Last Measurement Q.A. Report 5/4/17 7:18:25 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 5/4/17 7:18:25 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 0504201 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 5/4/17 7:10:51 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.8 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.5916E-001 -2.2689E-001

[SD: 9.6189E-001+/- 0.012] < : : : >

Peak FWHM (1407.92 keV) 2.6177E+000 1.6698E+000

[SD: 2.3882E+000+/- 0.137] < : : : >

Peak FWHM (964.13 keV) 2.0178E+000 1.1760E+000

[SD: 1.9188E+000+/- 0.084] < : : : >

Peak energy (964.13 keV) 9.6411E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2195E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4084E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4270E-001 -1.0054E+000

[SD: 5.4902E-001+/- 0.006] < : : : >

Corrected activity (964. 4.8612E-001 2.2128E-001

[SD: 4.8455E-001+/- 0.007] < : : : >

Corrected activity (1407 4.8218E-001 -3.9170E-001

[SD: 4.8485E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)