Report of Analysis/Certificate of Conformance, L38101
ML100740159
Person / Time
Site: Oyster Creek
Issue date: 04/28/2009
From: Jeter K
Teledyne Brown Engineering, Teledyne Technologies
To: Vouglitois J
Exelon Nuclear, Office of Information Services
References
FOIA/PA-2009-0214
Download: ML100740159 (8)


Text

A*A 6M, TELEDYNE PPN BROWN ENGINEERING, INC.

A Teledyne Technologies Company 2508 Quality Lane Knoxville, TN 37931-3133 865-690-6819 James J. Vouglitois Exelon Nuclear (Oyster Creek)

P.O. Box 388 US Route 9 Forked River, NJ 08731

~&7?~,o 32~,c1pc(i £2A~i'vK'4c iA~

Report of Analysis/Certificate of Conformance 04/28/2009 LIMS #:

L38101 Project 1D#

EX00I-3EREMPOC-05 Received:

03/30/2009 Delivery Date 05/04/2009 P.O.#:

01000298 REL.#00010 Release #

3 DG#:

This is to certify that Teledyne Brown Engineering - Environmental Services located at 2508 Quality Lane, Knoxville, Tennessee, 37931, has analyzed, tested and documented samples, as received by the laboratory, as specified in the applicable purchase order.

This also certifies that requirements of applicable codes, standards and specifications have been fully met and that any quality assurance documentation which verified conformance to the purchase order is on file and may be examined upon request.

I hereby certify that the above statements are true and correct.

Keith Jeter Operations Manager Cross Reference Table Client ID Laboratory'I)

Station ID (if applicable) 33 L38101-1 33 94 L38101-2 94 I

L38101-3 1

37 L38101-4 37 38 L38101-5 38 39

~L38101-6 3

QCA L38101-7 QCA 4'ý Page 1 of 8

BROWN/EN'iGINEERING, INC.

A Teledyne Technologies Company 2508 Quality Lane Knoxville, TN 37931-3133 865-690-6819 Method Reference Numbers Matrix Analysis Method Reference WD GAMMA EPA 901.1 WD)

H-3 (DIST) f EPA 906.0 WT

[

GAMMA EPA 901.1 WT H-3 (DIST)

EPA 906.0 This report shall not be reproduced or distributed'except in its entirety.

Page 2 of 8

UI1..OIu v,'J4, A Teledyne Technologies Company L38101 Exelon Nuclear (Oyster Creek)

EX001-3EREMPOC-05 Sample ID: 33 Collect Start: 03/05/2009 10:15 Matrix:

Surface Water (WT)

Station:

33 Collect Stop: 03/25/2009 10:25 Volume:

==

Description:==

ReceiveDate: 03/30/2009

% Moisture:

LIMS Number:

L38101-1 1op I Activity jUncertainty IRun Aiut lqot Rfrne Cnt Count C0n Radionuclide j SO Conc 2 Sigma MDC j

Units,
  1. 1 Volume Units Date Date Time j Units Flag Values 4-3(DIST) 2011 T -2.64E+01 1.19E+02 1.98E+02 I pCi/L 1

1 10 J 2 m1 I 03/31/09F 80 T M

I U I 3E-7 2007 8.93E+00 1 2.46E+01 4.06E+01 I pCi/L 11 3502.26 1 ml 1 03/25/09 10:25 1 04/08/091 25665 1 See I U I I No I C-40 F 2007 1 2.84E+02 1 4.62E+0l 1 3.00E+01 3

pCi/L 1-1 3502.26 1 ml 1 03/25/09 10:25 1 04/08/091 25665 1 Sec [ +

I Yes I IN-54 F 2007 1 2.07E+00 2.32E+00 1 3.94E+00 F pCi/L

{V7 3502.26" ml 103/25/09 10:25 1 04/08/091 25665 1 See U I No 1

ý0-58 2007 T -2.84E+00I 2.64E+00 I 4.11E+00 I pCi/L 3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 See

[ U I No E-59 F 2007 1 -148E+00I 5:65E+00 1 9.13E+00'E pCi/L 1

1 3502.26 1 ml 103/25/09 10:25 1 04/081091 25665 1 Sec U r No 20-60 F 2007 1 8.76E-0l I 2.34E+00 0 3.92E+00 r pCi/L 1

