Last Measurement Q.A. Report 5/11/17 8:55:30 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 5/11/17 8:55:30 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 051 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 5/11/17 7:48:30 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.4 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.7312E-001 1.0053E+000

[SD: 9.6143E-001+/- 0.012] < : : : >

Peak FWHM (1407.92 keV) 2.5398E+000 1.0997E+000

[SD: 2.3850E+000+/- 0.141] < : : : >

Peak FWHM (964.13 keV) 1.9660E+000 5.8901E-001

[SD: 1.9163E+000+/- 0.084] < : : : >

Peak energy (964.13 keV) 9.6403E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2193E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4083E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4210E-001 -4.1715E-001

[SD: 5.4730E-001+/- 0.012] < : : : >

Corrected activity (964. 4.8317E-001 -1.5639E-001

[SD: 4.8427E-001+/- 0.007] < : : : >

Corrected activity (1407 4.8204E-001 -4.3989E-001

[SD: 4.8521E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)