL)......... ~.................................. 37 H.

Nuclide Identification and Quantification................................... ~.. 37

1.

Test for Detection.................................................. 37

2.

Negative Results.................................................... 38 3*.

Sample Result Recalculation......................................... 38

  • I.

Instrument-Specific Sample Considerations.................................... 38

l.

Sample Analytical Parameters for Gas Proportional Counting............... 39

2.

Spectral Interpretation (alpha and ganuna spectrometry)................... 39 VI.

APPENDIXES........................................................,.......... 41

,;~,.. '

"t.,*~...* *:

Evaluation of Radiochemical Data Usability April 7, 1995 I.

PURPOSE This procedure provides a framework for implementation of radiochemical data* verification and validation for environmental remediation activities. It has been developed through participation of many individuals cunently involved in analytical radiochemistry, radiochemical validation, and validation program development throughout the DOE complex. It should be regarded as a guidance to use in developing an implementable radiochemical validation strategy.

This procedure provides specifications for developing and implementing a radiochemical validation methodology flexible enough to allow evaluation of data useability for project-specific Data Quality Objectives (DQO). Data produced by analytical methods for which this procedure provides limited guidance are classified as "non-routine" radionuclides and methods, and analyses by these methods may necessitate adoption of modified criteria from this procedure.

This procedure is applicable to radionuclide contaminants routinely analyzed by common radioanalytical methods primarily in aqueous and soil/sediment matrices for Environmental Restoration (ER) activities.

This procedure is applicable to radionuclide data produced through radioanalytical methods which use instrumentation for detecting activity. This procedure is not applicable to mass spectrometric or fluorimetric methodologies. An example listing of radionuclides and general inethodologies to which this procedure applies is provided in Appendix D.

Specifications in this procedure should be incorporated into appropriate project documentation such as the Sampling and Analysis Plan or Project Work Plan and into contractual Statements of Work (SOW) between the project and the analytical laboratories. This procedure shall be used as a baseline used to create project-specific procedures and checklists needed to perform radiochemical data verification and validation.

In this procedure, the word "shall" is used to denote a requirement; the word "should" is used to denote a recommendation; and the word "may" is used to denote permission, neither a requirement nor a recommendation. In conformance to this procedure, all analyses shall be perfo.rmed in accordance with its requirements, but not necessarily with its recommendations; however, justification must be documented for deviations from recommendations.

II.

REFERENCES ANSIN13.30, Performance Criteriafor Radio/Jioassay. 1/15/93.

ANSI N4~.14, Calibration and Use of Germanium Spectrometers for the Measurement of Gamma-Ray Emission Rates of Radionuclides. October 30, 1991.

ANSIN42.12-1980,American National Standard Calibration and Usage of Sodium Iodide Detector Systems. April 28, 1980.

ANSI N42. l 5-1990, American National Standard Performance Verification of Liquid-Scintillation Counting Systems. April 23. 1990.

Page 1 of41

Evaluation ofRadiilchemical Data Usability April 7, 1995 ANSI N42.2, Measurement Quality Assurance for Radioassay Laboratories, Rev. 12, Draft. June 1, 1993 Bechtel National, Inc. Review and Verification of QC Level L *IL and HI Radiological Data, PP:

  • 1503.2, Rev. 2.

Currie, L.A. "Limits for Qualitative Detection and. Quantitative Determination, Application to Radiochemistry." Analytical Chemistry. 40, 3 (1968).

DOE Methods for Evaluating Environmental and Waste Management Samples DOEIEM-0089T, Rev.

2, October, 1994.

FEMP Data Validation i'rocedure, SSOP-1004. 06/23/93 EG&G Idaho National Engineering Laboratory. Standard Operating Procedure for Radiological Data Validation. MSO-SOP-12.1.2. January, 1993.

EG&G Rocky,Flats, Environmental Monitoring and Assessment Division. Radiochemical Data Validation ~idelines. May; 1990.

IEEE Standard for Software Verificati9n and Validation Plans. IEEE Std. 1012-:1986.

  • Kanipe, L.H Handbook/or Analytical Quality Control in Radioanalytical Laboratories. EPA-600/7-77-088.

Karnofsky; J. MDAsfor Radiochemistry Data Validation, Draft #2, October, 1994.

Karnofsky, J. Radiochemistry Data Validation, Draft #4. November, 1993.

Knoll, G. F. Radiation Detection and Measurement, second edition. John Wiley and Sons. 1989 Montgomery.. D., and D.E. McCurdy. Specifications for WEMCO Environmental Radioanalytical Services Contract. October 23, 1992.

  • Science Applications International Corporation. Laboratory Data Validation ~idelines for Evaluating Radionuclide Analyses. Re¥. OSA.

Taylor, J.K.,Quality Assurance of Chemical Measurements. Lewis Publishers, 1987.

United States Environmental Protection Agency. Manual for the Certification of Laboratories Analyzing Drinking Water, Criteria and Procedures Quality Assurance, Third Edition. EP A/570/9-90/008.

United States Environmental Protection Agency. Data Quality Objectives Process for Superfund.

EP A/540/G-93/071 Page2 of41


'---"----"'-..;._---"--'--'-~------'-~--'~*~~---~-------:-~--*~**--**

-~-..

. ~.,* *,*

Evaluation ofRadiochemical Data Usability April 7, 1995 United ~fates Environmmtal Protection Agency. Guidance for the Data Quality Objectives Process.

EPA QA/G-4. August, 1994.

United States Environmental Protection Agency. Prescribed Procedures for Measurement of Radioactiv~ty in Drinking Water. EPA-600/4-80-032.

United States Environmental Protection Agency. Region NHazardous Waste Site Field Sampling Workshop. June 1993.

United States Environmental Protection Agency. "Radiochemical Determination o6ross Alpha and Gross Beta Particle ActiVity in Water, "Radiochemical Procedures Manual EPA-520/5-84-006.

United States Environmental Protection Agency.Upgrading environmental Radiation Data. Health Physics Society Committee Report HPSR-1 (1980). EPA 520/1-80-012.

ID.

DEFINITIONS Aff~cted Sample Result: a sample result that is considered to be significantly influenced by a quality deficiency, and is qualified, accordingly, through analytic~} data validation.

Analytical Batch: An analytical batch is a group of sample aliquots analyzed together on the same instrument detector system.

Analytical Data Validation: a technically based analyte and sample specific process that extends the qualification i:rocess beyond method or cqntractual compliance and provides level of confidence in the data that an analyte is presentor absent and if present, the associated variability. Data validation is a*

systematic process, performed external from the data generator, Which applies a defined set of performance-based criteria to a body of data that may result in physical qualification of the data. Data validation occurs prior to drawing a conclusion from the body of data.

Analytical Data Verification: a process of evaluating the completeness, correctness, consistency, and compliance of a set of facts against a standard or contract. Data verification is.defined as a systematic process, performed by either the data generator* or by an entity exter;nal to the data generator.

Calibration Verification:

Calibration verification, as described in this procedure, is defined as a periodic evaluation ofinstrumentstandardaation established during initial calibration. Using tolerance or statistical control charts, calibration verification Bil alert the instrument User of the occurrence of out-of-control instrumental conditions.

Carrier: A carrier is a stable eiement/compound, introduced into the sample preparation/analysis p1>cess, that will behave chemically similar to the analyte isotope(s). It is by virtue of this cheinical similarity that the carrier will "carry" tlie analyte isotope(s) through the samplepreparation/analysis process. The amount of the carrier recovered at the end of the analysis compared to that added iitially is often used in the calculation of the final result.

Page 3 of41

. ** :* ~

<';~. *~.*

Evaluation of Radiochemical Data Usability April 7, 1995 Correctable Problem: Correctable problems are deficiencies within data packages which may be rectified through consultation with the laboratory. Correctable problems may be revealed during both data verification and data validation.

correctable problems revealed during verification are those deficiencies that can be addressed by.obtaining additional information from the laboratory conectable problems revealed during vaidation are those deficiencies with analyses that can be solved by either a second preparation and/or analysis of a sample.

  • Counting uncertainty: Counting uncertanty, as described in this procedure, is defined as the statistical sample stan4ard deviation, which is an approximation of the population standard deviation, and is numerically defined as the square ropt of the number of counts obtained from a detector. This relationship holds true, provided that thedistribution that the counts foilows the Poisson distribution.
  • Units for counting uncertainty are the same as for the reported result and the MDC.

Initial Calibration: Initial calibration, as described in this procedure, is defjned as the standardization of an instrument used in radioactivity detection against a traceable radioactive source(s) of known identity and quantity. liis standardization prevails until such time as analytical conditions are deemed out of acceptable tolerance or statistical control limits.

Holding Time: Holding time, as described in this procedure, is defined as the period of time between sample collection and sample activity detection.

Minimum Detectable Activity (MDA) : The amount of a radionuclide, which if present in a sample, would be detected with a$ probability ofom-detection while accepting a probability,", of erroneously detecting that radionuclide in a appopriate blank sample. For this procedure, the" and$ probabilities are both set at 0.05.

Minimum Detectable Concentration (MDC): The MDA expressed in concentration units relative to the sample weight or volume.

Non-correctable problem: Non-correctable problems are those deficiencies, within data packages that cannot be addressed through additional laboratory submittals, and sample results must stand as-is.

Non-Correctable Problem: Non-correctable problems _are deficiencies within data packages which preclude the evaluation of data quality by predefined criteriaNon-coqectable problems may be revealed during both data verification and data validation.

Preparation Batch: A preparation batch is a groupofsample aliquots prepared together at the same time using the same method and related to the same quality-indicator samples.

Quality Control Chart: For purposes of this procedure, a quality control chart is used to determine if the response of the instrument has changed statistically; the magnitude the statistical response change Page4 cif 41 I

.. :-:~-: :**. -.::-:." :-.. -

-.. -T*.,.,

Evaluation of Radiochemical Data Usability April 7, 1995 may or may not be significant when compared to the required precision and*accuracy criteria for the overall analytical technique.

Quality-indicator Sample: Quality-indicator samples arethose samples made ready in the laboratory which provide direct or inlirect evaluation of the status of analytical system and resulting data quality.

Collectively, qµality indicator samples arethe laboratory control sample, laboratory duplicate, matrix spike, and method blank.

Reporting Batch: A reporting batch is a group of sample results reported together in a single data package. The ~g batch may be comprised of saiples prepared and analyzed together in the same preparation batch or samples prepared and analyzed in different preparation or analytical batches..

Required Detection Limit (RDL): The RDL is a contradually-specified detection limit (MDA or MDC) which, under typical analytical circUm&tances, should be achievable.

Standard Reference Material (SRM): A material or substance of one or more properties of which are sufficiently well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials. The SRM is characterized by the U.S.

National Institute of Standards and Technology (NIST) or other certified testing authority, and issued with a certificate providing the results of the clµuacterization.

Tolerance Chart: For pmposes of this irocedure, a tolerance chart is based upon maintaining a change of ipstrument response to.a tolerance level judged acceptableto meet overall quality requirements for the technique; a tolerance level should never be more restrictive than what is statistically possible.

Total Propagated Uncertainty (/'PU): The addition of the square root of the sum of the squares of random components of the individual uncertainties, plus the magnitude of the estimated individual systematic relative unceitainties: TPU may include ilncertainties introduced through field sampling and analytical labaatory procedures. For the purposes of this procedure, TPU includes only those random and systematic uncertainties associated only with laboratory preparation and analysis.

  • Refer to Appendix C for a full description ofTPU.

R s

TPU = Jr.Ri.

2+'£.s/

random components of individual relative uncertainties magnitude of the estimated individual systematic relative uncertainties Traceable Reference Material (/'RM): A NIST prepared standard reference material or a sample of known activity or concentration prepared from a NIST standard reference material (derived standard material).

Tracer: A tracer is a radioactive isotope, introduced into the sample preparation/analysis process, that will behave chemically similar to the analyte ietope( s ). The tracer isotope is of the same element as the analyte isotope(s) except where the decay mode, half-life, or availability dictate the use of the isotope Page 5 of41

  • ,,..i.****

Evaluation of Radiochemical Data Usability' April 7, 1995 of a different elmient. The activity of tracer detected at the end of the analysis compared to that added initially is_ used~ the calculation of the final result.

Turn-around Time: Tum-around time is contractually-specified as the amount of time which elapses between laboratory receipt of the raw samples arid subsequent data receipt by the client.

Well Characterized Reference Material (WCRM): The WCRM may be derived from a field sample which has been well chara~ through multiple analyses providing a high level of confidence of the activity level in the sample. The WCRM may be submitted to NIST for characterization and classification as a TRM.

IV.

GENERAL INFORMATION A.

Relation to Laboratory Statement of Work hnplementation* of this procedi.ire depends on the agreeinent of work to be performed by analytical laboratory jn the form ofa project-specific laboratory Statement of Work (SOW). While it is outside the scope of this procedure to provile requirements for the radioanalytical laboratory, validation will be expedited if laboratory deliverables are consistent With the evaluatia.n requirements of this procedure.

B.

Reporting Recommenda~ons All samples should be iqnrted with a background-subtracted sample result. The counting uncertainty, TPU, and MDC must be reported at the same level of confidence (e.g. +/-2s at 95%).

Page6 of41

    • ~- *.

I 1;.

~?.':,-

~ I

        • .... **-,*,.=*... * :~--:~':'*:*,.. *

~J.*

,\\..... ~ **

Evaluation.of Radiochemical Data Usability April 7, 1995 C.

Qualification of Data*

When qualification conditions are met, a qualifier and reason code shall ht: physically or electronically associated \\Vith theiiffected result(s). If more than one reason code is applied, each must be separated by commas. The listing of available codes is preseµted in Appendix B. If data are not qualified, a character (e.g. an_ equals sign, orother alphanumeric symbol different from a qualifier) is entered on the

  • sample result line. As direeted or as appropriate, reason codes may be applied without qualifiers.

D.