3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 See I U I NoF

,FN-65 1 2007 1 -l.55E+00 5.20E+00 1 8.39E+00 F pCi/L 11 3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 Sec I U I I No I B-95 1 2007 1 -6.91E-01 I 2.61E+O0 1 4.22E+00 F pCi/L I

I 3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 Sec I U I _

No I ER-95 1 2007 1 -2.64E+00.I 4.52E+00 1 7.19E+00 F pCi/L 11 3502.26 1 ml 103/25/09 10:25 1. 04/08/091 25665 1 See U I FNoF_

-131 1 2007 -I-1.22E+00 8.77E+00 1 1.39E+01 F pCi/L I

" 3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 See I U I No I S-134 1 2007 1 4.10E+00 5.57E+00 0 4.26E+00 I pCi/b I

I 3502.26 1 ml 103/25/09 10:25 6

04/08/09[

25665 1 See FFU I No

ýS-137 1 2007 1 2.38E+00 I 2.58E+00 1 4.37E+00 I pCi/L 1.1 3502.26 1 ml 103/25/09 10:25 1 04/08/091 25665 1 See FU I N 3A-140 1 2007 1 -3.21E+00 1.83E+01 I 2.97E+01 I pCi/L J

3502.26 1 ml 103/25/09 10:25 1 04/08/091.

25665 1 See

[F I

No_

LA-140, 2007 4.83E-01 F 5.08E+00 I 8.39E+00 pCi/L I

I 3502.26 1 ml 103/25/09 10:25 1 04/08/09 25665 1 See

[ U No_

PA-226 1 2007 5.53E+00 I 7.53E+01 I 1.07E+02 V pCi/L I

3502.26 i ml 103/25/09 10:25 1 04/08/091 25665.1 See FU Yes_

TI-232 1 2007

-1.88E+00 9.04E+00 I 1.44E+01 V pCi/L I

I 3502.26 I

ml 03/25/09 10:25 1 04/08/091 25665 1 See U I_____

Flag Values U

=

Compound/Analyte not detected or less than 2 sigma No =Peak not identifiedin gamma spectrum

+

=

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

Yes = Peak identified in gamma spectrum U*

=

Compound/Analyte not detected. Peak not identified, but forced activity concentration exceeds MDC and 2 sigma

      • Results are reported on an as received basis High

=

Activity concentration exceeds customer reporting value unless otherwise noted Spec

=

MDC exceeds customer technical specification L

Low recovery MDC - Minimum Detectable Concentration H

=

High recovery Bolded text indicates reportable value..

PAe I of R

U4LnI. JY V)1"j

.A Teledyne Technologies Company L38101 Exelon Nuclear (Oyster Creek)

James J. Vouglitois EX001-3EREMPOC-05 Sample ID: 94 Collect Start: 03/05/2009'09:35 Matrix:

Surfice Water (WT)

Station: 94 Collect Stop: 03/26/2009 10:15

/

Volume:

==

Description:==

Receive Date: 03/30/2009

% Moisture:

LIMS Number:

L38101-2 So# Activity UncertaintyRu Alqo Alqo Rernc Cut Cut Cut Radionuclide SOP#

Cone 2 Sigma MDC Units Volume Units Date

  • Date Time Units Flag Values H-3 (DIST) 1 2011 1-7.66E+01 1.13E+02 I 1.92E+021 pCi/L F

F 10 1

ml 1

103/31/091 80 1 M I 1 1

3E-7 1 2007 1 -8.45E+00 F 8.45E-E+00 1.36E+01 1 pCi/L 1F 3547.87 1,

ml 103/26/09 10:15 1 04/08/091 47007 1 See U I I N N o

-40 1 2007 1 2.76E+02.1 2.69E+01 I 1.32E+01 I pCi/L f7F 3547.87 ml" 03/26/09 10:15 1 041081091 47007 1 Sec 1 +

__ TYesl MN-54 1 2007 1 4.90E-01 I 9.61E-0I I 1.59E+00 pCi/L T1 3547.87

  • . ml

[03/26/09.10:15 1 04/08/091 47007 1 Sec I U I _

NoT I 30-58 1 2007 1 -2.69E-02I1.1.9.84E-01 I 1.59E+00 1 pCi/L II 3547.87 1 ml

[03/26/09 10:15 1 04/08/091 47007 1 Sec I.U I No I FE-59 1 2007 1 -2.19E+00 2.08E+00 I 3.25E+00 1 pCi/L

