Last Measurement Q.A. Report 4/21/17 8:47:53 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 4/21/17 8:47:53 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 042120 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/21/17 7:49:33 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.8 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.6654E-001 3.6201E-001

[SD: 9.6184E-001+/- 0.013] < : : : >

Peak FWHM (1407.92 keV) 2.4682E+000 7.3386E-001

[SD: 2.3730E+000+/- 0.130] < : : : >

Peak FWHM (964.13 keV) 1.9122E+000 9.4646E-003

[SD: 1.9115E+000+/- 0.079] < : : : >

Peak energy (964.13 keV) 9.6414E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2195E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4084E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4334E-001 -1.0204E+000

[SD: 5.4990E-001+/- 0.006] < : : : >

Corrected activity (964. 4.8470E-001 6.5529E-002

[SD: 4.8420E-001+/- 0.008] < : : : >

Corrected activity (1407 4.8455E-001 -9.5247E-002

[SD: 4.8525E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)