7.0 CROSS CHECK PROGRAM
This page intentionally left blank.
Laboratories LLc 2014 ANNUAL QUALITY ASSURANCE REPORT FOR THE RADIOLOGICAL ENVIRONMENTAL MONITORING PROGRAM (REMP)
GEL LABORATORIES, LLC P.O. Box 30712, Charleston, SC 29417 843.556.8171
RJ ij I laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 2 of 64 2014 ANNUAL QUALITY ASSURANCE REPORT FOR THE RADIOLOGICAL ENVIRONMENTAL MONITORING PROGRAM (REMP)
Approved By:
February 15. 2015 Rev. 1 Robert L. Pullano Date Director, Quality Systems
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 3 of 64 TABLE OF CONTENTS
- 1. INTRODUCTION..................................................................................................................... 5
- 2. QUALITY ASSURANCE PROGRAMS FOR INTER-LABORATORY, INTRA-LABORATORY AND THIRD PARTY CROSS-CHECK............................................................................................. 6
- 3. QUALITY ASSURANCE PROGRAM FOR INTERNAL AND EXTERNAL AUDITS................ 7
- 4. PERFORMANCE EVALUATION ACCEPTANCE CRITERIA FOR ENVIRONMENTAL SAMPLE ANALYSIS............................................................................................................................... 8
- 5. PERFORMANCE EVALUATION SAMPLES........................................................................... 8
- 6. QUALITY CONTROL PROGRAM FOR ENVIRONMENTAL SAMPLE ANALYSIS................ 8
- 7.
SUMMARY
OF DATA RESULTS............................................................................................ 9
- 8.
SUMMARY
OF PARTICIPATION IN THE ECKERT & ZIEGLER ANAL YTICS ENVIRONMENTAL CROSS-CHECK PROGRAM................................................................................................... 9
- 9.
SUMMARY
OF PARTICIPATION IN THE MAPEP MONITORING PROGRAM................... 10
- 10.
SUMMARY
OF PARTICIPATION IN THE ERA MRAD PT PROGRAM.............................. 10
- 11.
SUMMARY
OF PARTICIPATION IN THE ERA PT PROGRAM.......................................... 10
- 12. CORRECTIVE ACTION REQUEST AND REPORT (CARR).............................................. 10
- 13. REFERENCES.................................................................................................................... 12
M¥111 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 4 of 64 TABLE OF CONTENTS (CONTINUED)
TABLES Table 1 2014 Radiological Proficiency Testing Results and Acceptance Criteria..................... 13 Table 2 2014 Eckert & Ziegler Analytics Performance Evaluation Results............................... 29 Table 3 2014 Department of Energy Mixed Analyte Performance Evaluation Program (MAPEP)
Results............................................................................................................... 31 Table 4 2014 ERA Program Performance Evaluation Results................................................. 36 Table 5 2014 ERA Program (MRAD) Performance Evaluation Results................................... 38 Table 6 REMP Intra-Laboratory Data Summary: Bias and Precision By Matrix..................... 53 Table 7 All Radiological Intra-Laboratory Data Summary: Bias and Precision By Matrix........................................................................................................ 55 Table 8 2014 Corrective Action Report Summary........................................................... 62 FIGURES Figure 1 C~balt-60 Performance Evaluation Results and % Bias............................................. 44 Figure 2 Cesium-137 Performance Evaluation Results and % Bias......................................... 45 Figure 3 Tritium Performance Evaluation Results and % Bias.................................................. 46 Figure 4 Strontium-90 Performance Evaluation Results and % Bias........................................ 4 7 Figure 5 Gross Alpha Performance Evaluation Results and % Bias......................................... 48 Figure 6 Gross Beta Performance Evaluation Results and % Bias........................................... 49 Figure 7 lodine-131 Performance Evaluation Results and % Bias............................................. 50 Figure 8 Americium-241 Performance Evaluation Results and % Bias..................................... 51 Figure 9 Plutonium-238 Performance Evaluation Results and % Bias...................................... 52
1!'1~!1 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 5 of 64 2014 ANNUAL QUALITY ASSURANCE REPORT FOR THE RADIOLOGICAL ENVIRONMENTAL MONITORING PROGRAM (REMP)
- 1. Introduction GEL Laboratories, LLC (GEL) is a privately owned environmental laboratory dedicated to providing personalized client services of the highest quality.
GEL was established as an analytical testing laboratory in 1981. Now a full service lab, our analytical divisions use state of the art equipment and methods to provide a comprehensive array of organic, inorganic, and radiochemical analyses to meet the needs of our clients.
At GEL, quality is emphasized at every level of personnel throughout the company. Management's ongoing commitment to good professional practice and to the quality of our testing services to our customers is demonstrated by their dedication of personnel and resources to develop, implement, assess, and improve our technical and management operations.
The purpose of GEL's quality assurance program is to establish policies, procedures, and processes to meet or exceed the expectations of our clients. To achieve this, all personnel that support these services to our clients are introduced to the program and policies during their initial orientation, and annually thereafter during company-wide training sessions.
GEL's primary goals are to ensure that all measurement data generated are scientifically and legally defensible, of known and acceptable quality per the data quality objectives (DQOs), and thoroughly documented to provide sound support for environmental decisions.
In addition, GEL continues to ensure compliance with all contractual requirements, environmental standards, and regulations established by local, state and federal authorities.
GEL administers the QA program in accordance with the Quality Assurance Plan, GL-QS-B-001. Our Quality Systems include all quality assurance (QA) policies and quality control (QC) procedures necessary to plan, implement, and assess the work we perform. GEL's QA Program establishes a quality management system (QMS) that governs all of the activities of our organization.
This report entails the quality assurance program for the proficiency testing and environmental monitoring aspects of GEL for 2014. GEL's QA Program is designed to monitor the quality of analytical processing associated with environmental, radiobioassay, effluent (1 0 CFR Part 50), and waste (1 0 CFR Part 61) sample analysis.
This report covers the category of Radiological Environmental Monitoring Program (REMP) and includes:
Intra-laboratory QC results analyzed during 2014.
Inter-laboratory QC results analyzed during 2014 where known values were available.
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 6 of 64
- 2. Quality Assurance Programs for Inter-laboratory, Intra-laboratory and Third Party Cross-Check In addition to internal and client audits, our laboratory participates in annual performance evaluation studies conducted by independent providers. We routinely participate in the following types of performance audits:
Proficiency testing and other inter-laboratory comparisons Performance requirements necessary to retain Certifications Evaluation of recoveries of certified reference and in-house secondary reference materials using statistical process control data.
Evaluation of relative percent difference between measurements through SPC data.
We also participate in a number of proficiency testing programs for federal and state agencies and as required by contracts. It is our policy that no proficiency evaluation samples be analyzed in any special manner. Our annual performance evaluation participation generally includes a combination of studies that support the following:
US Environmental Protection Agency Discharge Monitoring Report, Quality Assurance Program (DMR-QA). Annual national program sponsored by EPA for laboratories engaged in the analysis of samples associated with the NPDES monitoring program. Participation is mandatory for all holders of NPDES permits. The permit holder must analyze for all of the parameters listed on the discharge permit. Parameters include general chemistry, metals, BOD/COD, oil and grease, ammonia, nitrates, etc.
Department of Energy Mixed Analyte Performance Evaluation Program (MAPEP). A semiannual program developed by DOE in support of DOE contractors performing waste analyses.
Participation is required for all laboratories that perform environmental analytical measurements in support of environmental management activities. This program includes radioactive isotopes in water, soil, vegetation and air filters.
ERA's MRAD-Multimedia Radiochemistry Proficiency test program. This program is for labs seeking certification for radionuclides in wastewater and solid waste. The program is conducted in strict compliance with US EPA National Standards for Water Proficiency study.
ERA's lnterlaB RadCheM Proficiency Testing Program for radiological analyses. This program completes the process of replacing the USEPA EMSL-LV Nuclear Radiation Assessment Division program discontinued in 1998. Laboratories seeking certification for radionuclide analysis in drinking water also use the study. This program is conducted in strict compliance with the USEPA National Standards for Water Proficiency Testing Studies. This program encompasses Uranium by EPA method 200.8 (for drinking water certification in Utah/Primary NELAP), gamma emitters, Gross Alpha/Beta, lodine-131, naturally occurring radioactive isotopes, Strontium-89/90, and Tritium.
ERA's Water Pollution (WP) biannual program for waste methodologies includes parameters for both organic and inorganic analytes.
l3 3
- Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 7 of 64 ERA's Water Supply (WS) biannual program for drinking water methodologies includes parameters for organic and inorganic analytes.
Environmental Cross-Check Program administered by Eckert & Ziegler Analytics, Inc.
This program encompasses radionuclides in water, soil, milk, naturally occurring radioactive isotopes in soil and air filters.
GEL procures single-blind performance evaluation samples from Eckert & Ziegler Analytics to verify the analysis of sample matrices processed at GEL. Samples are received on a quarterly basis.
GEL's Third-Party Cross-Check Program provides environmental matrices encountered in a typical nuclear utility REMP. The Third-Party Cross-Check Program is intended to meet or exceed the inter-laboratory comparison program requirements discussed in NRC Regulatory Guide 4.15.
Once performance evaluation samples have been prepared in accordance with the instructions provided by the PT provider, samples are managed and analyzed in the same manner as environmental samples from GEL's clients.
- 3. Quality Assurance Program for Internal and External Audits During each annual reporting period, at least one internal assessment of each area of the laboratory is conducted in accordance with the pre-established schedule from Standard Operating Procedure for the Conduct of Quality Audits, GL-QS-E-001. The annual internal audit plan is reviewed for adequacy and includes the scheduled frequency and scope of quality control actions necessary to GEL's QA program.
Internal audits are conducted at least annually in accordance with a schedule approved by the Quality Systems Director. Supplier audits are contingent upon the categorization of the supplier, and may or may not be conducted prior to the use of a supplier or subcontractor.
Type I suppliers and subcontractors, regardless of how they were initially qualified, are re-evaluated at least once every three years.
In addition, prospective customers audit GEL during pre-contract audits. GEL hosts several external audits each year for both our clients and other programs.
These programs include environmental monitoring, waste characterization, and radiobioassay. The following list of programs may audit GEL at least annually or up to every three years depending on the program.
NELAC, National Environmental Laboratory Accreditation Program DOECAP, U.S. Department of Energy Consolidated Audit Program DOELAP, U.S. Department of Energy Laboratory Accreditation Program DOE QSAS, U.S. Department of Energy, Quality Systems for Analytical Services ISO/IEC 17025:2005 A2LA, American Association for Laboratory Accreditation DOD ELAP, US Department of Defense Environmental Accreditation Program NUPIC, Nuclear Procurement Issues Committee South Carolina Department of Heath and Environmental Control (SC DHEC)
The annual radiochemistry laboratory internal audit (13-RAD-001) was conducted in July, 2014. One (1) finding, four (4) observations, and eight (8) recommendations resulted from this assessment.
By September, 2014, the finding was closed and appropriate laboratory staff addressed each observation and recommendation.
te;l 311 laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 8 of 64
- 4. Performance Evaluation Acceptance Criteria for Environmental Sample Analysis GEL utilized an acceptance protocol based upon two performance models. For those inter-laboratory programs that already have established performance criteria for bias (i.e., MAPEP, and ERA/ELAP),
GEL will utilize the criteria for the specific program. For intra-laboratory or third party quality control programs that do not have a specific acceptance criteria (i.e. the Eckert-Ziegler Analytics Environmental Cross-check Program), results will be evaluated in accordance with GEL's internal acceptance criteria.
- 5. Performance Evaluation Samples Performance Evaluation (PE) results and internal quality control sample results are evaluated in accordance with GEL acceptance criteria. The first criterion concerns bias, which is defined as the deviation of any one result from the known value. The second criterion concerns precision, which deals with the ability of the measurement to be replicated by comparison of an individual result with the mean of all results for a given sample set.
At GEL, we also evaluate our analytical performance on a regular basis through statistical process control (SPC) acceptance criteria. Where feasible, this criterion is applied to both measures of precision and accuracy and is specific to sample matrix. We establish environmental process control limits at least annually.
For Radiochemistry analysis, quality control evaluation is based on static limits rather than those that are statistically derived. Our current process control limits are maintained in GEL's AlphaLIMS. We also measure precision with matrix duplicates and/or matrix spike duplicates. The upper and lower control limits (UCL and LCL respectively) for precision are plus or minus three times the standard deviation from the mean of a series of relative percent differences. The static precision criteria for radiochemical analyses are 0 - 20%, for activity levels exceeding the contract required detection limit (CRDL).
- 6. Quality Control Program for Environmental Sample Analysis GEL's internal QA Program is designed to include QC functions such as instrumentation calibration checks (to insure proper instrument response), blank samples, instrumentation backgrounds, duplicates, as well as overall staff qualification analyses and statistical process controls. Both quality control and qualification analyses samples are used to be as similar as the matrix type of those samples submitted for analysis by the various laboratory clients. These performance test samples (or performance evaluation samples) are either actual sample submitted in duplicate in order to evaluate the precision of laboratory measurements, or fortified blank samples, which have been given a known quantity of a radioisotope that is in the interest to GEL's clients.
Accuracy (or Bias) is measured through laboratory control samples and/or matrix spikes, as well as surrogates and internal standards. The UCLs and LCLs for accuracy are plus or minus three times the standard deviation from the mean of a series of recoveries. The static limit for radiochemical analyses is 75 - 125%. Specific instructions for out-of-control situations are provided in the applicable analytical SOP.
GEL's Laboratory Control Standard (LCS) is an aliquot of reagent water or other blank matrix to which known quantities of the method analytes are added in the laboratory. The LCS is analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control, and whether the laboratory is capable of making accurate and precise measurements. Some methods may refer to these
1!3 #I' Laboratories LLc P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 9 of 64 samples as Laboratory Fortified Blanks (LFB). The requirement for recovery is between 75 and 125%
for radiological analyses excluding drinking water matrix.
Bias (%) = (observed concentration)
- 1 00 %
(known concentration)
Precision is a data quality indicator of the agreement between measurements of the same property, obtained under similar conditions, and how well they conform to themselves. Precision is usually expressed as standard deviation, variance or range in either absolute or relative (percentage) terms.
GEL's laboratory duplicate (DUP or LCSD) is an aliquot of a sample taken from the same container and processed in the same manner under identical laboratory conditions. The aliquot is analyzed independently from the parent sample and the results are compared to measure precision and accuracy.
If a sample duplicate is analyzed, it will be reported as Relative Percent Difference (RPD). The RPD must be 20 percent or less, if both samples are greater than 5 times the MDC. If both results are less than 5 times MDC, then the RPD must be equal to or less than 100%. If one result is above the MDC and the other is below the MDC, then the RPD can be calculated using the MDC for the result of the one below the MDC. The RPD must be 100% or less. In the situation where both results are above the MDC but one result is greater than 5 times the MDC and the other is less than 5 times the MDC, the RPD must be less than or equal to 20%. If both results are below MDC, then the limits on % RPD are not applicable.
Difference(%) = (high duplicate result-low duplicate result)
- 100 %
(average of results)
- 7. Summary of Data Results During 2013, forty-four (44) radioisotopes associated with seven (7) matrix types were analyzed under GEL's Performance Evaluation program in participation with ERA, MAPEP, and Eckert & Ziegler Analytics.
