7 % 6

  • ' 'wg UNITED ST ATES

,g NUCLEAR REGULATORY COMMISSION s ' ';;[;/. C y-f Q REGloN lli 7'J9 ROOSEVELT HOAD

'A 2/ f GLEP, E LLYN, It.LINOIS 6013 7 W.Y 1 4198)

MEMORANDDI FOR:

L. J. Cunningham, Chief, Renetor 1:adiological Safety Sectlon, II:

FROM:

L. R. Greger, Acting Chief, Facilities Radiation Protection Section SUL.1ECT:

EBERLINE MIS-3 CONTIN 0US AIR MONITOR Attached you will find two letters from Dow Chemical dated March 24, 1981 and April 20, 1981, which describe an apparent problem with the o-ring seals of their MIS-3 continuous air monitor.

'lhis matter was discussed with Eberline's regional representative on May 12, 1981. 11e agreed that o-ring maintenance was not mentioned in the technical manual but felt that inspection and maintenance of the o-rings was a matter of common sense. lie also felt that the filter lock-in device (as Dow calls it) was probably not the cause of the problem since the filter holder will stay in place without it, if the o-rings are in good condition.

In our review of this matter, we have determined that o-ring inspection and maintenance may be a generic problem.

k'e request that you examine this in-f3rmation to determine if it should be passed on to other users of this eq uiprrent.

If you have any questions, please contact Paul Lovendale at 384-2638 or me.

?

p L. R. Greger, Acting Chief Facilities Radiation Protection Section ALLa.hmento:

As stated i

i-i I

j.- M 8508270306 850712 PDR FOIA KOHN85-256 PDR