. I 3547.87 1 ml 103/26/09 10:15 1 04/08/091 47007 1 Sec I U I NoI "0-60 1 2007

-6.70E-01 I 9.71E-01 I 1.51E+00 I pCi/L T

3547.87 -1 ml

[03/26/09 10:15 F 04/08/091 47007 1 See I

I No I

,N-65 1 2007 F -2.42E+001 2.17E+00 I 3.39E+00 I pCi/L 11 3547.87 1 ml F 03/26/09 10:15 1.04/08/091 47007 1 Sec I UJ I --- No 414-95 1 2007 F 7.95E-01 1 1.04E+00 F 1.74E+00 I pCi/L 11 3547.87 F ml F03/26/09 10:15 1 04108/091 47007 1 Sec I U. I l

No I

  • R 1 2007 F 5.20E-01 1 1.76E+00 F 2.90E+00 I pCi/L 1

) 3547.87 F ml F03/26/09 10:15 1 04/08/091 47007 1 See I U I __

NoT

-131 1 2007 F 1.25E+00 1. 3.06E+00 F 4.98E+00 I pCi/L 11 3547.87 F ml

" 03/26/09 10:15 1 04/08/091 47007 1 Sec I U I No I-

,S-134-1 2007 F 1.81E-01 I 9.65E-01 I 1-39E+00 I pCi/L I

I 3547.87 F ml 03/26/09 10:15 I 04/08/091 47007 1 See IuI I No I "S-137 I 2007 F. -4.55E-01 I.9.67E-01 I 1.55E+00 I pCi/L 11 3547.87 1 ml 103/26/09 10:15 1 04/08/091 47007 1 Sec I U I NoF 3A-140 1 2007 I--2.1IE-01 I 6.10E+00 I 1.01E+01 I pCi/L 11 3547.87 F ml 103/26/09 10:15 1 04/08/091 47007 1 Sec I U I l&N A-140 I 2007

-1.10E+00 2.20E+00 I 3.56E,+00 p pCi/L 1

3547.87 U ml 103/26/09 10:15 1 04/08/091 47007 1 See I U I-TNoI

-226 I 2007 I -4.56E+01 I 2.82E+01 I 4.28E+01 I pCi/L I

1 3547.87 F ml 1 03/26/09 10:15 1 04/08/091 47007 1 Sec J U I I_-No F

-232 I 2007.

1.28E+00 3.76E+00 I 6.71E+00 I pCi/L I '

3547.87 I

ml 103/26/09 10:15 1 04/081091 47007 1 See

-IU I YesF Flag Values U

=

Compound/Analyte not detected or less than 2 sigma No = Peak not identified in gamma spectrum

-+.

=

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

Yes = Peak identified in gamma spectrum U.

=

Compound/Analyte not detected. Peak not identified, but forced activity concentration exceeds MDC and 2 sigma

      • Results are reported on an as received basis High

=

Activity concentration exceeds customerreporting value unless otherwise noted Spec

=

MDC exceeds customer technical specification L

Low recovery MDC - Minimum Detectable Concentration H

=

High recovery Bolded text indicates reportable value.

Paize 4 of 8

U4ILO/UY U*J."-t A lefedyno IeCnnologles Lompany L38101 Exelon Nuclear (Oyster Creek)

James J. Vouglitois EX0O1-3EREMPOC-05 Sample ID:

I Collect Start: 03/06/2009 15:30 Matrix:

Drinking Water (WD)

Station:

1 Collect Stop: 03/24/2009 10:54 Volume:

==

Description:==

Receive Date: 03/30/2009

% Moisture:

LIMS Number:

L38101-3 I

Conc 2 S Volume units Date Date Time Units Fla

-3 (DIST) 2011 1 -6.50E+001 1.18E+02 I 1.95E+02 r pCi/L 1 1 10 1

ml 1

1 04/01/091 80 f

M YU E-7 I 2007 1 -4.44E+00 I 1.19E+01 I 1.93E+01 pCi/L 1

1 3503.38 1 ml 1 03/24/09 10:54 1 04/08/091 46995 T See I U I 1 No

-40 I

2007 1 1.35E+01 2.24E+0l I 4.15E+01 I pCi/b 1.1 3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I _

No 1

-54 I 2007 1 -3.0713-01 1.35E+00 I 2.23E+00 [

pCi/L i

3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I No I 0-58 I 2007 1 4.95E-01 1.32E+00 1 2.25E+00 pCi/L 1