Matrix types were representative of client analyses performed during 2014. Of the four hundred forty-five (445) total results reported, 98.6% (439 of 445) were found to be acceptable. The list below contains the type of matrix evaluated by GEL.
Air Filter Cartridge Water Milk Soil Liquid Vegetation Graphs are provided in Figures 1-9 of this report to allow for the evaluation of trends or biases. These graphs include radioisotopes Cobalt-60, Cesium-137, Tritium, Strontium-90, Gross Alpha, Gross Beta, lodine-131, Americium-241, and Plutonium-238.
- 8. Summary of Participation in the Eckert & Ziegler Analytics Environmental Cross-Check Program
M d II Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 10 of 64 Eckert & Ziegler Analytics provided samples for sixty-nine (69) individual environmental analyses.
The accuracy of each result reported to Eckert & Ziegler Analytics, Inc. is measured by the ratio of GEL's result to the known value. All results fell within GEL's acceptance criteria (1 00%).
- 9. Summary of Participation in the MAPEP Monitoring Program MAPEP Series 30 and 31 were analyzed by the laboratory.
Of the one hundred thirty-eight (138) analyses, 97.8o/o (135 out of 138) of all results fell within the PT provider's acceptance criteria. Three analytical failures occurred: Uranium-234/233 and Uranium-238 in Soil and Uranium-238 in vegetation.
For the corrective actions associated with MAPEP Series 30, refer to CARR 140605-879 which is detailed in Table 8.
- 10. Summary of Participation in the ERA MRaD PT Program The ERA MRad program provided samples (MRAD-20 and MRAD-21) for one hundred eighty-eight (188) individual environmental analyses. One hundred eighty-seven (187) of the 188 analyses fell within the PT provider's acceptance criteria (99.4%). One analytical failure occurred: Americium-241 in water.
For the corrective actions associated with MRAD-20, refer to CARR140520-874 which are detailed in Table 8.
- 11. Summary of Participation in the ERA PT Program The ERA program provided samples (RAD-96, RAD-98, and 011 014L) for fifty (50) individual environmental analyses.
Of the 50 analyses, 96.0% (48 out of 50) of all results fell within the PT provider's acceptance criteria. One isotope failure occurred: Strontium-89 in water.
For the corrective actions associated with RAD-98 refer to corrective actions CARR140825-902 (Table 8).
- 12. Corrective Action Request and Report (CARR)
There are two categories of corrective action at GEL. One is corrective action implemented at the analytical and data review level in accordance with the analytical SOP. The other is formal corrective action documented by the Quality Systems Team in accordance with GL-QS-E-002. A formal corrective action is initiated when a nonconformance reoccurs or is so significant that permanent elimination or prevention of the problem is required.
Formal corrective action investigations include root cause analysis.
GEL includes quality requirements in most analytical standard operating procedures to ensure that data are reported only if the quality control criteria are met or the quality control measures that did not meet the acceptance criteria are documented. A formal corrective action is implemented according to GL-QS-E-002 for Conducting Corrective/Preventive Action and Identifying Opportunities for Improvement.
Recording and documentation is performed following guidelines stated in GL-QS-E-012 for Client NCR Database Operation.
W::t* Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 11 of 64 Any employee at GEL can identify and report a nonconformance and request that corrective action be taken. Any GEL employee can participate on a corrective action team as requested by the QS team or Group Leaders. The steps for conducting corrective action are detailed in GL-QS-E-002. In the event that correctness or validity of the laboratory's test results in doubt, the laboratory will take corrective action. If investigations show that the results have been impacted, affected clients will be informed of the issue in writing within five (5) calendar days of the discovery.
Table 8 provides the status of CARRs for radiological performance testing during 2014. It has been determined that causes of the failures did not impact any data reported to our clients.
nJ 3 !I laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 12 of 64
- 13. References
- 1. GEL Quality Assurance Plan, GL-QS-B-001
- 2. GEL Standard Operating Procedure for the Conduct of Quality Audits, GL-QS-E-001
- 3. GEL Standard Operating Procedure for Conducting Corrective/Preventive Action and Identifying Opportunities for Improvement, GL-QS-E-002
- 4. GEL Standard Operating Procedure for AlphaLIMS Documentation of Nonconformance Reporting and Dispositioning and Control of Nonconforming Items, GL-QS-E-004
- 5. GEL Standard Operating Procedure for Handling Proficiency Evaluation Samples, GL-QS-E-013
- 6. GEL Standard Operating Procedure for Quality Assurance Measurement Calculations and Processes, GL-QS-E-014
- 7. 40 CFR Part 136 Guidelines Establishing Test Procedures for the Analysis of Pollutants
- 8. ISO/IEC 17025-2005, General Requirements for the Competence of Testing and Calibration Laboratories
- 9. ANSI/ASQC E4-1994, Specifications and Guidelines for Quality Systems for Environmental Data Collection and Environmental Technology Programs, American National Standard
- 10. 2003 NELAC Standard, National Environmental Laboratory Accreditation Program
- 11. 2009 TNI Standard, The NELAC Institute, National Environmental Accreditation Program
- 12. MARLAP, Multi-Agency Radiological Laboratory Analytical Protocols
- 13. 10 CFR Part 21, Reporting of Defects and Noncompliance
- 14. 10 CFR Part 50 Appendix B, Quality Assurance Criteria for Nuclear Power Plants and Fuel Reprocessing Plants
- 15. 10 CFR Part 61, Licensing Requirements for Land Disposal and Radioactive Waste
- 16. NRC REG Guide 4.15 and NRC REG Guide 4.8
1!13!1 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 13 of 64 TABLE 1 2014 RADIOLOGICAL PROFICIENCY TESTING RESULTS AND ACCEPTANCE CRITERIA Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation 1st I RAD-ERA 2014 02/24114 96 Water pCi!L Barium-133 80.6 76.2 63.8-83.8 Acceptable 1st I RAD-ERA 2014 02/24114 96 Water pCiiL Cesium-134 64.7 66.8 54.4-73.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi!L Cesium-137 112.0 109 98.1 -122 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCi/L Cobalt-60 95.0 88.7 79.8-99.9 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Zinc-65 200 185 166-218 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCi/L Gross Alpha 34.8 36.1 18.6-46.4 Acceptable 1st I RAD-ERA 2014 02/24114 96 Water pCiiL Gross Beta 19.6 22.3 13.5-30.4 Acceptable 1st I RAD -
ERA 2014 02/24114 96 Water pCiiL Gross Alpha 34.6 36.1 18.6-46.4 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Radium-226 16.2 16.8 12.5-19.2 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi!L Radium-228 4.62 5.04 3.01-6.67 Acceptable 1st I RAD-Uranium ERA 2014 02124114 96 Water pCiiL (Nat) 7.39 7.23 5.51-8.53 Acceptable 1st I RAD-Uranium ERA 2014 02124114 96 Water ugll (Nat) mass 11.00 10.6 8.07-12.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi!L Radium-226 15.10 16.8 12.5-19.2 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Radium-228 4.66 5.04 3.01 -6.67 Acceptable 1st I RAD-Uranium ERA 2014 02124114 96 Water pCi!L (Nat) 7.47 7.23 5.51-8.53 Acceptable 1st I RAD -
Uranium ERA 2014 02124114 96 Water ugll (Nat) mass 11.4 10.6 8.07-12.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Tritium 3320 3580 3030-3950 Acceptable 1st I RAD-Strontium-ERA 2014 02124114 96 Water pCiiL 89 44.1 44.4 34.4-51.6 Acceptable 1st I RAD -
Strontium-ERA 2014 02124114 96 Water pCiiL 90 34.2 30.3 22.1-35.2 Acceptable 1st I RAD-Strontium-ERA 2014 02124114 96 Water pCiiL 89 38.9 44.4 34.4-51.6 Acceptable 1st I RAD -
Strontium-ERA 2014 02124114 96 Water pCiiL 90 27.1 30.3 22.1-35.2 Acceptable 1st I Strontium-ERA 2014 02/06114 011014L Water pCi!L 89 42.3 38.7 29.3-45.7 Acceptable 1st I Strontium-ERA 2014 02/06114 011014L Water pCiiL 89 42.2 38.7 29.3-45.7 Acceptable 1st I RAD-ERA 2014 02/24114 96 Water pCiiL lodine-131 25.2 24.4 20.2-28.9 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL lodine-131 22.4 24.4 20.2-28.9 Acceptable EZA 1sU2014 05116114 E10846 Cartridge jJCi lodine-131 7.83E+01 7.50E+03 1.04 Acceptable Strontium-EZA 1sU2014 05116114 E10847 Milk pCiiL 89 9.14E+01 9.17E+01 1
Acceptable Strontium-EZA 1sU2014 0511 6114 E10847 Milk pCi/L 90 1.27E+01 1.51E+01 0.84 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 14 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider Ye~r Date Number
~edia Unit Nuclide Value value Ratio Evalu_ation EZA 1st/2014 05/16/14 E10848 Milk pCi/L lodine-131 9.84E+01 9.85E+01 1
Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cerium-141 1.21 E+02 1.19E+02 1.02 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi!L Cr-51 5.19E+02 4.91E+02 1.06 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi!L Cesium-134 1.79E+02 2.10E+02 0.85 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cesium-137 2.55E+02 2.53E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cobalt-58 2.58E+02 2.68E+02 0.96 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Mn-54 3.01E+02 2.97E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi!L Iron-59 2.24E+02 2.19E+02 1.02 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi!L Zinc-65 3.45E+02 3.23E+02 1.07 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cobalt-60 3.39E+02 3.37E+02 1.00 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi!L lodine-131 9.24E+01 8.99E+01 1.03 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi!L Cerium-141 8.19E+01 7.71E+01 1.06 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cr-51 3.32E+02 3.19E+02 1.04 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cesium-134 1.27E+02 1.36E+02 0.93 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cesium-137 1.69E+02 1.64E+02 1.03 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cobalt-58 1.75E+02 1.74E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Mn-54 2.08E+02 1.93E+02 1.08 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi!L Iron-59 1.68E+02 1.42E+02 1.18 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi!L Zinc-65 2.25E+02 2.10E+02 1.07 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cobalt-60 2.31E+02 2.19E+02 1.02 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 GrF30 Filter Bq/sample Gross Alpha 1.980 1.77 0.53-3.01 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 GrF30 Filter Bq/sample Gross Beta 0.823 0.77 0.39-1.16 Acceptable MAPEP-14-Americium-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 241 65 68 47.6-88.4 Acceptable MAPEP-14-False Pas MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Cesium-134 5.44 0
Test Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Cesium-137 1270 1238 867-1609 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Cobalt-57 947 966 676-1256 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Cobalt-60 0.581 1.220 Sens. Eval.
Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Iron-55 580 643 444-824 Acceptable MAPEP-14-Manganese-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 54 1470 1430 1001-1859 Acceptable MAPEP-14-False Pas MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Nickel-63 6.95 0
Test Acceptable MAPEP-14-Plutonium-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 238 89.7 96.0 67-125 Acceptable MAPEP-14-Plutonium-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 239/240 69.80 76.8 53.8-99.8 Acceptable MAPEP-14-Potassium-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 40 703 622 435-809 Acceptable MAPEP-Strontium-False Pas MAPEP 2nd/2014 06/05/14 14-Soil Bq/kg 90 1.48 0
Test Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 15 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
. PrO~!...
Year Date Number Media Unit Nuclide Value value I*J Ratio
~ ~ ~ Evaluation MaS30 MAPEP-14-Technetium-False Pas MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg 99 37.1 0
Test Acceptable MAPEP-14-Not MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg U-234/233 30.5 81.0 57-105 Accept.
MAPEP-14-Uranium-Not MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/k~
238 35 83 58-108 Accept.
MAPEP-14-MAPEP 2nd/2014 06/05/14 MaS30 Soil Bq/kg Zinc-65 766 695 487-904 Acceptable MAPEP-14-Americium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 241 0.759 0.720 0.504-0.936 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Cesium-134 21.4 23.1 16.2-30.0 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water B_q/L Cesium-137 29.70 28.9 20.2-37.6 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Cobalt-57 28.0 27.5 19.3-35.8 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Cobalt-60 16.6 16.0 11.2-20.8 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Hydro~en-3 308 321 225-417 Acceptable MAPEP-14-False Pas MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Iron-55 0.3 0.0 Test Acceptable MAPEP-14-Manganese-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 54 14.4 13.9 9.7-18.1 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Nickel-63 31.4 34.0 23.8-44.2 Acceptable MAPEP-14-Plutonium-MAPEP 2nd/2014 06/05/14 MaW30 Water B_g/L 238 0.764 0.828 0.580-1.076 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Pu-239/240 0.6590 0.6760 0.473-0.879 Acceptable MAPEP-14-Potassium-False Pas MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 40 0.460 0
Test Acceptable MAPEP-14-Strontium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 90 8.32 8.51 5.96-11.06 Acceptable MAPEP-14-Technetium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 99 9.5 10.3 7.2-13.4 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water B_q/L U-234/233 0.210 0.225 0.158-0.293 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 238 1.41 1.45 1.02-1.89 Acceptable MAPEP-False Pas MAPEP 2nd/2014 06/05/14 14-Water Bq/L Zinc-65
-0.126 0.0 Test Acceptable
W¥111 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 16 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte/
GEL Known Range/
Provider Year,_
Date Number Media Unit *-
Nuclide 1
Value value _
Ratio
~-
Evaluation ~
MaW30 MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Gross Alpha 0.96 0.85 0.255-1.443 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Gross Beta 4.7 4.2 2.10-6.29 Acceptable MAPEP-14-False Pos MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L lodine-129 0.0227 0.00 Test Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample 235 0.018 0.020 0.014-0.026 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample 238 8.77 10.4 7.3-13.5 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample Total 8.80 10.4 7.3-13.5 Acceptable MAPEP-14-Americium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample 241 0.086 0.090 0.063-0.117 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cesium-134 1.85 1.91 1.34-2.48 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cesium-137 1.81 1.76 1.23-2.29 Acceptable MAPEP-14-False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cobalt-57 0.0757 0.00 Test Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cobalt-50 1.490 1.39 0.97-1.81 Acceptable MAPEP-14-Manganese-False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 54 0.0138 0.00 Test Acceptable MAPEP-14-Plutonium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 238 0.000819 0.00090 Sens. Eval.
Acceptable MAPEP-14-0.054-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Pu-239/240 0.071 0.7720 0.1004 Acceptable MAPEP-14-Strontium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 90 1.19 1.18 0.83-1.53 Acceptable MAPEP-14-0.0137-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample U-234/233 0.0159 0.0195 0.0254 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 238 0.118 0.129 0.090-0.168 Acce~table MAPEP-14-False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Zinc-65 0.246 0.00 Test Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Gross Alpha 0.656 1.20 0.36-2.04 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Gross Beta 0.95 0.85 0.43-1.28 Acceptable MAPEP-Americium-MAPEP 2nd/2014 06/05/14 14-Filter Bq/sample 241 0.106 0.104 0.073-0.135 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 17 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte/
GEL Known Range/
Provider -
Year Date Number Media l!nl!.__
Nuclide I*_.Y~l!!L._ 1 value Ratio Evaluation...
RdF30 MAPEP-14-Uranium-0.0188-Not MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample 235 0.261 0.0268 0.0348 Accept.
MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample 238 12.7 13.3 9.3-17.3 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample Total 12.7 13.3 9.3-17.3 Acceptable MAPEP-14-Americium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample 241 0.1100 0.108 0.076-0.140 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cesium-134 5.65 6.04 4.23-7.85 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cesium-137 4.98 4.74 3.32-6.16 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cobalt-57 11.1 10.1 7.1 -13.1 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cobalt-60 7.21 6.93 4.85-9.01 Acceptable MAPEP-14-Manganese-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 54 9.24 8.62 6.03-11.21 Acceptable MAPEP-14-Plutonium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 238 0.116 0.121 0.085-0.157 Acceptable MAPEP-14-0.108-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Pu-239/240 0.134 0.154 0.0200 Acceptable MAPEP-14-Strontium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 90 1.580 1.46 1.02-1.90 Acceptable MAPEP-14-0.0177-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample U-234/233 0.2640 0.2530 0.0329 Acceptable MAPEP-14-Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 238 0.174 0.165 0.116-0.215 Acceptable MAPEP-14-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation B_q/sample Zinc-65 8.87 7.00 4.38-8.13 Acceptable MRAD-Actinium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 228 1140 1240 795-1720 Acceptable MRAD-Americium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 241 418 399 233-518 Acceptable MRAD-Bismuth-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 212 976 1240 330-1820 Acceptable r
MRAD-Bismuth-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 214 2290 1960 1180-2820 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi!kg Cesium-134 3080 3390 2220-4070 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi!kg Cesium-137 8310 8490 6510-10900 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil
_pCi/kg Cobalt-60 6570 6830 4620-9400 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Lead-212 1330 1240 812-1730 Acceptable ERA 2nd/2014 05/16/14 MRAD-Soil pCi/kg Lead-214 2800 2070 1210-3090 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 18 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte/
GEL Known Range/
Provider Year
. Date Number Media Unit Nuclide Value value Ratio
- Evaluatio!!
20 MRAD-Manganese-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 54
<44.3
<1000 0-1000 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 238 579 578 348-797 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Soil pCi/kQ 239 488 471.00 308-651 Acceptable MRAD-Potassium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 40 10500 10500 7660-14100 Acceptable MRAD-Strontium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 90 2500 2780 1060-4390 Acceptable MRAD-Thorium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 234 3420 3360 1060-6320 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Zinc-65 5700 5400 4300-7180 Acceptable MRAD-Strontium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 90 6730 8530 3250-13500 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 234 2602 3390 2070-4350 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 238 2425 3360 2080-4260 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Total 5027 6910 3750-9120 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil ug/kg Totallmass}_
7110 10100 5570-12700 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 234 3440 3390 2070-4350 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 238 3680 3360 2080-4260 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Total 7310 6910 3750-9120 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil ug/kg Total(mass) 11000 10100 5570-12700 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 234 3740 3390 2070-4350 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg 238 3780 3360 2080-4260 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Total 7683 6910 3750-9120 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil ug/kg Total(mass) 11300 10100 5570-12700 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Soil ug/kg Total(mass) 11200 10100 5570-12700 Acceptable MRAD-Americium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 241 1670 1490 911-1980 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Cesium-134 657 646 415-839 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Cesium-137 861 880 638-1220 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kQ Cobalt-60 997 926 639-1290 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Curium-244 514 516 253-804 Acceptable MRAD-Manganese-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 54
<62.2
<300 0.00-300 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 238 2230 2110 1260-2890 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 239 3810 3740 2300-5150 Acceptable MRAD-Potassium-23000-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 40 30800 31900 44800 Acceptable ERA 2nd/2014 05/16/14 MRAD-Vegetation pCi/kg Strontium-2330 2580 1470-3420 Acceptable
li13!1 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 19 of64 Report Acceptance PT Quarter I Received Sample Sample Analyte/
GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation 20 90 MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kQ 234 1920 1760 1160-2260 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 238 1970 1750 1170-2220 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Total 4025 3580 2430-4460 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation ug/kg Total(mass) 5920 5240 3510-6650 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 VeQetation pCi/kg Zinc-65 1030 919 663-1290 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 234 1730 1760 1160-2260 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg 238 2000 1750 1170-2220 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kQ Total 3817 3580 2430-4460 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 VeQetation ug/kg Total(mass) 5990 5240 3510-6650 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Vegetation ug/kg Total(mass) 5620 5240 3510-6650 Acceptable MRAD-Americium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 241 60.2 59.7 36.8-80.8 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cesium-134 920 1010 643-1250 AcceQ_table MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cesium-137 816 828 622-1090 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cobalt-60 1130 1120 867-1400 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Iron-55 254 240 74.4-469 Acceptable MRAD-Manganese-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 54
<6.64
<50.0 0-50.0 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 238 51.3 56.3 38.6-74.0 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 239 47.5 48.6 35.2-63.5 Acceptable MRAD-Strontium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 90 76.7 78.9 38.6-118 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 234 33.8 36.4 22.6-54 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 238 34.5 36.1 23.3-49.9 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Total 70.3 74.3 41.1-113 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter UQ/Filter Total(mass) 104 108 69.1-152 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Zinc-65 737 667 478-921 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 234 35.5 36.4 22.6-54 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter 238 35.3 36.1 23.3-49.9 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Total 72.4 74.3 41.1-113 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter ug/Filter Total(mass) 105 108 69.1-152 Acceptable MRAD-Uranium-ERA 2nd/2014 05/16/14 20 Filter ug/Filter Total( mass) 100 108 69.1-152 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Gross Alpha 60.9 46 15.4-71.4 Acceptable ERA 2nd/2014 05/16/14 MRAD-Filter pCi/Filter Gross Beta 58.9 53.8 34.0-78.4 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 20 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
1 _.Provider -
Year 1-Date NumbE!L_
Media Unit Nuclide Value value Ratio Evaluation 20 MRAD-Americium-Not ERA 2nd/2014 05/15/14 20 Water pCi/L 241 185 114 75.8-153 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Cesium-134 1540 1550 1220-1910 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Cesium-137 2750 2590 2280-3220 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Cobalt-50 1320 1270 1100-1490 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Iron-55 1230 1200 715-1530 Acceptable MRAD-Manganese-ERA 2nd/2014 05/15/14 20 Water pCi/L 54
<7.54
<100 0.00-100 Acceptable MRAD-Plutonium-ERA 2nd/2014 05/15/14 20 Water pCi/L 238 37 44 32.5-54.9 Accep_table MRAD-Plutonium-ERA 2nd/2014 05/15/14 20 Water pCi/L 239 124 150 124-202 Acceptable MRAD-Strontium-ERA 2nd/2014 05/15/14 20 Water pCi/L 90 95 890 580-1180 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 234 77.8 82.4 51.9-105 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 238 50.8 48.4 35.9-59.4 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L Total 155 158 123-217 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water ug/L Total(mass) 233 245 195-295 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Zinc-55 2030 1800 1500-2270 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 234 82.1 82.4 51.9-105 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 238 84.5 48.4 35.9-59.4 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L Total 170 158 123-217 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water ug/L Total(mass) 253 245 195-295 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 234 80.5 82.4 51.9-105 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water J2Cill 238 90.0 48.4 35.9-59.4 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L Total 175 158 123-217 AcceRtable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water ug/L Total(mass) 259 245 195-295 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 234 77.8 82.4 51.9-105 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L 238 78.3 48.4 35.9-59.4 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water pCi/L Total 155 158 123-217 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water ug/L Total(mass) 233 245 195-295 Acceptable MRAD-Uranium-ERA 2nd/2014 05/15/14 20 Water ug/L Total(mass) 232 245 195-295 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water RCi/L Gross Alpha 141.0 133 47.2-205 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Gross Beta 172 174.0 99.5-258 Acceptable MRAD-ERA 2nd/2014 05/15/14 20 Water pCi/L Tritium 5280 5580 3740-7950 Acceptable EZA 2nd/2014 08/08/14 E10897 Cartridge pCi lodine-131 8.73E+01 8.54E+01 1.02 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 21 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media
.. Unit *-*
~ Nuclide Value value Ratio Evaluation Strontium-EZA 2ndl2014 08108114 E10898 Milk pCiiL 89 9.84E+01 9.13E+01 1.08 Acceptable Strontium-EZA 2ndl2014 08108114 E10898 Milk pCiiL 90 1.44E+01 1.45E+01 0.99 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCiiL lodine-131 9.89E+01 9.09E+01 1.09 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCi/L Cerium-141 1.38E+02 1.24E+02 1.12 Acceptable Chromium-EZA 2ndl2014 08108114 E10899 Milk pCi!L 51 2.68E+02 2.53E+02 1.06 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCi!L Cesium-134 1.58E+02 1.62E+02 0.97 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCi!L Cesium-137 1.27E+02 1.20E+02 1.06 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCiiL Cobalt-58 1.20E+02 1.12E+02 1.07 Acceptable Manganese-EZA 2ndl2014 08108114 E10899 Milk pCiiL 54 1.67E+02 1.56E+02 1.07 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCiiL Iron-59 1.02E+02 1.02E+02 1.00 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCiiL Zinc-65 2.68E+02 2.52E+02 1.06 Acceptable EZA 2ndl2014 08108114 E10899 Milk pCiiL Cobalt-60 2.42E+02 2.24E+02 1.08 Acceptable EZA 2ndl2014 08108114 E10900 Water pCiiL lodine-131 1.13E+02 9.83E+01 1.15 Acceptable EZA 2ndl2014 08108114 E10900 Water pCi!L Cerium-141 1.52E+02 1.43E+02 1.06 Acceptable Chromium-EZA 2ndl2014 08108114 E10900 Water pCi!L 51 3.62E+02 2.94E+02 1.23 Acceptable EZA 2ndl2014 08108114 E10900 Water pCi!L Cesium-134 1.69E+02 1.88E+02 0.90 Acceptable.
EZA 2ndl2014 08108114 E10900 Water pCiiL Cesium-137 1.48E+02 1.39E+02 1.06 Acceptable EZA 2ndl2014 08108114 E10900 Water pCiiL Cobalt-58 1.34E+02 1.30E+02 1.03 Acceptable Manganese-EZA 2ndl2014 08108114 E10900 Water pCi!L 54 1.88E+02 1.80E+02 1.04 Acceptable EZA 2ndl2014 08108114 E10900 Water pCiiL Iron-59 1.29E+02 1.19E+02 1.09 Acceptable EZA 2ndl2014 08108114 E10900 Water pCiiL Zinc-65 3.29E+02 2.93E+02 1.12 Acceptable EZA 2ndl2014 08108114 E10900 Water pCi!L Cobalt-60 2.74E+02 2.60E+02 1.05 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Barium-133 67.8 68.7 57.3-75.6 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi/L Cesium-134 71 72.3 59.0-79.5 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi!L Cesium-137 161 163 147-181 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi/L Cobalt-60 76.7 75.5 68.0-85.5 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi!L Zinc-65 92 82 73.8-98.5 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Gross Alpha 45.3 45.4 23.6-57.4 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Gross Beta 32.3 33.4 21.7-41.1 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi/L Gross Alpha 48.6 45.4 23.6-57.4 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi!L Radium-226 8.26 9.06 6.80-10.6 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Radium-226 8.54 9.06 6.80-10.6 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Radium-226 9.7 9.06 6.80-10.6 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Radium-228 5.07 5.07 3.03-6.79 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCi!L Radium-228 5.74 5.07 3.03-6.79 Acceptable 3rd I RAD-Uranium ERA 2013 08125114 98 Water pCiiL (Nat) 13.9 13.5 10.7-15.4 Acceptable 3rd I RAD-Uranium ERA 2013 08125114 98 Water ugll (Nat) mass 22.25 19.8 15.6-22.6 Acceptable 3rd I RAD-Uranium ERA 2013 08125114 98 Water pCiiL (Nat) 13 13.5 10.7-15.4 Acceptable ERA 3rd I 08125114 RAD-Water U!=)IL Uranium 20.7 19.8 15.6-22.6 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 22 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider 1
Year Date Number Media Unit Nuclide Value value __
Ratio Evaluation 2013 98 (Nat) mass 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Tritium 10200 11200 9750-12300 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL Tritium 10400 11200 9750-12300 Acceptable 3rd I RAD-Strontium-Not ERA 2013 08125114 98 Water pCiiL 89 56.3 42.7 32.9-49.8 Acceptable 3rd I RAD-Strontium-ERA 2013 08125114 98 Water pCiiL 90 28.2 31.7 23.1-36.7 Acceptable 3rd I RAD-Strontium-Not ERA 2013 08125114 98 Water pCi/L 89 56.5 42.7 32.9-49.8 Acceptable 3rd I RAD-Strontium-ERA 2013 08125114 98 Water pCi/L 90 26 31.7 23.1-36.7 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL lodine-131 28.6 26.1 21.7-30.8 Acceptable 3rd I RAD-ERA 2013 08125114 98 Water pCiiL lodine-131 22.3 26.1 21.7-30.8 Acceptable EZA 3rdl2014 11122114 E10993 Cartridge pCi lodine-131 9.47E+01 8.99E+01 1.05 Acceptable Strontium-EZA 3rdl2014 11122114 E10994 Milk pCiiL 89 9.73E+01 9.69E+01 1.00 Acceptable Strontium-EZA 3rdl2014 11/22/14 E10994 Milk pCi/L 90 1.31 E+01 1.64E+OO 0.80 Acceptable EZA 3rd/2014 11/22114 E10995 Milk pCi/L lodine-131 1.04E+02 9.76E+01 1.07 Acceptable EZA 3rd/2014 11/22114 E10995 Milk pCi/L Cerium-141 1.28E+02 1.26E+02 1.01 Acceptable Chromium-EZA 3rd/2014 11/22114 E10995 Milk pCi/L 51 3.12E+02 2.88E+02 1.08 Acceptable EZA 3rd/2014 11/22114 E10995 Milk pCi/L Cesium-134 1.51 E+02 1.58E+02 0.96 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cesium-137 2.03E+02 1.93E+02 1.05 Acceptable EZA 3rd/2014 11/22114 E10995 Milk pCi/L Cobalt-58 1.44E+02 1.43E+02 1.01 Acceptable Manganese-EZA 3rd/2014 11/22/14 E10995 Milk pCiiL 54 1.49E+02 1.42E+02 1.05 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCiiL Iron-59 1.82E+02 1.58E+02 1.15 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Zinc-65 7.41E+01 7.30E+01 1.01 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cobalt-60 3.14E+02 2.94E+02 1.06 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L lodine-131 1.02E+02 9.88E+01 103 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCiiL Cerium-141 1.30E+02 1.25E+02 104 Acceptable Chromium-EZA 3rd/2014 11/22/14 E10996 Water pCi/L 51 2.75E+02 2.86E+02 0.96 Acceptable EZA 3rd/2014 11/22114 E10996 Water pCi/L Cesium-134 1.45E+02 1.56E+02 0.93 Acceptable EZA 3rd/2014 11122/14 E10996 Water pCi/L Cesium-137 1.94E+02 1.92E+02 1.01 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCiiL Cobalt-58 1.43E+02 1.42E+02 1.01 Acceptable Manganese-EZA 3rd/2014 11/22/14 E10996 Water pCi/L 54 1.46E+02 1.41E+02 1.04 Acceptable EZA 3rd/2014 11122/14 E10996 Water pCi/L Iron-59 1.66E+02 1.57E+02 1.06 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Zinc-65 7.55E+01 7.24E+01 1.04 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cobalt-60 3.09E+02 2.95E+02 1.05 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 GrF31 Filter Bq/sample Gross Alpha 0.433 0.530 0.16-0.09 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 GrF31 Filter Bq/sample Gross Beta 1.060 1.060 0.53-1.59 Acceptable MAPEP-14-Americium-MAPEP 4th /2014 01/09/15 MaS31 Soil BqiKg 241 88.4 85.5 59.9-111.2 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil BqiKg Cesium-134 588 622 435-809 Acceptable MAPEP-14-False Pas MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cesium-137 1.67 Test Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 23 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cobalt-57 1160 1116 781 -1451 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cobalt-60 821 779 545-1013 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil B_g/Kg Iron-55 796 680 476-884 Acceptable MAPEP-14-Manganese-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 54 1060 1009 706-1312 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Nickel-63 924 980 686-1274 Acceptable MAPEP-14-Plutonium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 238 0.92 0.48 Sens. Eval.
