Recent Misadministrations Caused by Incorrect Calibrations of Strontium-90 Eye ApplicatorsML031050494 |
Person / Time |
---|
Site: |
Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant ![Entergy icon.png](/w/images/7/79/Entergy_icon.png) |
---|
Issue date: |
12/13/1996 |
---|
From: |
Cool D NRC/NMSS/IMNS |
---|
To: |
|
---|
References |
---|
IN-96-066, NUDOCS 9612090360 |
Download: ML031050494 (15) |
|
Similar Documents at Beaver Valley, Millstone, Hatch, Monticello, Calvert Cliffs, Dresden, Davis Besse, Peach Bottom, Browns Ferry, Salem, Oconee, Nine Mile Point, Palisades, Palo Verde, Perry, Indian Point, Fermi, Kewaunee, Catawba, Harris, Wolf Creek, Saint Lucie, Point Beach, Oyster Creek, Watts Bar, Hope Creek, Grand Gulf, Cooper, Sequoyah, Byron, Pilgrim, Arkansas Nuclear, Three Mile Island, Braidwood, Susquehanna, Summer, Prairie Island, Columbia, Seabrook, Brunswick, Surry, Limerick, North Anna, Turkey Point, River Bend, Vermont Yankee, Crystal River, Haddam Neck, Ginna, Diablo Canyon, Callaway, Vogtle, Waterford, Duane Arnold, Farley, Robinson, Clinton, South Texas, San Onofre, Cook, Comanche Peak, Yankee Rowe, Maine Yankee, Quad Cities, Humboldt Bay, La Crosse, Big Rock Point, Rancho Seco, Zion, Midland, Bellefonte, Fort Calhoun, FitzPatrick, McGuire, LaSalle, Fort Saint Vrain, Shoreham, Satsop, Trojan, Atlantic Nuclear Power Plant |
---|
Category:NRC Information Notice
MONTHYEARInformation Notice 2020-02, Flex Diesel Generator Operational Challenges2020-09-15015 September 2020 Flex Diesel Generator Operational Challenges ML20225A0322020-09-0303 September 2020 NRC Choice Letter to NAC International with Attached Safety Inspection Report, IR 0721015/2020201, February 24-27, 2020 and July 22, 2020, Inspection of NAC International in Norcross, Georgia Information Notice 2012-09, PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs.2019-11-30030 November 2019 PWROG-16043-NP-A, Revision 2, PWROG Program to Address NRC Information Notice 2012-09: Irradiation Effects on Fuel Assembly Spacer Grid Crush Strength for Westinghouse and CE PWR Fuel Designs. Information Notice 2011-20, NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011)2019-07-24024 July 2019 NRC060 - NRC Information Notice 2011-20: Concrete Degradation by Alkali-Silica Reaction (Nov. 18, 2011) ML19196A2452019-07-15015 July 2019 Public Notice - Sequoyah Nuclear Plant, Unit 2 - Exigent Amendment to Facility Operating License Information Notice 2019-01, Inadequate Evaluation of Temporary Alterations2019-03-12012 March 2019 Inadequate Evaluation of Temporary Alterations ML16028A3082016-04-27027 April 2016 NRC Information Notice; IN 2016-05: Operating Experience Regarding Complications From a Loss of Instrumentation Air Information Notice 2015-05, Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability of Auxiliary and Emergency Feedwater Auto Start Circuits on Loss of Main Feedwater Pumps Information Notice 2015-05, Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps2015-05-12012 May 2015 Inoperability Of Auxiliary And Emergency Feedwater Auto Start Circuits On Loss Of Main Feedwater Pumps Information Notice 2013-20, OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 OFFICIAL EXHIBIT - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-20, Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143)2013-10-0303 October 2013 Official Exhibit - NYS000538-00-BD01 - NRC Information Notice 2013-20: Steam Generator Channel Head and Tubesheet Degradation (October 3, 2013) (ML13204A143) Information Notice 2013-11, OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 OFFICIAL EXHIBIT - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2013-11, Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013)2013-07-0303 July 2013 Official Exhibit - NYS000551-00-BD01 - NRC Information Notice 2013-11: Crack-Like Indication at Dents/Dings and in the Freespan Region of Thermally Treated Alloy 600 Steam Generator Tubes (July 3, 2013) Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Contain2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Con2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend and/or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2010-12, Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notic2012-08-17017 August 2012 Intervenors' Fifth Motion to Amend And/Or Supplement Proposed Contention No. 5 (Shield Building Cracking). Appendix VI: NRC FOIA Responses (B-51 Through B-53); Turkey Point Event Report; NRC Information Notice 2010-12: Containment Liner Cor Information Notice 2012-13, Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs and Corrective Actions in the Spent Fuel Pool Information Notice 2012-13, Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool2012-08-10010 August 2012 Boraflex Degradation Surveillance Programs And Corrective Actions In The Spent Fuel Pool Information Notice 2012-11, Age Related Capacitor Degradation2012-07-23023 July 2012 Age Related Capacitor Degradation ML12031A0132012-02-0606 February 2012 U.S. Nuclear Regulatory Commission Investigation Report No. 2-2010-058, Cpn International, Inc Information Notice 2011-19, Licensee Event Reports Containing Information Pertaining to Defects to Basic Components2011-09-26026 September 2011 Licensee Event