ML072000423
| ML072000423 | |
| Person / Time | |
|---|---|
| Site: | Rancho Seco |
| Issue date: | 04/17/2007 |
| From: | Rosalyn Jones Sacramento Municipal Utility District (SMUD) |
| To: | John Hickman NRC/FSME/DWMEP/DURLD/RDB |
| References | |
| 151732 | |
| Download: ML072000423 (61) | |
Text
Rancho Seco Well Sample Analysis John Hickman - Rancho Seco Well Sample Analysis From:
"Robert E. Jones" <RJones2@smud.org>
To:
<jbh@nrc.gov>
Date:
04/17/2007 7:34 PM
Subject:
Rancho Seco Well Sample Analysis Page 1 of 1
- John, Attach is the case narrative for one of our quarterly well samples at Rancho Seco. The one I sent you earlier was generic and probably not as helpful to you.
- Thanks, Bob
<<Case Narrative.pdf>>
file://C:\\temp\\GW}00001.HTM 07/17/2007
6-: V, e-m-p, \\GWJ00, 01., T, MP Page 11[
[c\\tri....OO
.M Pg 1
Mail Envelope Properties (46255987.2B9 : 5:49849)
Subject:
Creation Date From:
Created By:
Rancho Seco Well Sample Analysis 04/17/2007 7:34:07 PM "Robert E. Jones" <RJones2(@smud.org>
RJones2(@smud.org Recipients nrc.gov OWGWPOO4.HQGWDOO1 JBH (John Hickman)
Post Office OWGWPO04.HQGWDO0 1 Route nrc.gov Files MESSAGE TEXT.htm Case Narrative.pdf Mime.822 Options Expiration Date:
Priority:
ReplyRequested:
Return Notification:
Concealed
Subject:
Security:
Size 205 852 1204193 1651335 None Standard No None No Standard Date & Time 04/17/2007 7:34:07 PM Junk Mail Handling Evaluation Results Message is eligible for Junk Mail handling This message was not classified as Junk Mail Junk Mail settings when this message was delivered Junk Mail handling disabled by User Junk Mail handling disabled by Administrator Junk List is not enabled Junk Mail using personal address books is not enabled Block List is not enabled
TABLE OF CONTENTS Case Narrative.............................
1 C hain of C ustody.............
..........................,............. 3 Cooler Receipt Checklist.............................................
5 Radiological Analysis...................................................
7 Sample Data Summary...........................................
28 Quality Control Data 38
CASE NARRATIVE ED -I
Receipt Narrative for Rancho Seco Nuclear Generating Station Work Order: 151.732 January 03, 2006 Laboratory Identification:
General Engineering Laboratories, LLC 2040 Savage Road Charleston, South Carolina 29407 (843.) 556-8171 Summary_
Sample receipt The samples arrived at General Engineering Laboratories, LLC, Charleston, South Carolina on December 12, 2005 for analysis.
Sample Identification The laboratory received the following samples:
Laboratory ID Client ID 151732001
- 1 MWIC 151732002
- 2 MW2A 151732003
- 3 MW3A 151.732004
- 4 MW4A Case Narrative Sample analyses were conducted using methodology as outlined in General Engineering Laboratories (GEL) Standard Operating Procedures. Any technical or administrative problems during analysis, data review, and reduction are contained in the analytical case narratives in the enclosed data package.
Data Package_.
The enclosed data package contains the following sections: Case Narrative, Chain of Custody, Cooler Receipt Checklist, Data Package Qualifier Definitions and data from the following fr-actions:
Radiochemistry.
Project Manager GENERAL ENGINEERING LABORATORIES. LLC a Member of THE GEL GROUP, INC.
Page 2 of 58 P.O. Box 30712 ° Charleston, SC 2941.7 - 2040 Savage Road (29407)
Phone (843) 556-8171 - Fax (843) 766-1178 - www.gel.com
CHAIN OF CUSTODY Palcyp 3
20060997M Page:
of t
General Engineering Laboratories. LLC Project #:
GEL Chain of Custody and Analytical Request 21040 Savage Road GEL Quote 4:G
.L P7-Charleston, SC 29407 COC Number GEL Work Order Number:
Phone: 1243) 556-8171 PO Number:
.Fax:
(943) 766-1178 C
(.L2 Phone #:
7/
32 Sample Analysis Requested ( (Fill in the number of containers for each test)
- oject/Site Name:,12ý Sold2-Preservati e Type' (6)
~~~~~~~S uld-this satinpiee be
' r V
osord by:
Sen Reslt To 4
Comments
(&Ilected by:-~
Sen Reut To:t t&
2C Note: extra sample is Ti5e 75->'Sml I
Clets o
I Xd tld ai4 required for sample 1 Z Sample ID' t
N,,,.
