Change 0 to Procedure CP/O/B/8100/20, Chemistry Procedure for Determination of Ammonia
ML20080P598
Person / Time
Site: Catawba  
Issue date: 10/06/1983
From: Bolin C, Evans L, Tuckman M
DUKE POWER CO.
To:
Shared Package
ML20080P419 List:
References
CP-O-B-8100-20, NUDOCS 8402220593
Download: ML20080P598 (17)


Text

\\

Form 34731 (10-81)

(Formerly SPD-1002-1)

DUKE POWER COMPANY (1)

ID No: CP/0/B/8100/20 PROCEDURE PREPARATION Change (s) O to PROCESS RECORD D Incorporated (2) STATION:

Catawba (3) PROCEDURE TITLE:

Chemistry Procedure for the Determination of Ammonia (4) PREPARED BY:

/%MM DATE:

7 e29-fd (5) REVIEWED BY:

ebva DATE: /v - J- [3 Cross-Disciplinary Review By:

h Md (6) TEMPORARY APPROVAL (IF NECESSARY):

By:

(SRO) Date:

By:

Date:

l (7) APPROVED BY: N$.7h e e-

/ MC. Date:

/8"[*k3 1

t l

(8) MISCELLANEOUS:

l l

Reviewed / Approved By:

Date:

1 Reviewed / Approved By:

Date:

l MASTER FILE B402220593 840215 PDR ADOCK 05000413 PDR E

4

~

Frrm 34634 (4 81)

SPD-10012 DUKE POWER COMPANY NUCLEAR SAFETY EVALUATION CHECK LIST (1) STATION:

Catawba UNIT: 1 2

3 OTHER:

Shared (2) CHECK LIST APPLICABLE TO:

CP/0/B/8100/20

.(3) SAFETY EVALUATION - PART A The item to which this evaluation is applicable represents:

Yes No / A change to the station or procedures as described in the FS, or a test or experinent not described in the FSAR?

If the answer to the above is "Yes", attach a detailed description of the ite:

being evaluated and an identification of the affected section(s) of the FSAR.

(4) SAFETY EVALUATION - PART B

[ Will this item require a change to the station Technical Yes No Specifications?

If the answer to the above is "Yes," identify the specification (s) aff ected and/or attach the applicable pages(s) with the change (s) indicated.

(5) SAFETY EVALUATION - PART C As a result of the item to which this evaluation is applicable:

Yes No / Will the probability of an accident previously evaluated in the FSAR be increased?

Yes No

/ Will the consequences of an accident previously evaluated in the FSAR be increased?

Yes No / May the possibility of an accident which is different than any already evaluated in the FSAR be created?

Yes No

/ Will the probability of a malfunction of equipment important to safety previously evaluated in the FSAR be increased?

Yes No

/ Will the consequences of a malfunction of equipment important to safety previously evaluated in the FSAR i

be increased?

l Yes No

/ May the possibility of malfunction of equipment important to safety different than any already evaluated in the FSAR be created?

Will the margin of safety as defined in the bases to any Yes No Technical Specification be reduced?

If the answer to any of the preceding is "Yes", an unreviewed safety question is involved. Justify the conclusion that an unreviewed safety question is or is not involved. Attach additional pages as necessary.

)

(6) PREPARED BY:

MM DATE:

f-2 9 - M (7) REVIEWED BY:

J sCww DATE:

/o O (8) Page 1 of

/

Form 18855 (3-80)

DUKE POWER COMPANY ALARA EVALUATION CHECKLIST (1) Station Catawba Unit:

1 2

3 Other: Shared (2) Checklist Applicable to:

CP/0/B/8100/20 (3) ALARA Evaluation Check those items below which were considered applicable during the preparation and review of this document.

Flushing and draining were used to minimize source - strength and con-tamination levels prior to performing an operation.

Permanent and/or movable shielding was specified for reduction of levels.

Use of permanent or temporary local exhaust ventilation systems was used for control of airborne contamination.

Operation was designed to be completed with the least pra:ticable time spent in the radiation field.

Appropriate tools and equipment were specified for the operation to be performed.

The operation was designed considering the minimum ' number of people necessary for safe job completion.

Remote handling equipment and other special tools were specified to reduce external dose.

Contamination - control techniques were specified.

The operation was designed to be conducted in areas of as low an exposure as practicable.

Additional ALARA considerations were:

!i v/

ALARA Principles were not considered since the procedure did not involve work in a radiation area.