Reports Containing Information Pertaining to Defects to Basic Components Information Notice 2011-15, Steel Containment Degradation and Associated License Renewal Aging Management Issues2011-08-0101 August 2011 Steel Containment Degradation and Associated License Renewal Aging Management Issues Information Notice 2011-17, Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping2011-07-26026 July 2011 Calculation Methodologies for Operability Determinations of Gas Voids in Nuclear Power Plant Piping Information Notice 2011-13, Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 Official Exhibit - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-13, Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13)2011-06-29029 June 2011 Official Exhibit - Nys000329-00-Bd01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (Nrc in 2011-13) Information Notice 2011-13, OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13)2011-06-29029 June 2011 OFFICIAL EXHIBIT - NYS000329-00-BD01 - NRC Information Notice 2011-13, Control Rod Blade Cracking Resulting in Reduced Design Lifetime (Jun 29, 2011) (NRC in 2011-13) Information Notice 2011-04, IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 IN: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 In: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2011-04, in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors2011-02-23023 February 2011 in: Contaminants and Stagnant Conditions Affecting Stress Corrosion Cracking in Stainless Steel Piping in Pressurized Water Reactors Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review2010-12-30030 December 2010 New England Coalition'S Motion for Leave to Reply to NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 and Entergy'S Response to the Supplement to Nec'S Petition for Commission Review of LBP-10-19 Information Notice 2010-26, 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-262010-12-21021 December 2010 2010/12/21-NRC Staff'S Objection to Nec'S Notification of Information Notice 2010-26 ML13066A1872009-12-16016 December 2009 Draft NRC Information Notice 2009-xx - Underestimate of Dam Failure Frequency Used in Probabilistic Risk Assessments ML1007804482009-11-23023 November 2009 Email from Peter Bamford, NRR to Pamela Cowan, Exelon on TMI Contamination Control Event Information Notice 2009-11, NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-112009-07-0707 July 2009 NSP000059-Revised Prefiled Testimony of Northard/Petersen/Peterson-NRC Information Notice 2009-11 Information Notice 2009-10, Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10)2009-07-0707 July 2009 Official Exhibit - NYS000019-00-BD01- NRC Information Notice 2009-10, Transformers Failures - Recent Operating Experience (Jul. 7, 2009) (NRC in 2009-10) Information Notice 2009-09, Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify2009-06-19019 June 2009 Improper Flow Controller Settings Renders Injection Systems Inoperable and Surveillance Did Not Identify Information Notice 2008-12, Reactor Trip Due to Off-Site Power Fluctuation2008-07-0707 July 2008 Reactor Trip Due to Off-Site Power Fluctuation Information Notice 2008-11, Service Water System Degradation at Brunswicksteam Electric Plant Unit 12008-06-18018 June 2008 Service Water System Degradation at Brunswicksteam Electric Plant Unit 1 Information Notice 2008-04, Counterfeit Parts Supplied to Nuclear Power Plants2008-04-0707 April 2008 Counterfeit Parts Supplied to Nuclear Power Plants Information Notice 1991-09, Counterfeiting of Crane Valves2007-09-25025 September 2007 Counterfeiting of Crane Valves Information Notice 2007-28, Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls2007-09-19019 September 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Due to Inadequate Chemistry Controls Information Notice 2007-29, Temporary Scaffolding Affects Operability of Safety-Related Equipment2007-09-17017 September 2007 Temporary Scaffolding Affects Operability of Safety-Related Equipment Information Notice 2007-14, Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station2007-03-30030 March 2007 Loss of Offsite Power and Dual-Unit Trip at Catawba Nuclear Generating Station Information Notice 2007-06, Potential Common Cause Vulnerabilities in Essential Service Water Systems2007-02-0909 February 2007 Potential Common Cause Vulnerabilities in Essential Service Water Systems Information Notice 2007-05, Vertical Deep Draft Pump Shaft and Coupling Failures2007-02-0909 February 2007 Vertical Deep Draft Pump Shaft and Coupling Failures Information Notice 2006-31, Inadequate Fault Interrupting Rating of Breakers2006-12-26026 December 2006 Inadequate Fault Interrupting Rating of Breakers Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves as a Result of Stem Nut Wear Information Notice 2006-29, Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear2006-12-14014 December 2006 Potential Common Cause Failure of Motor-operated Valves As a Result of Stem Nut Wear Information Notice 2006-13, E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination2006-07-13013 July 2006 E-mail from M. Mclaughlin on NRC, Regarding NRC Information Notice 2006-13: Groundwater Contamination 2020-09-03 The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>.
