/03L-0:on 810224,oxygen Analyzer Declared Inoperable.Caused by Damaged Oxygen Cell Due to Collection of Condensed Moisture.Lines Cleared,Cell Replaced & Analyzer Recalibr.Addl Vents Will Be Added
ML19350C386
Person / Time
Site: Sequoyah Tennessee Valley Authority icon.png
Issue date: 03/24/1981
From: Harding M, Wilkey A
TENNESSEE VALLEY AUTHORITY
To:
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION II)
Shared Package
ML19350C385 List:
References
LER-81-028-03L, LER-81-28-3L, NUDOCS 8104010266
Download: ML19350C386 (2)


LER-1981-028, /03L-0:on 810224,oxygen Analyzer Declared Inoperable.Caused by Damaged Oxygen Cell Due to Collection of Condensed Moisture.Lines Cleared,Cell Replaced & Analyzer Recalibr.Addl Vents Will Be Added
Event date:
Report date:
3271981028R03 - NRC Website

text

.

m

. ~ - - -............

l Q )' '

..I l

1 l

l lh i,tcAss,RiNr OR Tyrs au. REcuinto in,onwarioN, i

CONrROt. 8i.ccic: l

~

i.

lo lil IT IN ls IN IP'!1 (h O l 0 !

! O ! O ! 0 l 0 l0 l-l0 IO lh 4 l 1 l'111111@{_ l' { g e s UCENsti CCOE I4

.ib UCENSENUMSER 26 24 UCENSE TYPE 40 Sp CAg g CON'T

[ TIT}

  • ,Cy dl0.15 10 IO l 0 I 31217 @l 012 l~2 I4 l811 @l0131214I811l@

3,

g 3,

s 60 Si CoCit ti NUuC EM OS 69 EvtNT DATE 14. 76 REPCRT DAT E so

~

' EVtNT DESCRIPTION AND PROB A8LE CONSEQUENCES h f'5 TTI I With unit 1 in Mode s. the oxveen analyzer used to monitor for an exolosive ens mixtura g-

~

ITTTII in the waste' holdup ' tank was declared inoperable. Gr'ab samples.were taken within foor l

I n l41 l hours.as required Ev action statement'43 of LCO 3.3.3.10, but the samples were not l

lol3l l analyzed for approx'imately twelve hours because both gas chromatographs used for*this

'g l o l61 l analysis were fnoperable. There was no ef fect on publik: health oY safety. No previous g

~

g j occurrences.

g l o la l l I

80 1

as SYSitu cAust CAU$t CCup.

VAtyt CCDE Coot *

$U8 Coot COU*CNt NT CO3t SusCoct sugCoat tonel I M l ~ c l@ LE_j h l E l h l T l N l 9 I T l *t lU lh W h d h io n

  • n is

. i.

i.

2C LE R/40 EvtNT YE AR mtponT NO.

CCDE TYPE ~

R EVt3tCN S t out NTI AL OCCunntNCE*

REPOmf No.

@, mig l2 B8l1l g

l 0 l2 l8 l l/l l0 ~l3 ]

y

[_.]

y 22 23 24 26 27 28 29 30

. 3, 22 T

Et A

Or PLANT MET o MOURS 5 a sf FCn s.

svPPtitm

  • u ANUF ACTun g a l Al h ! F l h lSZl@

[Z_j h I O l 0 l O l 0 l

{d@

ULl@

U._}@

lc!7I6Is!

3J Ja J

34 37 40 48 42 43 44 47 CAUSE DESCRIPTION AND CORRECTIVE ACTIONS 27 l i j o l l The" oxygen' cell was d'amaged due to the' collection of condomad mM dntra in the a alvzer. 1 li li j l The water was cleared from the lines, the cell was replaced, the analyzer was recalibratedi

,,,7, g and returned'to service. Subsequent analysis verified that the sampled gas was within the j l

,,,3j l acceptable limits. Additional vents will be ad'ded prior to un'it 2 fue loading to

,[

i eliminate excessive water collection.

,,,,g; s

s i so '

DasCovany otsCniPTsow h '

s'r U.S.'-

s Powtm 01HtnSTAfus sC NA l

l A l@l Operator observation l

I * ! S I l C l@ l 0I O l 0 l@l 8

' f,, v,,,

Co'",cNT AuoVNT OF ACTivtTY @ l A

LOCATiCN 07 ptLEAst stat t Aste or n:LtAst l

NA

! f l

! i lo l I z i @ I z_J@l NA I

2 s 3 to i.

44 45 60 Ptn50NNEL f xPC5UmES NUue t a TYPE OtsCniPTICN 30

.~

11 i I 71 1 O l 0 l 0 l@l z l@l NA

,c a:eNN E'L',N Jit s f

Nuwetm ot5CRsPfl0N l l L,i !.gj i 01010 l@l NA

.. i

,i n

l LC15 0F on OAuAct TO FAcauTY 4

l TV't C E SCRIPif 0N l

!1191 1 zl@l NA so R

io N R C USE ON L '

l PU A LICIT V I

. r i e l W@DESCmiPTION 151UtD I

11I!!tillt!!l l

l u

as es i

e s to Name of PreparerM. R. Unrdine/A. M. Wilkev Phone (615) 842-8317 __

l 8"a n au 200R ORSML

4 h i

- Tcnn22203 Vallcy Aut or ty Soqueyth Nuclear Plant LER SUPPLEMENTAL IhTORMATION SQRO-50-327/81028 Technical Specification Involved:

3.3.3.10 6.9.1.13.b Reported Under Technical Specification:

Date of Occurrence: 2-24-81 Time of Occurrence: 1250 CST Identification and Description of Occurrence:

The oxygen analyzer used to monitor for an explosive gas mixture in the waste holdup tank was declared inoperable when water damaged the oxygen Grab samples were taken as required by action statement 43 of cell.

LCO 3.3.3.10 but analysis of the samples was delayed for approximately 12 hours1.388889e-4 days <br />0.00333 hours <br />1.984127e-5 weeks <br />4.566e-6 months <br /> because both gas chromatographs used for the analysis were inoperable.

Conditions Prior to occurrence:

~

Unit 1 in Mods 5 Apparent Cause of Occurrence:

Condensed moisture collected in the analyzer and damaged the oxygen cell.

1

[

[

Analysis of Occurrence _:

Analysis of the gas sample was delayed until a loan unit could be obtaine,d from the Power Operations Training Center. Subsequent analysis verified that the sampled gas was within technical specification limits.

Corrective Action

A design modification will be implemented prior to fuel loading of Unit 2 to install additional low-point vents to eliminate the collection of excessite water in the lines connected to the analyzer.