Last Measurement Q.A. Report 6/22/17 8:05:15 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 6/22/17 8:05:15 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 0622201 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 2:00:00 PM Measurement Date: 6/22/17 7:58:22 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.8 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.7927E-001 1.1613E+000
[SD: 9.6388E-001+/- 0.013] < : : : >
Peak FWHM (1407.92 keV) 2.4824E+000 6.7218E-001
[SD: 2.3874E+000+/- 0.141] < : : : >
Peak FWHM (964.13 keV) 1.9222E+000 2.5618E-002
[SD: 1.9201E+000+/- 0.084] < : : : >
Peak energy (964.13 keV) 9.6389E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2191E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4081E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4699E-001 4.0681E-002
[SD: 5.4652E-001+/- 0.011] < : : : >
Corrected activity (964. 4.7997E-001 -5.3724E-001
[SD: 4.8377E-001+/- 0.007] < : : : >
Corrected activity (1407 4.8939E-001 6.5187E-001
[SD: 4.8502E-001+/- 0.007] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)