Last Measurement Q.A. Report 5/12/17 9:22:33 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 5/12/17 9:22:33 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 0512201 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 5/12/17 9:08:53 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.6 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.7365E-001 1.0314E+000
[SD: 9.6166E-001+/- 0.012] < : : : >
Peak FWHM (1407.92 keV) 2.5651E+000 1.2551E+000
[SD: 2.3880E+000+/- 0.141] < : : : >
Peak FWHM (964.13 keV) 1.9821E+000 7.7217E-001
[SD: 1.9173E+000+/- 0.084] < : : : >
Peak energy (964.13 keV) 9.6388E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2191E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4081E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4362E-001 -2.8912E-001
[SD: 5.4720E-001+/- 0.012] < : : : >
Corrected activity (964. 4.8362E-001 -9.0778E-002
[SD: 4.8425E-001+/- 0.007] < : : : >
Corrected activity (1407 4.7940E-001 -8.0532E-001
[SD: 4.8515E-001+/- 0.007] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)