[Table view]The following query condition could not be considered due to this wiki's restrictions on query size or depth: <code> [[:Beaver Valley]] OR [[:Millstone]] OR [[:Hatch]] OR [[:Monticello]] OR [[:Calvert Cliffs]] OR [[:Dresden]] OR [[:Davis Besse]] OR [[:Peach Bottom]] OR [[:Browns Ferry]] OR [[:Salem]] OR [[:Oconee]] OR [[:Nine Mile Point]] OR [[:Palisades]] OR [[:Palo Verde]] OR [[:Perry]] OR [[:Indian Point]] OR [[:Fermi]] OR [[:Kewaunee]] OR [[:Catawba]] OR [[:Harris]] OR [[:Wolf Creek]] OR [[:Saint Lucie]] OR [[:Point Beach]] OR [[:Oyster Creek]] OR [[:Watts Bar]] OR [[:Hope Creek]] OR [[:Grand Gulf]] OR [[:Cooper]] OR [[:Sequoyah]] OR [[:Byron]] OR [[:Pilgrim]] OR [[:Arkansas Nuclear]] OR [[:Three Mile Island]] OR [[:Braidwood]] OR [[:Susquehanna]] OR [[:Summer]] OR [[:Prairie Island]] OR [[:Columbia]] OR [[:Seabrook]] OR [[:Brunswick]] OR [[:Surry]] OR [[:Limerick]] OR [[:North Anna]] OR [[:Turkey Point]] OR [[:River Bend]] OR [[:Vermont Yankee]] OR [[:Crystal River]] OR [[:Haddam Neck]] OR [[:Ginna]] OR [[:Diablo Canyon]] OR [[:Callaway]] OR [[:Vogtle]] OR [[:Waterford]] OR [[:Duane Arnold]] OR [[:Farley]] OR [[:Robinson]] OR [[:Clinton]] OR [[:South Texas]] OR [[:San Onofre]] OR [[:Cook]] OR [[:Comanche Peak]] OR [[:Yankee Rowe]] OR [[:Maine Yankee]] OR [[:Quad Cities]] OR [[:Humboldt Bay]] OR [[:La Crosse]] OR [[:Big Rock Point]] OR [[:Rancho Seco]] OR [[:Zion]] OR [[:Midland]] OR [[:Bellefonte]] OR [[:Fort Calhoun]] OR [[:FitzPatrick]] OR [[:McGuire]] OR [[:LaSalle]] OR [[:Fort Saint Vrain]] OR [[:Shoreham]] OR [[:Satsop]] OR [[:Trojan]] OR [[:Atlantic Nuclear Power Plant]] </code>. |
UNITED STATES
NUCLEAR REGULATORY COMMISSION
OFFICE OF NUCLEAR MATERIAL SAFETY AND SAFEGUARDS
WASHINGTON, D.C. 20555 December 13, 1996 NRC INFORMATION NOTICE 96-66: RECENT MISADMINISTRATIONS CAUSED BY
INCORRECT CALIBRATIONS OF STRONTIUM-90
EYE APPLICATORS
Addressees
to use
All U.S. Nuclear Regulatory Commission Medical Use Licensees authorized
strontium-90 (Sr-90) eye applicators.
Purpose
The U.S. Nuclear Regulatory Commission (NRC) is issuing this information
notice to alert addressees to recent misadministrations caused by incorrect that
source strength determinations of Sr-90 eye applicators. It is expected
facilities
recipients will review the information for applicability to their
and consider actions, as appropriate, to avoid similar problems. However, suggestions contained in this information notice are not NRC requirements;
therefore, no specific action nor written response is required.
DescriDtion of Circumstances
The primary causes of two recent events have been problems with calibration the
and calculation of the dose rates from Sr-90 applicators. As a result, administered doses were not within 20 percent of the prescribed dose.
to use
Case 1. In October 1995, during an inspection of a licensee authorized the
a Sr-90 eye applicator in Mayag~ez, Puerto Rico, it was determined that A
calibration record for the licensee's Sr-90 eye applicator was missing.
and
previous owner of the source lost the original calibration certificate, This
his medical physics consultant performed a check of the source strength. resulted
check was performed with an inappropriate measurement instrument and 0
in an incorrect determination of the source strength as 0.24 Gray (Gy)/sec the
(24 rad/sec). The licensee obtained the source in 1994 and relied on source
erroneous source strength determination during treatments. Also, the
had not been decay-corrected by either owner since 1990. NRC urged the Fc2 licensee to have the eye applicator recalibrated and to be instructed in the
D
proper method to calculate the decay of the source.
National
The licensee subsequently had the eye applicator recalibrated by the a
Institute of Standards and Technology. The resultant recalibration revealed used
dose rate of 0.53 Gray (53 rad)/sec rather than the 0.24 Gy (24 rad)/sec to
10
by the licensee. The licensee had prescribed a total dose of between
25 Gy (1000 to 2500 rad) for approximately
9-0709036 f arr&0 C. Ilt, 2a
Dr
IN 96-66 December 13, 1996 70 patients, but actually administered about twice this amount. An NRC
medical consultant has reviewed the cases and has recommended that the
patients be evaluated for any adverse effects.
Case 2. In November 1995, while reviewing treatment records during a routine
NRC inspection of a licensee authorized to use a Sr-90 eye applicator in
Honolulu, Hawaii, an NRC inspector determined that the licensee had
incorrectly calculated the decay of the strength of a Sr-90 eye applicator and
subsequently delivered doses 20 percent over the prescribed doses to 16 patients. The root cause of this event was the licensee improperly
calculating the decay of the Sr-90 source. Specifically, the licensee
linearly decayed the strength of the source instead of logarithmically
decaying the source strength of the eye applicator. Before May 1995, a
technologist had correctly calculated the decay of the source; however, when
this technologist left, the physician incorrectly calculated the source decay
based on a linear decay from the previous calculations. In addition, the
incorrect half-life of Sr-90 was used in the calculation.
Following a correct determination of the source strength, the licensee
determined that 17 misadministrations involving 16 patients occurred, from
May 6, 1995 to November 16, 1995, including one patient who was treated on
both eyes. The delivered doses range from 21.1 to 22.7 percent greater than
the prescribed dose of 40 Gy (4000 rad).
