Last Measurement Q.A. Report 4/11/17 7:46:47 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 4/11/17 7:46:47 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 041 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/11/17 7:39:45 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 352.1 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.5951E-001 2.3035E-001
[SD: 9.5725E-001+/- 0.010] < : : : >
Peak FWHM (1407.92 keV) 2.3053E+000 -3.4336E-001
[SD: 2.3488E+000+/- 0.127] < : : : >
Peak FWHM (964.13 keV) 1.9746E+000 8.5674E-001
[SD: 1.9051E+000+/- 0.081] < : : : >
Peak energy (964.13 keV) 9.6378E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2196E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4079E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4304E-001 -1.1441E+000
[SD: 5.5114E-001+/- 0.007] < : : : >
Corrected activity (964. 4.7932E-001 -6.2119E-001
[SD: 4.8465E-001+/- 0.009] < : : : >
Corrected activity (1407 4.8032E-001 -5.9111E-001
[SD: 4.8529E-001+/- 0.008] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)