Problem Identification Problems *identified through data verification and validation are separated into two categories~.

correctable problems and non..:correctable problems.

i.

Correctable problems The first category contains those problems whth may be rectified through consultation with the laboratory.

The second subcategory contains.those problem5 are those that can be rectified by either a second preparation and/or analysis of a sample.

  • ii.

Non-Correctable Problem: Non~correctable problems are deficiencies within data packages which preclude the evaluation of data quality by predefined criteria.

E.

Data Verification Pata verification is defined as a systematic process, performed by either the data generator or by an entity external to the data generator, which results in evalilation of the completeness, correctness, consistency' and compliance of a data set against a standard or contract..

If verification is performed by the data generator, a project-level surveillance must be established by whfoh the performance of the verification process is evaluated...

Data verification, at the project level; may be conducted either by the data verifier or by the data validator, whichever expedites the review process. If verification is conducted independently of the validator, it includes two activities. The first activity entails inventory of the data package to*ensure compliance with the contract and statement of work, in terms of the required deliverables. The second activityentails to inform the validator that a non-correctable problem has occurrechnd that data may need qualification. Ifveification is conducted by the v"alidator, the first activity is conducted similarly, but the second activity may result in prompt qualification of data.

Data verification slnuld provide a mechanism for problem resolution with the laboratory; it should not be an after-the-fact identification of non-correctable. deficiencies.

Page 7 of41

~

... -. *. ~.'* ~-.

  • ~-. :;*

-~"-* *** *

-.,~.-. -~._,,..; *~.1,**. "!'*.*'*"-:*

~

...-.k.f<,ft.-'i: *'*'*.

Evaluation of Radiochemical Data Usability

  • April 7, 1995
1.

Verification Report A verification report is written by thedata verifier and takes, as input, the steps in this procedure that are listed as "Verification." The data verifiedoes not qualify data if verification criteria are not met but indicates in the verification report the circumstances surrounding correctable and non-correctable problems in the data package which will be transferred to the validator for possible qualification. The verification reportmust be in a standard format, and must remain a part of the analytical data package.

An example of a verification report is included in Appendix F.

F.

Analytical Data Validation Analytical data validation, including field anclaboratriry data review, is defmed as a systematic process, performed external from the data generator, which applies a defined set of performance-based criteria to a body of data whi:h may re8ult in qualificatiori of the data. Data validation is not performed by the analytical laboratory. Data validation provides a levebf assurance, based on a technical evaluation, that an analyte is present or absent,and if present, the level of uncertainty associated with the meaSu.rement, and occurs prior to drawing a conclusion from a body of data. Analytical data validation for radiochemistry includes a technical review of a laboratory data package covering the evaluation of quality-indicator samples, the identification and quantitation of analytes, and the effect of deficiencies in quality control on analytical sample data.

1.

Validation Report A validation report that includes the results of validation activities must be completed by the validator and takes, as input, the verific~on report and the steps in this procedure that are listed as "Validation."

A validation report must be completed for evecy reporting batch that passes through validation. To expedite writing the valilation report, comments must include explanations for qualification only *if the reason codes do not adequately ci:scribe justification for qualification. Comments on data qualification for which reason codes adequately describe qualification reason are not necessary.

Both the verification and valilation reports must be completed regardless of who performs the reviews.

  • V.

PROCEDURE The steps in this section describe the processes of radiochemical data verification and validation.

Refer to AJl>endix A for qualifier and reason code descriptions. Refer to Appendix B for guidance on qualification due to multiple quality deficiencies.

A Custody of Samples and Sample Documentation While verification/validation cannot asure a sample has truly been in custody from the field to the final result, an evaluation of fidl and laboratory chains of custody will provide the best indicator. A sample is defined as being in custody if any of the following conditions are met:

it is within one's.possession; Page 8 of41

Evaluation of Radiochemical Data Usability it is within one's view, after being in one's possession; it was in one's possession and. then was secured to prevent tampering; i~ is placed in a designat~ secure area.'.

I.

Verification April 7, 1995 Custodyofthe samples reported in the data package should be verified by tracing the signature record on both field and laboratory chains of custody. Verify that all samples on the chains Of custody are present in the reporting batch. If there is a break in custody of the sample( s ), representativeness of the samples may be in question, Indicate this problem in the verification report.

Verification ofsampe documentation includes result report header checks for accuracy from the COC.

If header information is incorrect when compared to the COC, the verifier shall mark through the incorrect header entry and pen in the correct entry, placing initials and date next to the correction. If

. sample idmtity is in question, every attempt should be made to verify the true identity of each sample, else sample representativeness is in doubt.

2.

Validation*

  • If sample origin and identity can not be substantiated, sample data may be qualified "R."

B.

Holding Time and Tum-around Time *

. Holding time is defined for radiohemical data verification *and validation as the period of time between sample collection and sample activity detection. Samples must be analyzed within a holding time precluding significant decay of short-lived target radionuclides. Holding time should be contractually specified to the laboratory. If holding times are not contractually satisfied, and if other technical corrective acfuns, such as increased count times or increased sample aliquot volumes or weights, have not been performed to compensate for long holding times,, the RDL may be exceeded Physical characteristics and matrix influences must illso b considered when setting holding times. These characteristics must be considered when planning for validation implementation. The following table presents commonly-analyzed radionuclides on the DOE Complex that have characteristics affecting holding time establishment decisions:

T bl 5 1 Ph. 1 d M a e tys1ca an 1 edCh atnx-re at aractenstics Nuclide Physical or MatrlX-related Characteristic Anionic Volatile when placed in acid solution 3H Volatile 99Tc Volatile,*

The following table presents commonly-analyzed rildionuclides on the DOE Complex particularly susceptible to holding time and RDL exceedences due to short half-life;

  • Page9of41

. - ='~ '.

Evaluation ofRadiochemiCal Data Usability April 7, 1995 Table 5.2 Commonly-Analvzed Short Half-Life radi onuclides Nuclide T%

1311 8.04 days 222Rn 3.82 days 21op0 138 days s9sr 50.5 days

1.

Verification If the sample MDC is reported at a level greaterthan the RDLwhen the sample result is less than the MDC or +/-2s counting uncertahty, and no justification is provided for not reanalyzing at a longer count time or greater sample aliquot, data are noncompliant with the contractual RDL. Note the occurrence ofRDL exceedence in the verification rei}ort.

Tum-around time is evaluated as strictlv a contractual issue An agreement should be established between the client and the laboratory concerning tum-around times for reported data. Ifturn-around times are not satisfied, a contractual mechanism should provide for appropriate action.

2.

Validation It should be realized that RDL exceedence may occur for reasons other than excessive radionuclide disintegration. Reference shodd be made to other potential factors such as count.time and aliquot size to apply the appropriate reasoncode(s). The affected results are not qualified, but are flagged only with the reason code "Q 10" if reanalysis can not be performed or deemed by the project as unnecessary.

C.

Sample Preservation.

When appropriate, or in the absence ofknlMl preservation techniques, the preservatives and container types listed in the following table should be used for aqueotis samples.

a e.

eservation an T bl 5 3 Pr dC on tamer eaurrements Nuclide or Group Preservative Container Gross Alpha or Beta Concentrated HCl or HNq to pH# 2 Plastic or Glass 226Ra or 22sRa Concentrated HCl or HNq to pH# 2 Pla8tic or Gla8s U,Th Concentrated HCl or HNq to pH# 2 Plastic or Glass 89Sr I 90Sr Concentrated HCl or HNq to pH# 2 Plastic or Gla8s 1311 HCl/ 2M NaHS03 Plastic or Glass JH None Glass Page 10 of41.

.-,.~*.
.... *.' :7..

.. ~

Evaluation of Radiochemical Data Usability April7, 1995 99Tc Concentrated HCl or aN<l.J to pH# 2 Plastic or Glass Gamma Emitters Concentrated HCl or HNq to pH 2 Plastic or Glass United States Environmental Protection Agency. Manual for the Certification of Laboratories Analyzing Drinking Water, Criteria andProcedures Quality Assurance, Third Edition. EPNS70/9-90/008.

I.

Verification If the proper preservative andbr container type have not been used, note the problem in the verification report.

2.

Validation.

If samples have n0t been preserved properly in the field or have been stored in an improper container, quaiify those sample results < MDC "UJ." Sample results $MDC may llot require qualification; however "J" may be placed if necessary.

If samples with radionuclides amenable to preserative with acid have not been acidified.in the field, but have been acidified in the laboratory prior to subsampling, qualification may not be necessary. The matrix and container type will not affect the radioactive charactesiics of the radionuclides m the sample.

For this reason, neglecting to acidify sa.mples prior to shipmerit to a laboratory should not necessarily result in qualification. However, as radionuclides will adhere to the container walls, acidification (of aqueous samples) either during the sampling event or at the laboratory prior to subsampling is critical to ensure that all radioactive components are in solution, and the representativeness of the sample is.

maintained.*

D.

Instrument Calibration I.

. Scope It is outside the scope of this document to prescnbecalibration requirements for the laboratory: This

  • section provides recommended frequencies, perfonilance and:valuaticin criteria, based on existing ANSI

. standards. Decision8 regarding the deliverable reqUirements for calibration data must be made during the project plamiing phase and communicated to the laboratory in the SOW. Decisions regarding calibration evaluation s~tegy should be influenced b)the strategy outlined In this ~ection in order to provide a consistent approach to data evaluation with respect* to calibration and to expedite the verification and validation processes.

2.

Mechanism for fuitial Calibration Evaluation All calibration data are essential for.,roject records* from the *standp9~t of defending the conditions under which samples were analyzed; however, evaluation of initial calibration data should not involve verification and validation with each reporting batch.

fage 11 of41

Evaluation ofRildiochemical Data Usability April 7, 1995 To provide a mechanism for initial calibration evaluation at the project-level, the follo\\\\'ing approach should be taken..

a.

Laboratory Statements of Work must incorpoae calibration definitions, frequency, and quality control criteria.

b.

Upon award of a contract, the laboratory must provide to the project all initial calibration data for all detecta systems to be used under the contract_ At this time, the control status of initial calibration can be evaluated. The contract must require that the laboratory update the project with changes in calibration status. Tracking numbenfor the initial calibration must be included in all reporting batches for reference purposes.

c.

If the necessary deliverables for initial calibration cannot be provided by the laboratory, tWo potential optfons exist:

i.

if missing deliverables do not significantly affecthe ability to evaluate sample data, the project may acceptthe initial calibration data

- ii.

if missing deliverableseither present a significant nonconformance for project records retention or preclude the ability to evaluate sample data, the project may request the laboratory to perform a new initial calibration in accordance with the deliverables in this section.

3.

General Technical Specifications for Initial Calibration The folloWing technical specificatiorts apply to all initial calibrations and calibration verifications, independent of counting instrumentation category.

a.

Check Source Statistics The.instrument calibration sources should provide adequate counting statistics (less than 1 %

Poisson statistic uncertainty) over the time period for which the source is to be counted.

However, the source shall not be so radioactive as to cause 1) pulse pileups, 2) dead time that is significantly different from that to be expected :from routine analysis, or 3) gain shift in the case of pulse height analyzer systems. *

b.

Radioactive Sources Connnercially pre{>ared and sealed standards Shall not be usd after their stated expiration dates, which are based on radionuclidehalf-life or physical form of the standard, e.g. sealed sorirce or plated planchet. Standards prepared at the laboratory or those purchased without expiration dates should be replaced yearly.

The standard source(s) used in initial calibration shall be NIST-traceable Standard Reference Materials, or equivaknt; however, source(s) used in calibration verification are not required to Page 12 of41


*----------------""--~------~---------------------'---

Evaluation of Radiochemical Data Usability Apri17,1995 be NIST-traceable, Unless measurements of these sources are directly used in calculation of analytical sample data results.

c.

Control Criteria The scope ~fthis document does not include presriptive requirements for calibration, however, quality of analytical ruita is highly depentht on control of the calibration process. To facilitate a framework for defining control of the calibration process, the three following strategies may be incorporated dependant on what instrumentation is being used:

i.

Tolerance charts may be established based on consideration of specific performance characteristics of the instrument and complexity of the matrix of samples that will be analyzed. The required precision of toleranee charts must never be more restrictive than that ofa,quality control chart.

n.

Statistical quality control charts may be eslblished based only on a level of confidence considered necessary for statistical quality control.

111.

Fixed limits may be used by consideration of percent deviation from a known value.

With some radiochemical methodologies (e.g. alpha and gamma spectroscopy) establislment of tolerance or statistical quality control charts may provide unrealistic precision goals (e.g. 5% RPD may exceed a +/-3s control chart limit but still provide adequate instrumental precision). In these cases evaluation *of measured values using a percent deviation approach may provide realistic evaluation of detector precision.

d. *
  • Establishment of control points Establishment of wntrol points for use with a tolerance or statistical quality control chart may.

be approached m two differing strategies, fixed range or moving range.

i.

Fixed range control charts are established by acquiring a predeti;mnined number of poin1s, with associated mean and standard deviation, and comparing subsequent data point acquisitions to those statistical descriptors. This allows for evaluation of instrumental control over time, but may riot represent true precision over time.

ii.

Moving range control charts are established by acquiring a predetermined number of points, with associated mean and standard deviation, and as subsequent points are acquired, trey are included for an up-to-date evaluation of system precision. In using a moving range control chart,. only the most recent 20 points are considered in establishing statEtical descriptors. The use of moving range control charts allows for real-time evaluation of detector control, but does not allow for evaluation of detector control in relation to initial calibration.

e.

Control of Background Page 13 of41

I I*

  • -",,~:*.

~.

... \\... **'.-..

  • Evaluation of Radiochemical Data Usability April 7, 1995 The control limits for.check sources and backgrounds (where applicable) shall be established using a minimum of 20 sequentially measured data points.. For extended background count periods, a series of at least 10 single background measurements is accei}table. No samples, subject to these specifications may ba:ounted until these warning and control limits have been established.

