Safety Concerns Related to Repeated Control Unit Failures of the Nucletron Classic Model High-Dose-Rate Remote Afterloading Brachytherapy DevicesML031040399 |
Person / Time |
---|
Site: |
Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Mcguire, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant |
---|
Issue date: |
07/06/1999 |
---|
From: |
Cool D NRC/NMSS/IMNS |
---|
To: |
|
---|
References |
---|
IN-99-023, NUDOCS 9907010001 |
Download: ML031040399 (9) |
|
Similar Documents at Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Mcguire, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant |
---|
Category:NRC Information Notice
MONTHYEARInformation Notice 2020-02, Flex Diesel Generator Operational Challenges2020-09-15015 September 2020 Flex Diesel Generator Operational Challenges ML20225A0322020-09-0303 September 2020 NRC Choice Letter to NAC International with Attached Safety Inspection Report, IR 0721015/2020201, February 24-27, 2020 and July 22, 2020, Inspection of NAC International in Norcross, Georgia Information Notice 2012-09, PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs.2019-11-30030 November 2019 PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs. Information Notice 2011-20, NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011)2019-07-24024 July 2019 NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011) ML19196A2452019-07-15015 July 2019 Public Notice - Sequoyah Nuclear Plant, Unit 2 - Exigent Amendment to Facility Operating License Information Notice 2019-01, Inadequate Evaluation of Temporary Alterations2019-03-12012 March 2019 Inadequate Evaluation of Temporary Alterations ML16028A3082016-04-27027 April 2016 NRC Information Notice; IN 2016-05: Operating Experience Regarding Complications From a Loss of Instrumentation Air Information Notice 2015-05, Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps Information Notice 2015-05, Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps Information Notice 2013-20, OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-20, Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-11, OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2013-11, Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Contain2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Con2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notic2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2012-13, Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool Information Notice 2012-13, Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool Information Notice 2012-11, Age Related Capacitor Degradation2012-07-23023 July 2012 Age Related Capacitor Degradation ML12031A0132012-02-0606 February 2012 U.S. Nuclear Regulatory Commission Investigation Report No. 2-2010-058, Cpn International, Inc Information Notice 2011-19, Licensee Event Reports Containing Information Pertaining to Defects to Basic Components2011-09-26026 September 2011 Licensee Event Reports Containing Information Pertaining to Defects to Basic Components Information Notice 2011-15, Steel Containment Degradation and Associated License Renewal Aging Management Issues2011-08-0101 August 2011 Steel Containment Degradation and Associated License Renewal Aging Management Issues Information Notice 2011-17, Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping2011-07-26026 July 2011 Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping Information Notice 2011-13, Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-13, Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13)2011-06-29029 June 2011 Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13) Information Notice 2011-13, OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-04, IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-262010-12-21021 December 2010 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 ML13066A1872009-12-16016 December 2009 Draft NRC Information Notice 2009-xx - Underestimate of Dam Failure Frequency Used in Probabilistic Risk Assessments ML1007804482009-11-23023 November 2009 Email from Peter Bamford, NRR to Pamela Cowan, Exelon on TMI Contamination Control Event Information Notice 2009-11, NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-112009-07-0707 July 2009 NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-11 Information Notice 2009-10, Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10)2009-07-0707 July 2009 Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10) Information Notice 2009-09, Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify2009-06-19019 June 2009 Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify Information Notice 2008-12, Reactor Trip Due to Off-Site Power Fluctuation2008-07-0707 July 2008 Reactor Trip Due to Off-Site Power Fluctuation Information Notice 2008-11, Service Water System Degradation at Brunswicksteam Electric Plant Unit 12008-06-18018 June 2008 Service Water System Degradation at Brunswicksteam Electric Plant Unit 1 Information Notice 2008-04, Counterfeit Parts Supplied to Nuclear Power Plants2008-04-0707 April 2008 Counterfeit Parts Supplied to Nuclear Power Plants Information Notice 1991-09, Counterfeiting of Crane Valves2007-09-25025 September 2007 Counterfeiting of Crane Valves Information Notice 2007-28, Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls2007-09-19019 September 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls Information Notice 2007-29, Temporary Scaffolding Affects Operability of Safety-Related Equipment2007-09-17017 September 2007 Temporary Scaffolding Affects Operability of Safety-Related Equipment Information Notice 2007-14, Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station2007-03-30030 March 2007 Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station Information Notice 2007-06, Potential Common Cause Vulnerabilities in Essential Service Water Systems2007-02-0909 February 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Information Notice 2007-05, Vertical Deep Draft Pump Shaft and Coupling Failures2007-02-0909 February 2007 Vertical Deep Draft Pump Shaft and Coupling Failures Information Notice 2006-31, Inadequate Fault Interrupting Rating of Breakers2006-12-26026 December 2006 Inadequate Fault Interrupting Rating of Breakers Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear Information Notice 2006-13, E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination2006-07-13013 July 2006 E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination 2020-09-03 The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Mcguire]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>.