Acceptable MAPEP-14-Plutonium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 239/240 61.5 58.6 41.0-76.2 Acceptable MAPEP-14-Potassium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 40 879 824 577-1071 AcceR_table MAPEP-14-Strontium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 90 891 858 601-1115 Acceptable MAPEP-14-Technetium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 99 466 589 412-766 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg U-234/233 905 89 62-116 Acceptable MAPEP-14-Uranium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 238 257 259 181-337 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Zinc-65 605.0 541 379-703 Acceptable MAPEP-14-Americium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 241 0.915 0.880 0.62-1.14 Acceptable MAPEP-14-False Pos MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cesium-134
-0.06 Test Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cesium-137 18.4 18.4 12.9-23.9 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cobalt-57 25 24.7 17.3-32.1 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cobalt-60 12.5 12.4 8.7-16.1 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Hydrogen-3 216 208 146-270 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Iron-55 34.0 31.5 22.1-41.0 Acceptable MAPEP-14-Manganese-MAPEP 4th/2014 01/09/15 MaW31 Water Bq/L 54 14.2 14.0 9.8-18.2 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 24 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider _
Year _
Date Number Media _ -
Unit Nuclide Value value Ratio Evaluation MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Nickel-63 23.6 24.6 17.2-32.0 Acceptable MAPEP-14-Plutonium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 238 0.547 0.618 0.433-0.803 Acceptable MAPEP-14-Plutonium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 239/240 0.015 0.005 Sens. Eval.
Acceptable MAPEP-14-Potassium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 40 174 161 113-209 Acceptable MAPEP-14-Strontium-False Pos MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 90 0.03 Test Acceptable MAPEP-14-Technetium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 99 6.92 6.99 4.89-9.09 Acceptable MAPEP-14-Uranium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 234/233 0.206 0.205 0.144-0.267 Acceptable MAPEP-14-Uranium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 238 1.280 1.420 0.99-1.85 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Zinc-65 11.900 10.90 7.6-14.2 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Gross Alpha 0.793 0.701 0.201-1.192 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Gross Beta 6.220 5.94 2.97-8.91 Acceptable MAPEP-14-Uranium-0.0278-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample 235 0.040 0.040 0.0516 Acceptable MAPEP-14-Uranium-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample 238 19.3 20.3 14.2-26.4 Acceptable MAPEP-14-Uranium-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample Total 19.00 20.4 14.3-26.5 Acceptable MAPEP-14-Americium-0.0472-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample 241 0.0561 0.067 0.0876 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cesium-134 0.8640 0.96 0.67-1.25 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cesium-137 1.190 1.20 0.84-1.56 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15
. RdF31 Filter Bq/sample Cobalt-57 1.540 1.43 1.00-1.86 Acceptable MAPEP-14-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cobalt-60 1.200 1.10 0.77-1.43 Acceptable MAPEP-14-Manganese-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 54 0.808 0.75 0.53-0.98 Acceptable MAPEP-14-Plutonium-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 238 0.155 0.107 0.075-0.139 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 25 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte/
GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP-14-Plutonium-0.0328-MAPEP 4th 12014 01109115 RdF31 Filter Bqlsample 2391240 0.048 0.0468 0.0608 Acceptable MAPEP-14-Strontium-MAPEP 4th 12014 01109115 RdF31 Filter Bqlsample 90 0.762 0.70 0.492-0.914 Acceptable MAPEP-14-Uranium-0.0251 -
MAPEP 4th 12014 01109115 RdF31 Filter Bqlsample 2341233 0.037 0.0358 0.0465 Acceptable MAPEP-14-Uranium-MAPEP 4th 12014 01109115 RdF31 Filter Bqlsample 238 0.227 0.253 0.177-0.329 Acceptable MAPEP-14-MAPEP 4th 12014 01109115 RdF31 Filter Bqlsample Zinc-65 0.779 0.76 0.53-0.99 Acceptable MAPEP-14-Americium-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 241 0.226 0.19 0.135-0.251 Acceptable MAPEP-14-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample Cesium-134 4.750 5.20 3.64-6.67 Acceptable MAPEP-14-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample Cesium-137 6.910 6.60 4.62-8.58 Acceptable MAPEP-14-False Pos MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample Cobalt-57
-0.002 0.00 Test Acceptable MAPEP-14-False Pas MAPEP 4th 12014 01109115 RdV31 Vegetation B_g/sample
- Cobalt-60 0.008 0.00 Test Acceptable MAPEP-14-Manganese-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 54 7.980 7.88 5.52-10.24 Acceptable MAPEP-14-Plutonium-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 238 0.001 0.001 Sens. Eval.
Acceptable MAPEP-14-Plutonium-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 2391240 0.1510 0.171 0.120-0.222 Acceptable MAPEP-14-Strontium-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 90 2.330 2.32 1.62-3.02 Acceptable MAPEP-14-Uranium-0.0326-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 2341233 0.046 0.047 0.0606 Acceptable MAPEP-14-Uranium-MAPEP 4th 12014 01109115 RdV31 Vegetation Bqlsample 238 0.332 0.324 0.227-0.421 Acceptable MAPEP-14-MAPEP 4th 12014 01109115 RdV31 Vegetation B_g/sample Zinc-65 2.850 2.63 1.84-3.42 Acceptable MAPEP-14-SrF-Strontium-MAPEP 4th 12014 01109115 31 Filter Bqlsample 89 3.62 3.79 2.65-4.93 Acceptable MAPEP-14-SrF-Strontium-MAPEP 4th 12014 01109115 31 Filter Bqlsample 90 3.62 3.79 2.65-4.93 Acceptable MAPEP-14-MAPEP 4th 12014 01109115 XaW-31 Water Bqll lodine-129 4.56 4.55 3.19-5.92 Acceptable 3rd I MRAD-Actinium-ERA 2014 11125114 21 Soil pCilkg 228 1280 1240 795-1720 Acceptable ERA 3rd I 11125114 MRAD-Soil p_Cilkg Americium-825 763 431-956 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 26 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
1 _Pro~
1
__ Year Date
~
Number Media Unit Nuclide Value va~ue __
Ratio
- Evaluation_
2014 21 241 3rd I MRAD-Bismuth-ERA 2014 11125114 21 Soil pCi!kg 212 1620 1240 330-1820 Acceptable 3rd I MRAD-Bismuth-ERA 2014 11125114 21 Soil pCilkg 214 2900 2810 1690-4040 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Cesium-134 1960 2140 1400-2570 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi!kg Cesium-137 6760 6550 5020-8430 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/k~
Cobalt-60 4480 4260 2880-5860 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Lead-212 1260 1240 812-1730 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Lead-214 3480 2750 1610-4100 Acceptable 3rd I MRAD-Manganese-ERA 2014 11125114 21 Soil pCi/kg 54
<30.0
<1000 0-1000 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Soil pCilkg 238 732 739 444-1020 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Soil pCilkg 239 281 309 202-427 Acceptable 3rd I MRAD-Potassium-ERA 2014 11125114 21 Soil pCilkg 40 11500 10700 7810-14400 Acceptable 3rd I MRAD-Strontium-ERA 2014 11125114 21 Soil pCilkg 90 8790 8420 3210-13300 Acceptable 3rd I MRAD-Thorium-ERA 2014 11125114 21 Soil pCilkg 234 2000 2350 743-4420 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi!k~
Zinc-65 3910 3270 2600-4350 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Soil pCilkg 234 2280 2370 1450-3040 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Soil pCilkg 238 2340 2350 1450-2980 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Soil pCilkg Total 4762 4540 2360-6390 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Soil uglkg Total(mass) 7020 7050 3890-8870 Acceptable 3rd I MRAD-Americium-ERA 2014 11125114 21 Ve~etation pCi/k~
241 2260 2290 1400-3505 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation
~Ci!kg Cesium-134 837 849 545-1100 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCi/kg Cesium-137 729 644 467-896 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Cobalt-60 818 784 541-1100 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Veqetation pCi!k~
Curium-244 361 367 180-572 Acceptable 3rd I MRAD-Manganese-ERA 2014 11125114 21 Vegetation pCilkg 54
<25.3
<300 0-300 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Vegetation pCilkg 238 886 862 514-1180 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Vegetation pCilkg 239 675 701 430-965 Acceptable 3rd I MRAD-Potassium-22300-ERA 2014 11125114 21 Veqetation pCi/k~
40 35300 30900 43400 Acceptable 3rd I MRAD-Strontium-ERA 2014 11125114 21 Vegetation pCilkg 90 1230 1710 975-2270 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation pCilkg 234 1980 1780 1170-2290 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation pCi/kg 238 1970 1760 1170-2240 Acceptable ERA 3rd I 11125114 MRAD-Vegetation pCilkg Uranium-4038 3620 2450-4510 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 27 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider_. _ -Year
-. Date Number..*
Media Unit n
Nuclide __
Value value Ratio Evaluation 2014 21 Total 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation uglkg Total(mass) 5910 5280 3540-6710 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation pCi/kg 234 1670 1780 1170-2290 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation pCilkg 238 1800 1760 1170-2240 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation pCilkg Total 3556 3620 2450-4510 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation uglkg Total(mass) 5390 5280 3540-6710 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Vegetation uglkg Total(mass) 5860 5280 3540-6710 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Zinc-65 1930 1570 1130-2200 Acceptable 3rd I MRAD-Americium-ERA 2014 11125114 21 Filter pCiiFilter 241 41.4 38.6 23.8-52.2 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Cesium-134 742 765.0 487-949 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Cesium-137 677 647 486-850 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Cobalt-60 543 523 405-653 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Iron-55 117 120.0 37.2-234 Acceptable 3rd I MRAD-Manganese-ERA 2014 11125114 21 Filter Q_Ci/Filter 54
<5.87
<50 0.00-50.0 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Filter ugiFilter 238 32.9 35.7 24.5-46.9 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Filter pCiiFilter 239 26.8 29.1 21.1-38.0 Acceptable 3rd I MRAD-Strontium-ERA 2014 11125114 21 Filter pCiiFilter 90 187 168 82.1-252 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter
_p_CiiFilter 234 26 28 27.8-41.9 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter pCiiFilter 238 28 27.60 17.8-38.2 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter pCiiFilter Total 56 57 31.4-86.3 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter ugiFilter Total(mass) 82.6 82.7 52.9-116 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Zinc-65 629 547 392-755 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter pCiiFilter 234 28 28 27.8-41.9 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter pCiiFilter 238 25 27.60 17.8-38.2 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter pCi/Filter Total 55 57 31.4-86.3 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter ugiFilter Tota\\(mass) 75.1 82.7 52.9-116 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Filter ugiFilter Total(mass) 90.7 82.7 52.9-116 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Gross Alpha 47.4 36.9 12.4-57.3 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Gross Beta 27.2 21.1 13.3-30.8 Acceptable 3rd I MRAD-Americium-ERA 2014 11125114 21 Water pCiiL 241 72.4 68.6 46.2-92.0 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Cesium-134 816.0 850 624-977 Acceptable ERA 3rd I 11125114 MRAD-Water pCi/L Cesium-137 1310 1240 1060-1490 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 28 of 64 Report Acceptance PT Quarter I Received Sample Sample Analyte I GEL Known Range/
Provider Year -
Date Number Media Unit Nuclide Value value _
Ratio 1 _Evaluation 2014 21 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Cobalt-60 1130 1070 930-1250 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Iron-55 130 134 79.9-182 Acceptable 3rd I MRAD-Manganese-ERA 2014 11125114 21 Water pCi/L 54
<6.34
<100 0.00-100 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Water
!JCiiL 238 35 33 24.6-41.4 Acceptable 3rd I MRAD-Plutonium-ERA 2014 11125114 21 Water pCiiL 239 46.4 51 39.7-64.4 Acceptable 3rd I MRAD-Strontium-ERA 2014 11125114 21 Water pCi/L 90 300 254 165-336 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCiiL 234 42 44 32.9-56.5 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCiiL 238 50 43.50 33.2-53.4 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L Total 92 89 65.5-115 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water ugll Total(mass) 137 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Zinc-65 1070 921 768-1160 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L 234 43 44 32.9-56.5 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L 238 45 43.50 33.2-53.4 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L Total 90 89 65.5-115 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water ugll Total(mass) 134 130 104-157 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCiiL 234 49 44 32.9-56.5 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L 238 42 43.50 33.2-53.4 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water pCi/L Total 93 89 65.5-115 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water ugll Total(mass) 126 130 104-157 Acceptable 3rd I MRAD-Uranium-ERA 2014 11125114 21 Water ugll Total(mass) 144 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Gross Alpha 96.2 98 34.8-152 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Gross Beta 86.1 77.5 44.4-115 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Tritium 5490 5500 3680-7840 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 29 of 64 TABLE 2 2014 ECKERT & ZIEGLER ANAL YTICS PERFORMANCE EVALUATION RESULTS Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation EZA 1st/2014 05/16/14 E10846 Cartridge pCi lodine-131 7.83E+01 7.52E+01 1.04 Acceptable Strontium-EZA 1st/2014 05/16/14 E10847 Milk pCi/L 89 9.14E+01 9.17E+01 1
Acceptable Strontium-EZA 1st/2014 05/16/14 E10847 Milk pCi/L 90 1.27E+01 1.51 E+01 0.84 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L lodine-131 9.84E+01 9.85E+01 1
Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cerium-141 1.21E+02 1.19E+02 1.02 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cr-51 5.19E+02 4.91E+02 1.06 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cesium-134 1.79E+02 2.10E+02 0.85 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cesium-137 2.55E+02 2.53E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cobalt-58 2.58E+02 2.68E+02 0.96 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Mn-54 3.01E+02 2.97E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Iron-59 2.24E+02 2.19E+02 1.02 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Zinc-65 3.45E+02 3.23E+02 1.07 Acceptable EZA 1st/2014 05/16/14 E10848 Milk pCi/L Cobalt-60 3.39E+02 3.37E+02 1.00 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L lodine-131 9.24E+01 8.99E+01 1.03 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cerium-141 8.19E+01 7.71 E+01 1.06 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cr-51 3.32E+02 3.19E+02 1.04 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cesium-134 1.27E+02 1.36E+02 0.93 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cesium-137 1.69E+02 1.64E+02 1.03 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cobalt-58 1.75E+02 1.74E+02 1.01 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Mn-54 2.08E+02 1.93E+02 1.08 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Iron-59 1.68E+02 1.42E+02 1.18 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Zinc-65 2.25E+02 2.10E+02 1.07 Acceptable EZA 1st/2014 05/16/14 E10849 Water pCi/L Cobalt-60 2.31E+02 2.19E+02 1.02 Acceptable EZA 2nd/2014 08/08/14 E10897 Cartridge pCi lodine-131 8.73E+01 8.54E+01 1.02 Acceptable Strontium-EZA 2nd/2014 08/08/14 E10898 Milk pCi/L 89 9.84E+01 9.13E+01 1.08 Acceptable Strontium-EZA 2nd/2014 08/08/14 E10898 Milk pCi!L 90 1.44E+01 1.45E+01 0.99 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L lodine-131 9.89E+01 9.09E+01 1.09 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Cerium-141 1.38E+02 1.24E+02 1.12 Acceptable Chromium-EZA 2nd/2014 08/08/14 E10899 Milk pCi/L 51 2.68E+02 2.53E+02 1.06 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Cesium-134 1.58E+02 1.62E+02 0.97 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L
