Strontium-90 Eye Applicators: Submission of Quality Management Plan (Qmp), Calibration, and UseML031060614 |
Person / Time |
---|
Site: |
Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Mcguire, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant |
---|
Issue date: |
03/11/1994 |
---|
From: |
Paperiello C NRC/NMSS/IMNS |
---|
To: |
|
---|
References |
---|
IN-94-017, NUDOCS 9403080349 |
Download: ML031060614 (7) |
|
Similar Documents at Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Mcguire, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant |
---|
Category:NRC Information Notice
MONTHYEARInformation Notice 2020-02, Flex Diesel Generator Operational Challenges2020-09-15015 September 2020 Flex Diesel Generator Operational Challenges ML20225A0322020-09-0303 September 2020 NRC Choice Letter to NAC International with Attached Safety Inspection Report, IR 0721015/2020201, February 24-27, 2020 and July 22, 2020, Inspection of NAC International in Norcross, Georgia Information Notice 2012-09, PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs.2019-11-30030 November 2019 PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs. Information Notice 2011-20, NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011)2019-07-24024 July 2019 NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011) ML19196A2452019-07-15015 July 2019 Public Notice - Sequoyah Nuclear Plant, Unit 2 - Exigent Amendment to Facility Operating License Information Notice 2019-01, Inadequate Evaluation of Temporary Alterations2019-03-12012 March 2019 Inadequate Evaluation of Temporary Alterations ML16028A3082016-04-27027 April 2016 NRC Information Notice; IN 2016-05: Operating Experience Regarding Complications From a Loss of Instrumentation Air Information Notice 2015-05, Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps Information Notice 2015-05, Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps Information Notice 2013-20, OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-20, Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-11, OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2013-11, Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Contain2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Con2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notic2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2012-13, Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool Information Notice 2012-13, Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool Information Notice 2012-11, Age Related Capacitor Degradation2012-07-23023 July 2012 Age Related Capacitor Degradation ML12031A0132012-02-0606 February 2012 U.S. Nuclear Regulatory Commission Investigation Report No. 2-2010-058, Cpn International, Inc Information Notice 2011-19, Licensee Event Reports Containing Information Pertaining to Defects to Basic Components2011-09-26026 September 2011 Licensee Event Reports Containing Information Pertaining to Defects to Basic Components Information Notice 2011-15, Steel Containment Degradation and Associated License Renewal Aging Management Issues2011-08-0101 August 2011 Steel Containment Degradation and Associated License Renewal Aging Management Issues Information Notice 2011-17, Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping2011-07-26026 July 2011 Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping Information Notice 2011-13, Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-13, Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13)2011-06-29029 June 2011 Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13) Information Notice 2011-13, OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-04, IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-262010-12-21021 December 2010 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 ML13066A1872009-12-16016 December 2009 Draft NRC Information Notice 2009-xx - Underestimate of Dam Failure Frequency Used in Probabilistic Risk Assessments ML1007804482009-11-23023 November 2009 Email from Peter Bamford, NRR to Pamela Cowan, Exelon on TMI Contamination Control Event Information Notice 2009-11, NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-112009-07-0707 July 2009 NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-11 Information Notice 2009-10, Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10)2009-07-0707 July 2009 Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10) Information Notice 2009-09, Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify2009-06-19019 June 2009 Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify Information Notice 2008-12, Reactor Trip Due to Off-Site Power Fluctuation2008-07-0707 July 2008 Reactor Trip Due to Off-Site Power Fluctuation Information Notice 2008-11, Service Water System Degradation at Brunswicksteam Electric Plant Unit 12008-06-18018 June 2008 Service Water System Degradation at Brunswicksteam Electric Plant Unit 1 Information Notice 2008-04, Counterfeit Parts Supplied to Nuclear Power Plants2008-04-0707 April 2008 Counterfeit Parts Supplied to Nuclear Power Plants Information Notice 1991-09, Counterfeiting of Crane Valves2007-09-25025 September 2007 Counterfeiting of Crane Valves Information Notice 2007-28, Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls2007-09-19019 September 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls Information Notice 2007-29, Temporary Scaffolding Affects Operability of Safety-Related Equipment2007-09-17017 September 2007 Temporary Scaffolding Affects Operability of Safety-Related Equipment Information Notice 2007-14, Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station2007-03-30030 March 2007 Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station Information Notice 2007-06, Potential Common Cause Vulnerabilities in Essential Service Water Systems2007-02-0909 February 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Information Notice 2007-05, Vertical Deep Draft Pump Shaft and Coupling Failures2007-02-0909 February 2007 Vertical Deep Draft Pump Shaft and Coupling Failures Information Notice 2006-31, Inadequate Fault Interrupting Rating of Breakers2006-12-26026 December 2006 Inadequate Fault Interrupting Rating of Breakers Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear Information Notice 2006-13, E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination2006-07-13013 July 2006 E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination 2020-09-03 The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Mcguire]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>.
