UNITED STATES
NUCLEAR REGULATORY COMMISSION
OFFICE OF NUCLEAR MATERIAL SAFETY AND SAFEGUARDS
WASHINGTON, D.C. 20555-0001
June 12, 2006
NRC INFORMATION NOTICE 2006-11:
APPLICABILITY OF PATIENT INTERVENTION IN
DETERMINING MEDICAL EVENTS FOR GAMMA
STEREOTACTIC RADIOSURGERY AND OTHER
THERAPY PROCEDURES
ADDRESSEES
All medical licensees.
PURPOSE
The U.S. Nuclear Regulatory Commission (NRC) is issuing this information notice (IN) to inform
addressees of an NRC concern that has arisen, regarding licensees assessing the contribution
of patient activities to errors, in medical administrations, when determining whether the events
constitute reportable medical events under 10 CFR 35.3045, Report and notification of a
medical event. It is expected that recipients will review this information for general applicability
to all their licensed medical use operations and consider actions, as appropriate, to avoid
similar problems. The information contained in this IN does not constitute new NRC
requirements; therefore, no specific action or written response is required.
DESCRIPTION OF CIRCUMSTANCES
Event 1 During a routine inspection of an NRC medical use licensee, NRC inspectors discovered
records of a medical administration involving the licensees gamma stereotactic radiosurgery
(Gamma Knife) unit that should have been reported as a medical event under §35.3045.
Specifically, following a Gamma Knife treatment, the licensee noted that the z-axis (up and
down) coordinate of the head frame had been displaced 7 centimeters (cm) (2.8 inches) during
the course of treatment. The x- and y-coordinates, however, had remained unchanged. The
licensee believed the misalignment occurred when the patient moved vigorously more than
half way through the procedure. The misalignment of the z-axis coordinate of the head frame
resulted in an estimated absorbed dose of 35 Gray (Gy) (3500 rads) [or 35 Sievert (Sv) (3500
rem) dose equivalent] to an unintended site--a dose that was greater than 50 percent of the
dose expected to that unintended site from the administration defined in the written directive.
The patient had complained of discomfort and back pain and had asked attending staff if he
could move his legs to a more comfortable position. Permission was granted to move a little, but the licensee noted that the patient moved vigorously. This occurred approximately 30 minutes into a 51-minute treatment. Although the patient moved vigorously, treatment
continued until completion of the procedure. The licensee stated that there had been no
observable reorientation of the patient in the Gamma Knife after the movement and that no
permanent functional damage to the patient had occurred from the dose delivered to the wrong
site, after the movement of the z-axis coordinate. The latter conclusion was based on the
licensees analysis of subsequent magnetic resonance images.
The licensee inspected the head frame and stated that there was no observable damage to the
z-bar, which controls the positioning of the z-axis coordinate. As a corrective action, the
licensee replaced the z-bars for that particular head frame. Although the z-bars were removed
from service, the licensee did not return them to the manufacturer for component failure
evaluation. Other corrective actions taken or planned included: (1) upgrading to the Model C
head frame with Automatic Positioning System; (2) instructing patients not to move; and (3)
increasing monitoring of patients during treatments that last 30 minutes or longer.
The licensee believed that the 7-cm (2.8-inch) change in the z-axis coordinate was caused by
the patients vigorous movement. Accordingly, the licensee believed that the patients
movement qualified as patient intervention. Since the licensee also determined that the dose
delivered to the wrong site had not resulted in permanent functional damage, the licensee
concluded that the criteria for reporting an event, in §35.3045(b), had not been met, so the
event was not reported to NRC. NRC, however, concluded that this occurrence should have
been reported under §35.3045.
Event 2 During a routine inspection of another NRC medical use licensee, NRC inspectors discovered
records of a medical administration, involving the licensees Gamma Knife unit, that also should
have been reported as a medical event, under §35.3045. Specifically, after an 11-exposure
Gamma Knife treatment, the licensee noted that the left anterior pin attaching the head frame to
the patient's head had been displaced laterally, resulting in a shifting of the isocenter an
estimated 6 millimeters (mm) (0.24 inches) during the course of the treatment. The licensee
initially believed that the movement of the head frame occurred when the patient coughed at the
start of the 11th exposure. The movement of the head frame resulted in a licensee initially
estimated additional absorbed dose of approximately 5 Gy (500 rads), or an additional dose
equivalent of approximately 5 Sv (500 rem), to an unintended site. This movement resulted in a
licensee initially estimated total absorbed dose of approximately 7.6 Gy (760 rads), or a dose
equivalent of approximately 7.6 Sv (760 rem), to an unintended site, which was greater than 50
percent above the approximately 2.6 Gray (260 rads) absorbed dose [or 2.6 Sv (260 rem) dose
equivalent] expected to that unintended site from the administration defined in the written
directive.