- Cesium-137 1.27E+02 1.20E+02 1.06 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Cobalt-58 1.20E+02 1.12E+02 1.07 Acceptable Manganese-EZA 2nd/2014 08/08/14 E10899 Milk pCi/L 54 1.67E+02 1.56E+02 1.07 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Iron-59 1.02E+02 1.02E+02 1.00 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Zinc-65 2.68E+02 2.52E+02 1.06 Acceptable EZA 2nd/2014 08/08/14 E10899 Milk pCi/L Cobalt-60 2.42E+02 2.24E+02 1.08 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L lodine-131 1.13E+02 9.83E+01 1.15 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 30 of 64 Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation EZA 2nd/2014 08/08/14 E10900 Water pCi/L Cerium-141 1.52E+02 1.43E+02 1.06 Acceptable Chromium-EZA 2nd/2014 08/08/14 E10900 Water pCi/L 51 3.62E+02 2.94E+02 1.23 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Cesium-134 1.69E+02 1.88E+02 0.90 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Cesium-137 1.48E+02 1.39E+02 1.06 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Cobalt-58 1.34E+02 1.30E+02 1.03 Acceptable Manganese-EZA 2nd/2014 08/08/14 E10900 Water pCi/L 54 1.88E+02 1.80E+02 1.04 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Iron-59 1.29E+02 1.19E+02 1.09 Acce_ptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Zinc-65 3.29E+02 2.93E+02 1.12 Acceptable EZA 2nd/2014 08/08/14 E10900 Water pCi/L Cobalt-60 2.74E+02 2.60E+02 1.05 Acceptable EZA 3rd/2014 11/22/14 E10993 Cartridge pCi lodine-131 9.47E+01 8.99E+01 1.05 Acceptable Strontium-EZA 3rd/2014 11/22/14 E10994 Milk pCi/L 89 9.73E+01 9.69E+01 1.00 Acceptable Strontium-EZA 3rd/2014 11/22/14 E10994 Milk pCi/L 90 1.31E+01 1.64E+01 0.80 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L lodine-131 1.04E+02 9.76E+01 1.07 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cerium-141 1.28E+02 1.26E+02 1.01 Acceptable Chromium-EZA 3rd/2014 11/22/14 E10995 Milk pCi/L 51 3.12E+02 2.88E+02 1.08 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cesium-134 1.51 E+02 1.58E+02 0.96 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cesium-137 2.03E+02 1.93E+02 1.05 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cobalt-58 1.44E+02 1.43E+02 1.01 Acceptable Manganese-EZA 3rd/2014 11/22/14 E10995 Milk pCi/L 54 1.49E+02 1.42E+02 1.05 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Iron-59 1.82E+02 1.58E+02 1.15 AcceQ_table EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Zinc-65 7.41 E+01 7.30E+01 1.01 Acceptable EZA 3rd/2014 11/22/14 E10995 Milk pCi/L Cobalt-60 3.14E+02 2.94E+02 1.06 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L lodine-131 1.02E+02 9.88E+01 103 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cerium-141 1.30E+02 1.25E+02 104 Acceptable Chromium-EZA 3rd/2014 11/22/14 E10996 Water pCi/L 51 2.75E+02 2.86E+02 0.96 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cesium-134 1.45E+02 1.56E+02 0.93 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cesium-137 1.94E+02 1.92E+02 1.01 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cobalt-58 1.43E+02 1.42E+02 1.01 Acceptable Manganese-EZA 3rd/2014 11/22/14 E10996 Water pCi/L 54 1.46E+02 1.41E+02 1.04 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Iron-59 1.66E+02 1.57E+02 1.06 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Zinc-65 7.55E+01 7.24E+01 1.04 Acceptable EZA 3rd/2014 11/22/14 E10996 Water pCi/L Cobalt-60 3.09E+02 2.95E+02 1.05 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 31 of 64 TABLE 3 2014 DEPARTMENT OF ENERGY MIXED ANALYTE PERFORMANCE EVALUATION PROGRAM (MAPEP) RESULTS Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio MAPEP MAPEP 2nd/2014 05/05/14 GrF30 Filter Bq/sample Gross Alpha 1.980 1.77 0.53-3.01 MAPEP MAPEP 2nd/2014 05/05/14 GrF30 Filter Bq/sample Gross Beta 0.823 0.77 0.39-1.15 MAPEP Americium-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 241 55 58 47.5-88.4 MAPEP False Pas MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Cesium-134 5.44 0
Test MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Cesium-137 1270 1238 857-1509 MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Cobalt-57 947 955 575-1255 MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Cobalt-50 0.581 1.220 Sens. Eval.
MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Iron-55 580 543 444-824 MAPEP Manganese-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 54 1470 1430 1001-1859 MAPEP False Pas MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Nickel-53 5.95 0
Test MAPEP Plutonium-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 238 89.7 95.0 57-125 MAPEP Plutonium-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 239/240 59.80 75.8 53.8-99.8 MAPEP Potassium-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 40 703 522 435-809 MAPEP Strontium-False Pas MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 90 1.48 0
Test MAPEP Technetium-False Pas MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 99 37.1 0
Test MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil B_q/kg U-234/233 30.5 81.0 57-105 MAPEP Uranium-MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg 238 35 83 58-108 MAPEP MAPEP 2nd/2014 05/05/14 MaS30 Soil Bq/kg Zinc-55 755 595 487-904 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Am-241 0.759 0.720 0.504-0.935 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Cesium-134 21.4 23.1 15.2-30.0 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Cesium-137 29.70 28.9 20.2-37.5 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Cobalt-57 28.0 27.5 19.3-35.8 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Cobalt-50 15.5 15.0 11.2-20.8 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Hydrogen-3 308 321 225-417 MAPEP False Pas MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Iron-55 0.3 0.0 Test MAPEP Manganese-MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L 54 14.4 13.9 9.7-18.1 MAPEP MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L Nickel-53 31.4 34.0 23.8-44.2 MAPEP Plutonium-MAPEP 2nd/2014 05/05/14 MaW30 Water Bq/L 238 0.754 0.828 0.580-1.075 Evaluation Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Not Accept.
Not Accept.
Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable Acceptable
M#UI Laboratories LLc P.O. Box 30712, Charleston, SC 29417 201 4 ANNUAL QUALITY ASSURANCE REPORT Page 32 of 64 Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Pu-239/240 0.6590 0.6760 0.4 73-0.879 Acceptable MAPEP Potassium-False Pos MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 40 0.460 0
Test Acceptable MAPEP Strontium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 90 8.32 8.51 5.96-11.06 Acceptable MAPEP Technetium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 99 9.5 10.3 7.2-13.4 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L U-234/233 0.210 0.225 0.158-0.293 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L 238 1.41 1.45 1.02-1.89 Acceptable MAPEP False Pos MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L
\\
Zinc-65
-0.126 0.0 Test Acceptable MAPEP MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Gross Alpha 0.96 0.85 0.255-1.443 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L Gross Beta 4.7 4.2 2.10-6.29 Acceptable MAPEP False Pos MAPEP 2nd/2014 06/05/14 MaW30 Water Bq/L lodine-129 0.0227 0.00 Test Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample 235 0.018 0.020 0.014-0.026 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter u~/sam[Jie 238 8.77 10.4 7.3-13.5 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample Total 8.80 10.4 7.3-13.5 Acceptable MAPEP Americium-MAPEP 2nd/2014 06/05/14 RdF30 Filter ug/sample 241 0.086 0.090 0.063-0.117 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cesium-134 1.85 1.91 1.34-2.48 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cesium-137 1.81 1.76 1.23-2.29 Acceptable MAPEP False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cobalt-57 0.0757 0.00 Test Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Cobalt-60 1.490 1.39 0.97-1.81 Acceptable MAPEP Manganese-False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 54 0.0138 0.00 Test Acceptable MAPEP Plutonium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 238 0.000819 0.00090 Sens. Eval.
Acceptable MAPEP 0.054-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Pu-239/240 0.071 0.7720 0.1004 Acceptable MAPEP Strontium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 90 1.19 1.18 0.83-1.53 Acceptable MAPEP 0.0137-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample U-234/233 0.0159 0.0195 0.0254 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 238 0.118 0.129 0.090-0.168 Acceptable MAPEP False Pos MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Zinc-65 0.246 0.00 Test Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Gross Alpha 1.980 1.77 0.53-3.01 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample Gross Beta 0.83 0.77 0.39-1.16 Acceptable MAPEP Americium-MAPEP 2nd/2014 06/05/14 RdF30 Filter Bq/sample 241 0.106 0.104 0.073-0.135 Acceptable MAPEP Uranium-0.0188-Not MAPEP 2nd/2014 06/05/14 RdV30 Ve~etation ug/sample 235 0.261 0.0268 0.0348 Accept.
MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Veqetation ug/sample 238 12.7 13.3 9.3-17.3 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample Total 12.7 13.3 9.3-17.3 Acceptable MAPEP Americium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation ug/sample 241 0.1100 0.108 0.076-0.140 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 33 of 64 Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cesium-134 5.65 6.04 4.23-7.85 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cesium-137 4.98 4.74 3.32-6.16 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cobalt-57 11.1 10.1 7.1-13.1 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Cobalt-60 7.21 6.93 4.85-9.01 Acceptable MAPEP Manganese-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 54 9.24 8.62 6.03-11.21 Acceptable MAPEP Plutonium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 238 0.116 0.121 0.085-0.157 Acceptable MAPEP 0.108-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Pu-239/240 0.134 0.154 0.0200 Acceptable MAPEP Strontium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 90 1.580 1.46 1.02-1.90 Acceptable MAPEP 0.0177-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample U-234/233 0.2640 0.2530 0.0329 Acceptable MAPEP Uranium-MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample 238 0.174 0.165 0.116-0.215 Acceptable MAPEP MAPEP 2nd/2014 06/05/14 RdV30 Vegetation Bq/sample Zinc-65 8.87 7.00 4.38-8.13 Acceptable MAPEP MAPEP 4th /2014 01/09/15 GrF31 Filter Bq/sample Gross Alpha 0.433 0.530 0.16-0.09 Acceptable MAPEP MAPEP 4th /2014 01/09/15 GrF31 Filter Bq/sample Gross Beta 1.060 1.060 0.53-1.59 Accep_table MAPEP Americium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 241 88.4 85.5 59.9-111.2 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cesium-134 588 622 435-809 Acceptable MAPEP False Pas MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cesium-137 1.67 Test Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bg/Kg Cobalt-57 1160 1116 781-1451 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Cobalt-60 821 779 545-1013 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Iron-55 796 680 476-884 Acceptable MAPEP Manganese-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 54 1060 1009 706-1312 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Nickel-63 924 980 686-1274 Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 238 0.92 0.48 Sens. Eval.
Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 239/240 61.5 58.6 41.0-76.2 Acceptable MAPEP Potassium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 40 879 824 577-1071 Acceptable MAPEP Strontium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 90 891 858 601-1115 Acceptable MAPEP Technetium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 99 466 589 412-766 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg U-234/233 905 89 62-116 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg 238 257 259 181-337 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaS31 Soil Bq/Kg Zinc-65 605.0 541 379-703 Acceptable MAPEP Americium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 241 0.915 0.880 0.62-1.14 Acceptable MAPEP False Pos MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cesium-134
-0.06 Test Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cesium-137 18.4 18.4 12.9-23.9 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 34 of 64 Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cobalt-57 25 24.7 17.3-32.1 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Cobalt-60 12.5 12.4 8.7-16.1 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Hydrogen-3 216 208 146-270 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Iron-55 34.0 31.5 22.1-41.0 Acceptable MAPEP Manganese-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 54 14.2 14.0 9.8-18.2 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Nickel-63 23.6 24.6 17.2-32.0 Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 238 0.547 0.618 0.433-0.803 Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 239/240 0.015 0.005 Sens. Eval.
Acceptable MAPEP Potassium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 40 174 161 113-209 Acceptable MAPEP Strontium-False Pos MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 90 0.03 Test Acceptable MAPEP Technetium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 99 6.92 6.99 4.89-9.09 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 234/233 0.206 0.205 0.144-0.267 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L 238 1.280 1.420 0.99-1.85 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Zinc-65 11.900 10.90 7.6-14.2 Acceptable MAPEP-1 4-MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Gross Alpha 0.793 0.701 0.201-1.192 Acceptable MAPEP MAPEP 4th /2014 01/09/15 MaW31 Water Bq/L Gross Beta 6.220 5.94 2.97-8.91 Acceptable MAPEP Uranium-0.0278-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample 235 0.040 0.040 0.0516 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 RdF31 Filter UQ/sample 238 19.3 20.3 14.2-26.4 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample Total 19.00 20.4 14.3-26.5 Acceptable MAPEP Americium-0.0472-MAPEP 4th /2014 01/09/15 RdF31 Filter ug/sample 241 0.0561 0.067 0.0876 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cesium-134 0.8640 0.96 0.67-1.25 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cesium-137 1.190 1.20 0.84-1.56 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdF31 Filter Bg/sample Cobalt-57 1.540 1.43 1.00-1.86 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Cobalt-60 1.200 1.10 0.77-1.43 Accep_table MAPEP Manganese-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 54 0.808 0.75 0.53-0.98 Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 238
. 0.115 0.107 0.075-0.139 Acceptable MAPEP Plutonium-0.0328-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 239/240 0.048 0.0468 0.0608 Acceptable MAPEP Strontium-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 90 0.762 0.70 0.492-0.914 Acceptable MAPEP Uranium-0.0251-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 234/233 0.037 0.0358 0.0465 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample 238 0.227 0.253 0.177-0.329 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdF31 Filter Bq/sample Zinc-65 0.779 0.76 0.53-0.99 Acceptable MAPEP Americium-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 241 0.226 0.19 0.135-0.251 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 35 of 64 Acceptance PT Quarter I Report Sample Sample Analyte I GEL Known Range/
Provider Year Date Number Media Unit Nuclide Value value Ratio Evaluation MAPEP MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample Cesium-134 4.750 5.20 3.64-6.67 Acceptable MAPEP MAPEP 4th /2014 01/09/15 RdV31 VeQetation Bq/sample Cesium-137 6.910 6.60 4.62-8.58 Acceptable MAPEP False Pas MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample Cobalt-57
-0.002 0.00 Test Acceptable MAPEP False Pas MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample Cobalt-60 0.008 0.00 Test Acceptable MAPEP Manganese-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 54 7.980 7.88 5.52-10.24 Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 238 0.001 0.001 Sens. Eva!.
