ML19099A088

From kanterella
Jump to navigation Jump to search
Appendix K, Qaqc Records, 2017, Isos QCs - Daily QC 05032017-1
ML19099A088
Person / Time
Site: 07000371
(SNM-0368)
Issue date: 03/01/2018
From:
Cabrera Services
To:
NRC Region 1
Shared Package
ml051740240 List: ... further results
References
Download: ML19099A088 (2)


Text

Last Measurement Q.A. Report 5/3/17 8:00:53 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 5/3/17 8:00:53 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 0503201 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 5/3/17 7:52:58 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.7 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.5854E-001 -2.8221E-001

[SD: 9.6196E-001+/- 0.012] < : : : >

Peak FWHM (1407.92 keV) 2.4426E+000 4.0096E-001

[SD: 2.3869E+000+/- 0.139] < : : : >

Peak FWHM (964.13 keV) 1.9256E+000 8.1808E-002

[SD: 1.9187E+000+/- 0.085] < : : : >

Peak energy (964.13 keV) 9.6426E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2197E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4086E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4169E-001 -1.1973E+000

[SD: 5.4918E-001+/- 0.006] < : : : >

Corrected activity (964. 4.8401E-001 -7.6975E-002

[SD: 4.8456E-001+/- 0.007] < : : : >

Corrected activity (1407 4.8031E-001 -6.7672E-001

[SD: 4.8495E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)