Background count time should be equal to or greater than sample count time unless precluded by extended low level *sample count times, in which case background count time may be less than sample count time.

f.

Geometry With all methods of detection, thecalibration counting geometry used should be the same as that used with the analytical samples.

g.

Background subtraction Calibration data should be background-subtracted, whether data is used in generation of efficiencies, cross-talk, or.resolution evaluation.

h.

Recalibration Recalibration should only be necessary in the event that the instrument/system has malfunctioned and the repaired equipment has responded to.a QC test in ~*fashion that the tolerance level of a control chart has been exceeded, i.e.the operating or response characteristics of the instrument/systemha1!: changed more than the tolerance/control limits permit. Detector calibration is verified according to :frequencies that will satisfy contractual criteria, and

  • according to criteria defining the warrant of corrective action.
4.

Specific Technical Specifications for Calibration Verification

a.

Calibration verification data must btsubmitted with each reporting batch and will be evaluated at that frequency. If samples within a reporting batch are from separate counting batches, calibration verification documentation must be included for all counting batches.

b.

Each reporting batch submitted from the laboratory to the project should contain control charting dilta related to calibration verification for all detectors. used in the analyses of the analytical samples.

c.

Calibration verification is performed. and monitored with tolerance or QC charts for instrumental parameters specific to each typoof detector. If the daily check source count result exceeds the tolerance limits or +/-3F control limit of the appropriate 20 points, the laboratory should recount the check source to verify the out of control condition. If the recount again Page 14 of41

Evaluation ofRadiochemical Data Usability.

April7, 1995 exceeds the control limit, the system is considered out of control, and no samples shall be run on that system until it is brought back into control. ~

recount is in control a third count shall be done and if in contol. analytical sample counting may continue, otherWise no samples shall be analyzed on that system mtil it is brought back into control. Any samples counted after the last in-control check standard must be recounted except for those where decay has eliminated that radionuclide.

d.

If calibration verification data exceed thetolerance limits or the +/-3F control limits, reference must be made to quality control sample data 'in the data package to detennine the extent of calibration nonconformance on the analytiCal batch. Exceedence of control limits may not constitute qualfication of data; but conversely, excessive control limit exceedence may affect all data in an analytical batch, justifying qualification.

5.

Initial Calibration for Gross Alpha and Gross Beta by Gas Proportional Systems Initial calibration review consists of evaluations of efficiency, background, and cross talk.

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

1.*

Summary report(s) for both gross alpha and gross beta including the following:

dates of calibration and planchet preparation; mass attenuation curves including mas of salts, planchet diameter, efficiencies at each mass, and mathematical relationships developed fr~m self absorption curves;

  • matrix residue identity used for curve establishment-e.g. ASTM Type II water; identities of nuclides used for calibration; geometries; backgrounds; charts of voltage plateaus; number of counts accumulated in each channel for each standard; count times for all analyses; calculated activities for all analyses; cross talk factors.

ii.

Evidence of decay correction of standardS prior to calculation of efficiencies, if appropriate; iii.

All raw data supporting initial calibration; iv.

Certificates for standards used in calibration.

b.

Frequency An initial calibration for gas proportional counters sbuld be performed when out-of-control conditions, indicated from control charting, necessitate recalibration of the instrument or if the project.requests recalibration based on deliverable deficiencies.

c.

Performance Criteria Page 15 of41

Evaluation of Radiochemical Data Usability April7, 1995 The following criteria are recommerided in the establishment of initial calibration* mass attenuation curV'es for gas proportional counters:

i.

Each alpha and beta calibraion standard shall be counted to an accumulatiori of I 0,000 counts ii.

Each mass attenuation curve should consii of at least 10 points well distributed throughout the mass range unless instrument specific programs designate otherwise iii.

Operating voltage on the plateau shail be established to achieve a cross talk of' into the $

channel of< 10%, and$ into the " channel of <1 %

iv.

The instrument backgroilnd must be checked at the time of initially calibrating the detector.

v.

The matrix residue used in establishment of the mass attenuation curves must be reasonably well matched to the expected sample matrix.

vi.

Plated planchets used must be less than 3 years old vii.

The maximum calibration plan:het density for alpha and beta counting should bt# 5 mg/ crri.

viii.

The sources used for tle determination of cross talk should be of similar geometry and isotope content to that of the analytical samples; however, when perfomung analyses for gross measurement, a standard isotope source is acceptable (e.g.241Am for gross alpha, and 137Cs, 90Sr/90Y for beta calibration).

ix.

Standard activity shall be decay corrected (if applicable) prior to calculation of instrument efficiency.

6.

Calibration Verification for Gross Alpha and Gross Beta by Gas Proportional Systems Calibration verification consists of acceptable efficiency and background data

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

i.
  • Matrix residue identity used for check source; ii; Date of preparation of planchets used in calibration verification; iii.

Number of counts in each channel for each mass-efficiency calibration 'standard; iv.

Calibration points including mass on planchet and attenuation factor;

v.

Tolerance chart or statistical control chart of the the appropriate 20 efficiencies with +/-lf limits with residue weigh~ for each efficiency verification; v1.

Tolerance chart or statistcal control chart of the appropriate 20 instrument backgrounds with

+/-3F limits; vii.

Listing of XIY coordinates used in constructing the control charts; viii.

Evidence of decay correction of standards prior to calculation of efficiencies, if appropriate; ix.

Geometries used in analysis.

b.

Frequency *

1.

Calibration verificationmust occur on a daily basis, or prior to use. The only exceptfon to this specification is when perfonning low level counts with extended count times precluding the verification of calibration on a dail)l>asis. In this case, calibration may be verified on a weekly basis.

Page 16 of41

11fY.* *~:r*

  • Evaluation of Radiochemical Data Usability
  • April 7, 1995 ii.

Both the operating voltage (on the plateau) and the ~lk shall be checked on a semi-annual basis.

c.

Performance Criteria

1.

Each alpha and beta verification standard shall be counted to an excess of 10,000 counts ii.

  • The check source count result for both alpha and beta should be maintained on a control chart with the mean and +/-3F limits plotted; If the daily check source count result, for either alpha or beta, exceoos the tolerance or statistical F control linit, the checks source must be recounted to verify the measurement was correct. If the recount gain exceeds the control limit, the system is consiclerCd out of control, no amples shall be run on that system until it is brought back into control. If the recount is in control a third count shall be done and, if in control, analytical sample counting may continue, otherwise no samples shall be run on that system until it is brought'back into control. Any samples counted after the last in-control check standard shall be recounted.

iii.

  • Efficiency for both alpha and beta must remain within turance or statistical +/-3F control limits of the appropriate 20 efficiencies.

iv.

Instrument backgromd for both alpha and beta must be maintained on a control chart with the

  • mean and +/-3F limits plotted.
v.

Following gas changes, gas should be allowed to flow for a period of time necessary to purge the system; cht;ck souic~ and background counts must be performed following a gas change

d.

Verification Verificatim of completeness of deliverables must be performed prior to validation. If deliverables are found to be inconsistent with the listing in this section, contact the laboratory to request the additional infonnation. If the information can not be obtained, ihdicate this in the verification report as a non-correctable *problem

e.

Validation If the frequency of calibration verification does not meet the frequency specifications, place the appropriate reason code on the affected data. Control charts shall be evaluated for out-of-control condition;. If any of these conditions are met, reference must be made to the labOratory case narrative for justification for analyzing analytical samples under non-Compliant conditions; QC samples should be evaluated*t> determine if the non-compliant check source is indicative of a systemic problem or if it is a chance occurrence. If the occurrence is considered a chance occurrence, and samples do not seem affected, no qualification action is necessary. If the occurrence is considered systemic, and sample results seem to be affected, sample results may be qualified "J" or "R" based on specific conditions in the analytical batch.

Page 17 of41


,----------,----,--------,--,-.,..--------------:----cc-;-------c-- ------ -

~~-
  • i Evaluation of Radiochemical Data Usability April 7, 1995 If standard source(s) have aged greater than the expiration date on the certificate(s), affected sample results should be qualified "J," and qualified "R" using professional judgement.
7.

Initial Calibration for Liquid SCintillation Initial calibration review consists of acceptable efficiency and background data

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

i.

Summary report(s) including dates of calibration, geometries, efficiency, background, quench levels, count times for all amlyses, number of counts accumulated for each standard, measured activities for all analyses; ii.

QC source identity; iii.

Matrix used for efficiency curve establishment (e.g. ASTM Type Il water);

iv.

Evidence of decay correction of standards prior to calculation of efficiencies, if appropriate;

v.

Calibration points including quench level and measured results; v1.

All raw data supporting initial calibration; vii. \\ Certificates for standards used in calibration.

b.

Frequency An initial calibration for liquid scintillation systems should occur when out-of-control conditions, indicated from control charting, necessitate recalibration of the instrument or if the project requests recalibration based on deliverable deficiencies. *

c.

Performance Criteria

i. --

The quench curve standard's vial characteristics (glass or plastic) and volume shall be similar to that of the samples to be counted.

ii.

The effi:iency standard shall be counted prior to each analytical batch and shall be counted to a low counting error Qess than l'/o at 3F, or 100,000 counts) at least twice. The average of the efficiencies shall be used to calculate the activity of the samples. The standard need not be prepared daily, but shall be replaced if a decrease in efficiency is noted or phase separation is apparent in the cocktail.

iii.

The tritiated water solution prepared from the flame-sealed NIST standard reference material (or equivalent) shall be replaced or recalibrated every 6 months with a fresh standard.

iv:

For those labs using the internal standard method of quench cm:ction, the tritium standard used to spike the samples shall be recalibrated or replaced on an annual basis.

v.

For those labs using an extemalquench monitor and a variable quench protocol, a minimum of ten standards, with quench's well distributed over the appropriate range, shall be USed to

  • Page 18 of41

~.

I._.. f.r:l'*

I

. *~.-

Evaluation of Radiochemical Data Usability

  • April 7, 1995 detenirine the quench curve. Each standard in the quench set will be counted to accumulate a minimum of 100,000 counts.
8.

Calibration Verification for Liquid Scintillation Calibration verification consists of acceptable efficiency and background data.

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

1.

Matrix used for check source (e.g. ASTM Type II water);

11.

Calibration point(s) including quench level, number of counts for each point, and measured result(s);

iii.

Tolerance chart or statistical control chart of the appropriate 20 efficiencies with +/-8' limits iv.

Tolerance chart or statisfoal control chart of the appropriate 20 instrument backgrounds with

+/-3F limits;

v.

Listing ofXN coordinates used in constructing the control charts; vi.

Geometries used in. analysis; vii.

Evidence of decay correction of standards prior to calculation of efficiencie.s, if appropriate.

b.

Frequency Calibration verification must occur on a daily basis, or prior to use. The only exception to this

mecificationis when perfomiinglow level counts with extended count times precluding the verification
  • of calibration on a daily basis; In this case, calibration may be verified on a weekly basis.
c.

Performance Criteria

i.

On each day that samples willbe counted,. an unquenched flame-sealed check source vial, shall be counted prior to sample counting. In excess of l 0,000 counts of the respective activity shall be accumulated.

ii.

On each day that samples will be counted, the background count rate shall be deterrniiied by counting a vial free of the analyte(s) of interest. The driration of this background count shall be as long, if not longer, than thatfor the analytical samples. This detel-m.ination shall be.made separate from anyvials that are counted to meet method blank requirements of the respective statement of work.

iii.

For those labs that quench correct the background activity, each day thaf samples will be counted, a vial ofthe quenched backgroud shall be counted to e11sure control of the instrument background. The quench level shall lie within the tolerance limits of the set of 20 used in establishing initial calibration.

iv.

For tritium analyses, the laboratory shall show that the.vater used for the background, to be free of tritium activity (by comparison to an EPA blank water or through other means).

Page 19 of41

.. ~.*.. _:.. **...

Evaluation of Radiochemical Data Usability April7, 1995

. v.

If the ~ily check sourcecount result or the background count rate result exceeds the 3 sigma limit, recount to verify the out of control condition. If the recorint again exceeds the control limit, the system is out of control, no samples shall btrun on that system. until it is brought back into control. If the recount is in control a third count shall be done and ifin control analytical sample counting may continue, otherwise ~o samples shall be analyzed on that system until it is brought intocontrol. Any samples analyzed after the last in-control check standard shall be reanalyzed.

d.

Verification Verificatim of completeness of deliverables must be performed prior to validation. If deliverables are found to be inconsistent with the listing in this section, contact the laboratory to request the additional information. If the information can not be obtained, indicate this in the verification report.

e.

Validation If the frequency of calibration verification does not meet the frequency specifications, place the appropriate reason code on the affected data. Control charts shall be evaluated for of out-of-control condition;. If any of these conditions are met, reference mti,st be made to the laboratory case narrative for justification for analyzing analytical samples under non-compliant conditions; QC samples should be evaluated b determine ifthe non-compliant check source is indicative ofa systemic problem orifit is a chance occurrence. If the occui'rence is considered a chance occurrence, and samples do not seem affected, no qualification action is necessary. If the occurrence is considered systemic, and sample results seem to be.affected, sample results may be qualified "J" or "R" based on specific* conditions in the analytical batch.

If standard source(s) have aged greater than the expii;ation date on the certificate(s), affected sample results should be qualified "J," and qualified "R" using professional judgement.

9.

Initial Calibration for Alpha Spectrometry Special no1e for alpha spectrometry calibration. Where a method or Statement of Work do not require the detenninationoftracer recovery, the determination of detector countiri.g efficiency is not necessary.

In this circumstance, the following specifications for efficiency calibration are not ap~licable.

  • Initial calibration review consists of acceptable efficiency, background, and peak resolution data.
a.
  • Deliverables The basic deliverables required to complete the checks are described in this section:
i.

Summary report(s) including dates of calibration,efficiency, background, peak(s) resolution, count times for all analyses, number of counts accumuiated in each channel for each standard, measured activities for all analyses;

11.