Acceptable MAPEP Plutonium-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 239/240 0.1510 0.171 0.120-0.222 Acceptable MAPEP Strontium-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 90 2.330 2.32 1.62-3.02 Acceptable MAPEP Uranium-0.0326-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 234/233 0.046 0.047 0.0606 Acceptable MAPEP Uranium-MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample 238 0.332 0.324 0.227-0.421 AcceiJtable MAPEP MAPEP 4th /2014 01/09/15 RdV31 Vegetation Bq/sample Zinc-65 2.850 2.63 1.84-3.42 Acceptable MAPEP Strontium-MAPEP 4th /2014 01/09/15 SrF-31 Filter Bq/sample 89 3.62 3.79 2.65-4.93 Acceptable MAPEP Strontium-MAPEP 4th /2014 01/09/15 SrF-31 Filter Bq/sample 90 3.62 3.79 2.65-4.93 Acceptable MAPEP MAPEP 4th /2014 01/09/15 XaW-31 Water Bq/L lodine-129 4.56 4.55 3.19-5.92 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 36 of 64 TABLE 4 2014 ERA PROGRAM PERFORMANCE EVALUATION RESULTS Acceptance PT Quarter Report Sample Sample GEL Known Range/
Provider I Year Date Number Media Unit Analyte I Nuclide Value value Ratio Evaluation 1st I RAD-ERA 2014 02124114 96 Water pCiiL Barium-133 80.6 76.2 63.8-83.8 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Cesium-134 64.7 66.8 54.4-73.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi!L Cesium-137 112.0 109 98.1-122 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Cobalt-60 95.0 88.7 79.8-99.9 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCiiL Zinc-65 200 185 166-218 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Gross Alpha 34.8 36.1 18.6-46.4 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCiiL Gross Beta 19.6 22.3 13.5-30.4 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Gross Alpha 34.6 36.1 18.6-46.4 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Radium-226 16.2 16.8 12.5-19.2 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Radium-228 4.62 5.04 3.01-6.67 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Uranium (Nat) 7.39 7.23 5.51-8.53 Acceptable 1st I RAD -
Uranium (Nat)
ERA 2014 02124114 96 Water ugll mass 11.00 10.6 8.07-12.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Radium-226 15.10 16.8 12.5-19.2 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCiiL Radium-228 4.66 5.04 3.01 -6.67 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Uranium (Nat) 7.47 7.23 5.51 -8.53 Acceytable 1st I RAD-Uranium (Nat)
ERA 2014 02124114 96 Water ugll mass 11.4 10.6 8.07-12.5 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Tritium 3320 3580 3030-3950 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCi/L Strontium-89 44.1 44.4 34.4-51.6 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L Strontium-90 34.2 30.3 22.1-35.2 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL Strontium-89 38.9 44.4 34.4-51.6 Acceptable 1st I RAD -
ERA 2014 02124114 96 Water pCi!L Strontium-90 27.1 30.3 22.1-35.2 Acceptable 1st I ERA 2014 02106114 011014L Water pCiiL Strontium-89 42.3 38.7 29.3-45.7 Acceptable 1st I ERA 2014 02106114 011014L Water pCiiL Strontium-89 42.2 38.7 29.3-45.7 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCiiL lodine-131 25.2 24.4 20.2-28.9 Acceptable 1st I RAD-ERA 2014 02124114 96 Water pCi/L lodine-131 22.4 24.4 20.2-28.9 Acceptable 3rd I RAD -
ERA 2014 08125114 98 Water pCiiL Barium-133 67.8 68.7 57.3-75.6 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCiiL Cesium-134 71 72.3 59.0-79.5 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Cesium-137 161 163 147-181 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Cobalt-60 76.7 75.5 68.0-85.5 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 37 of64 Acceptance PT Quarter Report Sample Sample GEL Known Range/
Provider I Year Date Number Media Unit_ Anal~te I Nu~
Value value Ratio Eva! uatio!_L 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Zinc-65 92 82 73.8-98.5 Acceptable 3rd I RAD -
ERA 2014 08125114 98 Water pCiiL Gross Alpha 45.3 45.4 23.6-57.4 Acceptable 3rd I RAD -
ERA 2014 08125114 98 Water pCiiL Gross Beta 32.3 33.4 21.7-41.1 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Gross Alpha 48.6 45.4 23.6-57.4 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Radium-226 8.26 9.06 6.80-10.6 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCiiL Radium-226 8.54 9.06 6.80-10.6 Acceptable 3rd I RAD -
ERA 2014 08125114 98 Water pCiiL Radium-226 9.7 9.06 6.80-10.6 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Radium-228 5.07 5.07 3.03-6.79 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Radium-228 5.74 5.07.
3.03-6.79 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCiiL Uranium (Nat) 13.9 13.5 10.7-15.4 Acceptable 3rd I RAD-Uranium (Nat)
ERA 2014 08125114 98 Water ugll mass 22.25 19.8 15.6-22.6 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Uranium (Nat) 13 13.5 10.7-15.4 Accep_table 3rd I RAD-Uranium (Nat)
ERA 2014 08125114 98 Water ugll mass 20.7 19.8 15.6-22.6 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Tritium 10200 11200 9750-12300 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCi/L Tritium 10400 11200 9750-12300 Acceptable 3rd I RAD-Not ERA 2014 08125114 98 Water pCiiL Strontium-89 56.3 42.7 32.9-49.8 Acceptable 3rd I RAD-ERA 2014 08125114 98 Water pCiiL Strontium-90 14.3 31.7 23.1-36.7 Acceptable 3rd I RAD-Not ERA 2014 08125114 98 Water pCiiL Strontium-89 56.5 42.7 32.9-49.8 Acceptable 3rd l RAD-ERA 2014 08125114 98 Water pCi/L Strontium-90 26 31.7 23.1-36.7 Acceptable 3rd I RAD-ERA 2014 08125/14 98 Water pCi/L lodine-131 28.6 26.1 21.7-30.8 Acceptable 3rd I RAD-ERA 2014 08/25/14 98 Water pCi/L lodine-131 22.3 26.1 21.7-30.8 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 38 of 64 TABLE 5 2014 ERA PROGRAM (MRAD) PERFORMANCE EVALUATION RESULTS Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Provider Year Date Number Media Unit Analyte I Nuclide Value value Ratio Evaluation MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Actinium-228 1140 1240 795-1720 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Americium-241 418 399 233-518 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Bismuth-212 976 1240 330-1820 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Bismuth-214 2290 1960 1180-2820 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Cesium-134 3080 3390 2220-4070 Acceptable MRAD-6510-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Cesium-137 8310 8490 10900 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Cobalt-60 6570 6830 4620-9400 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Lead-212 1330 1240 812-1730 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Lead-214 2800 2070 1210-3090 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Manganese-54
<44.3
<1000 0-1000 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Plutonium-238 579 578 348-797 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Plutonium-239 488 471.00 308-651 Acceptable MRAD-7660-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Potassium-40 10500 10500 14100 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Strontium-90 2500 2780 1060-4390 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Thorium-234 3420 3360 1060-6320 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Zinc-65 5700 5400 4300-7180
. Acceptable MRAD-3250-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Strontium-90 6730 8530 13500 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-234 2602 3390 2070-4350 AccepJable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-238 2425 3360 2080-4260 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-Total 5027 6910 3750-9120 Acceptable MRAD-5570-ERA 2nd/2014 05/16/14 20 Soil ug/kg Uranium-Total(mass) 7110 10100 12700 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-234 3440 3390 2070-4350 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-238 3680 3360 2080-4260 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-Total 7310 6910 3750-9120 Acceptable MRAD-5570-ERA 2nd/2014 05/16/14 20 Soil ug/kg Uranium-Total(mass) 11000 10100 12700 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-234 3740 3390 2070-4350 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-238 3780 3360 2080-4260 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Soil pCi/kg Uranium-Total 7683 6910 3750-9120 Acceptable MRAD-5570-ERA 2nd/2014 05/16/14 20 Soil ug/kg Uranium-Total(mass) 11300 10100 12700 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 39 of 64 Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Provid~r Ye_ar _
Date Number Media Unit. __ _A_nal~t~ I Nuclide Value value Ratio Evaluation _
MRAD-5570-ERA 2nd/2014 05/16/14 20 Soil ug/kg Uranium-Total(mass) 11200 10100 12700 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Am-241 1670 1490 911-1980 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Cesium-134 657 646 415-839 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Cesium-137 861 880 638-1220 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Cobalt-60 997 926 639-1290 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Curium-244 514 516 253-804 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Manganese-54
<62.2
<300 0.00-300 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Plutonium-238 2230 2110 1260-2890 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Plutonium-239 3810 3740 2300-5150 Acceptable MRAD-23000-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Potassium-40 30800 31900 44800 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Strontium-90 2330 2580 1470-3420 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-234 1920 1760 1160-2260 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-238 1970 1750 1170-2220 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-Total 4025 3580 2430-4460 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation ug/kg Uranium-Total(mass)_
5920 5240 3510-6650 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Zinc-65 1030 919 663-1290 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-234 1730 1760 1160-2260 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-238 2000 1750 1170-2220 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation pCi/kg Uranium-Total 3817 3580 2430-4460 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation ug/kg Uranium-Total(mass) 5990 5240 3510-6650 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Vegetation ug/kg Uranium-Total(mass) 5620 5240 3510-6650 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Americium-241 60.2 59.7 36.8-80.8 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cesium-134 920 1010 643-1250 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cesium-137 816 828 622-1090 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Cobalt-60 1130 1120 867-1400 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Iron-55 254 240 74.4-469 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Manganese-54
<6.64
<50.0 0-50.0 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Plutonium-238 51.3 56.3 38.6-74.0 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Plutonium-239 47.5 48.6 35.2-63.5 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi!Filter Strontium-90 76.7 78.9 38.6-118 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi!Filter Uranium-234 33.8 36.4 22.6-54 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi!Filter Uranium-238 34.5 36.1 23.3-49.9 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 40 of 64 Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Proyider _
Year Date Number 1
_ _ Media Unit Analyte I Nuclide Value _ _ value Ratio Evaluation MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Uranium-Total 70.3 74.3 41.1-113 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter ug/Filter Uranium-Total(mass) 104 108 69.1-152 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Zinc-65 737 667 478-921 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Uranium-234 35.5 36.4 22.6-54 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Uranium-238 35.3 36.1 23.3-49.9 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi!Filter Uranium-Total 72.4 74.3 41.1-113 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter ug/Filter Uranium-Total(mass) 105 108 69.1-152 Acceptable
- MRAD-ERA 2nd/2014 05/16/14 20 Filter ug/Filter Uranium-Total( mass) 100 108 69.1-152 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Gross Alpha 60.9 46 15.4-71.4 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Filter pCi/Filter Gross Beta 58.9 53.8 34.0-78.4 Acceptable MRAD-Not ERA 2nd/2014 05/16/14 20 Water pCi/L Americium-241 186 114 76.8-153 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Cesium-134 1540 1660 1220-1910 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Cesium-137 2760 2690 2280-3220 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Cobalt-60 1320 1270 1100-1490 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Iron-55 1230 1200 716-1630 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Manganese-54
<7.54
<100 0.00-100 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Plutonium-238 37 44 32.6-54.9 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi!L Plutonium-239 124 160 124-202 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water
!JCiiL Strontium-90 95 890 580-1180 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-234 77.8 82.4 61.9-106 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-238 50.8 48.4 36.9-59.4 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-Total 156 168 123-217 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water ug/L Uranium-Total(mass) 233 245 195-296 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Zinc-65 2030 1800 1500-2270 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-234 82.1 82.4 61.9-106 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-238 84.6 48.4 36.9-59.4 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-Total 170 168 123-217 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water ug/L Uranium-Total(mass) 253 245 195-296 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-234 80.5 82.4 61.9-106 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-238 90.0 48.4 36.9-59.4 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water pCi/L Uranium-Total 175 168 123-217 Acceptable MRAD-ERA 2nd/2014 05/16/14 20 Water u~/L Uranium-Total(mass) 269 245 195-296 Acceptable
1!131 Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 41 of 64 Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Provider Year Date Number Media Unit Anal~te I Nuclide Value value Ratio Evaluatiof1 MRAD-ERA 2ndl2014 05116114 20 Water pCi/L Uranium-234 77.8 82.4 61.9-106 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water pCi/L Uranium-238 78.3 48.4 36.9-59.4 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water pCi/L Uranium-Total 156 168 123-217 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water ugll Uranium-Total(massl_
233 245 195-296 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water ugll Uranium-Total(mass) 232 245 195-296 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water pCiiL Gross Alpha 141.0 133 47.2-206 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water pCiiL Gross Beta 172 174.0 99.6-258 Acceptable MRAD-ERA 2ndl2014 05116114 20 Water pCiiL Tritium 5280 5580 3740-7960 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Actinium-228 1280 1240 795-1720 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Americium-241 825 763 431-956 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Bismuth-212 1620 1240 330-1820 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Bismuth-214 2900 2810 1690-4040 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Cesium-134 1960 2140 1400-2570 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Cesium-137 6760 6550 5020-8430 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Cobalt-60 4480 4260 2880-5860 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Lead-212 1260 1240 812-1730 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Lead-214 3480 2750 1610-4100 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Manganese-54
<30.0
<1000 0-1000 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Plutonium-238 732 739 444-1020 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCi/kg Plutonium-239 281 309 202-427 Acceptable 3rd I MRAD-7810-ERA 2014 11125114 21 Soil pCi/kg Potassium-40 11500 10700 14400 Acceptable 3rd I MRAD-3210-ERA 2014 11125114 21 Soil pCi/kg Strontium-90 8790 8420 13300 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Thorium-234 2000 2350 743-4420 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Zinc-65 3910 3270 2600-4350 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Uranium-234 2280 2370 1450-3040 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Uranium-238 2340 2350 1450-2980 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil pCilkg Uranium-Total 4762 4540 2360-6390 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Soil uglkg Uranium-Total(mass) 7020 7050 3890-8870 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation QCi/kg Am-241 2260 2290 1400-3505 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Cesium-134 837 849 545-1100 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Cesium-137 729 644 467-896 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Cobalt-60 818 784 541-1100 Acceptable
[ij#jll Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 42 of 64 Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Pr9vi<:Jm:.
Year Date Number Media Unit Analyte I Nuclide
~-W!!e_
1 - v~
Ratio Evaluation 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Curium-244 361 367 180-572 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCi/kg Manganese-54
<25.3
<300 0-300 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCi/kg Plutonium-238 886 862 514-1180 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Plutonium-239 675 701 430-965 Acceptable 3rd I MRAD-22300-ERA 2014 11125114 21 Vegetation pCilkg Potassium-40 35300 30900 43400 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCi/kg Strontium-90 1230 1710 975-2270 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCi/kg Uranium-234 1980 1780 1170-2290 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Uranium-238 1970 1760 1170-2240 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Uranium-Total 4038 3620 2450-4510 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation uglkg Uranium-Total(mass) 5910 5280 3540-6710 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Uranium-234 1670 1780 1170-2290 Acceptable 3rd I MRAD-ERA 2014 11125114
- 21.