[Table view]The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Mcguire]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>. |
UNITED STATES
NUCLEAR REGULATORY COMMISSION
OFFICE OF NUCLEAR MATERIAL SAFETY AND SAFEGUARDS
WASHINGTON, D.C. 20555 March 11, 1994- NRC INFORMATION NOTICE 94-17: STRONTIUM-90 EYE APPLICATORS:
SUBMISSION OF QUALITY MANAGEMENT
PLAN (QMP), CALIBRATION, AND USE
Addressees
All U.S. Nuclear Regulatory Commission Medical Use Licensees.
Purpose
The NRC is issuing this Information Notice to remind licensees authorized to
possess Sr-90 for authorized uses described in 10 CFR 35.400(e) of the need
includes
to: (1)implement and submit a quality management plan (QMP)inthat 10 CFR 35.32;
policies and procedures to meet the requirements described
use and
and (2) inform all licensees of recent information regarding the
calibration of Sr-90 eye applicators.
Description of Circumstances
There has been confusion as to the need for a QMP for Sr-90 eye applicators
used for patient treatment procedures. The NRC's Office of the General
staff that because the QM program requirements
Counsel advised the technical use of Sr-90 eye
in 10 CFR 35.32 apply to many brachytherapy dose," the
submission of a QMP.
applicators, as described in 10 CFR 35.400(e), requires of the
This Information Notice reminds Sr-90 eye applicator licensees of 10
requirement to submit and implement a QMP that meets the requirements
CFR 35.32.
Additionally, this letter alerts licensees to possible discrepancies licensees
associated with the calibration of Sr-90 eye applicators and reminds and use.
of the regulatory requirements associated with their calibration
Discussion- Submissi0n of a Oualitv Management Plan
If you are licensed forAand using a Sr-90 eye applicator, the NRC's
you to submit a QMP that meets the requirements of 10 CFR
regulations require from the Sr-90
35.32. The QMP should provide high confidence that radiation user. The
eye applicator will be administered as directed by the authorized that meet the
submitted QMP should include written policies and procedures
five objectives, as described in 10 CFR 35.32(a):
943 08034 P E 9q031
,,,9403080349 . 11 :
,,.%l
'
>XV1 IO
-M
.-- AIN 94-17 March 11, 1994 directive, signed and dated
1. That prior to administration, a written each
by an authorized user, is prepared for applicable
administration.
(A written directive for Sr-90 eye-applicators means an order, in
and signed by an authorized
writing, for a specific patient, dated must include the
It
user prior to administration of radiation.strength (corrected for
radioisotope, the treatment site, source
decay), and exposure time (or equivalently, the total dose).
is identified by more
2. That prior to each administration, the patient the written directive.
than one method as the individual named in
3. That final plans of treatment and related calculations are in
accordance with the respective written directive.
the written
4. That each administration is in accordance with
directive.
directive is
5. That any unintended deviation from the written
identified and evaluated, and appropriate action taken.