Matrix used for efficiency establishment (e.g. ASTM Type II water);

iii.

Evidence of decay correction of standards prior to calculation *of efficiencies, if appropriate; Page 20 of41

";.:r:::w.. :.:*,,,,..

....,-~.

  • .. ~ *, *.*.

Evaluation of Radioehernical Data Usability April 7, 1995 iv.

Calibratjon points including efficienty, energy, and peak resolution for relevant peaks;

v.

Alliaw data supporting initial calibration;

  • vi.

Certificates for standards used in calibration;

b.

Frequency An initial calibration for alpha spectrometry systems should occur when out-of-control conditions, indicated from control charting, necessitate recalibration of the instrument or if the project requests recalibration based on deliverable deficiencies.

\\

c.

Performance Criteria

i.

A *single check source may be used for efficiency calibrations, and should consist ol"1 Am, 238Pu, 239Pu, or 244Cm. A running average ()f up to 5.efficiencies for one or more of these radionuclides may be used as the efficiency for sample calculations.

ii.

At least two, preferably three, isotopes shall be used for energy calibrations. *The isotopes used should be from those listed above.

iii.

Radionuclides should be selected Vhich have energies that will limit the detection of attenuated alpha events ofhigherenergies. The sources should be of a high enough quality as to limit the amoi.int of tailing created by atteriuation. Peaks outside the ROI ~eed not be identified.

10.

Calibration Verification for Alpha Spectrometry Calibration verification consists of acceptable efficiency, background, and peak resolution data.

a.

Deliverables The basic deliverables required to complete the checks are described in. this section:

i.

Tolerance chart or statistical control chart of the appropriate 20 efficiencies, peak eneriies or cqannel numbers with +/-3F limits; ii.

Background results; iii.

Resohition demonstration of relevant peak(s);

iv.

Listing of peak energies;

v.

Listing ofX/Y coordinates used in constructing tlie control charts; vi.

Evidence of decay correction of standards prior to calculation of efficiencies, if appropriate; vii.

Geometries used in analysis.

b.

Frequency Calibration verifimtion will be performed weekly prior to analytical sample counting, unless analytical conditions warrant more frequent verification.

Page2lof41

~-.-... '
'

~ :.. ; *'. '

I.

. *~.::....

Evaluation of Radiochemical Data Usability April7, 1995 The background count rate shall be detennilledat least once per month by counting.a clean mount of the same geometry to sample mounts. The, duration of this background count shaUbe at least 12 hours1.388889e-4 days <br />0.00333 hours <br />1.984127e-5 weeks <br />4.566e-6 months <br />.

c.

Performance Criteria

1.

The check source shall re counted a sufficient period of time to accumulate an excess of 2,000 counts in each region of interest. This check source count will be used to monitor counting efficiency, peak energy, and peak resolution. The region of illtest of a minimum of one isotope from the following (241Am, 238Pu, 239J>u, or 244Cm -

234U and 238U are also recommended) shall be used to monitor these perfonnance parameters.

II.

iii.

If the daily check source count result exceeds the 3 sigma limit, recount to verify the out of control conrution. If the recount again e&eeds the control limit, the system is out of control, no samples shall be run on that system until it is brought back into control. If the recount is in control a third count shall be done and if in control analytical sample counting may continue, otherwise no samples shall be analyzed on that system until it is brought into control. Any samples analyzed after the last in-control check standard shall be reanalyzed.

The isotope(s) radionuclides/isotopes chosen for evaluation of peak resolution should be the same as used in the irtitial calibration. The resolution of the detector shall be sufficient to minimize the tailing of counts from peaks of higher energy into regions of interest (ROI) of lower energy peaks.

iv.

The background count rate of a detector used for alpha spectrometry will naturally increase in those regiom of interest corresponding to radionuclides present in the counted samples. This

  • background count rate is corrected for iithe result calculations and thus has little impact on the quality ofthe reported data. However, it is possible that over time this background count rate could increase to a level where the RDL. of some analytes are impacted, unless the lab has corrected for this with longer counting times.If this problem should be observed, the corrective action of cleaning or replacing the detec~or should be discussed with the laboratory.
d.

Verification Verificatim of completeness of deliverables must be performed prior to validation. If deliverables are found to be inconsistent with the listing in this section, contact the laboratory to request the additional information. If the infonnatioii can not be pbtained, indicate this in the verification report.

e.

Validation If the frequency of calibration verification does not meet the frequency specification~ place the appropriate reason code on the affected data. Control charts shall be evaluated for of out-of-control condition;. If any of these conditions are met, reference must be made to the laboratory case narrative for justification for analyZing analytical samples under non-compliant conditions; QC. samples should be evaluated t> determine if the non-compliant check source is indicative ofa systemic problem or if it is a chance occurrence. If the occurrence is considered a chance occurrence, and samples do not seem affected, no qualification action is necessary. If the occurrence is considered systemic,. and sample Page 22 of41

JJ

'l";wj'

    • lr
  • ~...

Evaluation of Radiochemical Data Usability April7,1995 results seem to be affected, sample results may be qualified "J" or "R" based on specific conditions in the analytical batch.

If standard source(s) have aged.greater than the expiration date on the certificate(s), affected sample'

  • results should be qualified "J," and qualified "R" using professional judgement.
11.

Initial Calibration for Gamma Spectrometry Initial calibration review consists of aceptable peak energy, efficiency, background and peak resolution data

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

i.

Summary report(s) including dates of calibration (energy, efficiency, background, peak(s) resolution), geometry, count times for all analyses, number of counts accumulated in each channel for each standard,.measured activities for all analyses;.

u.

.Matrix used in the geometry standard (e.g. epoxy);*

iii.

Density of standard; iii.

Evidence of decay correction of standards prior to calculation of efficiencies, if appropriate; iv.

Calibration points including'*efficiency, energy, and peak resolution for relevant peaks;

v.

All raw data supporting initial calibration; vi.

Certificates for standards used in calibration.

b.

Frequency An initial calibratipn for gamma spectrometry systems should occur when out-of-control conditions, indicated from control charting, necessitate recalibration of the instrument or if the project requests recalibration based on deliverable deficiencies. An initial emgy/efficiency calibration for each geometry in conjunction with the preparation of a tolerance or QC chart shall be performed, thus linking the calibration to the tolerance/QC chart.

The energy/efficiency calibration of gamma spectrometers shall be performed at least semi-annually.

J The energy/effi.ciency calibration standard shall have at least three gammzmitting energies.. Ifthere are only 3 energies inthe calibration standard, then the difference between the energies shall be at least 500 keV with one energy being less than 300 keV. Enerw calibration photopeaks shall have 1332 keV

( cobalt-60) Full *Width at Half Maximum (FWHM) values of less than 4 keV for HPGe and Ge(Li) detectors. Where Na! detecbrs are used for an analysis, as permitted by the method/SOW,* the FWHM shall be less than 1.5 times the mstrument manufacturer's stated specification for FWHM.

c.

Performance Criteria Page23 of41

Evaluation of Radioohemical Data Usability April 7, 1995

i.

Where the method is not specific to a single radi01_1uclid using a single energy (e.g. cesium-137 using mly 661 keV), the calibration source shall have several emissions over an energy range of about 0 -.2000 keV.

The source shall not be used for calibration for more than 4 half-lives beyond the calibration date indicated on the certificate of calibration.

ii.

The calibraticn source shall be counted to accumulation of an excess of20,000 counts in.each

.region of interest.

iii.

The energy calibration standard shall have at least three gamma emitting energies. The difference between the energies shall be at least 500 ke V. with one energy being less than 300 keV. Energy calibrationphotopeaks shall have Full Width at Half Maximum (FWHM) values ofless than 4 keV.

12.

Calibration Verification for Gamma Spectrometry Calibration verification consistsof acceptable peak energy, efficiency, background and peak resolution data.

a.

Deliverables The basic deliverables required to complete the checks are described in this section:

i.

Tolerance chart or statistical control chart of the appropriate 20 efficiencies with +/-8' limits.

H.

Tolerance chart or statistical control chartof the appropriate 20 relevant peak energies with

+/-3F limits; iii.

.Resolution demonstration ofrelevant peak(s);

iv.

Listing ofX/Y coordinates used in constructing the control charts;

v.

Evidence of decay correction of standards prior to calculation of efficiencies, if appropriate; vi.

Geometries used in analysis.

b.

Frequency

i.

Energy, efficiency and resolution:alibration verification must occur on a daily ba!iis or prior to use. The only exception to this specification is when performing low level counts with extended courtt times precluding the verification of calibration on a daily basis. In this case, calibration may be verified on a weekly basis.

ii.

The background count rate for the representative geometries shall be determined at least once per week.. Two alternatives should be consideredin counting the background rate: by counting each representative geometry filled to the respective volume* with distilled or deionized water, or counting without a representafre geometry, an empty cave. The duration of the background count shall be as long, if not longer, as that of the respective samples.

c.

Performance Criteria Page24 of41

.**.\\':..

'*'**',.'°.v*.'

... ~*..., *-*: *; *'

Evaluation of Radiochemical Data Usability April 7. 1995

i.

The check source used to verify energy calibration should include radionuclides with energies covering the expected energy range of the radionuclides ofitcrest. It is recommended thaflow, mid. and highenergy radionucli.des be included for verification of energy, efficiency, and peak resolution.

ii.

Energies of the low, mid and high energyradionudlles should fall within+/- 0.2keV of the initial calibration energies.

iii.

The check source shall be counted to accumulation of an excess of 20,000 countm the low, mid. andhigh energy ranges of interest. Examples of checksources which will cover common radionuclide energy ranges are listed in the previous section.

These emissions shall not be usm for calibration verification for more than 4 half-lives beyond the calibration date indicated on the certificate of calibration.

1v.

If the daily check source count result for counting efficiency at a low, mid, and high points exceeds.the i>lerance chart or statistical 3 F control limit, recount to verify the out of control condition. If the recount again exceeds the control limit, th system is out of control, no samples shall be run on that system until it is brought back into control. If the recount is in control a third coµnt shall be done and if Ji. control sample counting may continue, otherwise no samples shall be run on that systtm until it is brought back into control. Any samples counted after the last in-control check standard shall be recounted.

.. v.

Peak resolution from the low, mid, and high energy ranges should be<+/- 1 FWHM from the respective peaks in the initial calibration. *

d.

. Verification Verificatim of completeness of deliverables must be performed prior to validation. If deliverables are found to be inconsistent with the listing in this section, contact the laboratory to request the additional information. If the information can not be obtained, indi.cate this in the verification report.

e.

Validation If the frequency of calibration verification does not meet the frequency specification~ place the appropriate reason code on the affected data. Control charts shall be evaluated for of out-of-control condition;. If any of these conditions are met, reference must be made to the laboratory case narrative for justification for analyzing analytical samples under non-compliant conditions; QC samples should be evaluated i> determine if the non-compliant check source is indicative of a systemic. problem or if it is a chance occurrence. If the occurrence is considered a chance occurrence, and samples do not seem affected, no qualification action is necessary. If the occurrence is considered systemic, and sample results seem to be affected, sample results may be qualified "J" or "R" based on specific conditions in the analytical batch.

If standard source(s) have aged greater than the expiration date on *the certificate(s), affected sample results should be qualified "J," and qualified "R" using professional judgement.

Page 25 of41

Evaluation of Radiochemical Data Usability April 7, 1995 E.

Quality-indicator samples quality-indicator samples are evlluated during data validation to determine the control of the analytical method, and under what conditions the usability of data has been affected.

The strategy by which quality-indicator samples are evaluated involves an eva,luation of whether the difference between expected and measured results is statistically significant when compared to their Total Propagated Uncertainties (TPU). The mathematical relationships presented in the following sections are compared to a factor corresponding to a statisticalevel of confidence. When the relationship exceeds the factor, the two results differ at that statistical level of confidence wheuompared to their TPU.

The statistical assumption inherent in these tests is that sample results are drawn from nonnally-distributed populations with estimated means and known varianeesFactors in the TPU relationship may originate from populations which are not necessarily normally distributed (e.gcounting uncertainty).

However, use of sample reailts and TPU, assuming approximation to the normal distribution, provides a reasonable ard appropriate approach to evaluating control. of analytical conditions. Presented in this procedure are statistical decision-maling levels at 95% and 99% levels of confidence (decision-making factors are 1.96 and 2.58, respectively). Projects may choose other levels of confidence and decision-making factors based on project DQOs, with the realization that qualification decisions made through validation will be at differing levels of confidence and conservatism. The following table provides examples of these decision-making factors which are applied as decision-making tools through this procedure:

Table 5.4 Examples of Confidence Levels for Qualification Decision-Makin g Confidence Level, %

Decision-making Decision-making level factor 50 0.68 more conservative 67 1.00 75 1.15 90 1.645 95 1.960 99 2.575 less conservative adapted from Taylor, J.K. Quality Assurance of Chemical Measurements. Lewis Publishers, 1987.

Listed in this section are guidance for qualification for single quality-indicator samples being outside control criteria based on a 95% and 9~/o level of confidence. Analytical samples should not be rejected Page 26 of41

  • Evaluation of Radiochemical Data Usability.

April 7, 1995 based on a singular quality control sampe. Effects of other QC sainple deficiencies must be considered to evaluate whether conditions are such to justify rejection of data. Appendix B provides analytical decision-making guidarice for situations wbu: multiple quality _deficiendes are encountered. A strategy to incorporate into laboratory SOWs is to require corrective action ifthe1:2.58F limits are exceeded.

If the limits are exceeded again, data may be reported, ht are subject to qualification through validation.

The laboratory case narrative should reflect the corrective action.

The tests in thi.s section are meaningful only if the radioanalytical method functions properly. If a method is deemed seriously out of control, the tests in this sectiomre not appropriate, and no further validation needs to be done; all results may be considered unusable.

  • 1.