Discussion
10 CFR 35.32, *Quality Management Program," requires licensees to develop
written procedures to meet five objectives associated with any brachytherapy
dose, including assurance that the prescribed dose is the administered dose.
A necessary part of this is to ensure that the dose rate emitted from an
applicator is correct. If the manufacturer's certificate of calibration or
original activity/dose rate name plate is missing, the licensee should arrange
with a qualified expert to determine the dose rate from the Sr-90 source.
(For additional details, see NRC IN 94-17, NSr-90 Eye Applicators: Submission
of Quality Management Plan, Calibration and Use," dated March 11, 1994.)
Failure to implement the Quality Management Rule and make required reports may
subject a licensee to orders, civil penalties, and notices of violation.
In view of the nature of the cited events, it appears worthwhile to review
some of the properties of Sr-90 eye applicators and the related processes for
decay correction over time. New Sr-90 eye applicators typically contain a
2 gigabecquerel (GBq) [54 millicurie(mCi)] source, exhibiting a surface dose
rate of about 0.50 Gy (50 rad)/sec. The half-life of the parent Sr-90 is
28.5 yrs [maximum beta energy equal to 0.54 mega-electron volts (MeY)], and
the yttrium-90 (Y-90) daughter half-life is 64.2 hrs (beta-max, 2.27 MeV);
therefore, both isotopes are in equilibrium on the eye applicator. Since
Sr-90 and Y-90 are in equilibrium, emissions from both isotopes must be
accounted for in dosimetry calculations.
IN 96-66 December 13, 1996 rate
The dose rate (D,) at a time (t) can be calculated from the initial dose
(De) at the time of the most recent calibration with the following formula:
D = D - (e-^t) (FI)
-Xt
where e, the base of the natural logarithm (2.718), is raised to the powerT. is
and is referred to as the 'decay factor' (df), where -X - -0.693/Tx, the half-life of the isotope, and t is the elapsed time. For Sr-90, TX-28.5 can
years; therefore, X - (0.693)/(28.5 yrs) = 0.0243 yr'. The values for df
after t
be used to determine the fraction of original activity remaining
years. These values can then be used to calculate dose rates by using the
following formula:
D,= Dao df (F2)
The fraction of activity remaining after a given number of years from the
original measurement date is given in Table 1 (Attachment 1).
For example
An eye applicator calibrated by the manufacturer on January 1, 1978, If, delivered
on contact. on
an initial dose rate ( Do) of 0.75 Gy/sec (75 rad/sec)
(in
January 1, 1996, the eye applicator needs to be used, the elapsed time
years) since calibration is 18, and from Table 1, df = 0.646 for Sr-90.
Using formula F2:
D,= bo. df
D18yaM - (0.75 Gy/sec)(0.646)
bigyea - 0.485 Gy/sec
(48.5 rad/sec)
IN 96-66 December 13, 1996 Table I may be extended using formula F1 and the value X - 0.0243 yr' for
Sr-90. Other values of df that need to be evaluated by licensees will need to
be re-calculated by using the appropriate half-life for that isotope.
Alternatively, a semi-logarithmic plot of the above data will yield a straight
line, which may be extended beyond 25 years.
This information notice requires no specific action nor written response. If
you have any questions about the information in this notice, please contact
one of the technical contacts listed below or the appropriate regional office.
Donald A. Cool, D rector
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 E-mail: jxd2@nrc.gov
Emilio M. Garcia, RIV
(510) 975-0239 E-mail: emg@nrc.gov
James A. Smith, NMSS
(301) 415-7904 E-mail: jas4@nrc.gov
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
42 T7SSI (/4e
Attachment I
IN 96-66 December 13, 1996 Page 1 of I
TABLE 1 FRACTION (EXPRESSED AS DECIMAL) OF ORIGINAL
SR-90 ACTIVITY REMAINING AFTER (t) YEARS
df Years (t) df Years (t) df Years (t) df
Years (t)
0.994 6.5 0.854 12.75 0.734 19 0.63
.25
6.75 0.849 13 0.729 19.25 0.626
.5 0.988
0.982 7 0.844 13.25 0.725 19.5 0.623
.75
0.976 7.25 0.838 13.5 0.72 19.75 0.619
1
0.97 7.5 0.833 13.75 0.716 20 0.615
1.25
0.964 7.75 0.828 14 0.712 20.25 0.611
1.5
0.958 8 0.823 14.25 0.707 20.5 0.608
1.75
0.953 8.25 0.818 14.5 0.703 20.75 0.604
2
0.947 8.5 0.813 14.75 0.699 21 0.6
2.25
0.941 8.75 0.808 15 0.695 21.25 0.597
2.5
0.935 9 0.804 15.25 0.69 21.5 0.593
2.75
0.93 9.25 0.799 15.5 0.686 21.75 0.589
3
0.924 9.5 0.794 15.75 0.682 22 0.586
3.25
9.75 0.789 16 0.678 22.25 0.582
3.5 0.918