In a standard Gamma Knife procedure, the head frame is secured to the patients head using
four sharp pins screwed in place and tightened sufficiently to embed the point of the pin into the
table of the patients skull. During patient preparation, the neurosurgical team performed
physical tests and measurements to determine if there would be collisions between either the
patient's head or frame and the collimator helmet. In this specific case, there would have been
a collision with the right anterior pin, and the licensee made a decision to remove this pin and
proceed with the Gamma Knife procedure, using only three pins (three-pin technique). The
licensee indicated that after the other three pins were tightened, the right anterior pin
was removed. The procedure then continued without event until the final 11th exposure, when the patient
coughed, initially reported as occurring at the beginning of the exposure. It was not until after
the treatment was completed and the patient was removed from the unit that the staff noted the
patient was bleeding because the left anterior pin had moved from its original position.
The licensee determined the shift of the head frame from the movement of the pin from its
original position on the skull. From this observation, the licensee estimated the isocenter
shifted laterally by 6 mm (0.24 inches) and reviewed earlier magnetic resonance images to
approximate the location of the new isocenter and the wrong treatment site. Based on this
analysis, the licensee believed the location of the isocenter for the misdirected final exposure
was inside the auditory canal.
As corrective action, the licensee initially prohibited use of the three-pin technique for Gamma
Knife treatments. No other corrective actions were initially taken or planned. In later discussion
with NRC staff, the licensee indicated it was reevaluating its prohibition of the three-pin
technique.
The licensee believed that the movement of the left anterior pin was caused by the patient's
cough, and that the patients coughing movement constituted patient intervention. The
licensees staff concluded that there was ...no harm to the patient, since the patient received
almost the complete dose to the treatment site, and the wrong treatment site for the one
exposure was in the auditory canal, which did not result in permanent functional damage. The
licensee therefore believed that the criteria for reporting an event, in §35.3045(b), had not been
met, so the event was not reported to NRC. NRC, however, concluded that this occurrence
should have been reported, under §35.3045.
DISCUSSION
In each of the two events discussed, the licensee asserted that the patients movement
constituted patient intervention. Each licensee also decided that each event was not
reportable, because §35.3045(b) only requires reporting of an event resulting from intervention
of a patient ...in which the administration of byproduct material or radiation from byproduct
material results or will result in unintended permanent functional damage to an organ or
physiological system, as determined by a physician, and this condition did not occur. However, NRC concluded that neither licensee provided sufficient evidence to exclude equipment setup
as the cause of its medical event, rather than patient movement. Therefore, NRC concluded
that these occurrences should have been reported to NRC as medical events, under
§35.3045(a)(3). This, in part, requires that the licensee report any event (except for an event
that results from patient intervention) in which the administration of byproduct material, or
radiation from byproduct material, results in a dose to the skin, or to an organ or tissue other
than the treatment site, that exceeds by 0.5 Sv (50 rem) or more, and 50 percent or more, of
the dose to the skin, or organ or tissue other than the treatment site, that was expected from
the administration defined in the written directive. See Enclosure 1 for discussion of these two
Gamma Knife events.
For each of these events, the licensees corrective actions are silent about stopping treatments
when a patient moves, in order to ensure that the movement did not result in patient position
changes that could result in a medical event. In fact, the licensee for Event 1, through one of its corrective measures, implies that it is unnecessary for it to increase patient monitoring during
treatments that last less than 30 minutes.
As a measure for prevention of patient movement during Gamma Knife treatment, a licensee
could respond to a patients expression of discomfort by: (1) halting the treatment; (2) assisting
the patient in moving to become comfortable; (3) checking the head frame for correct
positioning; and (4) then resuming the treatment. We also believe that, as a potentially
corrective measure, regardless of the treatment time, a licensee authorized for Gamma Knife
treatments, or other high dose-rate treatments, should monitor the patient and stop the
treatment when a patient moves, in order to ensure that the movement did not result in a
patient position change that could result in a medical event.
The licensees believed that both of these Gamma Knife events resulted from patient
intervention. However, NRC views these as resulting primarily from patient equipment setup.
Similarly, incorrect decisions as to causes of events when patient actions are involved have
also been made by medical use licensees employing other treatment modalities, such as
temporary implant brachytherapy. Medical use licensees employing any treatment modality in
which patient actions may potentially interfere with licensees properly implementing physicians
intentions, as expressed in prescribed doses or dosages, should be aware that patient
movement or other involvement in an occurrence or event alone is not sufficient to rule out the
need to report the occurrence as a medical event. NRCs position is that a medical event has
occurred, even when the occurrence had patient movement or other involvement, if the licensee
has not followed appropriate preventative and corrective procedures for usage, and if the
criteria specified in §35.3045(a) or (b) are met.