3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I_

No E-59 I 2007 1 1.85E+00 3.20E+00 1 5.40E+00 I pCi/L 11 3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec

[ U

_I No 0-60 I 2007 1 3.73E-01 1 1.32E+00 1 2.17E+00 0 pCi/L 1

3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I F No_

N-65 I

2007 I -5.70E+00 3.18E+00 1 4.66E+00 1 pCi/L 1.1 3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I No_ I B-95 2007

[ 8.25E-01 I 1.39E+00 1 2.39E+00 1 pCi/L I I

3503.38 1 ml 103/24/09 10:54 1 04/081091 46995 1 Sec I U I No I R-95 I 2007 I 2.21 E-02 1 2.43E+00 I 4.09E+00 1 pCi/L 11 3503.38 1 ml 103/24/09 10:54 1 04/08/09I 46995 1 Sec I U I __

No 1

-131 1 2007 1 2.67E+00 I 4.3113+00 1 7.37E+00 pCi/L 11 3503.38.

ml 1 03/24/09 10:54 1 04/08/091 46995 1 Sec

[OI U1 No S-134

. 2007

[ 4.06E-01 1 1.36E+00 1 1.95E+001 pCi/L IT 3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec

[ U I No S-137 1 2007 1 3.63E-01 1 1.3213+00 1 2.17E+00

-pCi/L 11 3503.38 1 ml 103/24/09.10:54 1 04/08/091 46995,1 Sec I U I

No I A-140 1 2007 I -3.59E+00I 9.59E+00 I 1.55E+01 pCi/L 1 3503.38 1 ml 103/24/09 10:54 1 04/08/091 46995 1 Sec I U I

.__ No I A-140 1 2007'

-1.28E-01 I 3.27E+00 5.39E+00 pCi/L

[

3503.38 1 ml

[03/24/09 10:54. 1 04/08/091 46995 1 Sec U 1

[No 1

-226 1 2007 1 -3.69E+01I 3.35E+01. F 5.04E+01 1 pCi/L

[

3503.38 1 ml

[03/24/09 10:54 1 04/08/091 46995 j Sec I U I I No 1

-232 1 2007 F 2.51 E+00 5.53E+00 F 8.84E+00 I pCi/L I

3503.38 1 ml

[03/24/09 10:54 1 04/08/091 46995 1 Sec I U I

__No_

Flag Values N

U

=

Compound/AMalyte not detected or less than 2 sigma No Peak not ident-ified in gamma spectrum

+

=

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

Yes =Peak identified in gamma spectrum U*

=

Compound/Analyte not detected. Peak not-identified, but forced activity concentration exceeds MDC and 2 sigma

        • Results are reported on an as received basis High

=

Activity concentration exceeds customer reporting value unless otherwise noted Spec

= x MDC exceeds customer technical specification L

=

Low recovery MDC -Minimim Detectable Concentration H

-High recovery Bolded text indicates reportable value.

Page 5 of 8

VtI,.OIU7 U7.jt L38101 Exelon Nuclear (Oyster Creek)

F.XO01-I APRR.MPOC-05 A Teledyne Technologies Company T-cp*

Vn1icrl tnie Sample ID: 37 Collect Start: 03/05/2009 10:45 Matrix:

Drinking Water (WD)

Station:

37 Collect Stop: 03/25/2009 11:35 Volume:

==

Description:==

Receive Date: 03130/2009

% Moisture:

LIMS Number:

L38101-4 I S

. Activity Uncertainty Ru~n A

Aliquot Reference Count Count Count Radionuclide SOP#

Conc 2 Sigma MDC Units Volume Units Date Date Time Units Flag Values H-3 (DIST) '

2011 T-6.47E+01 1.15E+02 1.94E+02 pCi/L T

10 1

ml I

T04/01/091 80 1 M I uI BE-7 F 2007.

9.26E+001 1.01E+01 1.75E+01 F pCi/L T 3501.82 1 ml 103/25/09.11:35 1 04/08/091 47008 1 Sec I U I 1 No K-40.