Additionally, to meet the requirements of 10 CFR 35.32(b), you need policies
to verify compliance of all
and procedures for conducting a review of the QMPthan 12 months. The review is
aspects of the program at intervals no greater administrations, all recordable
to include a representative sample of patient actions taken.
events and all misadministrations, and any corrective
the program's
Licensees may make modifications to their QMPs toisincrease
not decreased.
efficiency provided the program's effectiveness to their QMP to the
10 CFR 35.32(e) requires licensees to submit modifications
the modification has been
appropriate NRC regional office within 30 days after
made.
are not, or do not intend
If you are licensed for a Sr-90 eye applicator, but in lieu of a QMP. This
to use it,you may submit a "negative declaration' of a QMP to NRC prior to
written declaration must commit to your submission procedures.
use of the Sr-90 eye applicator for patient treatment
IN 94-17 March 11, 1994 Calibration and Use of Strontium-90 Eve aDplicators
correspondence from
The NRC staff has had discussions with and received and Technology (NIST) about
researchers at the National Institute of Standards to Sr-90 eye
large discrepancies among calibrated outputs assigned may have been in units
applicators. The manufacturer's original calibrations
to System Internationale
that are no longer used or have little relationshipunits from different
units in use today. Further, comparisons between
discrepancies
manufacturers may be meaningless and misleading. Finally, output measurements between
larger than 10 percent still exist when comparing
competent measurement labs using state-of-the-art techniques.
Committee on Medical Uses
The above information was provided to NRC's Advisorymeeting. The ACMUI members
of Isotopes (ACMUI) at its November 1 and 2, 1993, factor in the use of
advised the staff that calibration was not critical
a
licensees treat for
Sr-90 eye applicators for treating pterygium because
recommended that NRC inform
response rather than to tolerance. The ACHUI with calibration of Sr-90 eye
licensees of the potential problems associatedcautioning licensees that if they
applicators. The Committee also recommended possession or buy a
use an applicator, other than the one currently in theirto another device
new one, their current technique may not be applicable for the differenti
because of variances in stated and actual exposure rates
applicators.
eye applicators that
NIST offers a service for the calibration of Sr-90
rate, and (2)a mapping of the
provides: (1)the calculation of the emission in order to ascertain
Sr-90 distribution across the face of the applicator
on the NIST calibration service, uniformity of dose. For further information 975-5589 or Dr. Bert Coursey at
contact either Dr. Christopher Soares at (301)
knowledge, the
(301) 975-5584. Although NIST is, to the best of the staff'sNRC is not requiring
only provider of such a service, there may be others.
and accepts calculations of
that licensees participate in the NIST program,calibration, corrected for
exposure time based on the original vendor's
radioactive decay.
as a brachytherapy
Since 10 CFR 35.400(e) identifies the Sr-90 eye applicator
Requirements for possession
device, licensees are reminded that 10 CFR 35.59, does apply to Sr-90 eye
of sealed sources and brachytherapy sources," applicable leak tests and
applicators. Specifically, licensees must perform
surveys 4escrlbed in 10 CFR 35.59. Additionally, 10 CFR 35.400(e) limits the
eye conditions."
use of S"90-veye applicators to "treatment of superficial of a Sr-90 eye
LicenseebAtL6'reminded that NRC authorization for use
authorize its use on treatment
applicator'for'pattent procedures does not
sites other than the eye.
V. IN94-17 March 11, 1994 written response. If
This Information Notice requires no specific action or please contact the
you have questions about the information in this notice, appropriate-technical contact listed below.
Carl J. Paperiello, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical contacts: Jim Dwyer, RI Jacqueline Burks, RIV
(215) 337-5309 (817) 860-8132 John Pelchat, RII James Montgomery, RV
(404) 331-5083 (510) 975-0249 Cassandra Frazier, RIII Sally L. Merchant, NMSS
(708) 790-5704 (301) 504-2637 Attachments: Notices
1. List of Recently Issued NMSS InformationNotices
2. List of Recently Issued NRC Information
l IAB 'IMAB TechEd* IMOB* IMOB l
OI l IMAB
KRamsey FCombs
NAME SMerchantt 2 j AMCiamper JEGlenn EKraus
DATE l02/24/94C 12 Z - 24/94 12/29L93 0113/94 02/20/9 OiFC- ZE WBrach MI iMN
CP aiejiello _
NAME STreby
DATE 02/04/94 02/07/94 03/0/940
DOC NAME: 94-17.IN
OFFICIAL RECORD COPY
- S\f
s._J Attachment 1 IN 94-17 March 11, 1994 LIST OF RECENTLY ISSUED
NMSS INFORMATION NOTICES
Unformation Date of
Notice No. Subject Issuance Issued to
Recent Incidents Resulting 03/03/94 All U.S. Nuclear Regulatory
94-16 Commission material and fuel
in Offsite Contamination
cycle licensees.
