ML19099A065
| ML19099A065 | |
| Person / Time | |
|---|---|
| Site: | 07000371 |
| Issue date: | 03/01/2018 |
| From: | Cabrera Services |
| To: | NRC Region 1 |
| Shared Package | |
| ml051740240 | List:
|
| References | |
| Download: ML19099A065 (2) | |
Text
Last Measurement Q.A. Report 3/30/17 7:24:50 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 3/30/17 7:24:50 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 033 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 3/30/17 7:15:33 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 354.5 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.6991E-001 1.4049E+000
[SD: 9.5518E-001+/- 0.010] < : : : >
Peak FWHM (1407.92 keV) 2.4632E+000 1.3393E+000
[SD: 2.3223E+000+/- 0.105] < : : : >
Peak FWHM (964.13 keV) 1.8985E+000 1.4959E-002
[SD: 1.8975E+000+/- 0.067] < : : : >
Peak energy (964.13 keV) 9.6378E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2195E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4079E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.6436E-001 1.6964E+000
[SD: 5.5080E-001+/- 0.008] < : : : >
Corrected activity (964. 5.0045E-001 1.7661E+000
[SD: 4.8356E-001+/- 0.010] < : : : >
Corrected activity (1407 4.9997E-001 1.6550E+000
[SD: 4.8429E-001+/- 0.009] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)