2007 T 3.01E+00 1.96E+01 I 3.46E+01 F pCi/L 1 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 I Sec JU 1 No M-54 r 2007 T -5.53E-01 1 1.13E+00 1.84E+00 I pCi/L 1

1 3501.82 1

ml

[03/25/09 11:35 1 04/08/091 47008 1 Sec JU INoTF CO-58 F 2007 1 4.41E-01 I I.19E+00 1 2.03E+00 F pCi/L 11 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec I U I _

No 1,

_E-59 i 2007 1 6.27E-01 1 2.80E+00 0 4.59E+00 r pCi/L 1T 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec I U I _

NoFi

,0-60 r 2007 1 9.48E-01 1.62E+00 1 2.17E+00 F pCi/L J

f 3501.82 T ml 103/25/09 11:35 1 04/08/091 47008 1 Sec J U I IYesF

?ýN-65 F 2007 1 -1.35E+00[

2.64E+00 1 4.15E+00 r pCi/L 1

3501.82 1 ml 103/25/0911:35 1 04/08/091 47008 I Sec I U I Nor qB,95 r 2007 1 4.23E-01 1 1.20E+00 1 2.04E+00 F pCi/L 1

3501.82 T ml 03/25/09 11:35 1 04/08/091 47008. 1 Sec

-I U I I NoI

_R-95 2007 1 1.08E+00 1 2.13E+00 1 3.51E+00 0 pCi/L 11 3501.82 1 ml 103/25/09 11:35 1 04/08/09§ 47008 1 Sec I U I I No I

-131 r 2007 1 -2.3 1E+001 3.53E+00 1 5.58E+00 r pCi/L 1*1 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec I U I _

NoF "S-134 F 2007

-1.13E+001 1.29E+00 1 1.74E+00 0 pCi/L 11 3501.82 1 ml 1 03/25/09 11:35 1 04/08/091 47008 1 Sec I U I I

No__

ýS-137 I 2007 1 1.37E-01 1.21E+00 1 1.98E+00 pCi/L 1

3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec I U I I No I 3A-140 i 2007 1 -1.20E+001 8.01E+00 1 1.32E+01 i pCi/L iT 3501.82 1 ml 103/25/09 11:35 1.04/08/091 47008 1 Sec I U I 1 No

,A-140 2007 1 2.85E-01 I 2.87E+00 1 4.69E+00 0 pCi/L 11 3501.82 1 ml 1 03/25/09 11:35 1 04/08/091 47008 1 Sec I U I No I RA-226 1 2007 1 3.55E+01 I 3.65E+01 1 3.64E+01 I pCi/L 11 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec U

U IYes TH-228 1 2007 1 4.37E+00 1 2.62E3+0 0 3.12E+00 I pCi/L 11 3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 1 Sec

+[

+

_I__

Y*

TH-232 I 2007 1 2.89E+00 4.00E+00 1 6.30E+00 0 pCi/L 1

3501.82 1 ml 103/25/09 11:35 1 04/08/091 47008 I Sec I U

_I YesJ Flag Values U

=

Compound/Analyte not detected or less than 2 sigma

+

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

U*

=

Compound/Analyte not detected. Peak not identified, but forced activity concentration exceeds MDC and 2 sigma High

=

Activity concentration exceeds customer reporting value Spec

=

MDC exceeds customer technical specification L

=

Low recovery H

=

High recovery Bolded text indicates reportable value.

-Page 6 of 8 No = Peak not identified in gamma spectrum Yes. = Peak identified in gamma spectrum

  • Results are reported on an as received basis unless otherwise noted MDC - Minimum Detectable Concentration

A Teledyne Technologies Company L38101 Exelon Nuclear (Oyster Creek)

James J. Vouglitois EX001-3EREMPOC-05.

Sample ID: 38 Collect Start: 03/05/2009 10:00 Matrix:

Drinking Water (WD)

-Station: 38 Collect Stop: 03/26/2009 13:45 Volume:

==

Description:==

Receive Date: 03/30/2009

% Moisture:

LIMS Number:

L38101-5 Activity Uncertainty Run Aliquot Aliquot Reference Count Count nt Radionuclide SOP#

Conc 2 Sigma MDC Units Volume Units Date Date Time Units Flag Values

[I-3 (DIST)