(5) Prepared by:

[2zdM Date i ~ c) 9 " [d (6) Reviewed by:

(%

Date

/o - T f)

CP/0/B/8100/20 DUKE P0kTR COMPANY CATAk'BA NUCIIAR STATION CHEMISTRY PROCEDURE FOR THE DETERMINATION OF AMMONIA 1.0 DISCUSSION 1.1 Scope This procedure describes the two methods (gas diffusion electrode and direct Nesslerization) for the determination of ammonia.

Ammonia is monitored in the Condensate System because it leads to the corrosion of copper metal.

At high temperatures and in the presence of oxygen, ammonia has the ability to dissolve copper. The moisture separators will be an area subject to attack due to its 90-10 Copper-Nickel piping.

?.

Principle 1.2.1 Gas Diffusion Electrode The ammonia electrode consists of a silver-silver chloride reference electrode and a glass pH electrode immersed in a filling solution of ammonium chloride. The electrodes and filling solution are separated from the sample solution by a hydrophobic gas-permeable men.brane. Dissolved ammenia in the sample solution can diffuse through the membrane until the partial pressure of ammonia is the same on both sides in any given sample. The partial pressure of ammonia will be proportional to its concentration.

Ammonia diffusing through the membrane dissolves in the internal filling solution where it reacts reversibly with water in the filling solution:

^

NH3+HO W

+ OH" Equation 1 2

4 The relationship between the concent ntions cf ammonia, ammonium and hydroxide is given by the following equation:

+

~

[NH ]

[0H ] = constant Equation 2 4

[NH ]

3

[0H']

Equation 3 or constant [NH ]

=

3

+

[NH;]

CP/0/B/8100/20 Peas 2 of 11 The internal filling solution contains ammonium chloride

+

at a sufficiently high level so tha: the [NH;] can be considered constant. This allows Equation 3 to be simplified as follows:

~

[0H ]

[NH ]

  • constant Equation 4

=

3

~

Equation 4 shows that the [0H ] in the filling solution is directly related to the [NH ] in the sample. Thus, when 3

the electrode is immersed in an alkaline test solution, ammonia can diffuse through the membrane, changing the

[NH ) in the filling solution.

3

~

This in turn changes the [0H ] in the filling solution which is monitored by the glass electrode, i.e. pH.

The potential of the glass electrode (with respect to the reference electrode) varies in a Nernstian manner with changes in the hydroxide level:

~

E=E

- S 1 g [0H ]

O Where:

S = the electrode slope (Enclosure 6.1).

Substitution in the above question shows that the electrode response to ammonia is Nernstian:

E=E

- S log [NH ]

g 3

1.2.2 Direct Nesslerization Ammonia can be determined in high purity water by direct Nesslerization. Direct Nesslerization is only applicable to condensate and domineralized water.

In some cases drinking water and highly purified wastewater may be used if errors 1 to 2 mg/L are acceptable.

Pre:reatment before direct Nesslerization with zine sulfate and alkali precipitates calcium, iron, magnesium i

and sulfide, which torm turbidity when treated with i

Nessler reagent. The floc also removes suspended matter and sometimes colored matter.

Nesslers reagent consists of an alkaline solution of mercuric iodide in potassium iodide.

It renc:s with ammonia according to the following equation:

2K2 [HgI;)

' 2 NH

  • NH E23+

4 3

2

CP/0/B/8100/20 s

- Pag 2 3 of 11 The ammonia reacts fairly rapidly, but not instantaneously, to form an orange-brown product which remains in colloidal suspension, but flocculates on long standing. A colormetric analysis is made before flocculation occurs.

Rochelle salt solution is added to inhibit the precipitation of residual calcium and magnesium in the presence of the.

Nessler reagent and for buffering action to maintain pH.

1.3 Precisions and Interferences l

1.3.1 Ammonia Probe The accuracy and precision of this procedure will be determined by Q-Sum data. Volatile amines (derivatives of ammonia) may interfere with the electrode selectivity.

Hydrazine up to 20 ppe does not interfere; however, > 50 ppe hydrazine shows a slight positive interference.

During S/G or Condensate System wet layup, the hydra:ine level is

> 75 ppe. Therefore, these samples must be diluted with Super Q Water to bring the hydrazine concentration below the interference level.

For solutions with a boric acid concentration of 13,500 ppm or greater, it becomes necessary to dilute the sample (1:1) with Super Q Water to i

achieve a pH of 11 to 13 with the addition of 1.0 ml.of alkaline reagent.