Radiation Exposures during 03/02/94 All U.S. Nuclear Regulatory
94-15 Commission licensees author- an Event Involving a Fixed
Nuclear Gauge ized to possess, use, manu- facture, or distribute
industrial nuclear gauges.
Release of Patients with 02/03/94 All U.S. Nuclear Regulatory
94-09 Commission medical
Residual Radioactivity
from Medical Treatment and licensees.
Control of Areas due to
Presence of Patients Con- taining Radioactivity
Following Implementation
of Revised 10 CFR Part 20
Solubility Criteria for 01/28/94 All byproduct material and
94-07 fuel cycle licensees with
Liquid Effluent Releases
to Sanitary Sewerage under the exception of licensees
the Revised 10 CFR Part 20 authorized solely for
sealed sources.
Reporting Requirements 12/22/93 All U.S. Nuclear Regulatory
93-100 Commission licensees.
for Bankruptcy
Implimentation of the 10/08/93 All byproduct, source, and
93-80 special nuclear material
dRevised-10 COR Part 20
licensees.
27i a'h rsrt-that Result 10/04/93 All nuclear fuel cycle
93-77 in InadVertent Transfers licensees.
of Special Nuclear Material
at Fuel Cycle Facilities
K- I ..JAttachment 2 IN 94-17 March 11, 1994 LIST OF RECENTLY ISSUED
4_' NRC INFORMATION NOTICES
Information Date of
Subject Issuance Issued to
Notice No.
Recent Incidents Resulting 03/03/94 All U.S. Nuclear Regulatory
94-16 in Offsite Contamination Commission material and fuel
cycle licensees.
Radiation Exposures during 03/02/94 All U.S. Nuclear Regulatory
94-15 Commission licensees author- an Event Involving a Fixed
Nuclear Gauge ized to possess, use, manu- facture, or distribute
industrial nuclear gauges.
Failure to Implement 02/24/94 All holders of OLs qr CPs
94-14 for nuclear power and non- Requirements for Biennial
Medical Examinations and power reactors and ll
Notification to the NRC licensed reactor operators
of Changes in Licensed and senior reactor
Operator Medical Conditions operators.
Intersystem LOCA 02/22/94 All holders of OLs or CPs
92-36, for nuclear power reactors.
Supp. 1 Outside Containment
Unanticipated and Un-
- 02/22/94 All holders of OLs or CPs
94-13 for nuclear power reactors.
intended Movement of
Fuel Assemblies and
Other Components due to
Improper Operation of
Refueling Equipment
02/09/94 All holders of OLs or CPs
94-12 Insights Gained from for nuclear power reactors.
Resolving Generic
4Ijsue.57; Effects of
i~jNmrotecton System
- ~n. 6n Safety- t 4tqp ent
Turbine Overspeed and 02/08/94 All holders of OLs or CPs
94-11 for nuclear power reactors.
Reactor Cooldown during
Shutdown Evolution
OL - Operating License
CP - Construction Permit
K../VIN 94-17 March 11, 1994 or written response. If
This Information Notice requires no specific action notice, please contact the
you have questions about the information in this
appropriate'technical contact listed below.