Total Propagated Uncertainty (TPU)"

The tests pres".llted in this section rely heavily upon evaluation of uncertainty as~ociated with radioanalytical results. *The factors* presented in the TPU relationship.in* ~ppendix D must be communicated to the laboratory prior to sample analysis in order for the laboratory :to provide the relevant components of the TPU. The random factors in the TPU relationship include counting uncertainty, and net counlrate uncertainty; the remaining terms comprise the systematic factors. Many laboratories choose to report uncertainties separately astotal _random and total systematic...These factors are acceptable to use in thetests in this section providing that the components of the uncertainties are recognized.

Jn the event that not all the requested uncertainties are available, the magnitude of*TPU must be eValuated consi~ which (£)mponents are the dominant factors in the relationship. At relatively low

  • count rates, the random components will likely be the dominant factors; and at high rates, systematic components may be dominant. Communicatin with the laboratory in the evaluation of TPU in absence ofall components is cnicial in evaluating the tests presented in this section.
2.

Standard Traceability

  • Standards used in the preparation of qualit)control samples (laboratory control samples, lJll!,trix spikes) or sample-specific spikes (tracers or carriers) shall be shown to be traceable to a reliable source (e.g.

NIST, IAEA).

a.

Verification Verification of the identity of the standard used in11J1lity control sample preparation, or sample-specific spiking shall be performed by tracing the standard control number from the certificate to the quality

  • control sample p~eparation documentation~
b.

Validation If a standard is determined to be not traceable, a qualifier i!llot required, but the reason code "E02" mtist be placed next to sample results that are in the same preparation batch with the affected QC sample.

3.

Laboratory Control Sample (LCS)

Page27 of41

. ; *:. ***~.

1** *.

Evaluation of Radiochemical Data Usability April 7, 1995 The purpose of an LCS is to monitor the accuracy of the preparation and analysis of the analytical samples, proVided that LCS is fully homogenized prior to preparation and analysis. The LCS mu8t be the same matrix type as the analytical samples (e.g., water, soil). Exceptions should be made in cases of novel matrices (e.g. sludge, oil, biota). The spike in the LCS should be of a level near that of the analytical sunples, or contractually specified at a predetermined level (e.g. 20 times the MDC may be appropriate).

Three types of LCS material have been identified as being appropriata:o evaluate laboratory process accuracy:

Traceable Reference Material (TRM)

Standard Reference Material (SRM)

Well Characterized Reference Material (WCRM)

a.

Deliverables The basiC deliverables necessary to complete the checks are described in this section:

  • Traceability certificae(s) for TRM and/or SRM with uncertainty associated with the standard.

Measured result of LCS expressed in activity/unit weightor volume.

TPU for LCS expressed in activity/unit weight or volume.

MDC ofLCS expressed in activity/unit weight or volume.

Expected result ofLCS expressed in activity/unit weight or volume.

If samples within a reporting batch originate from separate preparation batches, validation requires that laboratory control sample results be included from all appropriate preparation batches.

b.

Frequency The preparation of the LCS occurs in the laboratory simultaneously with the preparation batch of analytical samples. The LCS must be ana!Jed on the same detection system as the s~ples with which it was prepared and using the same analysis cod.itions as with the samples. The MDC of the LCS must be less than the specified RDL. The LCS should be prepared at a frequency of one per batch of up to 20 analytical samples.

c.

Verification Verify that one LCS was prepared at the same time with the analytical samples, is of the same matrix, and has been aruiyzed with an analytical batch of up to 20 analytical samples. If these criteria are not met, state as a non-correctable problem in the verification report.

d.

Validation IfLCS dataare not reported for a sample(s), place the reason code "LOS" on the sample results for the affected preparation batch.

Page28 of41

\\

I I I

I, I I I!

__ /

Evaluation of Radiochemical Data Usability April 7, 1995 LCS - Test for bias The normali7.ed difference between the me:.1ured LCS and expected LCS results given by the following relationship is used in testing the null hypothesis that the 111easured and true results of the LCS do not differ significantlywhen compared to their respective TPU..

LCSmeas.=

LCScxp. =

TPUmeas.=

TPUcxp. =

LCS

- LCS Measured.LCS result

  • Expected result ofLCS meas.

exp.

Total Propagated Uncertainty of measured result Total Propagated Uncertainty ofexpected result If the normalized differenceis between 1.96 and2.58, or between -1.96 and -2.58, qualify sample results

$ MDC "J, 11 as the conclusion is reached that the spiked and expected resultS differ at the 5% ievel of significance. Qualify results < MDC 11 J 11 only if the normalized difference shows a negative bias. If the

. nonnalizeddifferenceis greater than 2.58 or Ie8s than -2.58, consider the effects of deficiencies in other quality-indicatorsamples prior to qualifying sample results "R, 11 as at the 1 % level of significance, the

  • . conclusionis reached that the spiked and expected results are different enough to indicate a* significant positive or negative bias. If multiple quality deficiencies are encountered, qualify using.the guidance
  • provided in Appendix B.
4.

Laboratory Duplicate.

The P\\UPOSe of a laboratory duplicate is to mo~or the precision of the analytical method, provided the.

sample is fullyhomogenized prior to preparation and analysis. The laboratory duplicate is a randomly chosen split of an analytical sample into two aliquots prior to sample preparation. To provide for relevancyofla,boratory duplicate data, the sample chosen should have measurable activity(i.e. >MDC and 2F countinguncertainty); however, the test provided in this section may be performed. on results <

MDC or 2F countinguncertainty. In analytical methods where no sample preparation or separation is performed (e.g. gamma spectrometry), a sample recount tnay be performed in lieu of'a laboratory duplicate, although qualification under thes~ conditions should be based on instrumental performance, as most gamma spectrometry entails minimal sample preparation.

a, Deliverables The basic deliverables necesi;ary to complete the checks are described in this section:

Measured result of sample expressed in activity/unitweight or volume.

Measured result of duplicate expressed in activity/unitweight or volume.

MDCs of sample and duplicate expressed in activity/unitweight or volume.

TPU of sample and duplicate expressed in activity/unitweight or volume.

~. :

Page 29 of41

.'~..... -..

Evaluation of Radiochemical i:>ata UsabilitY April 7, 1995 If samples within a reporting batch originate frc)m separate preparatfon batches, validation reqwres that laboratory dttplicate results be included from all appropriate preparation batches.

b.

Frequency The preparation of the laboratory duplicate occurs in the laboratory simultaneouslywith the preparation batch of analyticalsamples. Analyses of the laboratory duplicate must proceed using the same analysis conditionsused with the samples. The laboratory duplicate shouldbe prepared at a frequency of one per preparation bat<;h ofup to 20 analytical samples.

c.

Verification Verify that one laboratory duplicate was prepared at the same time with the analytical samples, is of the same matrix, and has been analyzed with a preparation batch of up to 20 analyticalsamples. If there criteria are not met, state this aS a non-correctable problem In the verification report.

d.

Validation Iflaboratoryduplicatedaµtare not reported for a sample(s), plaee the reason code "D04" on the sample results for the affected preparation batch.

Duplicate - Statistical test for difference between sample and duplicate The normalized absolute difference between the sample and laboratqry duplicate, given by the relationship below is used in testing the null hypothesis that the results do not differ significantlywhen compared to their respective TPU.

s D

TPU, TPU0 S-D Sample result.

Laboratory duplicate result Total Propagated Uncertainty of the sample Total Propagated Uncertainty of the duplicate If the sample or laboratory duplicate results are less than their respective MDC, the results may still be used in this relationship to determine precision.

If the nonnalized absolute difference is greater than l. 96, qualify the affected sample results "J," as at the 5% level of significance, the conclusion is reached that the sample and laboratory duplicate differ.

If multiple quality deficiencies are encountered, qualify using the guidance provided in Appendix B.

5.

Matrix Spike (MS)

  • Page 30 of 41

'**I::

Evaluation of Radiochemical Data Usability April 7, 1995 The purpose of a matrix spike is to measure the effect of interferences*from the sample matrix that will preclude accurate quantitationby the instrumentation, provided that sa'mples:are fully homogenized prior to preparation and analysis. The matrix spike is a split ofa field~originatinganalytical sample in which one half of the split is spiked with a known amount of radionuclide of interest prior to sample preparation.

Due to difficulties in homogenization of solid samples for gamma analyses, a matrix spike may not present useful information.

Matrix spikes may present useful information for aqueous samples; however, should not be used for energies <l OOKeV.

Matrix spikes may not be required for methods where a carrier or tracer is used, provided that the tracer chosen is chemically similar to the radionuclide of interest. Matrix effects will be detected through tracer*

recovery; however, difficulty may be experienced in ascertaining that poor recovery is due to matrix effect or through losses in separation.

a.

Deliverables The basic deliverables necessary to complete the checks are described iri this section:

Traceability certificate(s) for standard with uncertainty associated with the standard.

Measured result of the unspiked sample in activity/unitweightor volume.

Measured result of MS expressed in activity/unitweight or volume or volume Expected result ofthe MS expressed in activity/unitweight or volume.

MD Cs of unspiked sample and MS expressed in activity/unitweight or volume.

TPU ofunspiked sample, and MS expressed in activity/unitweight or volume.

If samples within a reporting batch originate from separate preparation batches, validation requires that matrix spike results be included from all preparation hatches.

b.

Frequency The preparation of the matrix spike occurs in the laboratory simultaneously with the preparation batch.

of analytical samples. tf notably different matrices are present within the same preparation batch, matrix spikes should be prepared for each matrix.. Analyses of the matrix spike must proceed using the same analysis conditions used with the samples. The matrix spike should be prepared at a frequency of one per preparation batch ofup to 20 analytical samples.

c.

Verification V erifythat one matrix spike was prepared at the same time with the analytical samples, was of the same matrix as the analytical samples, and has been analyzed with a preparation batch ofup to 20 analytical samples. If this frequency is not met, state this as a non-correctable problem.

d.

Validation Page 31of41

Evaluation of Radiochemical Data Usability April7, 1995 Ifmatrixspikedataarenotreportedfor a sample(s), place the rea8on code "MOS" on the sample results for the affected preparation batch.

Matrix Spike - Test for matrix-induced bias The nonnali7.eddifferencebetween the spike result (SSR-SR) and exp~ted v~ue (ER) of the MS given

  • by the following equation is used in testing the null hypothesis that the expected and measured results of the MS do not differ significantlywhen compared to their respective TPU.

SSR SR ER TPU (SSR - SR) - ER

.fTPU2 +TPU 2 +TPU2 V

SSR SR ER Spiked Sample Result Sample Result (unspiked)

Expected Result (spike amount)

Total Propagated Uncertainty Ifthenormali7£ddifferenceis between.l.96 and2.58, or between -1.96 and-2.58, qualify sample results

$ MDC "]," as the conclusion is reached that the spiked and expected results differ at the 5% level of significance. Quatify sample results < MDC "UJ" only ifthe normalized difference shows a negative bjas. If the normalized difference is greater than 2.58 or less than -2.58, consider the effects of deficienCiesin other quality-indicatorsamples prior to qualifying sample results "R," as at the level of significance of I%, the conclusion.is reached that the spiked and ex.pected results differ to a point indicative of a significantpositive or negative matrix-induced bias.,

Consideration should be given to the ~imilaricy in matrix type among samples in the preparation batch batch~. If the matrices differ notably (particularly in soil particle size) qualification may be placed on only the sample associated with the matrix spike. If matrices do not differ notably, qualificationmay be placed on all samples in the preparation batch. If multiple quality deficiencies are encountered, I

qualify using the guidance provided in Appendix B.

6.

Method Blank The purpose ofa method blank is to monitor the presence of contaminatio~of the analyte of interest in the sample preparation and analysis processes. The method/blank is a laboratory-generated sample of the same matrix as the analytical samples but-in absence of the ana~yte of interest.

a.

Deliverables The basic deliverables necessary to complete the checks are described in this section:

Measured result of method blank expressed in activity/unitweight or volume.

MDC of method blank expressed in activity/unitweight or volwne.

TPU of method blank expressed inactivity/unit-Weight or :volume.

Page32 of41

  • Evaluation of Radiochemical Data Usability April 7, 1995 If samples within a reporting batch originate from separate preparation batches, validation requires that method blank results be included from all preparation batches.
b.

Frequency The preparation of the method blank occurs in the labo:r;atory simultaiieouslyWith the preparation batch of analytical samples. Analyses of the method blank must proceed using the same analysis conditions used with the samples. The method blank should be prepared at a frequency of one per batch of up to 20 analytical samples.

c.

Verification Verify that all method blank activities aie less than their MDC and 2s counting uncertainty. If method blank activityis greater than the MDC or 2s countinguncertainty,note this in the verification report.

Verifythatonemethodbiank was prepared at the.same time with the ailalyticalsamples, is of the same matrix, and has been aDalyz.ed at a frequencyofone per preparation batch ofup to 20 analytical samples.

If these criteria are not met; state this as a non-correctable problem in the verification report.

d.

Validation If method blank data are not reported for a sample( s ), place the reason code "B06" on *the sample results for the affected preparation batch.

If either of the following conditions are met, no actions outlined in this section need to be performed:

method blank result is less than its MDC or l.ess than its 2s counting uncertainty method blank result is greater than its MDC with th~ sample result less than its MDC Validation of the method blank tests whether blank contamination is ofa level which may be indicative.

of systeinatic contamination through sample preparation.

  • If both the method blank and sample activity are greater than their respective MDC or 2s counting uncertainty, the following test shall be used* in determining the effect of possible blank contamination on sample results.

The normalized absolute difference between the method blank. and a sample result, given by the relationship below, is used in testing the null hypothesis that the sarriple and the method blank do not differsignificantlywhen compared to theirresi>ective TPU.. This tc;:st may be used as long as the method blank is reported in terms of activity per unit weight or volume consistent with the sample results.

IS-Bl Page33 of41

.: ; -~=:. "'

',~*..... ~.'. -

. Evaluation of Radiochemical Data Usability s

B TPff

=

Sample result Method blank result.

