ERG Certificate of Calibration Calibration and Voltage Plateau Meter:
Manufacturer:
Ludlum Model Ntunber:
222lr Detector: Manufacturer:
Alpha Spectra Model Number:
FIDLER Environmental Restoration Group, Inc.
8809 Washington St NE, Suite 150 Albuquerque, NM 87113 (505) 298-4224 www.ERGoffice.com Serial Number:
108859 Serial Number:
010807B
~ Mechanical Check I" THR/WTN Operation HV Check(+/- 2.5%): D 500 V D 1000 V 0 1500 V
~ F/S Response Check bl" Reset Check Cable Length:
LJ 39-it1ch LJ 72-inch
~ Other:
60"
~ Geotropism
~ Audio Check
~ Meter Zeroed
~ Banery Check (Min 4.4 VDC)
Source Distance: 0 Contact D 6 inches [2"( Other:
1/2" Threshold:
lOmV Source Geometry: D Side
~ Below D Other:
Window:
Instrument found within tolerance: ~ Yes D No Range/Multiplier Reference Setting "As Found Reading" Meter Reading x 1000 400 400 400 x 1000 100 100 100 x 100 400 400 400 x 100 100 100 100 x IO 400 400 400 x 10 100 100 100 x 1 400 400 400 x 1 100 100 100 High Voltage Source Counts
Background
700 30982 800 93547 900 130639 950 134090 1000 134583 4948 1050 135059 1100 134484 1150 134847 1200 134997 Comments: HV Plateau Scaler Count Time= 0.5-min. Recommended HV = 1000 Reference Instruments and/or Sources:
Barometric Pressure: 24.57 Temperature:
77 Relative Humidity:
20 inches Hg Of Integrated Log Scale Count 1-Min. Count 399229 400 100 39923 400 100 3994 400 100 399 400 100 Voltage Plateau 160000 140000 120000 100000 80000 60000 40000 20000 r
/
f I
I
- 0 Ludlum pulser serial number: D 97743
~ 201932 Fluke multimeter serial number: 0 87490128 D Alpha Source: Th-230@ 12,800 dpm (l /4/12) sn: 4098-03 D Beta Source:
~ -99 @ 17, 700 dpm (I /4/12) sn: 4099-03 D Gamma Source Cs-137@ 5.2 uCi (l/4/12) sn: 4097-03
~ Other Source:
Am-241 @ I uCi Calibrated By:
Calibration Due: (;, :£---17 Reviewed By: ~*-------
Calibration Date: ?*-£ ~ / (.,,
Date:
~ / <i /1 i, ERG Form ITC. 101.A This calibration conforms to the requirements and acceptable calibration conditions ofA1\\'Sl J\\'323A - 1997