1.3.2 Nesslerization The accuracy and precision of this procedure will be l

determined by Q-Sum data.

Hydrazine, even at low concentrations, interferes.

Iron, magnesium and sulfides can form pr~cipitates that cloud e

the colormetric analysis.

Pretreatment can usually remove these, but turbidity in the original samples can affect results.

Boric Acid 5 1500 ppm does not interfere.

1.4 Limits and Precautions 1.4.1 Ammonia Probe l

The concentration range of this procedure is 0.1 ppm to 2 ppm ammonia.

The samples and standards must be adjusted to a pH 11 to 13.

The electrode must be held at a 20' angle with respect to the vertical to prevent air bubble entrapment utAer the electrode.

Avoid touching the electrode membrane during assembly or other times because this can affect its hydrophobic 7

properties.

Membrane failure is characterized by a shift in electrode potential, drift and poor response.

Failure may also be apparent on visual inspection as dark spots or discoloration of the membrane.

CP/0/B/8100/20

,Pcg3 4 of 11 When handling or mixing sodium hydroxide, personnel shall wear a labcoat, eye protection and gloves.

1.4.2 Nesslorination This method is good on water of high purity only and has a range from 0.02 mg/L to 5 mg/L cmmonia nitrogen.

This method is not valid in the presence of more than 1500 ppm boric acid.

The Nessler reagent is poisonous.

Gloves should be worn throughout the analysis to insure the reagent is not absorbed.

When handling or mixe.g sodium hydroxide, personnel shall wear a labcoat, eye y otection and gloves.

2.0 APPARATUS 2.1 Ammonia Probe 2.1.1 pH/ loc meter 2.1.2 Orion model 95-10-00 ammonia electrode 0

2.1.3 Electrode holder (electrode must be held at a 20' angle) 2.1.4 Multielectrode selector 2.1.5 Magnetic stirrer and teflon-coated stirring bars 2.1.6 Timer 2.1.7 250 ml beaker; one beaker for each standard sample 2.2 Direct Nesslerization 2.2.1 Spectrophotometer NOTE:

Allow to warm up for at least 20 minutes.

2.2.2 Two 100mm and two 10mm light path optically matched sample cells.

2.2.3 100 m1, 50 m1, 10 m1, 5 ml pipets (Class A) 2.2.4 pH meter 2.2.5 Automatic pipets with required tips 2.2.5.1 1000p1 pipet (1.00 ml) 2.2.3.2 500ul pipet (0.50 ml)

CP/0/B/8100/20 s

Pogo 5 of 11 2.2.5.3 250ul pipet (0.25 ml) 2.2.5.4 100ul pipet (0.10 ml) 2.2.6 M1111 pore filtration assembly 2.2.7 Filters, 0.45u 3.0 REAGENTS 3.1 Ammonia Electrode 3.1.1 Orion ammonia electrode filling solution.

3.1.2 Alkaline Reagent (10M NaOH)

In a 100 ml volumetric flask dilute 80.0 1.0 ml of commercially available 50*4 NaOH to 100 ml with Super Q Water.

3.1.3 Stock Standard Solution (170 ppm as NH )

3 In a 1000 ml volumetric flask, dissolve 0.5349 t 0.0005 grams of ammonium chloride (NH C1) in Super Q Water and 4

dilute to volume.

Prepare fresh monthly.

Store in a polyethylene bottle.

3.1.4 Electrode Storage Solution (0.03M NH )

3 In a 1000 mi volumetric flask, dissolve 2,6750 i 0.0005 grams of ammonium chloride in Super Q Water.

Prepare fresh solution every six (6) months.

3.2 Direct Nesslerization 3.2.'

Nessler Reagent Dissolve 100 t 1.0g HgI and 70 t 0.!g KI in 200 ml Super 2

Q Water. Add this mixture slowly while stirring to a cool solution of 160 t lg NaOH in 300 ml of Super Q Water.

Dilute this solution to 1 liter, then store for one day and filter.

Store in pyrex glassware out of sunlight.

If a precipitate torms, remake this solution. This solution is stable for one year.

CAUTION:

The Nessler reagent is toxic; gloves shall be worn to avoid ingestion.

CP/0/B/8100/20 s

Pag 2 6 of 11 3.2.2 Rochelle Salt Solution Dissolve 50 0.5g of potassium sodium tartrate etrahydrate, KNaC H,0

  • 4H 0 in 100 ml Super Q Water. To remove the
  • 6 3

ammonia usually present in the salt, boil off 30 ml of the solution. After cooling, dilute to 100 ml.