Carl J. Paperiello, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical contacts: Jim Dwyer, RI Jacqueline Burks, RIV
(215) 337-5309 (817) 860-8132 John Pelchat, RII James Montgomery, RV
(404) 331-5083 (510) 975-0249 Cassandra Frazier, RIII Sally L. Merchant, NMSS
(708) 790-5704 (301) 504-2637 Attachments: Notices
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information
63j1 b
|
---|
|
list | - Information Notice 1994-01, Turbine Blade Failures Caused by Torsional Excitation from Electrical System Disturbance (7 January 1994)
- Information Notice 1994-02, Inoperability of General Electric Magne-Blast Breaker Because of Misalignment of Close-Latch Spring (7 January 1994)
- Information Notice 1994-03, Deficiencies Identified During Service Water System Operational Performance Inspections (11 January 1994, Topic: Biofouling)
- Information Notice 1994-04, Digital Integrated Circuit Sockets with Intermittent Contact (14 January 1994)
- Information Notice 1994-05, Potential Failure of Steam Generator Tubes with Kinetically Welded Sleeves (19 January 1994, Topic: Stress corrosion cracking)
- Information Notice 1994-06, Potential Failure of Long-Term Emergency Nitrogen Supply for the Automatic Depressurization System Valves (28 January 1994)
- Information Notice 1994-07, Solubility Criteria for Liquid Effluent Releases to Sanitary Sewerage Under the Revised 10 CFR Part 20 (28 January 1994)
- Information Notice 1994-08, Potential for Surveillance Testing to Fail to Detect an Inoperable Main Steam Isolation Valve (1 February 1994)
- Information Notice 1994-09, Release of Patients with Residual Radioactivity from Medical Treatment & Control of Areas Due to Presence of Patients Containing Radioactivity Following Implementation of Revised 10 CFR Part 20 (3 February 1994, Topic: Brachytherapy)
- Information Notice 1994-10, Failure of Motor-Operated Valve Electric Power Train Due to Sheared or Dislodged Motor Pinion Gear Key (4 February 1994)
- Information Notice 1994-11, Turbine Overspeed and Reactor Cooldown During Shutdown Evolution (8 February 1994, Topic: Overspeed trip)
- Information Notice 1994-12, Insights Gained from Resolving Generic Issue 57: Effects of Fire Protection System Actuation on Safety-Related Equipment (9 February 1994, Topic: Overspeed trip)
- Information Notice 1994-13, Unanticipated and Unintended Movement of Fuel Assemblies and Other Components Due to Improper Operation of Refueling Equipment (28 June 1994)
- Information Notice 1994-14, Failure to Implement Requirements for Biennial Medical Examinations and Notification to the NRC of Changes in Licensed Operator Medical Conditions (24 February 1994)
- Information Notice 1994-15, Radiation Exposures During an Event Involving a Fixed Nuclear Gauge (2 March 1994, Topic: Overspeed trip)
- Information Notice 1994-16, Recent Incidents Resulting in Offsite Contamination (3 March 1994, Topic: Overspeed trip)
- Information Notice 1994-17, Strontium-90 Eye Applicators: Submission of Quality Management Plan (QMP), Calibration, and Use (11 March 1994, Topic: Brachytherapy)
- Information Notice 1994-17, Strontium-90 Eye Applicators: Submission of Quality Management Plan (Qmp), Calibration, and Use (11 March 1994, Topic: Brachytherapy, Overspeed)
- Information Notice 1994-18, Accuracy of Motor-Operated Valve Diagnostic Equipment (Responses to Supplement 5 to Generic Letter 89-10) (16 March 1994)
- Information Notice 1994-19, Emergency Diesel Gemerator Vulnerability to Failure from Cold Fuel Oil (16 March 1994)
- Information Notice 1994-20, Common-Cause Failures Due to Inadequate Design Control and Dedication (17 March 1994)
- Information Notice 1994-21, Regulatory Requirements When No Operations Are Being Performed (18 March 1994)
- Information Notice 1994-22, Fire Endurance & Ampacity Derating Test Results for 3-Hour Fire-Rated Thermo-Lag 330-1 Fire Barriers (16 March 1994, Topic: Fire Barrier)