10 0.784 16.25 0.674 22.5 0.579
3.75 0.913
10.25 0.78 16.5 0.67 22.75 0.575
4 0.907
0.902 10.5 0.775 16.75 0.666 23 0.572
4.25
0.896 10.75 0.77 17 0.662 23.25 0.568
4.5
11 0.765 17.25 0.658 23.5 0.565
4.75 0.891
11.25 0.761 17.5 0.654 23.75 0.562
5 0.886
11.5 0.756 17.75 0.65 24 0.558
5.25 0.88
0.875 11.75 0.752 18 0.646 24.25 0.555
5.5
0.87 12 0.747 18.25 0.642 24.5 0.551
5.75
0.864 12.25 0.743 18.5 0.638 24.75 0.548
6
0.859 12.5 0.738 18.75 0.634 25 0.545
6.25
Attachment 2 IN 96-66 December 13, 1996 Page 1 of I
UST OF RECENTLY ISSUED
NMSS INFORMATION NOTICES
Information Date of
Notice No. Subject Issuance Issued to
96-63 Potential Safety Issue 12/05196 All U.S. Nuclear Regulatory
Regarding the Shipment Commission licensees
of Fissile Material authorized to possess special
nuclear material in unsealed
quantities greater than a
critical mass
96-57 Incident-Reporting Require- 10/30196 All U.S. Nuclear Regulatory
ments Involving Intakes, Commission licensees
During a 24-Hour Period
that May Cause a Total
Effective Dose Equivalent
in Excess of 0.05 Sv (5 rem)
96-54 Vulnerability of Stainless 10/17/96 All material licensees
Steel to Corrosion When
Sensitized
96-53 Retrofit to Amersham 660 10115/96 All industrial radiography
Posilock Radiography licensees
Camera to Correct Incon- sistency in 10 CFR Part 34 Compatibility
96-52 Cracked Insertion Rods 09/26/96 All U.S. Nuclear Regulatory
on Troxler Model 3400 Commission portable gauge
Series Portable Moisture licensees and vendors
Density Gauges
96-51 Residual Contamination 09/11196 All material licensees
Remaining in Krypton-85 Handling System After
Venting
' >
Attachment 3 IN 96-66 December 13, 1996 Page 1 of I
UST OF RECENTLY ISSUED
NRC INFORMATION NOTICES
Information Date of
Notice No. Subject Issuance Issued to
96-65 Undetected Accumulation 1211 1/96 All holders of OLs
of Gas in Reactor Coolant or CPs for nuclear
System and Inaccurate power reactors
Reactor Water Level
Indication During Shutdown
96-64 Modifications to Con- 12/10196 All holders of OLs
tainment Blowout Panels or CPs for nuclear
Without Appropriate reactors
Design Controls
96-63 Potential Safety Issue 12/05196 All U.S. Nuclear
Regarding the Shipment Regulatory Commission
of Fissile Material licensees authorized
to possess special
nuclear material in
unsealed quantities
greater than a critical
mass
96-62 Potential Failure of the 11120196 All holders of OLs
Instantaneous Trip Function and CPs for nuclear
of General Electric RMS-9 power plants
Programmers
96-61 Failure of a Main Steam Safety 11/20/96 All holders of OLs or
Valve to Reseat Caused by an CPs for nuclear power
Improperly Installed Release . reactors
Nut
96-60 Potential Common-Mode Post- 11/14/96 All holders of OLs or
Accident Failure of Residual CPs for nuclear power
Heat Removal Heat Exchangers reactors
OL = Operating License
CP = Construction Permit
IN 96-66
. December 13, 1996 Table 1 may be extended using formula FH and the value x - 0.0243 yr1 for
Sr-90. Other values of df that need to be evaluated by licensees will need to
be re-calculated by using the appropriate half-life for that isotope.
Alternatively, a semi-logarithmic plot of the above data will yield a straight
line, which may be extended beyond 25 years.
This information notice requires no specific action nor written response. If
you have any questions about the information in this notice, please contact
one of the technical contacts listed below or the appropriate regional office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 E-mail: jxd2@nrc.gov
Emilio M. Garcia, RIV
(510) 975-0239 E-mail: emg@nrc.gov
James A. Smith, NMSS
(301) 415-7904 E-mail: jas4enrc.gov
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
Official Record Copy S:\Jsmith\imns5350.JAS
OFC IMAB I E IMAB lIMAB I IMOB I Tech Ed I
KANE JASmith SLMerchant LWCamper o JMPiccone o EKraus o
DATE 1 11/26/96 11/26/96 110/28/96 10/29/96 8/01/96 OFC J IJE OGC I DD/IMNS I D/IMNS I
NAME JLiebermano SATreby o FCCombso DACool
DATE 08/07/96 11/07/96 11/12/96 I 11/29/96
- COWVER E = COVER & Attachment N - NO COPY o- See Previous Concurrence
eIN 96-##
'J November .1996 Table 1 may be extended using formula F1 and the value A = 0.0243 yr" for
Sr-90. Other values of df that need to be evaluated by licensees will need to
be re-calculated by using the appropriate half-life for that isotope.
Alternatively, a semi-logarithmic plot of the above data will yield a straight
line, which may extended beyond 25 years.