This solution is stable 90 days.

3.2.3 Zine Sulfate Solution, 10%

Dissolve 100 t 1.0g ZnSO, 7H O and dilute to 1 liter.

2 This solution is stable 90 days.

3.2.4 Stock Ammonium Solution, 1000 ppm NH

-N 3

Dissolve 3.819 0.001g of anhydrous NH C1, dried at 4

105' C for one (1) hour in water and dilute to 1,000 ml.

This solution is stable 6 mon:hs.

3.2.5 Standard Ammonium Solution, 10 ppm Dilute 10.00 mi stock solution to 1,000 ml with Super Q Water. This so.'ution is stable for 1 conth.

3.2.6 Sodium Hydroxide Solution Dissolve 240 i Ig of NaOH in 1 liter of Super Q Water.

4.0 PROCEDURE i

4.1 Ammonia Electrode 4.1.1 Preparation of Standards NOTE:

Two 0.85 ppm standards will be run daily (when i

in use) for Q-sum.

l Prepare a series of standards by diluting suitable volumes of the stock standard solution, Section 3.1.3, to 100 ml l

with Super Q Water in a volur.etric flask.

l ul of stock solution Cenc. orm 100 0.17 (170 ppb) 1000 1.70 (1700 ppb)

NOTE:

An alternate standard may be prepared by diluting 500ul of stock solution to 100 1.0 ml with Super Q Watar, this is equal to 0.85 ppm NH.

3 4.1.2 Preparation of the Electrode When initially placing the electrode into service, it should be assembled as shown in Enclosure 6.2.

CP/0/B/8100/20 Peg 2 7 of 11 NOTE:

The internal filling should be diluted 1 part filling solution to 9 parts Super Q Water.

The assembled elettrode is to be allowed to stand in 0.05M NH standard for at least one hour (preferably overnight) 3 prior to use. When the electrode is not in use, allow the electrode to stand ir. 0.05M NH standard. However, before 3

making measurements and between measurements, the probe should stand in pH 4 buffer for 4 - 10 minutes, until the reading has stabilized.

4.1.3 Calibration Set the pH meter function switch to "MVabs" and the electrode switen to the correct channel.

Allow samples and standards to adjust to room temperature.

4.1.3.1 Place 50 ml. of the 0.17 ppm NH standard in a 3

100 ml beake, with a stirring bar, on a magnetic stirrer.

Be sure the bar does not hit the electrode.

The electrode must be immersed one inch in the standard or sample.

o NOTE:

Rinse the electrode carefully with Super Q Water before placing in each solution.

4 1 3.2 Check to bs cure *here is not an air bubble under the memb.ne.

Adjust the stir rate until there is no vortex, then disengage the pH meter " Standby" pushbutton.

4.1.3.3 Add 1 ml alkaline reagent to the sample to adjust the pH between 11 and 13.

Immediately set the timer for 4 minutes.

I.1.3.4 After 4 minutes, set the 0.17 ppm NH solution 3

to read + 104 mv by adjusting the Channel calibration knob.

NOTE:

Do not adjust the calibration knob on the pH meter.

Let the probe stand in pH 4 buffer for several minutes before reading 1.7 ppm NH.

3 4.1.3.5 Repeat Sections 4.1.3.1, 4.1.3.2 and 4.1.3.3, substituting the 1.7 ppm NH solution for the 3

0.17 ppm NH solution.

3 4.1.3.6 After 4 minutes, check the 1.7 ppm NH solution J

mV reading. The mV reading should be + 45 mV @

25*C.

L._.

CP/0/B/8100/20 Pagi 8 of 11 NOTE:.1 shows the change in slope (mV) vs. temperature for a ten-fold change in the ammonia concentration.

If the reading is not + 45 mV, adjust the reading using the correct channel calibration knob of the electrode switch to reduce the error by one half.

EXAMPLES ERROR Case 1 - Actual reading:

'49 mV AmV Adjust to read:

+47 mV 2mV Case 2 - Actual reading:

43 mV 2mV Adjust to read:

44 mV 1mV By doing this, the reading for the 0.17 ppm standard will change as follows:

Case ~. from +104 mV to about +106 mV Case 2 from +104 mV to about +103 mV This reduces the error encountered over the concentration range to a minimum.

NOTE:

If the mV reading for the 1.7 ppm standard is not between +41 and +49 mV, either the membrane needs replacement or the sensing element needs cleaning (assuming the standard is correct).

