ML19099A099

From kanterella
Jump to navigation Jump to search
Appendix K, Qaqc Records, 2017, Isos QCs - Daily QC 04192017
ML19099A099
Person / Time
Site: 07000371
Issue date: 03/01/2018
From:
Cabrera Services
To:
NRC Region 1
Shared Package
ml051740240 List: ... further results
References
Download: ML19099A099 (2)


Text

Last Measurement Q.A. Report 4/19/17 9:06:51 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 4/19/17 9:06:51 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 041 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/19/17 8:51:44 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 352.2 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.5663E-001 -3.2220E-001

[SD: 9.6073E-001+/- 0.013] < : : : >

Peak FWHM (1407.92 keV) 2.2728E+000 -7.1762E-001

[SD: 2.3592E+000+/- 0.120] < : : : >

Peak FWHM (964.13 keV) 1.8020E+000 -1.3850E+000

[SD: 1.9101E+000+/- 0.078] < : : : >

Peak energy (964.13 keV) 9.6430E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2198E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4087E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.5314E-001 4.6293E-001

[SD: 5.5003E-001+/- 0.007] < : : : >

Corrected activity (964. 4.8545E-001 1.3679E-001

[SD: 4.8437E-001+/- 0.008] < : : : >

Corrected activity (1407 4.8894E-001 4.7033E-001

[SD: 4.8532E-001+/- 0.008] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)