ML19099A059

From kanterella
Jump to navigation Jump to search
Appendix K, Qaqc Records, 2017, Isos QCs - Daily QC 04102017
ML19099A059
Person / Time
Site: 07000371
(SNM-0368)
Issue date: 03/01/2018
From:
Cabrera Services
To:
NRC Region 1
Shared Package
ml051740240 List: ... further results
References
Download: ML19099A059 (2)


Text

Last Measurement Q.A. Report 4/10/17 8:19:47 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 4/10/17 8:19:47 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC UNC 041 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/10/17 8:13:06 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 352.2 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.5389E-001 -3.4884E-001

[SD: 9.5738E-001+/- 0.010] < : : : >

Peak FWHM (1407.92 keV) 2.2155E+000 -1.0965E+000

[SD: 2.3539E+000+/- 0.126] < : : : >

Peak FWHM (964.13 keV) 1.7203E+000 -2.6045E+000

[SD: 1.9122E+000+/- 0.074] < :In : : >

Peak energy (964.13 keV) 9.6401E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2199E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4082E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4966E-001 -2.1285E-001

[SD: 5.5120E-001+/- 0.007] < : : : >

Corrected activity (964. 4.8463E-001 -2.2655E-003

[SD: 4.8465E-001+/- 0.009] < : : : >

Corrected activity (1407 4.8315E-001 -2.5890E-001

[SD: 4.8537E-001+/- 0.009] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)