1 To clean the sensing element, carefully wipe its surface with a soft tissue.

4.1.*

Unknown Concentrations (samples te be analyzed)

Use the same procedure as with the standard, as described in Sections 4.1.3.1 through 4.1.3.3, substituting the sample to be analyzed for the standards.

Read the mV response for each sample after 4 minutes have elapsed.

I Rinse the electrode thoroughly after it has been in any solution. Always place the electrode in pH 4 buffer several minutes before placing in the sample.

4.1.5 Calculations Record the sample analyzed and the corresponding mV reading for the sample.

Compare the mV recding to a graph correlating mV to ppb and record the results in ppb SE.

3 4.2 Direct Nesslerization 4.2.1 Prepare two standard curves by diluting the following

. amounts of 10 mg/L standard ammonia solution in Section

-3.6 in 100 mi volumetric flasks.

i1

CP/0/B/8100/20 Pag: 9 of 11 NOTE:

Generation of a standard curve is no: required if this method is in current use; however. four standard (two for the high range and two for the low range) will be run daily.

4.2.1.1 For standards with a concentration of 0.25 mg/L or greater, a 10mm sample cell will be used.

The following standards should be used in i

preparing the high standard graphs.

Amount of 10 mg/L Final Conc.

Standard in mi mg/L NH3 -N 100.0 10.0 50.0 5.0 10.0 1.0 5.0 0.5 4.2.1.2 For standards with a concentration of < 0.25 mg/L, a 100mm sample cell will be used. The following standards should be used in preparing the low standard graph:

Amount of 10 mg/L Final Conc.

Standard in mi mg/L NH3 -N 1.0 0.1 0.50 0.05 0.25 0.025 0.10 0.010 4.2.2 For sample blanks and standards with low turb!.dity, calcium, magnesium and iron; i.e. does not produce turbidity when the Nessler reagent is added, go directly to 4.2.3.

Otherwise, pretreat samples, as in 4.2.2.1 through 4.2.2.5.

4.2.2.1 Add 1 ml of the sinc sulfate solution (Section 3.2.3) to 100 ml of the sample, 4.2.2.2 Add 0.4 to 0.5 ml of the NaOH solution (Section 3.2.o) and mix to obtain a pH of 10.5 0.3 as determined with a pH meter.

4.2.2.3 Let the solutions stand :nree (3) minutes.

4.2.2.4 Filter the sample and collect 50 t 0.5 ml of the solution.

4.2.2.5 Add 100ul of the Rochelle sal: solution (See:Icn 3.2.2) to the samples and mix.

l t

CP/0/3/8100/20 Prg2 10 of 11 4.2.3 Add 1 ml of Nessler Reagent (Section 3.2.1) to 50 e1 of untreated samples, blank, and standsrds or the tre. ed samples from 4.2.2.5 in a 100 m1 beai sr.

Mix well.

4.0.4 Let stand for ten minutes.

4.2.5 Instrument Calibration 4.2.5.1 Adjust spectrophotometer to a wavelength of 420 nm.

Allow spectrophotometer to warm up 20 minutes.

4.2.3.2 Set the Mode Selector Knob to "Absorbance".

4.2.5.3 Pour the blank into a 10mm sample cell, place in spectrophotometer and set zero absorbance.

Be sure to clean all cell faces with Kimwipes or bibulous paper.

4.2.6 Pour the high standards and/or the unknown samples into the 100mm sample cells.

Place in the spectrophotometer in order of increasing concentrations and read the absorbance.

Check the zero absorbance with the reagent blank before each sample.

4.2.7 Prepare the high standard curve plotting absorbance vs.

concentration.

4.2.8 Determine the concentration of the unknown sample by comparing the ' sample absorbance with the standard curve.

NOTE:

If the concentration of the sample is below 0.25 mg/L NH., go to 4.2.9.

D 4.2.9 Pour the low setndards and/or the unknown samples containing < 0.25 mg/L NH into a 100mm sample cell.

3 l

Place in the spectrophotometer in order of increasing concentration and read the absorbance. Chech the zero absorbance with the reagent blank before each sample.

4.2.7 Prepare the low standard curve plotting absorbance vs.

concentration.

