ML19099A102
| ML19099A102 | |
| Person / Time | |
|---|---|
| Site: | 07000371 (SNM-0368) |
| Issue date: | 03/01/2018 |
| From: | Cabrera Services |
| To: | NRC Region 1 |
| Shared Package | |
| ml051740240 | List:
|
| References | |
| Download: ML19099A102 (2) | |
Text
Last Measurement Q.A. Report 4/12/17 9:04:41 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 4/12/17 9:04:41 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 041 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/12/17 8:58:18 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 352.1 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.6382E-001 2.5794E-001
[SD: 9.6039E-001+/- 0.013] < : : : >
Peak FWHM (1407.92 keV) 2.3875E+000 2.4245E-001
[SD: 2.3576E+000+/- 0.123] < : : : >
Peak FWHM (964.13 keV) 1.9197E+000 1.2151E-001
[SD: 1.9102E+000+/- 0.078] < : : : >
Peak energy (964.13 keV) 9.6394E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2172E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4079E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4692E-001 -5.0090E-001
[SD: 5.5039E-001+/- 0.007] < : : : >
Corrected activity (964. 4.8132E-001 -3.7854E-001
[SD: 4.8446E-001+/- 0.008] < : : : >
Corrected activity (1407 4.9018E-001 6.4806E-001
[SD: 4.8501E-001+/- 0.008] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)