. Total Propagated Uncertainty April7,1995 If the nonnalized absolute difference is greater than 2.58, no qualification is necessary, as atthe I% level of significance, the conclusion is reached that the method blank and sample differ significantly. If the nonnalized ab8olute differenceis between 1.96 and 2.58, qualify sample results $ MDC "J, 11 the sample and method. blank differ at the 5 % level of significance (sample results < MDC do not require qualificatipn). If the normalized absolute difference is between 0 and 1.96, consider the effects of deficiencies in other quality-indicatorsamples prior to qualifying saJJ1ple results "R, 11 the conclilsiOn is reached that the method blank and sample results differ at the 1 % level of significance. If multiple quality deficiencies are enco\\Jntered, qualify using the guidance provided in Appendix B.

F.

Chemical Yield-Tracers and Carriers Tracers and earners.are used in radiochemical separations methods to provide evaluation of chemical separation. Chemical yield is evaluated through the recovery of chemical species spiked into samples.

Yield is evaluatedradiometricallywith a tracer and gravimetrically with a carrier. Each sample is spiked With either a carrier or tracer, and sample results are adjusted for yields greater or less than 100%.

Generally' a low yield is indicative of losses of tracer I and radionuclide of interest through sample separation, and recoveries greater than expected (> 100%) *are indicative of instrumental problems or contaminafon, as carriers fortified into saxrtples are not expected to be recovered at levels greater than spiked.

1.

Verification Verify that for applicable analyses, one carrier or tracer recovery is reported for each sample. If a carrier or tracer percent recovery is not reported for each sample, contact the laboratoryfor submittal of this data. If the data can not be provided, state this as a non-correctab1e problem in the verification report.

As yield decreases, the MDC may elevate to a point at which the RDL is exceeded, and analytical results are contractuallynoncompliant. If the laboratory has not initiated corrective action, for samples in which t~e MDC exceeds the RDL, the project may choose to contact the laboratory for sample r~ork. If rework is not feasible, indicate the noncompliant data in the verification report.

2.

Validation If yield is not reported for a sample(s), place the reason code "Y04" on the sample result.

Yield is validated based on percent recovery of the spiked nuclide. Low yield may be indjcative of.

increased uncertaintyin the sample result. Criteria for qualification should be based on what magnitude of correction has been applied to the sample result (e.g. 20% recovery implies a sample result correction

  • of 5), although a point of debate exists concerning useability of radionuclide data with yields near 0%.

Page 34 of41

~:'

-~... *~f:

. *:'"':~*

-~ ~.....

  • Evaluation of Radiochemical Data Usability April 7, 1995 Yie~d criteria may also be established from existing sample yield data from previous sampling at the site,.

if this data are available.

Sample results should not be qualified based on yield alone. Sample yield should be evaluated in reference to chemical yield. of quality-indicator samples. If yield is generally low throughout the preparation batch, but recoveries oftargetradionuclidesirfthe LCS are acceptable, data may be accepted withoutqualification;however, if quality control sample yieid is generally low, sample results with low yield may need qualification.

G.

Requ~ Detection Limits (RDL)

1.

Verification For all samples, the,MDC must be less than the contractual RDL. RDLs are identified and communicatedto the laboratory in the labof!\\tory SOW. If the MDC is greater thail the RDL, this may indicateuse ofa small sample size, inadequate count times, or matrix problems. All sample results shall be reviewed to determine RDL compliance. In cases where the MDC is greater than the.RDL, the verifier should consult the laboratory case narrati~e to evaluate the reason for the noncompliance; the proj~ct

.may request a reanalysis.

H.

Nuclide Identificationand Quantification 1..

Test for Detection

a.

Validation The test for detection of a radionuclide includes two. distinct steps.

1.

. *The first step is to evaluate whether a sample result is greater than its MDC.

iL

. The second step is to evaluate whether the sample result is greaf~ than its 2s counting uncertainty.

If the sample result is not greater than its MDC or less than its 2s counting.uncertainty, qualify the sample result "U."

If the sample result is less than its MDC (assuming that the MDC is rep.orted at the 95% confidence level) the data user is accepting the probability of 5% of a false negative result. If the sample result is less than its 2s counting uncertainty, at the 95% confidence level; the radionuclide result is not different from zero.

In placing the "U" qualifier, special attention must be paid to instances where nominal conditions are not followed (i.e. sample aliquot size, count time). Sample results which are less than their MDC, but greater than.their2s countinguncertaintymay have been counted for a period oftime long enough to be considered detected. Sample results which are greater than the MDC, but less than the 2s counting uncertaintymay not have been counted for a p'eriod.of time long enough to be considered detected.

Page 35 of41

... ~ *:*......

Evaluation ofRadii>chemical Data Usability April7, 1995

2.

Negative Results

a.

Validation Negative results that have abs.Olute values greater than their 2s coiinting uncertainty are an indication that the ins~ent background has.shifted: The implfoation of unstable background is a possible negative bias in the sample result. Qualification due to very negative sample resultS must take into account

. evaluation of calibration verification background and quality-indicatorsamples that may indicate a bias.

Qualifiers should be placed usingprofessionaljudgementwhen applied due to the occurrence of.negative results, and the rationale for placement must be fully described in the validation report.

3.

Sample Result Recalculation The accuracy and consistency of sample result calculation by the laboratory can be addressed through two different techniques. The application of each strategy depends on the laboratory's ability to minimize transcription during reporting~ *and how familiar the project is with the performance of the laboratory.

a.

If sample resu.lts are produced primarily through software processing and minimal transcription is performed in the la~oratory, the data system(s) can be evaluated during an audit or surveillance by perforining tw~ different tests on the software: l). supply the data system a consistent set of input designed to provide a consistent set of output, and 2) supply the data system a set of nonconforniingdata to test the error detection routines. An evaluation of the laboratory's software configuration control and security is also necessary. If reporting software has been developed fC!r cu8tomized use in the laboratory, addifunal confidence is gained thiough the laboratory's adherence to the IEEE Standard for Software Verification and Validation Plans 1012-1986 and through maintenance of a software verificatl.on and validation plan.

Through this technique, a high level of confidence can be gained in the laboratory's reporting techniques and will result in a minimal need for manual recalculation of sample results.

b.

If the laboratory has a high rate of manual transcription in generation of sample results, the project may choose to manually recalculate sample results at a detennined frequency. If sample

  • reslllts cannot be reproduc~ through manual calculation, contacting the* laboratory may be necessary to resolve the problem. Qualifiers should be placed using professional judgement.

Activity and TPU equations provided in appendix Care ~efulfor providing the basic structure for

  • calculating radioanalytical results. Modifications to the equations may be needed in method-specific eases. All calculations used must be contractually agreed upon between the project and the laboratory prior to sample analysis, and. iterated in the laboratory SOW.

I.

Iristrument-Specific Sample Considerations I.

Sample AnalYtical Parameters for Gas Proportional Counting

a.

Validation Page 36 of41

~*...

~*:. : *"': ~'.

.~*.;'. :.: ~ *.

.. :~:**..

Evaluation ofRadfochemical Data Usability,

.. April 7, 1995 Equations to approximate a parametric relationship between standard weight and activity are detector-

  • specific. They are developed from the ma8s attenuation cm-Yes by the lal>Qratory and are applied to calculate sample-specific efficiencies. If rec8Iculating sample results, efficiencies used must be sample-specific.

A representative s,ample aliquot must be chosen to ensure the dissolved solid content of the sample falls within the mass range of the appropriate curve. Sample results for which atiquot weight is outside the attenuation curve should be qualified" J" if not reanalyzed with a smaller aliquot.

2.

Spectral Interpretation (alpha and gamma spectrometry)

a.

Gamma Spectrometry Identification of the radionuclides associated with the measured gamma-ray energies and intensities is largely performed by gamma-ray spectral-analysis algorithms.' These algorithms automatically locate peaks in spectra by applying a user-definable set of criteria. Based upon this set of criteria, the peaks in a spectrum are distin~shed from the Compton edges, backscatter peaks, and the numerous featUres created by the background radiation and detection process. The algorithms automatically calculate the peak areas and correct them for the system's detection efficiencyas'.a function of gamma energy.. The identification of the radionuclide!) that produced the peaks is accomplished by comparison to the infonnationin a customiz.3bleradionuclide analysis library. For all peaks identifiedm a spectruin (that correspond to a radionuclide in the analysis library), the algorithms calculate the activities of the

. identifiedradionuclides in the sample. However, the use of peak search, identification, and quantification algorithms does not preclude the need for laboratory data review by an expenenced spectroscopist.

The resulting quality of the gamn;ia spectrometry data will largely depend upon the ability of the spectroscopist to establish a set of analysis criteria to be used by the algorithms with perform the location, identification and quantification of features within spectra. The.quality of this process is greatly influenced by the spectral library. For this reason, it is recommended that the library originate from a reliable source and software be controlled with respect to changes in identification criteria.

Multiple libraries are often constructed based upon expected interferences and the level of activity in the samples. The multiple libraries reduce the number of unidentified peaks due to summation events and interferences within the sample matrix. The goal of reducing the number.of unidentified peaks within a spec1rum, while not compromising the analysis routines by the inclusion of extremely low abundance branching probabilities, helps limit the uncertainty of detection of those radionuclides which emit more tha~ one photon.

The data package should include both identifiedand unideitified peaks. This will support the evaluation of instrumental system control.

b.

Evaluation of Gamma Spectra The assurance of radionuclideidentifieationand quantitationby properly functioning algorithms, through reasonable resolution criteria, and by an evaluation by an experienced radiochelllistwill provide a level

  • Page 37 of 41

f'"'~-*......

  • *.;.. ': !.\\:~.,,\\

\\:*~*...

Evaluation of Radiochemical Data Usability April 7, 1995 of certainty in the quality of the data. Validation of data prod~ced under these conditions should be minima, i.e. a thorough review of all spectra including radionuclide identification and recalclilation of sample results should not be necessary. A recommendation is to broWse sample-specific spectra for changes in energies positions of target radionuclides arid for significantpeak overlap.

Radiochemical data produced under substandard or unknown conditions for identification and quantification warrant a closer inspection. Recommended criteria for evaluating gamma spectra in relation to radionuclide identificationand resolution are as follows:

i.

Where more than on~ isotope of a single Naturally Occurring Radioactive Material (NORM) series are reported, reported results for that series should demonstrate secular equilibrium.

ii.

For soil samples, two peaks that ~e almost always observed are the 511 ke V annihilation peak and the 1460 ke V peak of<<K. The appearance of these peaks at the respective energy, and the respective peak shape should be checked.

m.

If 4°K. is quantitated in the analysis, the reported value should be checked against the activity expected in site's soil (if those data are available).

iv.

For isotopes identifiedand/or quantitated with two or moo: gamma energies (i.e. 1173 and 1332 keV for 60Co) the count rate at each energy can be observed to ensure that the count rate at eac~

respective energy is reasonable to confirm the presence of the isotope..

c.

Alpha Spectra As with gamma spectrometr}1, the alpha spectronietry can require the use of algorithms for the location, identification, and quantification of peaks created by radionuclides with spectra. While many of the analysis algorithms are the same for alpha spectrometry as gamma spectrometry, there are fundamental differences which effect the validation of data. These differences which are inherent to the mode of the

  • decay, sample preparation techniques, and phenomena associated to the detection and quantification.

require validation. As with gamma spectrometry, if algorithms and operational criteria are employed by the laboratory, then they should be reviewed by a competent spectroscopist and/or radiochemist.

Alpha spectrometry has limitations due to the alpha particles ability to transfer energy at an extremely high rate. This phenomenon is responsible for alpha particles being significantly attenuated by the sample matter, the mass of the radionuclile itself, and the molecules of air between the detector and the sample. If the mass of the samples is not reduced or the radionuclides are present in appreciable amounts, then likelihoodofreceivingmeaningfil data is reduced due to attenuation. Therefore, samples for alpha spectrometry frequently require extensive radiochemical preparations, with the sample being.

placed in vacuums and in close proximity to. the detectors. Typically, these separation/preparation tecbniques isolate one or two target nuclides, thus reducing the possible attenuation which occurs at high masses.

This separation process presents the possibility that contaminating elements could break through the chemical separation process. If chemical breakthrough were to occur, these interferences could erroneously be incorporated into the quantification of the radioisotope of interest.

Page38 of41

=;:-;-:
  • Evaluation ofRadiochemieal Data Usability April 7, 1995 Unlike gamma spectrometry, the spectra resulting from properly prepared alpha sources/samples ~

relatively simple, since the peaks are near Qaussian, the background is very low, and detector efficiem::y is almost uniform for the 3.0 MeV to 6.5 MeVenergy range. Frequently, the use of Range of Interest (ROI) are integrated to detemiine gross activltywhich can be at\\justed for background and then corrected for chcinical recovery, decay, aliquot, and detection efficiencyto proouce concentration. A ROI peak report, includingall informationneeded to mamiallyrecalculate results, is useful to evaluate data, as well as the spectra itself.

d.

Evaluation of Alpha Spectra Alpha spectra is not generated through the application of algorithms; consequently the approach to,

validationdiffeis from that of gamma spectrometr}r. The following two tests should be performed on

  • alpha spectral data:
i.

target peaks should be in the energy range of interest

u.

peak tailing should not significantlyoverlap peaks at lower energies.

Professional judgement should be used in qualification of sample data based on spectral interferences or peak misidentification. Communication with the laboratory is essential to obtaining reliable identification/quantitatiomf suspect radionuclides.

VI.

APPENDIXES Appendix A AppendixB

. AppendixC AppendixD AppendixE Qualifiers and Reason Codes

  • Qualification Tables for Multiple Quality Deficiencies Calculations and Equations Common Radionuclides and Analytical Methodologies Recommended Surveillance Considerations for Gamma Algorithms Page39 of41

'-~.*

~~"':

Radiochemical Data Verification and Validation

  • Appendix A

. :~**... ;*

.;:~

APPENDIX A*

  • ... '., *\\

DATA VALIDATION QUALIFIERS AND. REASON CODES Radiological Analytical Data Reason codes must be used. with all qualifiers placed on* analytical data.