[Table view]The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Mcguire]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>. |
UNITED STATES
NUCLEAR REGULATORY COMMISSION
OFFICE OF NUCLEAR MATERIAL SAFETY AND SAFEGUARDS
WASHINGTON, D.C. 20555-0001 July 6, 1999 NRC INFORMATION NOTICE 99-23: SAFETY CONCERNS RELATED TO REPEATED
CONTROL UNIT FAILURES OF THE NUCLETRON
CLASSIC MODEL HIGH-DOSE-RATE REMOTE
AFTERLOADING BRACHYTHERAPY DEVICES
Addressees
authorized to use
All U.S. Nuclear Regulatory Commission (NRC) medical licensees
remote afterloaders.
brachytherapy sources in Nucletron Classic Model high-dose-rate (HDR)
Purpose
unit failures with Nucletron
NRC is issuing this information notice to alert you to ongoing control to your facilities
for applicability
Classic Model HDR devices. You should review this information patient
failures on
and consider actions, as appropriate, to minimize the impact of such
suggestions contained in this information notice
treatments and personnel exposures. However, response is required.
are not NRC requirements; therefore, no specific action nor written
Description of Circumstances
Classic Model HDR
By January 1996, NRC was aware of three incidents where the Nucletron
unit failures exposed a
control units failed (locked up) for unexplained reasons. These control
Classic Model HDR.
deficiency in the design of the door interlock circuitry for the Nucletron
and providing
Nucletron issued a Safety Alert on March 4, 1996, describing this problem
This safety alert was also issued
information on the proper operator response to such failures. discussed
10, 1996, that
as an attachment to NRC Information Notice 96-21, issued on April
these three reported
both the door interlock problem and the control unit failures. Following
field modifications to these devices to
failures of the control unit, Nucletron Corporation made
unexplained control unit
correct the door interlock failure mechanism and to eliminate the
failures.
four in 1998, and three
Following these modifications, two similar failures were reported in 1997, since the 1996 corrective
to date in 1999. These additional nine control unit failures, reported
However, no further
action, indicate a ongoing problem with unexplained control unit failures. has continued
interlocks. Nucletron
problems have been reported with the operation of the door cause of
the root
to investigate the ongoing control unit failures and now believes it has found
We understand that
the failures and subsequently developed appropriate corrective measures.
before
the newly developed corrective measures are presently undergoing testing
implementation.
c
0t rn4-] dwotaqq-O2s q4o1o
eas
-o4 cW
'-I
IN 99-23 July 6, 1999 Discussion:
were reported. In
In all 12 reported control unit failures no patient or personnel overexposures
when communications with the
each case the treatment unit continued to operate, as designed, unit continues the
control unit was lost. In this event, the microprocessor in the treatment
the failure occurs, halting upon
preprogrammed treatment to the catheter being treated when
or during the last catheter in
completion. If the failure occurs during a single catheter treatment Otherwise, it will
a multi-catheter operation, then the treatment will proceed to completion. catheters
any remaining
terminate with completion of the catheter being treated, leaving on an inoperative
display
untreated. Most users, however, when faced with a blank or frozen
Stop" button on the control
control unit, immediately abort the treatment, using the "Emergency
on the Safety Alert procedures)
unit. The "Treatment Interrupt" button (called for in the 1" step
Irrespective of the method
on the control unit is rendered inoperative by the control unit failure.
can subsequently be read from
of terminating the patient's treatment, the actual treatment data
the treatment unit, using a handheld terminal device.
of a control unit failure and, at
Most patients receive less than the prescribed dose in the event
must be revised to reflect the
the discretion of the Authorized User, either the Written Directive
after repair or replacement of
actual dose given or the remaining portion of the treatment given
unit failures has resulted in
the defective control unit. To date, none of the reported control
the control unit removes the first
overexposures or misadministrations. However, the failure of
of excessive radiation exposures
layer of radiation safety protection. At this point, the avoidance
unit or operator
depends upon either the continued proper operation of the treatment
intervention.
and included a copy of the
NRC previously issued Information Notice 96-21 on April 10, 1996, as an attachment. This Safety
Nucletron issued Safety Alert, addressing the control unit failures, the Control Unit stops
Alert sets forth a series of four steps to be taken if, for any reason, should ensure that the treatment
updating the status of the treatment in progress. These steps
enter the treatment room. You
is interrupted and the source retracted before licensee personnel
vendor, as contained in its
are encouraged to follow these recommendations from the device
to these control unit failures
Safety Alert of March 4, 1996. Any questions you may have related
should be addressed to your
and related corrective actions, or to get a copy of the Safety Alert, Nucletron Corporation representative.