I 2011 1 -1.51E+01 1.17E+02 1.95E+02 1 pCi/L T 1 10 1

ml 1

I 04/01/09 80 1 M I U BE-7

.I 2007 1 6.50E+00 I l.00E+0 1 I 1.66E+01 1 pCi/L

[

1 3500.36 1

ml 1.03/26/09 13:45 1 04/08/09[

47003 1 Sec U lJ No

-40 2007 F 2.65E+00 I 1.77E+01 [13.17E+01 1 pCi/L F

3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I I No 1

-54

[ 2007 1 -6.94E-01 I 1.04E+00 I 1.65E+00 1 pCi/L r 1 3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I I No 1 0-58 1 2007 F -8.63E-01 I 1.09E+00 F 1.72E+00 1 pCi/L f7[ 3500.36 1 ml 103/26/09 13:45 F 04/08/09 47003 1 Sec lU I No 1 FE-59 2007 r 2.63E+00F 2.31E+00 F 4.04E+00 1 pCi/L I

  • F 3500.36 1 ml

[03/26/09 13:45 1 04/08/091 47003 F Sec I U I I No I 0-60 1 2007

[-7.56E-02 I 1.02E+00 F 1.66E+00 I pCi/L 1.

3500.36 1

ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I

No I

-65 F 2007 1 24.16E+00 2.24E+00 1 3.40E+00 I pCi/l I

1 3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I 1 No I

_B-95 2007.

6.51E-02 I 1.14E+00 0 1.88E+00 F pCi/L I

1 3500.36 F ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I 1No F ER-95 F 2007

  • 8.18E-01 1.99E+00 I 3.34E+00 F pCi/L 1

1 3500.36 T ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I I NoI R-131 1 2007 1 -7.I1E701 3.11E+00 1 5.10E+00 F pCi/L 1

T 3500.36 1 ml 103/26/09 13:45 I 04/08/091 47003 1 Sec I U [F I No I PS-134 1 2007 1 -3.87E-0l 1 1.49E+00 1 1.56E+00 r pCi/L 1

3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I No F CS-137 1 2007 1 6.24E-01 1 1.04E1+00 00 F pCi/L 1

3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec, I U I I No I BA-140 1 2007 1 7.07E-01 I 7.13E+00 1 1.15E+01 F pCi/L I

1 3500.36 1 ml 1 03/26/09 13:45 1 04/08/091 47003 1 Sec I U I I No I LA-140 2007 I -1.1OE+00 I 2.33E+00 0 3.64E+00 0 pCi/L 11 3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec U

No RA-226 1 2007 1 -2.94E+0 I 2.91E+01 I 4.30E+01 I pCi/L

[,1 3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U I I No I TH-232 1 2007 I 4.1OE-0l I 4.53E+00 I 7.15E+00 I pCi/L F

1 3500.36 1 ml 103/26/09 13:45 1 04/08/091 47003 1 Sec I U F I No I Sample ID: 39 Collect Start:

Matrix:

Drinking Water (WD)

Station:

39 Collect Stop:

Volume:

==

Description:==

Receive Date: 03/30/2009

% Moisture:

LIMS Number:

L38101-6 S1op I Activity jUncertaintyI Ait R

Aliquot 1 Aliquot Reference Count Count Count Radionuclide JSOPone 2 Sigma.

MDC Units Volume Units Date Date Time Units Flag Values SAMPLE FNA I

F I

F F

F F

F I

IUI I

I Comment:

I NO SAMPLE!! Station down for repairsl!

Flag Values U

=

, Compound/Analyte not detected or less than 2 sigma No Peak not identified in gamma spectrum

+

=

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

Yes = Peak identified in gamma spectrum U*

=

Compound/Analyte not detected. Peak not identified, but4forced activity concentration exceeds MDC and 2 sigma Results are reported on an as received basis High

=

Activity concentration exceeds customer reporting value unless otherwise noted Spec

=

MDC exceeds customer technical specification L

Low recov-ery.

MDC -Minimum Detectable Concentration H

=

High recovery Bolded text indicates reportable value.

Page 7 of 8

4

  • J A Teledyne Technologies Lompany" L38101 Exelon Nuclear (Oyster Creek)

James J. Vouglitois EX001-3EREMPOC-05 Sample ID: QCA Collect Start: 03/06/2009 15:30 Matrix:

Drinking Water (WD)

Station:

QCA Collect Stop: 03/24/2009 10:54 Volume:

==

Description:==

Receive Date: 03/30/2009

% Moisture:

LIMS Number:

L38101-7 So# Activity Uncertainty 1D t

Run Aliquot Aliquot.