U Nuclide considered not detected above the r~orted MDC or 'F counting uncertainty

. J Nuclide identified; the associated numerical value is approximated April 7, 1.995 UJ Nuclide not detected above the reported MDC or r counting uncertainty and a quality deficiency

  • affects the data making the reported data more uncertain
  • R Result is not usable for its intended purpose "Equals" sign, indicates that no qualifier is necessary DATA VALIDATION QUALIFICATION CODES Radiolo~ical Analytical Data Method Blarik BO 1 Concentration of cpntaminant in the method blank at a level => the qualification level.

B02 Method blank was not the same matrix as the analytical samples.

B03 Gross contamination exists.

B04 Blanks were not anal)'?.ed at appropriate frequency.

BOS Sample not significantly different than radiochemical method blank.

B06 Blank data not reported..

B07 Other (describe in comments)

  • Calibration CO 1 Initial calibration sequence was not followed as appropriate.

C02 Calibration was not performed at the appropriate frequency.

CQ3 Calibration data not reported C04 Calibration not performed.

CQS Chemical resolution criteria were not satisfied.

C06 Standard curve was established with fewer than the required number of standards C07 Instrumental system determined to be out of control Page 1 of3

i.

'. ;,*; '* :~;" ;'

Radi0cheniical Data Verification and Validation APJ?endixA cos Other (describe in comments).

Laboratory Duplicate

  • .*. /.!'
r:~... ".

DOI Significant difference between sample and duplicate.

D02 Laboratory duplicate was not analyzed at the appropriate frequency D03 Laboratory duplicate data not reported D04 Other (describe in comments)

Evidentiary Concem8

  • EOl Custody of Sample in Question.

E02 Standard not Traceable.

  • E03 Other (describe in comments)

General GO 1 P.rofessional judgement was used to qualify the data.

G02 Other (describe in comments)

Holding Times HO 1 Holding times were exceeded.

H02 Holding times were grossly exceeded.

H03 Samples were not preserved properly.

H04 Other (describe in comments)

Laboratory Control Sample LO 1 LCS recovery above upper control limit.

L02

. LCS recovery below lower control limit.

L03 LCS was not analyzed at appropriate frequency.

  • L04 LCS not the sam:e matrix as the analytical samples.

LOS LCS data not reported

  • L06 Other (describe in comments)

Matrix Spike and MS/MSD MOl M02 M03 M04 MOS MS recovery above upper control limit.

MS recovery below lower control limit.*

MS not analyzed at the appropriate frequency MS data not reported Other (describe in comments)

  • April 7, 1995 Page2 of3

I'

! l

~ <'!/

.. *, ~**

Radiochemical Data Verification and Validation Appendix A Instrument Performance PO 1 High background levels or a shift in the energy calibration were observed P02 Extraneous peaks were observed.

P03 Loss of resolution was observed.

P04 Peak-tailing or peak splitting that may result in inaccm:ate quantitation were observed

  • POS Instrument performance not analyzed at the appropriate frequency P06 Other (describe in comments)

Quantitation QO 1 Peak Misidentified Q02 Target analyte affected by interfering peak.

Q03 Qualitative criteria were not satisfied.

Q04 Cross contamination occurred.

QOS No raw data were provided to confirm quantitation.

Q06 MDC>RDL Q07 Inappropriate aliquot sizes were used.

Q08 Sample result < MDC.

Q09 Sample result < 2s uncertainty..

Q 10 Negative result.

Q 11 Compounds were not adequately resolved.

Q12 Sample geometry different from calibration geometry.

Q 13 Sample weight greater than greatest weight on mass attenuation curve.

Ql4 Other (describe in comments) '

Radiochemical Yield YOl Y02 Y03.

Y04 YOS Radiochemical tracer yield was above the upper control limit Radiochemical tracer yield was below the lower control limit Radiochemical tracer yield was zero.

Radiochemical yield data was not present.

Other (describe in comments)

    • ~**-*~

April7,1995 Page 3 of3

-~'

',I

.
.-. '~*~;*,.,. '

Radiochemical Data Verification and Validation AppendixB APPENDIXB Mapping for Multiple Quality Deficiencies Apri17, 1995 These tables represent validation qualification decisions made based on radioanalytical data quality considerations only. Data quality needs (e.g. risk ll;Ssessment, remediation technologies) must be considered when using the guidance in these tables. ~For example, quantitative data needs may necessitate that data be rejected due to multiple quality deficiencies, but qualititative data needs may indicate that the same data should only be qualified estimated.

Table table B.l provides a mapping scenario for qualification guidance. For example, if the Laboratory Control Sample shows a high bias, and chemical yield is also high, choose the letters "A" and "E" and reference Table B.2 for guidance on qualification based on those quality deficiencies. The differing separation lines in table B.1 indicate relationships among the quality indicators QC samples and yield.

Double lines indicate a:n "and" function, in that any combination is possible. Thick lines indicate that a single QC sample can only indicate a bias in one direction.

Table B.l Lab Control "fyfatrix Spike Chemical Yield Method Lab Sample Blank Duplicate High Low High Bias Low Bias High Low Bias Bias A

B c

D E

  • F G

H Page 1 of5

I; I\\

  • t,~~:*
    • .:fi~*p ~
_,.... *.~.1:.~:::::. ~:~... ' *. *. 1. ; i" Radiochemical Data Verification and Validation AppendixB
  • ~:.~~.f.:.~~*:t::}(,~*-r":t~***~!

Table B.2 Combination

' Qualification

<MDC Ar MhM~

AD UJ AE none AF UJ AG none AH none BC UJ BD R

BE UJ BF R

BG UJ BH

.UJ CE none CF UJ CG none CH none DE UJ DF R

DG UJ DH UJ EG none EH none FG

. UJ FH UJ GH UJ

': ~*-*

.-~~:

  • ~* -*.* -* *~:.
  • ~~-:1.~*-~

April 7, 1995

$MDC 11 J

R J

J J

J J

J J

J J

R J

R R

J J

R R

R J

J J

J Page 2 of5

Radiochemical Data Verification and Validation APJ?endixB Table B.3 Qualification Guidance for 3 Simultaneous Quality Deficiencies Combination Qualification

<MDC

$MDC At'~

1>

ACF UJ R

ACG none R

ACH none R

ADE UJ R

ADF R

R ADG UJ R

ADH UJ R

AEG none R

AEH none R

AFG UJ R

AFH none R

BCE UJ R

BCF UJ R

BCG UJ R

B(::H UJ R

BDE R

R BDF R

R BDG R

R BDH R

R BEG UJ R

BEH UJ R

BFG R

R BFH R

R CEG none R

CEH none R

CFG UJ R

April 7, 1995 Page 3 ofS

Radiochemical Data Verification and Validation AppendixB CFH DEG DEH DFG DFH EGH FGH 1

  • UJ UJ UJ R

R none UJ

~.

. *'... ~

.:.:~Tr:~:-:t?*

April 7, 1995 R

R R

R R

R R

Page4of5

'*
.:.~:~:-: < *' :, *.' : **

Radiochemical Data Verification and Validation ApPendixB 1.**;:

TableB.4 Qualification Guidance for 4 Simultaneous Quality Deficiencies Combination Qualification

<MDC

$MDC ACEG none R

ACFG UJ R

ACEH none R

ACFH UJ R

ADEG UJ R

ADFG R

R ADEH UJ R

ADFH R

R BCEG UJ R

BCFG R

R BCEH UJ R

"BCFH R

R BDEG R

R BDFG R

R BDEH R

R BDFH R

R CEGH none R

CFGH UJ R.

DEGH UJ R

DFGH R

R April 7, 1995 Pages ofS

-;:;::::::;::;:::======-------=====------~---:-=====-. ---- *--**--* *--*-.

'"<:~~,

Radiochemical Data Verification and Validation A endix C April7,1995 AppendixC Pages 1-6 Provide Equations For Method Blapk Corrected Sample Concentrations Pages 7-12 *Provide Equations For Sample Concentrations Without Blank Subtraction The equations in this appendix are meant to be used as the basis for developing inetbod-specific applications.

Sample Activity Concentration -

<Method Blank Corrected Sample Concentrations)

Where:

ACT

  • s. 2. 22
  • EFF *ALI
  • R
  • ABN5
  • e-M
  • CF ACT s = Sample Activity Concentration (pCi/g or pCi/L)

NCRs = Net Sample Count Rate in cpm 2.22 = Factor for Converting dpm to pCi EFF = Detector Efficiency (Fraction)

ALI= Sample Aliquot Volume or Mass (g or L)

ABN5 = Abundance Fraction of the Emissions Used for Analyte Identification/Quantification R = Sample Tracer/Carrier (Chemical) Recovery 8 = Analyte Decay Cmstant - In 2/(half-life) [Same units as the half-life used to compute 8]

t = Time from Sample Collection to Radionuclide Sepaation or Mid-point of Count Time (Same units as half-life)

CF = Other Correction Factors as Appropriate (i.e~ Ingrowth factor, Self-absorption Factor, etc.)

  • Page I of12

Radiochemical Data Verification and Validation A

endixC April 7, 1995 Net Sample Count Rate (NCR) and FNca- (Method Blank Corrected Sample Concentrations)

Where:

. ( CGS css) 1

  • t CGB CBB)

( EFFSD)

( Rs) l NCR*----

TGS TSB TGB TBB EFF BD RB Cos = Sample Counts T0s = Sample CountTime (minutes)

C58 = Background Counts T58 = Background CountTime (minutes)

C08 = Gross Method Blank Counts T08 = Gross Method Blank CountTime (minutes)

C88 =

  • Method Blank Background Counts T88 = Me*od Blank Background CountTime (minutes)

EFF50 = Efficiency of the Sample Detector EFF80 = Efficiency of the Method Blank Detector Rs = Sample Tracer/Carrier Recovery Fraction R8 = Method Blank Tracer/Carrier Recovery Fraction

~ * [ CGS

+ CsB] + {[ CGB + CBB] * ( EFFSD) 2 * ( Rs) 2}

N_CR T2 T2 T2 T2 EFF R

{ i(

) (

GS l S[B

) l GB~[

BB l 2 [ BD l 2. (B. l 2 ( l 2] }

  • c c

EFF R

2 (J

cr (J

(J GB BB SD s

EFFSIJ EFFBD Rs Rs

+

TGB -.. TBB EFF BD

  • . RB EFF50

+

EFF80

+

R5

+

RB Page 2of12

I I !

Radiochemical Data Verification and Validation Appendix C Calculation of Recovery - Radiometric- (Method Blank Corrected Sample Concentrations)

. Where:

R *

(

CGX _ CTB)

TGX TTB_

R = Tracer Recovery Car = Gross Count of Tracer Tar= Tracer CountTime (minutes)

Cm= Background Count of Tracer [Region of Interest (ROI)]

T TB = Background CountTime (minutes)

EFF = Detector Efficiency Fraction AMTT = Amount of Tracer Activity Added (dpm)

NCRT = Net Count Rate of Tracer (cpm)

April 7, 1995 ABNT = Abundance Fraction of the Tracer Emissions used for Quantification of the Tracer Note: It is assumed that the tracer half-life is long enough to be an insignificant uncertainty contributor. If the tracer has a relatively short half-life, then it must be considered and these equations modified. Likewise, uncertainty in the time is also considered to be an insignificant contributor.

Page 3of12

    • ***,'If,-

.::'... >: :*:: ~t.~- * *

. ** - *,\\ \\

' * : :>.':.. *..,. ~ *. ~. ~.

Radiochemical Data Verification and Validation Appendix C April 7, 1995

. Calculation ofRER - Radiometric- (Method Blank Comcted Sample Concentrations)

Where:

(RERY - (* (JR) 2

  • ( aNCR.,) 2 + (

(JEFF) 2

+ [

0 AMT.,l 2 R

NCRT EFF AMTT.

( c':. c~sJ TGlt TTB NCR 2 T

(

aNcR.,) 2 NCRT

( a AMT l 2

[

<JSTS l 2

(

(JHASS l 2

(

aVOL l 2 + l:

AMT:. *. STS:

MASS:

VOL:

(

O"DIL.,) 2 + E DILT RE,*~

RER = Relative Error of the Tracer Recovery FR = Standard Deviation of the Tracer Recovery FNcR = Standard Deviation of the Tracer's Net Count Rate F EF~ = Standard Deviation of the Detector Efficiency (JALI

(

]

2 A~I:

F AMT = Standard Deviation of the Amount of Tracer Activity Added F STS T. =:: Standard Deviation of the Amount of Trace Activity Taken for Stock Tracer Solution T

(provided with certificates received with standards)

FMAss = Standard Deviation of the Mass of Standard Solution Used to Prepare Stock Tracer T

Solution F0IL = Standard Deviation oftheVolume(s) of the Dilution(s) Made to Prepare the Working T

Tracer Solution F voL = Standard Deviation of the Volume of the Stock Tracer Solution F ALIT = Standard Deviation(s) of the Aliquot(s) of Tracer Solution(s) Diluted to Prepare T

Working Tracer Solution Car = Gross Count of Tracer Tk = Square ofTracer,.CountTime (minutes)

Cra = Background Count of Tracer [Area or Region of Interest (ROI)]

Tia= Square of Background CountTime (minutes)

R = Tracer Recovery NCRT = Net Count Rate of the Tracer

  • EFF = Detector Efficiency AMTT = Amount of Tracer Activity Added (dpm)

STST = Amount of Tracer Activity (dpm) in Stock Tracer Solution MASST = Mass (grams) of Standard Solution Used to Prepare Stock Tracer Solution VOLr = Volume of Tracer Solution Added DILr = Volume(s) of Dilution(s) Made to Prepare the Working Tracer Solution ALIT = Aliquot(s) of Tracer Solution(s) Taken to Prepare Serial Tracer Solution Dilution(s)

Note:

Certificates, such as those from NIST, may give two or even three sigma uncertainty. Only one sigma should be used f<<fi" STS T

Page4 ofl2

I !