Vegetation pCi!kg Uranium-238 1800 1760 1170-2240 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Uranium-Total 3556 3620 2450-4510 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation uglkg Uranium-Total(mass) 5390 5280 3540-6710 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation uglkg Uranium-Total(mass) 5860 5280 3540-6710 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Vegetation pCilkg Zinc-65 1930 1570 1130-2200 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Americium-241 41.4 38.6 23.8-52.2 Acceptable 3rd I MRAD-ERA 2014
- 11125114 21 Filter pCiiFilter Cesium-134 742 765.0 487-949 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Cesium-137 677 647 486-850 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Cobalt-60 543 523 405-653 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Iron-55 117 120.0 37.2-234 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi!Filter Manganese-54
<5.87
<50 0.00-50.0 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter ugiFilter Plutonium-238 32.9 35.7 24.5-46.9 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Plutonium-239 26.8 29.1 21.1-38.0 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Strontium-90 187 168 82.1-252 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Uranium-234 26 28 27.8-41.9 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Uranium-238 28 27.60 17.8-38.2 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Uranium-Total 56 57 31.4-86.3 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter ugiFilter Uranium-Total(mass) 82.6 82.7 52.9-116 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi!Filter Zinc-65 629 547 392-755 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Uranium-234 28 28 27.8-41.9 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Uranium-238 25 27.60 17.8-38.2 Acceptable
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 43 of 64 Acceptanc PT Quarter I Report Sample Sample GEL Known e Range/
Provider Year Date~
Number ~Media.. ~
Unit Analyte I Nu_clide 1.-Y~ Iue value R~li_<?___
Evaluation 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Uranium-Total 55 57 31.4-86.3 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter ugiFilter Uranium-Total(mass) 75.1 82.7 52.9-116 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter ugiFilter Uranium-Total(mass) 90.7 82.7 52.9-116 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCi/Filter Gross Alpha 47.4 36.9 12.4-57.3 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Filter pCiiFilter Gross Beta 27.2 21.1 13.3-30.8 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Americium-241 72.4 68.6 46.2-92.0 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Cesium-134 816.0 850 624-977 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Cesium-137 1310 1240 1060-1490 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Cobalt-60 1130 1070 930-1250 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Iron-55 130 134 79.9-182 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Manganese-54
<6.34
<100 0.00-100 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Plutonium-238 35 33 24.6-41.4 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Plutonium-239 46.4 51 39.7-64.4 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Strontium-90 300 254 165-336 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-234 42 44 32.9-56.5 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Uranium-238 50 43.50 33.2-53.4 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-Total 92 89 65.5-115 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water ug/L Uranium-Total(mass) 137 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Zinc-65 1070 921 768-1160 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-234 43 44 32.9-56.5 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-238 45 43.50 33.2-53.4 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-Total 90 89 65.5-115 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water ug/L Uranium-Total(mass) 134 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water Q_Ci/L Uranium-234 49 44 32.9-56.5 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCiiL Uranium-238 42 43.50 33.2-53.4 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Uranium-Total 93 89 65.5-115 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water ug/L Uranium-Total(mass) 126 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water ug/L Uranium-Total(mass) 144 130 104-157 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Gross Alpha 96.2 98 34.8-152 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Gross Beta 86.1 77.5 44.4-115 Acceptable 3rd I MRAD-ERA 2014 11125114 21 Water pCi/L Tritium 5490 5500 3680-7840 Acceptable
Laboratories LLC P. 0. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 44 of 64 FIGURE 1 COBALT -60 PERFORMANCE EVALUATION RESULTS AND % BIAS 100.0 75.0 50.0 25.0 U)
.~
0.0 OJ
~
0
-25.0
-50.0
-75.0
-100.0 2014 Cobalt-GO Performance Evaluation Results and % Bias January 2014-December 2014
- - - - Lower Control Limit
Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 45 of 64 FIGURE 2 CESIUM-137 PERFORMANCE EVALUATION RESULTS AND % BIAS 2014 Cesium-137 Performance Evaluation Results and% Bias 100.0.----------------------------------,
75.0 50.0 25.0
-25.0
-50.0
-75.0
-100.0...L.__ ___________________________
____J January 2014-December 2014
- - - - Lower Control Limit
Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 46 of 64 FIGURE 3 TRITIUM PERFORMANCE EVALUATION RESULTS AND% BIAS 100.0 75.0 50.0 25.0
(/)
cu iii 0.0
~
0
-25.0
-50.0
-75.0
-100.0 2014 Tritium Performance Evaluation Results and% Bias January 2014-December 2014
--H-3
- - - - Lower Control Limit
- - - - Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 47 of 64
(/)
ca iil
'cf!.
FIGURE 4 STRONTIUM-SO PERFORMANCE EVALUATION RESULTS AND% BIAS 2014 Strontium-90 Performance Evaluation Results and% Bias 100.0.---------------------------------,
50.0 0.0
-50.0
-100.0...L...------------------------------1 January 2014-December 2014
-- Strontium-90
- - - - Lower Control Limit
- --- Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 48 of 64 FIGURE 5 GROSS ALPHA PERFORMANCE EVALUATION RESULTS AND% BIAS
(/)
.5!1 2014 Gross Alpha Performance Evaluation Results and% Bias 100.0..,....-----------------------------,
50.0 m
o.o
";fl.
-50.0
-100.0 -'---------------------------=-----'
January 2014-December 2014
_._Gross Alpha
- - - - Lower Control Limit
Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 49 of 64 FIGURE 6 GROSS BETA PERFORMANCE EVALUATION RESULTS AND % BIAS 1/)
ns 2014 Gross Beta Performance Evaluation Results and% Bias 100.0 --.-------=---------:----:-----::---:-------:------::----::----------::-:--------,
50.0 iii 0.0
~
-50.0
-100.0 _.__ __
_;__....:........,__:---=.... _
January 2014-December 2014
__._Gross Beta
- - - - Lower Control Limit
Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 50 of 64 FIGURE 7 IODINE-131 PERFORMANCE EVALUATION RESULTS AND% BIAS 100.0 75.0 50.0 25.0 tiJ cu m 0.0
-25.0
-50.0
-75.0
-100.0 20141odine-131 Performance Evaluation Results and% Bias January 2014-December 2014
--lodine-131
-- - - Lower Control Limit
Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 51 of 64 100.0 75.0 50.0 25.0
(/)
<" iii 0.0 "eft.
-25.0
-50.0
-75.0
-100.0 FIGURE 8 AMERICIUM-241 PERFORMANCE EVALUATION RESULTS AND% BIAS 2014 Americium-241 Performance Evaluation Results and % Bias January 2014-December 2014
--Am-241
- - - - Lower Control Limit
-- - - Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 52 of 64 FIGURE 9 PLUTONIUM-238 PERFORMANCE EVALUATION RESULTS AND % BIAS II) ro 2014 Plutonium-238 Performance Evaluation Results and% Bias 100.0.--------------------------------,
75.0 50.0 25.0 iii 0.0
~
-25.0
-50.0
-75.0
-100.0 -'---------------------------------'
January 2014-December 2014
Plutonium-238
- - - - Lower Control Limit
- --- Upper Control Limit
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 53 of 64 TABLE 6 REMP INTRA-LABORATORY DATA
SUMMARY
- BIAS AND PRECISION BY MATRIX Bias Criteria { + I - 25°/o Precision Criteria {Note 1}
WITHIN OUTSIDE WITHIN OUTSIDE REMP 2014 CRITERIA CRITERIA CRITERIA CRITERIA MILK Gas Flow Sr 2nd count 36 0
36 0
Gas Flow Total Strontium 23 0
23 0
Gamma Spec Liquid RAD A-013 with Ba, La 48 0
109 0
SOLID LSC Iron-55 3
0 3
0 Gamma Spec Solid RAD A-013 30 0
43 0
0 3
0 Gas Flow Sr 2nd count 5
0 5
0 Gas Flow Total Strontium 5
0 5
0 Gamma Spec Solid RAD A-013 with Ba, La 2
0 8
0 Gamma Spec Solid RAD A-013 with Iodine 6
0 7
0 FILTER Gas Flow Sr 2nd Count 5
0 5
0 Gross A & B 429 0
429 0
Gas Flow Sr-90 1
0 1
0 Gamma Spec Filter 45 0
47 0
LIQUID Alpha Spec Uranium 1
0 2
0 Tritium 206 0
205 0
0 1
0 LSC Iron-55 12 0
12 0
13 0
Gamma Spec Liquid RAD A-013 4
0 4
0 Alpha Spec Am243 6
0 6
0 Gamma Iodine-131 28 0
28 0
Alpha Spec Plutonium 10 0
10 0
Gas Flow Sr 2nd count 15 0
15 0
Alpha Spec Am241 Curium 8
0 8
0 Gas Flow Total Strontium 30 0
31 0
Gross Alpha Non Vol Beta 45 0
45 0
Gamma Spec Liquid RAD A-013 with Ba, La 84 0
159 0
Gamma SQ_ec Liquid RAD A-013 with Iodine 40 0
40 0
TISSUE Gamma Spec Solid RAD A-013 48 0
46 0
Gas Flow Sr 2nd count 8
0 8
0 Gas Flow Total Strontium 17 0
17 0
Gamma Spec Solid RAD A-013 with Ba, La 10 0
10 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 54 of 64 Gamma Spec Solid RAD A-013 with Iodine I 23 I
0 I
22 I
0 SEA WATER LSC Iron-55 5
0 6
0 6
0 Gas Flow Total Strontium 6
0 6
0 Gross Alpha Non Vol Beta 6
0 6
0 Gamma Spec Liquid RAD A-013 with Iodine 7
0 11 0
VEGETATION Gas Flow Sr 2nd count 10 0
10 0
Gamma Spec Solid RAD A-013 with Iodine 86 0
96 0
AIR CHARCOAL Gamma Iodine 131 RAD A-013 560 0
606 0
Carbon-14 (Ascarite/Soda Lime Filter per Liter) 28 0
28 0
DRINKING WATER Tritium 39 0
40 0
LSC Iron-55 17 0
16 0
15 0
Gamma Iodine-131 27 0
26 0
Gas Flow Sr 2nd count 12 0
12 0
Gas Flow Total Strontium 19 0
18 0
Gross Alpha Non Vol Beta 72 0
73 0
0 Gamma Spec Liguid RAD A-013 with Ba La 35 0
75 Total 2200 2456 Note 1: The RPD must be 20 percent or less, if both samples are greater than 5 times the MDC. If both results are less than 5 times MDC, then the RPD must be equal to or less than 100%. If one result is above the MDC and the other is below the MDC, then the RPD can be calculated using the MDC for the result of the one below the MDC. The RPD must be 100% or less. In the situation where both results are above the MDC but one result is greater than 5 times the MDC and the other is less than 5 times the MDC, the RPD must be less than or equal to 20%. If both results are below MDC, then the limits on% RPD are not applicable.
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 55 of 64 TABLE 7 ALL RADIOLOGICAL INTRA-LABORATORY DATA
SUMMARY
BIAS AND PRECISION BY MATRIX:
Bias Criteria ( + I - 25°/o Precision Criteria (Note 1)
WITHIN OUTSIDE WITHIN OUTSIDE Total Radiological 2014 CRITERIA CRITERIA CRITERIA CRITERIA MILK Gamma Iodine-129 0
0 1
0 Gamma Iodine-131 36 0
110 0
Gas Flow Sr 2nd count 36 0
36 0
Gas Flow Strontium 90 5
0 5
0 Gas Flow Total Strontium 23 0
23 0
Gamma Spec Liquid RAD A-013 with Ba, La 48 0
109 0
Gamma Spec Liquid RAD A-013 with Iodine 3
0 4
0 SOLID Gamma Percent Leach 5
0 0
0 Gas Flow Radium 228 16 0
20 0
Tritium 211 0
247 0
Tritium by Combustion 1
0 1
0 Carbon-14 130 0
181 0
LSC Iron-55 103 0
121 0
Alpha Spec Polonium Solid 52 0
54 0
Gamma Nickel 59 RAD A-022 99 0
117 0
LSC Chlorine-36 in Solids 4
0 4
0 Gamma S12_ec Ra226 RAD A-013 21 0
24 0
Gamma Spec Solid RAD A-013 649 0
812 0
154 0
202 0
Technetium-99 224 0
250 0
Gamma Spec Liquid RAD A-013 2
0 2
0 ICP-MS Technetium-99 in Soil 61 0
60 0
11 0
Total Activity, 4
0
' 4 0
Tritium 16 0
17 0
Alpha Spec Am243 23 0
37 0
Gamma Iodine-129 100 0
120 0
Gas Flow Lead 210 6
0 6
0 Total Uranium KPA 7
0 10 0
Alpha Spec Uranium 214 0
309 0
LSC Promethium 147 2
0 2
0 LSC, Rapid Strontium 89 and 90 42 0
61 0
Alpha Spec Thorium 152 0
196 0
ICP-MS Uranium-233, 234 in Solid 49 0
47 0
Alpha Spec Plutonium 231 0
240 0
ICP-MS Technetium-99 Prep in Soil 62 0
61 0
Alpha Spec Neptunium 213 0
237 0
Alpha Sgec Plutonium 158 0
206 0
Gamma Spec Solid with Ra226 Ra228 9
0 13 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 56 of 64 Gas Flow Sr 2nd count 21 0
25 0
Gas Flow Strontium 90 195 0
201 0
Gas Flow Total Radium 2
0 3
0 Lucas Cell Radium 226 38 0
47 0
Total Activity Screen 9
0 10 0
Alpha Spec Am241 Curium 304 0
339 0
Alpha Spec Total Uranium 4
0 8
0 Gas Flow Total Strontium 43 0
46 0
Gross Alpha Non Vol Beta 1
0 1
0 ICP-MS Uranium-233, 234 Prep in Solid 49 0
48 0
ICP-MS Uranium-235 236, 238 in Solid 60 0
81 0
Gamma Spec Solid RAD A-013 with Ba La 2
0 8
0 Gamma Spec Solid RAD A-013 with Iodine 6
0 7
0 GFC Chlorine-36 in Solids 3
0 3
0 Gamma Spec Solid RAD A-013 _(pCi/Sample) 2 0
2 0
Tritium 8
0 8
0 Alpha Spec Am241 (pCi/Sample) 2 0
2 0
ICP-MS Uranium-234, 235, 236, 238 in Solid 148 0
132 0
ICP-MS Uranium-235, 236, 238 Prep in Solid 50 0
49 0
Alpha Spec Thorium 1
0 1
0 Alpha Spec Uranium 1
0 1
0 Gross Alpha/Beta 235 0
316 3
Alpha Spec Neptunium 1
0 1
0 Gas Flow Sr 2nd count 2
0 1
0 Gross Alpha/Beta (Americium Calibration) Solid 2
0 3
0 ICP-MS Uranium-234 235, 236 238 Prep in Solid 69 0