Reference Count CouniTCount Radionuclide SCon 2 Sigma Un Volume Units Date Date Time Units Flag Values 4-3 (DIST) 1 2011 1 1.08E+01 1 1.19E+02 1.95E+02 I pCi/L F

10 1

ml 1

04/01/091 80 1

M U

I BE-7 1 2007 1 5.97E+00 1.22E+01 I 2.02E+01 I pCi/L F

3505.52 1 ml 103/24/09 10:54 1 04/08/091 46992 1 See I U I I No I

-40 1 2007 1 3.14E+01 2.91E+01 I 1.75E+01 1 pCi/L F

3505.52 1 ml 03/24/09 10:54 1 04/08/091 46992 1 See 1 +

_ I Yes 1, MN-54 1 2007 1 -1.22E+00

1. 18E+00 1 1.87E+00 1 pCi/L 1

3505.52 i ml 03124109 10:54 F 04/08/091 46992 I See U I iNo I

,0-58 1 2007 1 -7.70E-01 r 1.25E+00 I 2.02E+0I pCi/L IF 3505.52 1 ml 103/24/09 10:54 I 04/08/091 46992 1 Sec 1 U INol E-59 1 2007 1 -8.62E-01 I 2.79E+00 F 4.52E+00 1 pCi/b F 3505.52 i ml

[03/24/09 10:54 1 04/08/091 46992 1 See 1 U 1 I No T-'

0-60

. 1 2007 F -7.05E-021 1.14E+00 F 1.85E+00 1 pCi/L TF [3505.52 i ml

[03/24/09 10:54. 04/08/091 46992 1 See JU 1 No.1 T-65 2007 F -3.55E+00 1 2.55E+00 4 3.82E+00 1 pCi/L T

3505.52 1 ml F 03/24/09 10:54 F 04/08/09 46992 1 Sec 1 U 1 1 No I N-95 2007 F 7.37E-02 I 1.35E+00 F 2.32E+00 I pCi/L IF 3505.52 1 ml 103/24/09 10:54 1 04/08/091 46992 F Sec l

U iNol R-95,

1 2007 F 2.14E+00 I 2.38E+00 I 4.17E+00 I pCi/L 1 F 3505.52 F ml

[03/24/09 10:54 1 04108/091 46992 1 See I U

_I No

[-131 2007 7 6.18E-01 1

4.48E+00 2 7.39E+00 I pCi/L i

3505.52 1 ml 103/24/09 10:54 1 04/08/091 46992 1 See I U I I Nol S-134 1 2007 F -2.53E+001 1.28E+00 F 1.87E+00 I pCi/b 1 _*

3505.52 [

ml 103/24/09 10:54 1 04/08/09 46992 F Sec l U INol S-137 1 2007 F -6.16E-02 4 1.28E+00

.9E+00 1 pCi/L I _

3505.52 F ml 103/24/09 10:54 1 04/08/091 46992 F Sec I U I Nol PA-140 F 2007 F -1.02E+00I 9.78E+00 F 1.57E+01 F pCi/L I

3505.52 I ml F 03/24/09 10:54 1 04/08/091 46992 1 See I U I No JA-140 I 2007 I 1.82E-01 I 3.15E+00 I 5.11E+00 0 pCi/L 1

3505.52 F ml 1.03/24/09 10:54 1 04/08/091 46992 1 Sec I U I I No

-226 2007 I 1.19E+01 I 4.25E+01 I 4.62E+01 I pCi/L 1

1 3505.52 1 ml 103/24/09 10:54 1 04/08/091

.46992 1 See I U I I Yes F

-232 J 2007, I -5.76E-01 I 5.51E+00 I 8.73E+00 I pCi/L I

3505.52 1 ml 103/24/09 10:54 1 04/08/091 46992 F See I U I

No H Flag Values U

=

Compound/Analyte not detected or less than 2 sigma No = Peak not identified in gamma spectrum

+

=

Activity concentration exceeds MDC and 2 sigma; peak identified(gamma only)

Yes = Peak identified in gamma spectrum U*

=

Compound/Analyte not detected. Peak not identified, but forced activity concentration exceeds MDC and 2 sigma

        • Results are reported on an as received basis High

=

Activity concentration exceeds customer reporting value unless otherwise noted Spec

=

MDC exceeds customer technical specification L

Low recovery MDC - Minimum Detectable Concentration H

=

High recovery Bolded text indicates reportable value.

Page 8 of 8