LER 82-041/03L-0:on 821011,during Calibr of Radioactive Liquid Effluent Monitors,Trip Point for RIS-6212 Found to Be Less Conservative than Required.Caused by Instrument Drift. High Voltage & Discriminator Settings Adjusted
ML20028A286
Person / Time
Site: Fort Saint Vrain Xcel Energy icon.png
Issue date: 11/10/1982
From: Fuller C
PUBLIC SERVICE CO. OF COLORADO
To:
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION IV)
Shared Package
ML20028A283 List:
References
LER-82-041-03L, LER-82-41-3L, NUDOCS 8211180048
Download: ML20028A286 (1)


Text

. ... - -- - . . . . .- .. . . . . . -.

  • LICENSEE EVENT REPORT (PLEAst PRINT OR TYPE ALL KEIUIRED INPDIAAATIONI CONTROL block: l l l l l l @

F o i 4 9 I C lLICENSEE 01 FICODE S l Vl l l@l 0101-10101010101-1010 le il LICENSE MvpSER 2' '

@l4lllll210 38 LICENSE TvPE JO

@l57 CAT I

SU l@

con'T 7

SU C80L'J@l$1 015101010 DOCKET NUMS4R

21617 0l SDi i 0 EVENT 81 i i l iDATE l 8121@l 75111111018121@

74 MEPORT DATE 50 EVENT DESCRIPTION AND pro 8ABLE CONSEQUENCES h O I During the quarterly calibration of the radioactive liouid effluent monitors. the I o l trip point for RIS-6212 was found to be less conservative than required and is m- I o . l portable as a degraded mode of LCO 4.8.2(d) per Fort St. Vrain Technical Specificatio..I o 3 lAC 7.5.2(b)2. No accompanying occurrences. No affect on public health or safety. I o o l Redundant system operable. Similar reports are R0's82-004, 81-034,81-036, 81-013. I o , l 00-67, 79-37, and 77-24. l ib F1 I I

' C$I'E *^"o'I CO sNC$5E COMPONENT CODE Sus "O E S E E I xI xl@ LEj@ [Ej@ l il NI Sl Tl Rl Ul@ LEJ@ W@

tER VENT y AR l'S"OT

    • EEE**'

"'E " '

@ ,ajg,RO 1812 l [ ,-j 1014 I I I 24 24 y

27 I 01 31 29 29 y3 [ ,-j 33 p3 2% 22 23 K N AC O ON PL NT ME HOURS 22 sei PO B. SU L MAN CTURER LEj@[Z_J@

E [ Z,,J@ 101010101 L.!d@ L.NJ@ l44G 10 16 13 l@'

33 34

[JSZ_J@ 36 3 40 @4@ 42 43 47 CAUSE DESCRIPTION AND CORRECTIVE ACTIONS g ; Instrument drift was the cause of he detector being out of calibration. High voltagei

, , land discriminator settings were adjusted during a calibration to yield the recommended l

, , , trip point. RIS-6212: Manufacturer - General Atomic Company, Model Number - ELD- i g ;185-3300-2C. No further corrective action is anticipated or required. l EI .J C '

sPOwER OTHER STATUS s@isEEv' oisCOvERv oEsCRiPTiON @

i s 'shT 'iVs*@ 10 l0 l0 lgl N/A l J@l Surveillance Test l

' JTi.viTv CO4 TENT AMOUNTOP ACTIVITv LOCATION OP RELEASE i 6 [Zj RELEASE @D OP RELEASE [Zj@l N/Al l N/A l 7 8 9 10 11 44 45 80 PERSONNEL EXPOSURES NUMB E R TYPE DESCRsPTION i 7 l010l0l@[Zj@l N/A i

' ' ' PERsONNE'L ' iN;U' lies

NuM.ER oEsCRiPTiON@

i R l0l0l0l@l N/A l ,

LOSS OP OR D AGE OPACILtTV

i 9 W @l N/A l 7 8 9 to 80 2 O t$dsUE l @DE N/A CRIPTION l l l li l l l I l l 11 lj

' 7 8 3 to 68 69 80 3

.. NAMF ns.paspacco ..

~

M / // ^ PHONE:

8211100048 821110

- PDR ADOCK 05000267 _. , ., - ,

m PDR