This information notice requires no specific action nor written response. If
you have any questions about the information in this notice, please contact
one of the technical contacts listed below or the appropriate regional office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 Internet: JXD2@NRC.GOV
Emilio M. Garcia, RIV
(510) 975-0239 Internet: EMG@NRC.GOV
James A. Smith, NMSS
(301) 415-7904 Internet:JAS4@NRC.GOV
Attachments:
1. Table of decay factors
2. List of Recently Issue(d NMSS Information Notices
3. List of Recently Issue(d NRC Information Notices
fffirial Rernrd COnY Svisrnith\imns5350JAS
OFC IMAB lE IMAB lIMAB lIMOB j Tech Ed
NAME JASmi SLM@f@I~nt LWCamper o JMPiccone o EKraus @
DATE it /96 i 10/8/
9 6 10/29/96 8/01/96 OFC OE I OGC DD/IMNS DlIMNS
NAME JLieberman@ SATreby @ FCCombs@ _____ __ _ _l
DATE 08/07/96 11/07/96 11/12/96 ///t11/96
= COVER E = COVER & Attachment N = NO COPY ?--See Previous Concurrence
N 96-Il
-November 91996 This information notice requires no specific action nor written respons . If
you have any questions about the information in this notice, please c tact
one of the technical contacts listed below or the appropriate reglo 1 office.
Donald A. Cool, Director, Division of Industrial and
Medical Nuclear Sa ty
Office of Nuclear terial Safety
and Safeguards
Contacts: Jose M. Diaz-Velez, RI
(404) 331-7438 Internet: JXD2@NRC.GOV
Emilio M. Garcia, RIV
(510) 975-0239 Internet: EMG@NRC.GOV
James A. Smith, NMSS
(301) 415-7904 Internet:JAS4@NRC.GOV
Attachments:
1. Table of decay factors
2. List of Recently Issued NMS Information Notices
3. List of Recently Issued MR Information Notices
Official Record Copy 5: \smitinins5350.JAS
OFC IMAB I E IMAB//1S IMAB I IMOB I Tech E
NAME JASmithm SL erc nt LWCamper o JMPiccone a EKraus o
DATE t//14/96 A/96 10/28/96 10/29/96 8/01/96 OFC OE l OGC DD/IMNS D/IMNS
NAME JLiebermanq/ SATreby o FCCombso DACool
DATE 08/07/9 11/07/96 11/12/96 / /96 C = COVER E M OVER h Attachment N a NO COPY o- See Previous Concurrence
N 96-##
- .
a *-i ovember .1996 D
18year (0.75 Gy/sec)(0.646)
"
blyrs, 0G0.485 Gy/sec
Table 1 may be extended using formula F1 and the value A 0.0243 yr-.
Alternatively, a semi-logarithmic plot of the above dat will yield a straight
line, which may extended beyond 20 years.
This information notice requires no specific actio nor written response. If
you have any questions about the information in t s notice, please contact
one of the technical contacts listed below or t appropriate regional office.
Donald . Cool. Director
Divis on of Industrial and
M ical Nuclear Safety
Ofice of Nuclear Material Safety
and Safeguards
Contacts: Jose M. Diaz-Velez,
(404) 331-7438 Internet:JMD2@NRC. V
Emilio M. Garci , RIV
(510) 975-0239/
Internet:EM C.GOV
James A. S ,thNMSS
(301) 415- 904 Internet- AS4@NRC.GOV
Attachments:
1. List of Recent y Issued NMSS Information Notices
2. List of Rece ly Issued NRC Information Notices
Official Recordo S:\jsmith\imns5350.JAS
OFC
NAME
DATE /
IMA
/1*196 E
I IMAB
le/6///96 ant
jI I
LWCamper o
10/28/96 II MOB
JMPiccone
10/29/96
@
Techi Ed
EKraus o
8/01/96 OFC/ OE I OGC DD/IMNS D/IMNS l
AM JLiebermano SATreby o FCCombso DACool
D0/07/96 11/07/96 11/12/96 / /96
,COVER E = COVER & Attachment N = NO COPY @- See Previous Concurrence
N 96-ft
<iovember ,1996 D (0.75 Gy/sec)(0.646)
.0.485 Gy/sec
1 Table 1 may be etended using formula F1 and the value X - 0.0243 yr .
Alternatively, a mi-logarithmic plot of the above data will yield a straight
line, which may ext nded beyond 20 years.
This information noti requires no specific action nor written response. If
you have any questions bout the information inthis notice, please contact
(one of) the technical c tact(s) listed below or the appropriate regional
office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 Internet:JMD2@NRC.GOY
Emilio M. Garcia, RIV
(510) 975-0239 Internet:EMG@NRC.GOV
James A. Smith, NMSS
(301) 415-7904 Internet:JAS4@NRC.GOV
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
Official Record Copy S:\Jsmith\imns5350.JAS \
OFC qIMABiLW B IM4AB - l
IMOB E\Tech Ed
NAME JASm1i SL chant LWCamper a JMPiccone c EKraus o
DATE I I 1/96 _ / j 10/28/96 10/29/96 8/01/96 OFC OE OGC DD/IMNS D/IMNS
NAME JLiebermano SATreby a F(Cm DACool
DATE 08/07/96 11/07/96 / /96 C a COVER E a CUVLK & Attacnment N = mU LUr1'u- aee rrevmuUs IA0ICUIIrCLM.
,N96-##
' "-'-6ctober ,1996 ble8.. - (0.75 Gy/sec)(0.646)
l 0.4 y/sec
Table 1 may be extended using fo ula F1 and the value X - 0.0243 yre.
Alternatively, a semi-logarithmic lot of the above data will yield a straight
line, which may extended beyond 20 ars.
This information notice requires no sp ific action nor written response. If
you have any questions about the infrma ion inthis notice, please contact
(one of) the technical contact(s) listed elow or the appropriate regional
office.