IN 99-23 July 6, 1999 If you have any
This information notice requires no specific action nor written response.
technical contact listed below
questions about the information in this notice, please contact the
or the appropriate regional office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contact: Robert L. Ayres, NMSS
(301) 415-5746 E-mail: rxal nrc.gov
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
ItWt4 a,-TV
K-' I
Attachment 1 IN 99-23 July 6, 1999 LIST OF RECENTLY ISSUED
NMSS INFORMATION NOTICES
Information Date of
Subject Issuance Issued to
Notice No.
10 CFR 34.43(a)(1): Effective 6/25/99 Industrial Radiography Licensees
99-22 Date for Radiographer Certification
and Plans for Enforcement Discretion
6/25/99 All material and fuel cycle
99-20 Contingency Planning for the
Year 2000 Computer Problem licensees and certificate holders
Update on NRC's Year 2000 6/14/99 All material and fuel cycle
99-18 Activities for Materials Licensees licensees and certificate
and Fuel Cycle Licensees and holders
Certificate Holders
Federal Bureau of Investigation's 5/28 All U.S. Nuclear Regulatory
99-16- Nuclear Site Security Program Commission fuel cycle, power
reactor, and non-power reactor
licensees
Incident Involving the Use of 4/16199 All medical use licensees
99-11 Radioactive lodine-1 31 Problems Encountered When 3/24/99 All medical licensees authorized
99-09 to conduct high-dose-rate (HDR)
Manually Editing treatment Data
on the Nucletron Microselectron- remote after loading
HDR (New) Model 105-999 brachytherapy treatments
1998 Enforcement Sanctions as 3/19/99 All U. S. Nuclear Regulatory
99-06 Commission licensees
a Result of Deliberate Violations
of NRC Employee Protection
Requirements
Inadvertent Discharge of Carbon 3/8/99 All holders of licenses for nuclear
99-05 power, research, and test reactor, Dioxide Fire Protection System
and fuel cycle facilities
and Gas Migration
Unplanned Radiation Exposures 3/8/99 All radiography licensees
99-04 to Radiographers, Resulting From
Failures to Follow Proper Radiation
Safety Procedures
Exothermic Reactions Involving 1/29/99 All operating uranium recovery
99-03 Dried Uranium Oxide Powder facilities that produce oxide
(Yellowcake) powder (U308) (yellowcake)
v> Attachment 2 IN 99-23 July 6, 1999 LIST OF RECENTLY ISSUED
NRC INFORMATION NOTICES
Information Date of
Subject Issuance Issued to
Notice No.
99-22 10 CFR 34.43(a)(1); Effective 7/6/99 Industrial Radiography Licensees
Date for Radiographer Certification
and Plans for Enforcement Discretion
Recent Plant Events Caused 6/25/99 All holders of licenses for nuclear
99-21 power, test, and research reactors
By Human Performance Errors
6/25/99 All material and fuel cycle
99-20 Contingency Planning for the
Year 200 Computer Problem licensees and certificate holders
Rupture of the Shell Side of a 6/23/99 All holders of operating licenses
99-19 or construction permits for nuclear
Feedwater Heater at the Point
Beach Nuclear Plant power reactors
Update on NRC's Year 2000 6/14/99 All material and fuel cycle
99-18 licensees and certificate holders
Activities for Materials Licensees
and Fuel Cycle Licensees and
Certificate Holders
Problems Associated with Post-Fire 6/3/99 All holders of OL for nuclear
99-17 Safe-Shutdown Circuit Analyses power reactors, except those who
have permanently ceased
operations and have certified that
the fuel has been permanently
removed from the reactor
Federal Bureau of Investigation's 5/28 All U.S. Nuclear Regulatory
99-16- Nuclear Site Security Program Commission fuel cycle, power
reactor, and non-power reactor
licensees
Misapplication of 10 CFR Part 71 5/27/99 All holders of operating licenses or
99-15 construction permits for nuclear
Transportation Shipping Cask
Licensing Basis to 10 CFR Part 50 power reactors
Design Basis
OL = Operating License
CP = Construction Permit
-1,-23
, 6, 1999 response. If you have any
This information notice requires no specific action nor written
the technical contact listed below
questions about the information in this notice, please contact
or the appropriate regional office.