- Information Notice 1994-23, Guidance to Hazardous, Radioactive and Mixed Waste Generators on the Elements of a Waste Minimization Program (25 March 1994, Topic: Fire Barrier)
- Information Notice 1994-24, Inadequate Maintenance of Uninterruptible Power Supplies & Inverters (24 March 1994, Topic: Safe Shutdown, Fire Barrier)
- Information Notice 1994-25, Failure of Containment Spray Header Valve to Open Due to Excessive Pressure from Inertial Effects of Water (15 March 1994, Topic: Fire Barrier, Water hammer)
- Information Notice 1994-26, Personnel Hazards and Other Problems from Smoldering Fire-Retardant Material in the Drywell of a Boiling-Water Reactor (28 March 1994, Topic: Fire Barrier)
- Information Notice 1994-27, Facility Operating Concerns Resulting from Local Area Flooding (31 March 1994, Topic: Fire Barrier)
- Information Notice 1994-28, Potential Problems with Fire-Barrier Penetration Seals (5 April 1994, Topic: Fire Barrier, Operability Determination)
- Information Notice 1994-29, Charging Pump Trip During a Loss-of-Coolant Event Caused by Low Suction Pressure (11 April 1994, Topic: Boric Acid)
- Information Notice 1994-30, Leaking Shutdown Cooling Isolation Valves at Cooper Nuclear Station (19 August 1994, Topic: Fire Barrier, Local Leak Rate Testing)
- Information Notice 1994-31, Potential Failure of Wilco, Lexan-Type HN-4-L Fire Hose Nozzles (14 April 1994, Topic: Hydrostatic)
- Information Notice 1994-32, Revised Seismic Estimates (29 April 1994, Topic: Earthquake)
- Information Notice 1994-33, Capacitor Failures in Westinghouse Eagle 21 Plant Protection Systems (9 May 1994, Topic: Stress corrosion cracking)
- Information Notice 1994-34, Thermo-LAG 330-660 Flexi-Blanket Ampacity Derating Concerns (13 May 1994, Topic: Fire Barrier, Stress corrosion cracking)
- Information Notice 1994-35, Niosh Respirator User Notices, Inadvertent Separation of the Mask-Mounted Regulator(Mmr) from the Facepiece on the Mine Safety Appliances (16 May 1994, Topic: Stress corrosion cracking)
- Information Notice 1994-35, Niosh Respirator User Notices, Inadvertent Separation of the Mask-Mounted Regulator(MMR) from the Facepiece on the Mine Safety Appliances (16 May 1994)
- Information Notice 1994-36, Undetected Accumulation of Gas in Reactor Coolant System (24 May 1994, Topic: Reactor Vessel Water Level, Stress corrosion cracking, Integrated leak rate test)
- Information Notice 1994-37, Misadministration Caused by a Bent Interstitial Needle During Brachytherapy Procedure (27 May 1994, Topic: Stress corrosion cracking, Brachytherapy)
- Information Notice 1994-38, Results of Special NRC Inspection at Dresden Nuclear Power Station, Unit 1 Following Rupture of Service Water Inside Containment (27 May 1994)
- Information Notice 1994-39, Identified Problems in Gamma Stereotactic Radiosurgery (31 May 1994)
- Information Notice 1994-40, Failure of a Rod Control Cluster Assembly to Fully Insert Following a Reactor Trip at Braidwood, Unit 2 (26 May 1994)
- Information Notice 1994-41, Problems with General Electric Type Cr124 Overload Relay Ambient Compensation (7 June 1994)
- Information Notice 1994-41, Problems with General Electric Type CR124 Overload Relay Ambient Compensation (7 June 1994)
- Information Notice 1994-42, Cracking in the Lower Region of the Core Shroud in Boiling-Water Reactors (7 June 1994)
- Information Notice 1994-43, Determination of Primary-to-Secondary Steam Generator Leak Rate (10 June 1994, Topic: Grab sample)
- Information Notice 1994-44, Main Steam Isolation Valve Failure to Close on Demand Because of Inadequate Maintenance and Testing (16 June 1994)
- Information Notice 1994-44, Main Steam Isolation Valve Failure to Close on Demand because of Inadequate Maintenance and Testing (16 June 1994)
- Information Notice 1994-45, Potential Common-Mode Failure Mechanism for Large Vertical Pumps (17 June 1994, Topic: Biofouling)
- Information Notice 1994-46, Nonconservative Reactor Coolant System Leakage Calculation (20 June 1994, Topic: Unidentified leakage)
... further results |
---|