Donald A. -ool, Director
Division o Industrial and
Medical Ntclear Safety
Office of N clear Material Safety
and Safegu rds
Contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 Internet:JMD2@NRC. GOY
Emilio M. Garcia, RIV
(510) 975-0239 Internet:EMG@NRC.GOV
James A. Smith, NMSS
(301) 415-7904 Internet:JAS4@NRC .GOV
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
Official Record Copy S:\Jsmithi\mns5350.JAS \
OFCIMAB AB IMAB I l IMOB l \l,TechEdl
NAME JASai R SUM nt LWCamper a JMPiccone o EKraus o
DATE / -J/ 10/28/96 10/29/96 8/01/96 OFC OE DD/IMNS D/IMNS
KANE JLiebermano Sireb NIL FCCombs DACool
DATE 08/07/96 l/ /1/96 / /96 / /96 aCOVER E a COVER & Attachment K = ND COPY o- See Previous Concurrence \
C,6 I < ~N96-#t
4 October ,1996 We have Dlew. - (0.75 Gy/sec)(0.646)
Table I may be extended using formula FH and the value X - 0.0243 yr1 .
Alternatively, a semi-icgarithmic plot of the above data will yield a straight
line, which may extended be~yond 20 years.
This information notice requires no specific action nor written response. If
you have any questions about the information in this notice, please contact
(one of) the technical contact(s) listed below or the appropriate regional
office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office cf Nuclear Material Safety
and Safrguards
Contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 Internet:JMD2@NRC.60V
Emilio M. Garcia, RIV
(510) 975-0239 Internet:EMG@NRC.GOY
James A. Smith, NMSS
(301) 415-7904 Internet:JAS4@NRC.GOV
Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
Official Record CODY SA:smithkimtns535GJAS
OFC IMAB E IMAB Tech
NAME JASmi S rchant C rer on EKraus o
DATE It/1 /96 /96 96 /09 /96 8/01/96 OFC OE lOGC lDD/IMNS l D/IMNS l~
KANE JLiebermano SATreby FCCombs DACool
DATE 08/07/96 / /96 I /96 / /96
- COVER E a COVER & Attachment N = NO COPY e- See Previous Concurrence
A '
IN 96-##
August ,1996 This informat n not1ce requires no specific action nor written response. If
you have any qu tions about the information in this notice, please contact
(one of) the techn cal contact(s) listed below or the appropriate regional
office.
Donald A. Cool, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Contacts: Jose M. Diaz-Velez, RII
(404) 331-7438 Emilio M. Garcia, RIV
(510) 975-0239 James A. Smith, NMSS
(301) 415-7904 Attachments:
1. List of Recently Issued NMSS Information Notices
2. List of Recently Issued NRC Information Notices
Official Record Copy G:\imns5350.JAS , )
OFC IMABl E IlTech
OGC Ed
NAME JASmi'W per SATreby EKraus a
DATE *in /96 __ /________ \___96 / /96 8/01/96 OFC \ IlI DD/IMNS l D/IMNS l
NAME Q jeerman FCCombs DACool
DATE ___9_
- w COVER
/ /96
'E i. COVER & Attachment N = NO COPY o- See Previous Concurrence
|
---|
|
list | - Information Notice 1996-01, Potential For High Post-Accident Closed-Cycle Cooling Water Temperatures to Disable Equipment Important to Safety (3 January 1996)
- Information Notice 1996-01, Potential for High Post-Accident Closed-Cycle Cooling Water Temperatures to Disable Equipment Important to Safety (3 January 1996)
- Information Notice 1996-02, Inoperability of Power-Operated Relief Valves Masked by Downstream Indications During Testing (5 January 1996, Topic: Stroke time)
- Information Notice 1996-03, Main Steam Safety Valve Setpoint Variation as a Result of Thermal Effects (5 January 1996)
- Information Notice 1996-03, Main Steam Safety Valve Setpoint Variation As a Result of Thermal Effects (5 January 1996)
- Information Notice 1996-04, Incident Reporting Requirements for Radiography Licensees (10 January 1996, Topic: Brachytherapy)
- Information Notice 1996-05, Partial Bypass of Shutdown Cooling Flow from Reactor Vessel (18 January 1996, Topic: Reactor Vessel Water Level)
- Information Notice 1996-06, Design & Testing Deficiencies of Tornado Dampers at Nuclear Power Plants (25 January 1996)
- Information Notice 1996-07, Slow Five Percent Scram Insertion Times Caused by Viton Diaphragms in Scram Solenoid Pilot Valves (26 January 1996)
- Information Notice 1996-08, Thermally Induced Pressure Locking of a High Pressure Coolant Injection Gate Valve (5 February 1996, Topic: Anchor Darling, Cold shutdown justification)
- Information Notice 1996-09, Damage in Foreign Steam Generator Internals (12 February 1996, Topic: Earthquake)
- Information Notice 1996-10, Potential Blockage by Debris of Safety System Piping Which Is Not Used During Normal Operation or Tested During Surveillances (13 February 1996)
- Information Notice 1996-10, Potential Blockage by Debris of Safety System Piping Which is Not Used During Normal Operation or Tested During Surveillances (13 February 1996)
- Information Notice 1996-11, Ingress of Demineralizer Resins Increases Potential For Stress Corrosion Cracking of Control Rod Drive Mechanism Penetrations (14 February 1996)