(ORIG. SIGNED BY)
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contact: Robert L. Ayres, NMSS
(301) 415-5746 E-mail: rxal nrc.gov
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
DOCUMENT NAME: A & G:%AYRES\IN99C-XX.WPD
MSIB* IN
OFFICE JMSIB C Editor* I MSIB* Ms N
EKraus FSturz LCamper DCool
NAME RAyres/liucy
6109 /99 6115/99 6/15/99 6/28/99 DATE 6/ 09 /99 n rn0PV
A4~w^=rloEIf
s
O
AI aMen
r K~~~AMLREGO~ I %L u
W
IN 99-XX
AJ June XX, 1999 This information notice requires no specific action nor written response. If you have any
questions about the information in this notice, please contact the technical contact listed below
or the appropriate regional office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contact: Robert L. Ayres, NMSS
(301) 415-5746 E-mail: rxal @nrc.gov /
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices A
</
/
DOCUMENT NAME: A & G:\AYRES\IN99C-XX.WPD
OFFICE MSIB IC Editor I MSIB* I lISIB l MSIB IN
NAME RAyrestic EKraus FSturz LCamper DC_______
DATE 6/ 09 /99 6/09 /99 6/15/99 6/15/99 6/.t1/99 OFFICIAL RECORD COPY
.z~ IN 99-XX
June xx, 1999 DOCUMENT NAME: A & G:\AYRES\IN99C-XX.WPD tfqff /
£fL411V5.--/0 :"//
OFFICE MSIB C Editor I MSIB MSIB N MSIB l NN/
NAME RAyres/jcy EKraus FSturz LCamper DCooI
,,
DATE 16/ 9 /99 6/ 09 /99 6/ IU< /99 6/ IS /99 16/ ,199 OFFICIAL RECORD COPY
I/
/'
,
,
Ii
6 I~- IN 99-XX
L/ June xx, 1999 DOCUMENT NAME: A & G:\AYRES\1N99C-XX.WPD
OFFICE _ SBIC Eio*I M~- S9fIN MI
NAME RAyres/kicy EKraus - erCool D
DATE 16/ 09 /99 6/09 /99 Is__ 199 6/
____99__ 6/ /99 OFFICIAL RECORD COPY
|
---|
|
list | - Information Notice 1999-01, Deterioration of High-Efficiency Particulate Air Filters in a Pressurized Water Reactor Containment Fan Cooler Unit (20 January 1999)
- Information Notice 1999-02, Guidance to Users on the Implementation of a New Single-Source Dose-Calculation Formalism and Revised Air-Kerma Strength Standard for Iodine-125 Sealed Sources (21 January 1999, Topic: Brachytherapy)
- Information Notice 1999-03, Exothermic Reactors Involving Dried Uranium Oxide Powder (Yellowcake) (29 January 1999, Topic: Brachytherapy)
- Information Notice 1999-04, Unplanned Radiation Exposures to Radiographers, Resulting from Failures to Follow Proper Radiation Safety Procedures (1 March 1999, Topic: Brachytherapy)
- Information Notice 1999-05, Inadvertent Discharge of Carbon Dioxide Fire Protection System and Gas Migration (8 March 1999, Topic: Brachytherapy)
- Information Notice 1999-06, 1998 Enforcement Sanctions as a Result of Deliberate Violations of NRC Employee Protection Requirements (19 March 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-06, 1998 Enforcement Sanctions As a Result of Deliberate Violations of NRC Employee Protection Requirements (19 March 1999, Topic: Enforcement Discretion)
- Information Notice 1999-07, Failed Fire Protection Deluge Valves & Potential Testing Deficiencies in Preaction Sprinkler Systems (22 March 1999, Topic: Safe Shutdown)
- Information Notice 1999-08, Urine Specimen Adulteration (26 March 1999, Topic: Brachytherapy)
- Information Notice 1999-09, Problems Encountered When Manually Editing Treatment Data on the Nucletron Microselectron-HDR (New) Model 105-999 (24 March 1999, Topic: Brachytherapy)
- Information Notice 1999-10, Degradation of Prestressing Tendon Systems in Prestresssed Concrete Containments (13 April 1999)
- Information Notice 1999-11, Incidents Involving the Use of Radioactive Iodine-131 (16 April 1999, Topic: Brachytherapy)
- Information Notice 1999-12, Year 2000 Computer Systems Readiness Audits (28 April 1999, Topic: Brachytherapy)
- Information Notice 1999-13, Insights from NRC Inspections of Low-and Medium-Voltage Circuit Breaker Maintenance Programs (29 April 1999, Topic: Brachytherapy)
- Information Notice 1999-14, Unanticipated Reactor Water Draindown at Quad Cities Unit 2, Arkansas Nuclear One Unit 2, & FitzPatrick (5 May 1999, Topic: Reactor Vessel Water Level, Brachytherapy)