Medical use licensees should review this IN and consider whether their procedures for use are
in accordance with the following recommended actions:
Monitor patient and/or source placement at reasonable frequencies;
Correctly identify patient and/or source displacement during monitoring;
Take prompt and appropriate actions should patient and/or source displacement occur;
Have trained personnel present or available to prevent or mitigate patient actions during
usage procedures that may impact treatment;
Promptly report all medical events to NRC; and
Promptly report all equipment malfunctions or problems to the vendors and, if required, to NRC (under 10 CFR 21.21 or §30.50(b)(2)) and the device licensing authorities (NRC
or Agreement States).
CONTACT
This IN requires no specific action nor written response. If you have any questions about the
information in this notice, please contact the technical contact below, or the appropriate
regional office.
/RA/
Charles L. Miller, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical Contact:
Ronald E. Zelac, Ph.D., NMSS
(301) 415-7635 E-mail: rez@nrc.gov
Enclosures:
1. List of recently issued NMSS Generic Communications
2. Discussion of Gamma Stereotactic Surgery Events 1 and 2
ML061360026 OFFICE
MSIB
MSIB
Tech. Ed.
IMNS
MSIB
IMNS
NAME
RZelac
SWastler
EKraus
AMcIntosh
CMiller
DATE
5 /18 /06
5 / 25 /06
5 / 22 /06
5/ 24 /06
5/30/06
6/9/06 Enclosure 1 Recently Issued NMSS Generic Communications
Date
GC No.
Subject
Addressees
04/23/06 RIS-06-10
Use of Concentration Control for
Criticality Safety
All licensees authorized to possess a
critical mass of special nuclear material.
01/26/06 RIS-02-15, Rev. 1
NRC Approval of Commercial Data
Encryption Products For the
Electronic Transmission Of
Safeguards Information
All authorized recipients and holders of
sensitive unclassified safeguards
information (SGI).
01/24/06 RIS-06-01 Expiration Date for NRC-Approved
Spent Fuel Transportation Routes
The U.S. Nuclear Regulatory Commission
(NRC) licensees who transport, or deliver
to a carrier for transport, irradiated
reactor fuel (spent nuclear fuel (SNF)).
01/13/06 RIS-05-27, Rev. 1
NRC Regulatory Issue Summary
2005-27, Rev. 1, NRC Timeliness
Goals, Prioritization of Incoming
License Applications and
Voluntary Submittal of Schedule
for Future Actions for NRC Review
All 10 CFR Parts 71 and 72 licensees
and certificate holders.
03/31/06 IN-06-07 Inappropriate Use of a Single- parameter Limit as a Nuclear
Criticality Safety Limit
All licensees authorized to possess a
critical mass of special nuclear material.
03/21/06 IN-02-23, Supl. 1
Unauthorized Administration of
Byproduct Material for Medical Use
All medical licensees.
01/19/06 IN-06-02
Use of Galvanized Supports and
Cable Trays with Meggitt Si 2400
Stainless- Steel-jacketed Electrical
Cables
All holders of operating licenses for
nuclear reactors except those who have
permanently ceased operations and have
certified that fuel has been permanently
removed from the reactor vessel; and fuel
cycle licensees and certificate holders.
Note: NRC generic communications may be found on the NRC public website at http://www.nrc.gov, under Electronic
Reading Room/Document Collections. Enclosure 2 Discussion of Gamma Stereotactic Surgery Events 1 and 2 In Event 1, the licensee believed that the change in z-bar position was caused by two factors:
(1) the large stature of the patient (mass); and (2) the vigorous movement of the patients
legs and torso (force-of-movement). However, historical evidence from the device
manufacturer has shown that z-bar movement has only been observed when the associated
screws were not properly tightened, or there was a lubricant on the z-bar. Although the patient
movement in this case may have contributed to the medical event, the purpose of the head
frame design and function is to keep the head from moving in spite of patient movement. The
head frame essentially immobilizes the patients head, limiting the patients degree of upper
body movement. For movement of a properly secured z-bar to occur, the patient would have to
exert an extreme force on the head frame. Since the pins that secure the head frame to the
patients head are screwed directly into the patients skull, the extreme amount of force that
must be exerted to move the z-bar would be expected to also cause one or more of the pins to
move. There was no indication that these pins moved during the vigorous movement.
Also, 7 cm (2.8 inches) of z-bar slippage at one time would be expected to result in an
observable reorientation of the patient, which should have prompted the licensee to stop the
treatment and recheck the coordinates before continuing treatment. As stated earlier, the
treatment was not interrupted, and the licensee claimed that there was no observable
reorientation of the patient after the vigorous movement. The licensees not noting observable
patient reorientation, after the vigorousmovement, is inconsistent with the licensees belief that
the patients vigorous movement caused the 7-cm (2.8-inch) change in the z-axis coordinate.