- Information Notice 1996-11, Ingress of Demineralizer Resins Increases Potential for Stress Corrosion Cracking of Control Rod Drive Mechanism Penetrations (14 February 1996)
- Information Notice 1996-12, Control Rod Insertion Problems (15 February 1996)
- Information Notice 1996-13, Potential Containment Leak Paths Through Hydrogen Analysis (26 February 1996)
- Information Notice 1996-14, Degradation of Radwaste Facility Equipment at Millstone Nuclear Power Station, Unit 1 (1 March 1996)
- Information Notice 1996-15, Unexpected Plant Performance During Performance of New Surveillance (8 March 1996)
- Information Notice 1996-16, BWR Operation with Indicated Flow Less than Natural Circulation (14 March 1996)
- Information Notice 1996-17, Reactor Operation Inconsistent with the Updated Final Safety Analysis Report (18 March 1996)
- Information Notice 1996-18, Compliance with 10 CFR Part 20 for Airborne Thorium (25 March 1996, Topic: Brachytherapy)
- Information Notice 1996-19, Failure of Tone Alert Radios to Activate When Receiving a Shortened Activation Signal (2 April 1996)
- Information Notice 1996-20, Demonstration of Associated Equipment Compliance with 10 CFR 34.20 (4 April 1996, Topic: Brachytherapy)
- Information Notice 1996-21, Safety Concerns Related to the Design of the Door Interlock Circuit on Nucletron High-Dose Rate and Pulsed Dose Rate Remote Afterloading Brachytherapy Devices (10 April 1996, Topic: Brachytherapy)
- Information Notice 1996-22, Improper Equipment Settings Due to Use of Nontemperature-Compensated Test Equipment (11 April 1996, Topic: Brachytherapy)
- Information Notice 1996-23, Fires in Emergency Diesel Generator Exciters During Operation Following Undetected Fuse Blowing (22 April 1996, Topic: Brachytherapy)
- Information Notice 1996-24, Preconditioning of Molded-Case Circuit Breakers Before Surveillance Testing (25 April 1996, Topic: Brachytherapy)
- Information Notice 1996-25, Traversing In-Core Probe Overwithdrawn at Lasalle County Station, Unit 1 (30 April 1996, Topic: Brachytherapy)
- Information Notice 1996-26, Recent Problems with Overhead Cranes (30 April 1996, Topic: Brachytherapy)
- Information Notice 1996-26, Recent Problems With Overhead Cranes (30 April 1996)
- Information Notice 1996-27, Potential Clogging of High Pressure Safety Injection Throttle Valves During Recirculation (1 May 1996, Topic: Brachytherapy)
- Information Notice 1996-28, Suggested Guidance Relating to Development and Implementation of Corrective Action (1 May 1996, Topic: Brachytherapy)
- Information Notice 1996-29, Requirements in 10 CFR Part 21 for Reporting and Evaluating Software Errors (20 May 1996, Topic: Brachytherapy)
- Information Notice 1996-30, Inaccuracy of Diagnostic Equipment for Motor-Operated Butterfly Valves (21 May 1996)
- Information Notice 1996-31, Cross-Tied Safety Injection Accumulators (22 May 1996)
- Information Notice 1996-32, Implementation of 10 CFR 50.55a(g) (6) (II) (A), Augmented Examination of Reactor Vessel (5 June 1996, Topic: Non-Destructive Examination)
- Information Notice 1996-32, Implementation of 10 CFR 50.55a(g) (6) (ii) (A), Augmented Examination of Reactor Vessel (5 June 1996, Topic: Nondestructive Examination)
- Information Notice 1996-33, Erroneous Data From Defective Thermocouple Results in a Fire (24 May 1996, Topic: Reverse polarity)
- Information Notice 1996-33, Erroneous Data from Defective Thermocouple Results in a Fire (24 May 1996, Topic: Reverse polarity)
- Information Notice 1996-34, Hydrogen Gas Ignition During Closure Welding of a VSC-24 Multi-Assembly Sealed Basket (31 May 1996)
- Information Notice 1996-35, Failure of Safety Systems on Self-Shielded Irradiators Because of Inadequate Maintenance and Training (11 June 1996)
- Information Notice 1996-36, Degradation of Cooling Water Systems Due to Icing (12 June 1996, Topic: High winds, Ultimate heat sink, Frazil ice)
- Information Notice 1996-37, Inaccurate Reactor Water Level Indication and Inadvertent Draindown During Shutdown (18 June 1996, Topic: Reactor Vessel Water Level)
- Information Notice 1996-38, Results of Steam Generator Tube Examinations (21 June 1996)
- Information Notice 1996-39, Estimates of Decay Heat Using ANS 5.1 Decay Heat Standard May Vary Significantly (5 July 1996)
- Information Notice 1996-40, Defciencies in Material Dedication and Procurement Practices and in Audits of Vendors (7 October 1996, Topic: Coatings, Troxler Moisture Density Gauge)
- Information Notice 1996-41, Effects of a Decrease in Feedwater Temperature on Nuclear Instrumentation (26 July 1996)
- Information Notice 1996-42, Unexpected Opening of Multiple Safety Relief Valves (5 August 1996, Topic: Reactor Vessel Water Level)
- Information Notice 1996-43, Failures of General Electric Magne-Blast Circuit Breakers (2 August 1996)
... further results |
---|