- Information Notice 1999-15, Misapplication for 10CFR Part 71 Transportation Shipping Cask Licensing Basis to 10CFR Part 50 Design Basis (27 May 1999, Topic: Brachytherapy)
- Information Notice 1999-16, Federal Bureau of Investigation'S Nuclear Site Security Program (28 May 1999, Topic: Brachytherapy)
- Information Notice 1999-17, Problems Associated with Post-Fire Safe-Shutdown Circuit Analyses (3 June 1999, Topic: Hot Short, Safe Shutdown, Temporary Modification, Emergency Lighting, Fire Protection Program)
- Information Notice 1999-18, Update on Nrc'S Year 2000 Activities for Material Licensees and Fuel Cycle Licensees and Certificate Holders (14 June 1999, Topic: Brachytherapy)
- Information Notice 1999-19, Rupture of the Shell Side of a Feedwater Heater at the Point Beach Nuclear Plant (23 June 1999)
- Information Notice 1999-20, Contingency Planning for the Year 2000 Computer Problem (25 June 1999, Topic: Brachytherapy)
- Information Notice 1999-21, Recent Plant Events Caused by Human Performance Errors (25 June 1999, Topic: Probabilistic Risk Assessment)
- Information Notice 1999-22, 10CFR 34.43(a)(1); Effective Date for Radiographer Certification and Plans for Enforcement Discretion (25 June 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-23, Safety Concerns Related to Repeated Control Unit Failures of the Nucletron Classic Model High-Dose-Rate Remote Afterloading Brachytherapy Devices (6 July 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-24, Broad-Scope Licensees' Responsibilities for Reviewing and Approving Unregistered Sealed Sources and Devices (12 July 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-25, Year 2000 Contingency Planning Activities (10 August 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-26, Safety and Economic Consequences of Misleading Marketing Information (24 August 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-27, Malfunction of Source Retraction Mechanism in Cobalt-60 Teletherapy Treatment Units (2 September 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-28, Recall of Star Brand Fire Protection Sprinkler Heads (30 September 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-30, Failure of Double Contingency Based on Administrative Controls Involving Laboratory Sampling and Spectroscopic Analysis of Wet Uranium Waste (8 November 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-30, Failure of Double Contingency Based On Administrative Controls Involving Laboratory Sampling and Spectroscopic Analysis of Wet Uranium Waste (8 November 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-31, Operational Controls to Guard Against Inadventent Nuclear Criticality (17 November 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-31, Operational Controls To Guard Against Inadventent Nuclear Criticality (17 November 1999, Topic: Enforcement Discretion, Brachytherapy)
- Information Notice 1999-32, Effect of Year 2000 Issue on Medical Licenseess (17 December 1999, Topic: Brachytherapy)
- Information Notice 1999-32, Effect of Year 2000 Issue on Medical Licensees (17 December 1999, Topic: Brachytherapy, Overdose, Underdose)
- Information Notice 1999-33, Management Of Wastess Contaminated with Radioactive Materialss (21 December 1999, Topic: Brachytherapy)
- Information Notice 1999-33, Management of Wastes Contaminated with Radioactive Materials (21 December 1999, Topic: Brachytherapy)
- Information Notice 1999-33, Management Of Wastes Contaminated with Radioactive Materials (21 December 1999, Topic: Brachytherapy)
- Information Notice 1999-33, Management Of Wastes Contaminated With Radioactive Materials (21 December 1999, Topic: Brachytherapy)
- Information Notice 1999-34, Potential Fire Hazard in the Use of Polyalphaolefin in Testing of Air Filter (28 December 1999)
- Information Notice 1999-34, PotentialPotentialPotential FireFireFire HazardHazardHazard ininIn thetheThe UseUseUse ofofOf PolyalphaolefinPolyalphaolefinPolyalphaolefin ininIn TestingTestingTesting ofofOf AirAirAir FilterFilterFilter (28 December 1999)
|
---|