4.2.8 Determine the concentration of the unknown sample by comparing the sample absorbance with the standard curve.

50 REFERENCES 3.1 Standard Methods for the Examination of Water and Wastewater 14th ed., 1976 5.2 Steam Production Department System Power Chemistry Procedure CP/16 5.3 McGuire Nuclear Station Chemistry Procedure CP/0/B/8100/04 5.4 Orion Research 1978 Instruction Manual

CP/0/B/8100/20 Pcg3 11 of 11

. 6.0 ENCLO5URES 6.1 Relative Millivol: Change vs. Temperature 6.2 Construction of the Ammonia Electrode 6.3 Calibration Curve and Data for the Ammonia Electrode ennimaisio n.

ENCLOSURE 6.1 CP/0/B/8100/20 RELATIVE MILLIVOLT CHANGE WITH TIMPERATJFI i

Values for the Nernst factor, 2.3 RT/F, the change in electrode potential for a ten-fold change in the ammonia concentration of the sample as a functica of temperature.

Temp.

Slope (mv)

Temp.

Slope (mv)

Temp.

Slope (mv)

^

O'C 54.20 15 57.18 30 60.16 5

55.20 20 58.17 35 61.15 l

10 56.19 25 59.16 40 62.14

ENCLOSURE 6.2 CP/0/B/8100/20 CONSTRUCTION OF THE AMMONIA ELECTRODE l

l N

N N

Reference elemen:

Outer bodv-N s

~

internal filling solution N

N N

N Inner Body

.e.

N

,s g

\\

S-W d

0-ring sy

s Spacer Bottom cap Sensing elemen-

"embrane I

ASSEMBLY DIRECTIONS:

This electrode iY shipped dry and without a membrane in place.

When initially placing the electrode into service, it should be assembled and allowed to stand in 0.05 MNH3*

"I***

        • 1

"" (F#*'"#*

7 **

8 *}

use.

STEP 1:

Remove top cap.

Lift out inner body.

Pour out old internal filling solution. Remove bottom cap.

STEP 2:

Remove "O" ring and old membrane.

l STEP 3:

Place a new membrane in the electrode, with the " dull" side toward the inner sensing element and the " shiny" side down, towards the l

sample solution.

1 STEP 4:

In addition to the red "O" ring, a second black "0" ring is an integral part of the spacer, trapped by a small groove on the bottom surface.

When assembling, be sure that the black "0" ring is firmly in place in this groove.

Place the spacer in the cap top of the new membrane with the black "O" ring down. Place red "0" ring on top of spacer.

Screw body into bottom cap.

1 STEP 5:

Fill the outer body with the dilute internal filling solution 1 + 9 to about icm above the line between the body and the bottom cap.

When the inner body is inserted into the outer body, any excess filling solution will flow out of the vent hole.

STEP 6:

Put inner body into outer body.

Screw top cap on.

s

T ENCLOSURE 6.3 CP/0/B/6100/20 2

CALIBRA

'N CURVE FOR THE A"MONIA ELECTE 7e

?~

C s

0 k

s J

L i

m 3

-1 1_

1 s

I

-~_~

1 4

W h.

e s

W J d 4

f i

e

's "3

f i <

a 1s' e

1 I

I i

1 x

t 1

1 M =

s E

1 k

i I

F k

i 1

s l'

]

1

'l 3

1 I

i b

E.

A i

1 i.

N i

I l

I I

1 I

1 f

N

\\i i-i 5

h I'

I L

e i

i t

t sg li I

\\

I 4w I

t I

I

.7 &

h(

I 1

[

sx i ;

i

\\ 'l

- 1:

i 1

- a 4,

- i l

E i

1

?) o t

'i.

{

l l'ii htr l_

' I'

+

I:

i

)-

m m\\.

i i"

L i

i E E l

r O

L s

1 x

1 l

T s

1 m

a 1

3 L

N t

I t

i

  • =-r

, x

-n

- z

=

=

C w

f I"

F 2

L

\\sK

=

i

{

x, 1

t I

r 1

I I

I k

1 7

=

s

{

I

\\

s i

\\

l 1

I i

i x.

E 1

1 i

e \\i I

I I

f l

k f

i i

!\\

N H

i k

i 6

! b L

l i

i i

i W i

i l

i

\\ i i

a i

=r i

t i

i

.\\l i

M i

i i.

i i

l.

I r

I i

a I

i I

!\\

p 1

o

}

l l

l l

l (

6 S

t l

}

N s

i i

I i

i l

i i

i i

I 0

0 C

C C

C 1

=

=

=

=

=

=

=

=

=

N a

c x

=

~

m f

(aanIosce) m

=

F

.