Text

Last Measurement Q.A. Report 4/25/17 8:22:26 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 4/25/17 8:22:26 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 042 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/25/17 7:58:28 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.4 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.5164E-001 -8.0515E-001

[SD: 9.6196E-001+/- 0.013] < : : : >

Peak FWHM (1407.92 keV) 2.1972E+000 -1.3822E+000

[SD: 2.3754E+000+/- 0.129] < : : : >

Peak FWHM (964.13 keV) 1.7870E+000 -1.6059E+000

[SD: 1.9115E+000+/- 0.078] < : : : >

Peak energy (964.13 keV) 9.6434E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2198E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4087E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.5027E-001 8.2825E-002

[SD: 5.4974E-001+/- 0.006] < : : : >

Corrected activity (964. 4.8911E-001 6.5675E-001

[SD: 4.8422E-001+/- 0.007] < : : : >

Corrected activity (1407 4.8666E-001 1.9682E-001

[SD: 4.8523E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)