65 0
FILTER Alpha Spec Uranium 14 0
18 0
Alpha Spec Polonium 1
0 5
0 Gamma I-131, filter 4
0 4
102 0
Tritium 76 0
112 0
Carbon-14 35 0
66 0
Nickel-63 0
0 8
0 LSC Iron-55 69 0
84 0
Gamma Nickel 59 RAD A-022 55 0
68 0
78 0
Technetium-99 51 0
75 0
Gamma Spec Filter RAD A-013 143 0
174 6
Alphaspec Np Filter per Liter 8
0 13 0
Alphaspec Pu Filter per Liter 11 0
22 0
Gamma Iodine-125 5
0 0
0 Gamma Iodine-129 46 0
60 0
Gross Alpha/Beta 5
0 5
0 Alpha Spec Am243 10 0
28 0
Gas Flow Lead 210 0
0 4
0 LSC Plutonium Filter per Liter 9
0 15 0
Total Uranium KPA 9
0 14 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 57 of 64 Alpha Spec Uranium 55 0
96 0
LSC Promethium 147 1
0 2
0 LSC Rapid Strontium 89 and 90 72 0
94 0
Alpha Spec Thorium 42 0
66 0
Gas Flow Radium 228 1
0 1
0 Alpha Spec Plutonium 81 0
98 0
ICP-MS Uranium-233, 234 in Filter 0
0 3
0 Alpha Spec Neptunium 62 0
83 0
Alpha Spec Plutonium 66 0
96 0
Alpha Spec Polonium,(Filter/Liter) 0 0
14 0
Alpha Spec Radium 226 0
0 2
0 Gas Flow Sr 2nd Count 72 0
81 1
Gas Flow Strontium 90 61 0
68 0
Lucas Cell Radium-226 1
0 1
0 Alpha Spec Am241Curium 95 0
117 0
Gas Flow Total Strontium 5
0 5
0 ICP-MS Uranium-233, 234 Prep in Filter 0
0 3
0 ICP-MS Uranium-235, 236, 238 in Filter 0
0 6
0 Total Activity in Filter, 1
0 10 0
Alphaspec Am241 Curium Filter per Liter 15 0
20 0
Tritium 86 0
89 0
Gamma Spec Filter RAD A-013 Direct Count 6
0 6
0 Carbon-14 12 0
12 0
GFC Chlorine-36 in Filters PL 1
0 1
0 Direct Count-Gross Alpha/Beta 48 0
1 0
Gross Alpha/Beta 48 0
60 0
ICP-MS Uranium-234 235 236, 238 in Filter 4
0 6
0 ICP-MS Uranium-235, 236, 238 Prep in Filter 0
0 3
0 Alpha Spec U 13 0
35 0
Gross A & B 497 0
473 0
LSC Iron-55 8
0 19 0
0 13 0
Gas Flow Sr-90 6
0 13 0
19 0
Gas Flow Pb-210 8
0 22 0
Gas Flow Ra-228 5
0 10 0
Gamma Iodine 129 8
0 8
0 ICP-MS Uranium-234, 235, 236 238 Prep in Filter 2
0 3
0 Gamma Spec Filter 97 0
117 0
Lucas Cell Ra-226 8
0 23 0
Alpha Spec Thorium 7
0 22 0
LIQUID Alpha Spec Uranium 390 0
553 0
Alpha Spec Polonium 4
0 7
0 Electrolytic Tritium 14 0
25 0
Tritium 1125 0
1177 0
Carbon-14 149 0
161 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 58 of 64 Plutonium 43 0
63 0
0 4
0 LSC Iron-55 192 0
233 0
Gamma Nickel 59 RAD A-022 18 0
21 0
Gamma Iodine 131 RAD A-013 2
0 2
0 Gamma Radium 228 RAD A-013 3
0 3
236 0
21 0
Technetium-99 377 0
425 0
Gamma Spec Liquid RAD A-013 702 0
732 0
Alpha Spec Total U RAD A-011 31 0
56 0
0 2
0 Alpha Spec Am243 17 0
18 0
Gamma Iodine-129 80 0
92 0
Gamma Iodine-131 28 0
28 0
ICP-MS Technetium-99 in Water 8
0 31 0
Gas Flow Lead 210 19 0
19 0
Total Uranium KPA 101 0
203 0
LSC Promethium 147 4
0 4
0 LSC, Rapid Strontium 89 and 90 7
0 8
0 Alpha Spec Thorium 145 0
186 0
Gas Flow Radium 228 171 0
206 0
Gas Flow Radium 228 40 0
37 0
Gas Flow Radium 228 1
0 1
0 Alpha Spec Plutonium 288 0
387 0
0 1
0 Alpha Spec Neptunium 90 0
141 0
Alpha Spec Plutonium 21 0
49 0
Alpha Spec Radium 226 7
0 7
0 Gas Flow Sr 2nd count 191 0
199 0
Gas Flow Strontium 90 365 0
422 0
Gas Flow Strontium 90 1
0 1
0 Gas Flow Total Radium 78 0
103 0
ICP-MS Technetium-99 PreQ in Water 8
0 32 0
ICP-MS Uranium-233 234 in Liquid 6
0 11 0
LSC Calcuim 45 1
0 1
0 Lucas Cell Radium 226 310 0
366 0
Lucas Cell Radium-226 10 0
10 0
Total Activity Screen 7
0 7
0 Chlorine-36 in Liquids 13 0
14 0
Alpha Spec Am241 Curium 217 0
333 0
Gas Flow Total Strontium 112 0
116 0
Gross Alpha Non Vol Beta 980 0
1167 0
LSC Phosphorus-32 2
0 3
0 Lucas Cell Radium 226 by Method Ra-04 2
0 2
0 ICP-MS Uranium-233, 234 Prep in Liquid 6
0 11 0
Tritium in Drinking Water by_ EPA 906.0 9
0 12 0
Gamma Spec Liquid RAD A-013 with Ba, La 84 0
159 0
Gamma Spec Liquid RAD A-013 with Iodine 162 0
189 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 59 of 64 Gas Flow Strontium 89 & 90 5
0 3
0 ICP-MS Uranium-235 236, 238 in Liquid 10 0
18 0
Gas Flow Total Alpha Radium 6
0 7
0 Gross Alpha Co-precipitation 3
0 13 0
ICP-MS Uranium-235, 236, 238 Prep in Liquid 6
0 11 0
ICP-MS Uranium-234, 235, 236, 238 in Liquid 31 0
74 0
Gross Alpha Beta (Americium Calibration) Liquid 32 0
46 0
ICP-MS Uranium-234 235, 236 238 Prep in Liquid 15 0
38 0
Alpha/Beta (Americium Calibration) Drinking Water 23 0
18 0
TISSUE Carbon-14 3
0 3
0 Gamma Spec Solid RAD A-013 76 0
78 0
0 4
0 Tritium 1
0 1
0 Alpha Spec Uranium 5
0 8
0 Alpha Spec Plutonium 5
0 10 0
Gas Flow Sr 2nd count 8
0 8
0 Gas Flow Strontium 90 11 0
12 0
Alpha Spec Am241 Curium 2
0 2
0 Gas Flow Total Strontium 17 0
17 0
Gamma Spec Solid RAD A-013 with Ba~_ La 10 0
10 0
Gamma Spec Solid RAD A-013 with Iodine 23 0
22 0
Gross Alpha/Beta 2
0 2
0 SEAWATER LSC Iron-55 5
0 6
0 6
0 Gas Flow Total Strontium 6
0 6
0 Gross Alpha Non Vol Beta 6
0 6
0 Gamma Spec Liquid RAD A-013 with Iodine 7
0 11 0
VEGETATION LSC Iron-55 2
0 2
0 Gamma Nickel 59 RAD A-022 1
0 0
0 Gamma Spec Solid RAD A-013 26 0
25 0
0 1
0 1
0 Technetium-99 4
0 3
0 Tritium 11 0
11 0
Gamma Iodine-129 1
0 0
0 Gas Flow Lead 210 2
0 3
0 Total Uranium KPA 4
0 4
0 Alpha Spec Uranium 22 0
22 0
Alpha Spec Thorium 5
0 5
0 Alpha Spec Plutonium 13 0
11 0
Alpha Spec Neptunium 1
0 1
0 Alpha Spec Plutonium 1
0 1
0 Gas Flow Sr 2nd count 10 0
10 0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 60 of 64 Gas Flow Strontium 90 12 0
11 0
Gas Flow Total Radium 2
0 2
0 Alpha Spec Am241 Curium 6
0 6
0 Gamma Spec Solid RAD A-013 with Iodine 86 0
96 0
Gamma Spec Solid RAD A-013 (pCi/Sample) 2 0
2 0
Alpha Spec Am241 (pCi/Sample) 1 0
2 0
ICP-MS Uranium-234, 235, 236, 238 in Solid 12 0
7 0
Alpha Spec Uranium 0
0 2
0 Gross Alpha/Beta 7
0 9
0 Alpha Spec Plutonium 0
0 2
0 Gas Flow Strontium 90 4
0 2
0 ICP-MS Uranium-234, 235 236 238 Prep in Solid 7
0 4
0 AIR CHARCOAL Gamma Iodine 131 RAD A-013 560 0
606 0
Gamma Iodine-129 7
0 6
0 Carbon-14 7
0 7
0 Carbon-14 (Ascarite/Soda Lime Filter per Liter) 28 0
28 0
Gamma Iodine 129 7
0 7
0 Gamma Spec Filter 7
0 7
0 DRINKING WATER Alpha Spec Uranium 4
0 5
0 Alpha Spec Polonium 1
0 25 0
Tritium 39 0
40 0
0 2
0 Iodine-131 2
0 2
0 LSC Iron-55 17 0
16 0
15 0
13 0
0 1
0 Gamma Spec Liquid RAD A-013 17 0
18 0
Gamma Iodine-129 2
0 4
0 Gamma Iodine-131 27 0
26 0
Gas Flow Lead 210 4
0 3
0 Total Uranium KPA 17 0
34 0
Alpha Spec Thorium 1
0 1
0 Gas Flow Radium 228 22 0
26 0
Alpha Spec Plutonium 3
0 3
0 Gas Flow Sr 2nd count 12 0
12 0
Gas Flow Strontium 90 20 0
22 0
LSC Calcuim 45 2
0 2
0 Lucas Cell Radium-226 23 0
49 0
Alpha Spec Am241 Curium 2
0 2
0 Gas Flow Total Strontium 19 0
18 0
Gross Alpha Non Vol Beta 247 0
214 0
Tritium in Drinking Water by EPA 906.0 28 0
26 0
Gamma Spec Liquid RAD A-013 with Ba, La 35 0
75 0
Gas Flow Strontium 89 & 90 17 0
11 0
Gas Flow Total Alpha Radium 1
0 1
0
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 61 of 64 Gross Alpha Co-precipitation 99 0
91 0
Alpha/Beta (Americium Calibration) Drinking Water 16 0
16 0
0 ECLS-R-GA NJ 48 Hr Rapid Gross Alpha 7
0 7
Total 16535 19734 Note 1: The RPD must be 20 percent or less, if both samples are greater than 5 times the MDC. If both results are less than 5 times MDC, then the RPD must be equal to or less than 100%. If one result is above the MDC and the other is below the MDC, then the RPD can be calculated using the MDC for the result of the one below the MDC. The RPD must be 100% or less. In the situation where both results are above the MDC but one result is greater than 5 times the MDC and the other is less than 5 times the MDC, the RPD must be less than or equal to 20%. If both results are below MDC, then the limits on% RPD are not applicable.
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 62 of 64 TABLE 8 2014 CORRECTIVE ACTION REPORT
SUMMARY
CORRECTIVE ACTION 10#
PE FAILURE CARR140605-879 ISO Documentation of PT Failures in MAPEP-14-RdV30 for Uranium 235 in Vegetation by ICP/MS and 14-MaS30 Uranium-233/234 and Uranium 238 by Alpha Spec.
DISPOSITION Root Cause Analysis of MAPEP-14-RdV28 in vegetation for Uranium-235 by ICP/MS The root cause of this failure was human error and inattention to detail. The QAO inadvertently entered the incorrect activity for this parameter when she was entering the results on the MAPEP website. 0.261 ug/sample instead of 0.0261 ug/sample was entered. The data entry error was not caught during the GL review process. MAPEP results only are peer reviewed by the GL of the applicable area to ensure that the data was entered correctly.
A second PT was successfully analyzed for this matrix.
Uranium-234/233, and Uranium-238 in soil by Alpha Spec:
Following reviews of our process and data and conversations with personnel from the affected laboratories, it was determined that all failures were due to an incomplete sample digestion. A total digestion technique using Hydrofluoric Acid was performed on the sample. However, this digestion was not vigorous enough to extract all the U-234 and U-238 from the soil because the analytes were fused into the soil at an extremely high temperature. Due to the high number of labs that received a Not Acceptable rating for this analysis, MAPEP has posted an explanation on the preparation of the Uranium Soil standard on their website.
Permanent Corrective/Preventive Actions or Improvements :
Upon notification of the failure, the sample was re-digested using a Sodium Hydroxide fusion method prior to ion-exchange separation chemistry. The results for both the U-234 and U-238 fall within acceptable range. In the future, all MAPEP soil samples will be analyzed with a NaOH fusion dissolution technique. Our analytical procedures provide the flexibility to perform different extraction techniques (leaching,
[!]#II Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 63 of 64 CARR140520-87 4 ISO Documentation of PT Failures in
-MRAD-20 for Americium-241 in water.
HF dissolution) based on client requests. For our DOE clients, complete dissolution using HF has been the approved method for Uranium. Some clients also ask for the Uranium analysis using a leach procedure. In all cases, GEL performs the required contractual procedure.for the analysis.
A second PT was successfully analyzed for this matrix.
Root Cause Analysis After a thorough review of all data, a definite reason for the failure could not be determined.
The following steps were taken to prove that this elevated bias was an isolated occurrence and that our overall process is within control.
- 1. The batch quality control samples were reviewed and found to be compliant. The recoveries in the Laboratory Control Sample (LCS) recovered at 98.2%. Two sample duplicates were also prepared in the batch. The RPDs were 4.8 and 8.6.
- 2. The sample was re-analyzed in duplicate after the report was received. One with our normal Am-243 tracer, and another with Cm-244 tracer. Both of the reanalysis confirm the original reported result (which is outside the range of acceptable results).
Control charts for all Am tracer recoveries were also reviewed to determine if there may be an issue with the tracers. While there is a slight bias in the average LCS recovery, it was not significant enough to consider abnormal, and did not come close to accounting for the high result on this analysis.
Additionally, since the sample was reanalyzed using two different tracers and achieved the same result, a tracer issue was ruled out as the potential culprit Permanent Corrective/Preventive Actions or Improvements :
The laboratory must assume unidentified random error caused the elevated bias because all quality control criteria were met for the batch. Additionally, a well characterized performance evaluation sample from another vendor was prepped and analyzed a few weeks after this sample. The Am-241 recovered at 105% for this sample and fell well within its acceptance range.
A second PT was successfully analyzed for this matrix.
Laboratories LLC P.O. Box 30712, Charleston, SC 29417 2014 ANNUAL QUALITY ASSURANCE REPORT Page 64 of 64 CARR140825-902 For Failures of RAD-98 for Strontium-Root Cause Analysis of Strontium-89 (Sr-89) 89 in Water After a review of the data, an apparent reason for this discrepancy could not be determined. The following steps were taken to prove that this high bias was an isolated occurrence and that our overall process is within control.
- 1. The batch quality control samples were reviewed and found to be compliant. The LCS recovered at 103%.
- 2. Laboratory control data were also reviewed for trends. None was noted.
- 3. The instrument calibrations were reviewed for positive biases that could have attributed to this failure. None were noted.
- 4. Sample duplicates were also prepared and counted along with the reported result. All results fell within the method's acceptance range for duplicates.
Permanent Corrective/Preventive Actions or Improvements The laboratory must assume an unidentified random error caused the high bias for this batch. While the LCS recovered outside to its acceptance range, the matrix spike (MS) recovery fell within both the acceptance range for the MS (80%-120%) and the acceptance range for the LCS (90%-
11 0%). The result was also confirmed using Method LAB PBMS-A-004. The lab will continue to monitor the recoveries of this radionuclide to ensure that there are no issues.
A second PT was successfully analyzed for this matrix.