Radiochemical Data Verification and Validation A

endbcC April 7, 1995 Calculation of Recovery and RER - Gravimetric -

(Method Blank Corrected Sample Concentrations)

Where:

WT R

  • ___

_...;;c __

a

) 2

(

PI!T. i'/'l' *.

PPT. WT.

+

'RE,*W R = Carrier Recovery RER = Relative Error in Recovery WT0 = Weight*ofCarrier Present in Finai1Precipitate CONC0* = Concentration of Carrier Solution VOL0* = Volume of Carrier Solution Added F PPT. WT.= Standard Deviation in Weight of Precipitate

. F coNc = Standard Deviation in Carrier Concentration F voL = Standard Deviation in Carrier Volume PPT. WT.= Weight of Final Carrier Precipitate*

Counting Uncertainty (CU) at the 95% Confidence Level Where:

1. 96 * (aNCRs) cu20 * ----------'--"'-------
2. 22
  • EFF *ALI* R
  • ABN5
  • e*At *CF FNcR = Standard Deviation of the Net Sample Count Rate EF~ = Detector Efficiency ALI= Sample Aliquot Volume or Mass R = Sample Tracer/Carrier Recovery ABNs = Abundance Fraction of the Emissions Used for Analyte Identification/Quantification 8 = Analyte Decay Cmstant - ln 2/(half-life) [Same units as the half-life used to compute 8]

)

t = Time from Sample Collection To Radionuclide Separatin or Mid-point of CountTime (Same units as half-life)

CF = Other Correction Factors as Appropriate (i.e, Ingrowth factor, Self-absorption Factor, etc.)

Pages of12

Radiochemical Data Verification and Validation AppendixC April 7, 1995 Sample Activity Concentration Total Propagated Uncertainty (TPU)-(Method Blank Corrected Sample Concentrations)

Where:

aNcR 2

+ NCR 2

  • REiFF + RE~Lr + RE!
  • I: RE;F TPUla
  • crACT
  • s _____________

2.22

  • EFF *ALI *-R
  • ABN5

.;. e-At *CF EFF = Detector Efficiency ALI= Sample Aliquot Volume or Mass

\\

R = Sample Tracer/Carrier Recovery ABN5 = Abundance Fraction of the Emissions Used for Identification/Quantification F~cR = Variance of the Net Sample Count Rate NC!t = Net Sample Count Rate RE~FF = Square of the Relative Error of the Efficiency Tenn

(

RE~u = Square of the Relative Error of the Aliquot REi = Square of the Relative Error of the Sample Recovery RE~F = Square of the Relative Error of Other Correction Factors 8 = Analyte Decay Cmstant - In 2/(half-life) [Same units as the half-life used to compute 8]

t = Time from Sample Collection to Radionuclide Separaion or Mid-Point of Count Time (Same units as half-life)

. CF = Other Correction Factors as Appropriate (i.e. Ingrowth factor, Self-absorption Factor, 1

etc,)

Page 6of12

Radiochemical Data Verification and Validation

  • AppendixC April 7, 1995 Sample Activity Concentration -

Sample Concentrations Without Blank Subtraction Where:

NBCR5 ACTB - -----------'-----......,----

2.22

  • EFF *ALI* R
  • ABN5
  • e-l\\.t *CF ACT 8== Sample Activity Concentration Without Method Blank Subtraction NBCRs== Net Sample Background-Corrected Count Rate 2.22== Factor for Converting dpm to pCi EFF== Detector Efficiency ALI== Sample Aliquot Volume or Mass ABNs== Abundance Fraction of the Emissions Used for Identification/Quantification R== Sample Tracer/Carrier Recovery
  • 8== Analyte Decay Cmstant - ln 2/(half-life) [Same units as the half-life used to compute 8]

t== Time from Sample Collection to Radionuclide Sepaation or Mid-point of Count Time (Same units as half-life)

CF== Other Correction Factors as Appropriate (i.e, Ingrowth factor, Self-absorption Factor, etc.)

Page 7of12

1*.

-~"'"'.

I I

.. ~ *.*

  • Radiochemical Data Verification and Validation AppendixC April 7, 1995 Net Sample Count Rate (NBCR8) and FNBCR Sample Concentrations Without Blank Subtraction s

Where:

FNBc~

(

CGS Css)

NBCR*---

s TGS TSB NBCRs = Net Background Corrected Count Rate Cas = Sample Counts Tas = Sample Count Time (minutes)

. Csa = Background Counts T58 = Background Count Time (minutes)

[ l 1

CGS CSB 2 cr NBCR5 T

2 T

2 GS SB Page 8of12

      • ~.

Radiochemical Data Verification and Validation AppendixC Calculation of Minimum Detectable Concentration (MDC) - general formula MDC

  • 4
  • 65 *.f£TT + 2
  • 71 K

K*T Where:

b= background count rate April 7, 1995 T =Sample Count Time (minutes)

K =instrument-specific and sample-specific co~ection factors (e.g. ALI* t'

  • R
  • EFFs
  • ABNs)

In using the above equation, the background and sample count times are eiter equivalent, or the background count time is greater than sample count time Calculation of Recovery and RE a - Radio~etric -

Sample Concentrations Without Blank Subtraction Where:

R = Tracer Recovery Ccrr = Gross Count of Tracer T crr = Tracer CountTime (minutes)

Crn = Background Count of Tracer Region of Interest (ROI)

  • Tre = Background CountTime (minutes)

EFF = Detector Efficiency AMTr = Amount of Tracer Activity Added (dpm)

NCRr = Net Count Rate of Tracer (cpm)

ABNr = Abundance Fraction of the Tracer Emissions used for Quantification of the Tracer Page 9of12

. __ ! __ --*-*~--...:.. __

Radiochemical Data Verification and Validation AppendixC April7, 1995 REa - Radiometric -

Sample Concentrations ~itllJW:Ul!J.wlk..~llllwmtn (RE.)' * ( ';:) 2

~: 2 * ( :;; )'. ::J Where:

RER = Relative Error of the Tracer Recovery FR*= Standard Deviation oftlie Tracer Recovery FNcR = Standard Deviation of the Tracer's Net Count Rate FEF! = Standard Deviation of the Detector Efficiency F AMT = Standard Deviation of the Amount of Tracer Activity Added F sTS T = Standard Deviation of the Amount of Trace Activity Taken for Stock Tracer Solution

  • T (provided with certificates received with standards)

FMASs = Standard Deviation of the Mass of Standard Sollltion Used to Prepare Stock Tracer T

  • Solution F0IL = Standard Deviation oftheVolume(s) of the Dilution(s) Made to Prepare the Working T

Tracer Solution F voL = Standard Deviation of the Volume of the *Stock Tracer Solution F ALIT = Standard Deviation(s) of the Aliquot(s) ofTracer Solution(s)Diluted to Prepare Tracer T

Working Solution Cm = Gross Count of Tracer.

Tk = Square of Tracer Count Time..

Cm= Background Count of Tracer ROI Tia = Square of Background Count Time R = Tracer Recovery NCRT = Net Count Rate of the Tracer EFF = Detector Efficiency AMTT= Amount of Tracer Activity Added (dpm)

STST = Amount of Tracer Activity (dpm) in Stock Tracer Solution MASST = Mass (grams) of Standard Solution Used to Prepare Stock Tracer Solution VOLr = Volume of Tracer Solution Added DILr = Volume(s) ofDilution(s) Made to Prepare.the Working Tracer Solution ALIT = Aliquot(s) of Tracer Solution(s) Taken to Prepare Serial Tracer Solution Dilution(s)

Note: Certificates, such as those from NIST, may give two or even three sigma uncertainty. Only one sigma should be used for F sTS T

Page 10of12

Radiochemical Data Verification and Validation AppendixC April 7, 1995 Calculation ofRecoveryand RER - Gravimetric.-Sample Concentrations WithoutBlank Subtraction Where:

WTC R:o -------

CONCcs

  • VOLcs (RER~ -( aRR) 2
  • ( aPPT. loll'.

)

2

.. (

PPT. WT.

RE,*ffi.

R== Carrier Recovery RER== Relative Error in Recovery WT c = Weight of Carrier Present in Final Precipitate CONCcs = Concentration of Carrier Solution VOLcs = Volume of Carrier Solution Added F PPT. wr. = Standard Deviation in Weight of Precipitate

. FcoNc = Standard Deviation in Carrier Concentration FvoL = Standard Deviation in Carrier Volume PPT. WT.= Weight of Final Carrier Precipitate Counting Uncertainty (CU) at the 95% Confidence Level Where:

cu

  • 82" 2.22. EFF. ALI. R. ABNS. e-At. CF FNBCR = Standard Deviation of the Net Background Corrected Count Rate.

EF~ = Detector Efficiency ALI = Sample Aliquot Volume or Mass R = Sample Tracer/Carrier Recovery ABNs = Abundance Fraction of the Emissions Used for Identification/Quantification 8 = Analyte Decay Cmstant - ln 2/(half-life) [Same units as the half-life used to compute 8]

t = Time from Sample Collection To Radionuclide Separation or Mid-point of Count Time (Same units as half-life)

CF= Other Correction Factors as Appropriate (i.e, Ingrowth factor, Self-absorption Factor, etc.)

Page 11of12

Radiochemical Data Verification and Validation AppendixC April 7, 1995 Sample Activity Concentration Total Propagated Uncertainty (TPU)-

Sample Concentrations Without Blank Subtraction Where:

NOTE:

2 R 2 2

E2 2

~

2 crNBCRs

  • NBC s
  • REEFF + R ALI + RER + ~ RECF 2.-22
  • EFF *ALI* R
  • ABN8
  • e*A.t *CF EFF = Detector Efficiency ALI= Sample Aliquot Volume or Mass R = Sample Tracer/Carrier Recovery ABNs = Abundance Fraction of the Emissions Used for Analyte Identification/Quantification F~ecR = Variance of the Net Background Corrected Count Rate s

NBCRs = Net Background Corrected Count Rate RE~FF = Square of the Relative Error of the Efficiency Term RE1_i = Square of the Relative Error of the Aliquot REi = Square of the Relative Error of the Sample Recovery RE~F = Square of the Relative Error of Other Correction Factors 8 = Analyte Decay Cmstant - ln 2/(half-life) [Same units as the half-life used to compute 8]

t = Time from Sample Collection to Radionuclide Separakm or Mid-Point of Count Time (Same units as half-life)

CF = Other Correction Factors as Appropriate (i.e, Ingrowth factor, Self-absorption Factor, etc.)

For methods ming a tracer or carrier~ the inclusi9n of efficiency and recovery terms in the equation above may result in overestimation of the TPU.

Page 12of12

Radiochemical Data Verification and Validation AppendixD

  • **,1 :,...,,;;:~* * * ::*.f"'°~,)~i.~*

APPENDIXD Listing of Common Nuclides/Methodologies July 10, 1998 Source of radionuclide identities: DOE National Sample Tracking System, FY 1994 Analysis Matrix Instrumentation Source Type

...c:

cc i:::

rn 0

i:::

'S i:::

L...

.s Q) 0

~

  • s 0

0 Q)

Q) en

~

r:n

~

E E

0 i:::

0 0

0.

"<:)

i:::

0 en c..>

  • c::;

c..>

0....

Q)

Q)

Cf) rn

-0 p.,

p.,

Cf) r:n

=

c:;T'

= -

.:3 Gross"/$

x x

x 2&4 JH x

x x

x 2

Pu isotopic x

x x

x x

x 2

U Isotopic x

x x

x x

x 2

241Am x

x x

x x

2 90Sr x

x x

x x

2&4 22J,224,226Ra x

x 3

226Ra x

x x

2 22sRa x

x 2

~c x

x 1

Th Isotopic x

x x

x x

x x

2 129J:

x x

4 Cm Isotopic x x

3 s9Sr x

x x

x x

2&4 Page 1 of2

Radiochemical Data Verification and Validation A endixD 14c x

mNp x

2:Z2Rn 210pb x

x 210p0 79Se x

Gamma x

Emitters Sources:

x

.' x x

x x

x x

1.

"Technetium-99 Analysis Using Extraction Chromatography" July 10, 1998 2

x 3

2 2

4 x

3

2.

"Eastern Environmental Radiation Facility Radiochemistry Procedures Manual"

3.

United States Environmental Protection Agency.Prescribed Procedures for Measurement of Radioactivity in Drinking Water. EPA-600/4-80-032.

4.

DOE Methods for Evaluating Environmental and Waste Management Samples DOEIEM-0089T, Rev. 2, October, 1994.

Page2 of2

Radiochemical Data Verification and Validation ApPendixE AppendixE Recommended Surveillance Considerations for Gamma Algorithms

  • January9, 1995 Radionuclide identification algorithms used by the laboratory should be evaluated during laboratory audits. Consideration should be give~ to the following:
i.

the width of the energy-search windows for radionuclide identification to ensure a minimum of peak misidentifications, ii.

the criteria by which the laboratory evaluates sample-specific peak resolution, iii.

documentation of the source of the isotope identification library used by the source

[Most software packages come with a master library which users can pare down to one more useful in day to day analyses.],

iv Training records of those persons(s) using the peak identification respective

software,
v.

SOP documentation of the parameters used by the software in the analysis of gamma spectra,

  • vi.

existence of SOP documentation for the internal review of the reported data, vii.

If the software package is capable of using pre-programmed macros to automate the spectral analysis, evaluate the eXistence of SOP documentation of these

macros, viii.

If the software package uses various flags to denote peak quality or analysis

  • results, the analyst using the software should be able to explain the meaning of the flags and their impact on data quality.

Page 1 ofl