This administration of radiation resulted in an unintended dose of 35 Sv (3500 rem), in an area
of the brain that was 7 cm (2.8 inches) away from the intended treatment site, and the patient
was expected to receive negligible dose in this area, which is distant from the treatment site.
Therefore, NRC staff has concluded that the licensee should have reported this event, under
At NRCs request, the licensee subsequently returned the z-bars to the manufacturer for
testing. The manufacturer determined that the z-bars failed (slipped) at 50 percent of their
designed locking force specification. After dismantling and thoroughly cleaning the z-bars, lubricating the locking screw and nut, and then reassembling the component, the z-bars
functioned at 100 percent of their design specification. The manufacturer concluded that the
slippage was caused by the reduced locking force of the z-bars, which was corrected by the
manufacturers dismantling, cleaning, and lubricating process. The licensee indicated that it
cleaned the z-bars before sending them to the manufacturer. But the licensees routine
cleaning, which involves soaking the intact components, is not as rigorous as the cleaning
performed by the manufacturer. The manufacturer revised its cleaning and lubricating
instructions, but those instructions do not include the dismantling and rigorous cleaning
performed by the manufacturer.
In Event 2, the licensee believed that the patients cough caused the movement of the left
anterior pin. An article from the 2002 Journal of Neurosurgery that the licensee provided to Enclosure 2 NRC, confirms that the three-pin technique the licensee used does not provide the same level
of immobility as the four-pin procedure. The manufacturer of the Gamma Knife offers a number
of different size screws and posts to permit repositioning of the patients head within the head
frame, if it appears a collision will occur. The manufacturer has also designed a front piece, specifically for a three-pin technique, that more evenly spaces the attachment pins and
distributes the forces on the pins around the skull. Further, historical evidence from the
manufacturer has shown that the attachment pins will move when the screws are not properly
tightened into the table of the skull. Movement of improperly tightened screws is also more
likely to happen in certain screw positions (e.g., when the screws are not positioned almost
perpendicularly to the tangential skull plane). Additionally, a loose screw may not be apparent
until the patient has gone through a number of preparation steps, such as pre-treatment
imaging or other movements that put dynamic stresses on the pin-skull interface. The
movement is usually detected by the presence of blood, caused as the sharp pin moves from its
initial location. The licensees failure to retighten the three remaining screws may have
contributed to the pin slippage; the angle of the screws and positioning of the head frame
medially, to accommodate the patients head size, may have also contributed to the slippage.
The licensee in this medical use occurrence did not provide sufficient evidence to justify the
claim of patient intervention. Specifically, the licensee did not provide sufficient evidence to
exclude improper tightening of the left anterior pin once the right anterior pin was removed or to
demonstrate that the three-pin technique used would provide the same immobilization provided
by either use of the fourth pin or equipment designed by the manufacturer specifically for a
three-pin head frame attachment. As noted previously, §35.3045 (a)(3) requires a license to
report any event (except for an event that results from patient intervention) where the
administration of byproduct material, or radiation from byproduct material, results in a dose to
the skin, or an organ or tissue other than the treatment site, that exceeds by 0.5 Sv (50 rem) or
more, and 50 percent or more, the dose to the skin, or organ or tissue other than the treatment
site, that was expected from the administration defined in the written directive. Based on the
licensees initial assessment, this administration resulted in an unintended dose of 7.6 Sv (760
rem) in an area of the brain that was 6 mm (0.24 inches) away from the intended treatment site
and was expected to receive approximately 2.6 Sv (260 rem). Therefore, at the time of the
licensees initial assessment, the licensee should have reported this event under
Subsequent to NRCs determination that the administration was a reportable medical event, the
licensee provided additional information that corrected errors in earlier information, concerning
the location of the unintended site and the dose to that site. The new information more
accurately set the time of the head frame slippage to halfway through the 11th exposure. This
reduced the estimated dose to the unintended site to 2.5 Sv (250 rem). The licensee also
corrected the location of the unintended site to the 30 percent isodose line, and not the 10
percent isodose line, as previously reported. The result of these corrections is that the event is
no longer considered a reportable medical event, under §35.3045(a)(3), because the additional
dose to the unintended site was 32 percent of the dose to that site (7.8 Sv, or 780 rem)
expected from the administration defined in the written directive, which is less than the 50
percent threshold for reporting a medical event under this criterion. However, the event is
included in the IN because the additional information the licensee subsequently provided did not
change NRCs conclusion that the event was not the result of patient intervention.