.i QC "c
Field,,,
specific QC For compxosiies, - 1ndicate start and stop doate/time tmm-d-,)
tlilttara(hhmm).
IL%
8i 1__
o q,
N11 GM "i
II S 4_
2* PA Wu fr j..-dc, wlo tk}
Nq eJ 6A-1 1 1 S TA eqet
-VNoml Rush:"
Specify:
Faxjec tResrults:t Yes I
K- )
Circle Deliverable of A
( QC Summar Level I /Level 2 ILevel?/Lel Remarks: Are there an' known hazards applicable to these samples? Ifso. please list the hazards Samtle Collection Tinme.Zne Eastern Pacific Central Other Mountain Chain of Custody Signatures Sample Shipping and Delivery Details Relinquished By (Signed)
Date Time Recalled by (signed)
Date Time (GEL PM:
I A~n.
stiLeA~L/
~ 4 %
j~D
('i~t{L~
L~i
~
10 LY Method of Shipment:
Date Shippd:
2 2
4:irbilt 0:
3 3
Airbilt 5:
I.) Chuin kJ`Custlkxk Number= Client Determined Z, )C Codes: N - Normal Sample. TB = Trip Blank. FD Field Duplicate. EB = Equipment Bla"k. MS = Matrix Spike Sample. MSD = Matrix Spike Duplicate Sample. G = Grob. C - Composite 3.I Vild Filtjtered: Frr liquid matrices. indicate stith a - Y - Ibr yes the sample was Iield filtered or - N - for sample was noI teld littered.
4.1t M;,rix Cdes: DWt)rinking Water. GW=Groundwaier. SW-Sutface Water. WW-)Vaste Water. W=Water. SO=Soil. SD=Sediment-SL=Sludge. SS-Solid Waste. O0Oil. F=Fttter. P=Wipe. Vt=Urine. F=Fecal. N-Nasal 3.t Sample Analysis Requested: Analytical method requested (i.e. 8260B. 60108/74"0A I and number ofcontainers provided for each ti.e. 8260B. 3. 6010.&77470A
.1 t..
6.1 Prcservative I'sype: HA = lisdrt,,hloric Acid. N1 = Nitric Acid, SH = Sodium Hydroxide. SA - Sulfuric Acid. AA = Ascorbic Acid. HX = I lexane. ST = Sodium Thkictlr.e, If no preservative is added - leave field blank WHITE = LABORATORY :YELLOW
= FILE PINK = CLIENT For Lab Receiving Use Only Custho." Seol Intact?
YES NO Cooler Tentp:
C
COOLER RECEIPT CHECKLIST PA-f
SAMPLE RECEIPT & REVIEW FORM PM me only
[Client:
[SDGARCOC/Work jDate Received:
\\
\\\\
y" (PM(A) Review (ensure non-conforming items are resolved prior to signing):
,Received By:
L M
Sample Receipt Criteria Comments/Qualifiers (Required for Non-Conforning Items)
Shipping containers received intact ircle Applicable: seals broken damaged container leaking container other (describe) and sealed?
Samples requiring cold ircie Coolant k
iee bags blue ice (try icC
(
otber descrbe) 2 preservation within (4 +/- 2 C)?
Record preservation method.
Chain of custody documents included with shipment?
inSample containers intact and Circle Applicable: seals broken damaged container leaking container oilier (descrbe) sealed?
5 Samples requiring chemical Sampl ID's, containers affected and observed pH:
preservation at proper pH?
6 VOA vials free of headspace,
Sample ID's and containers affected:
(defined as < 6mm bubble)?
Are Encore containers present?
7 (If yes, immediately deliver to
-VOA laboratory) 8 Samples received within holding Id's and tests affected:
time?
9 Sample ID's on COC match ID's Sample [Ys and containers affected:
9 on bottles?
10 Date & time on COC match date Sample ID's affecte:
& time on bottles?
Number of containers received Sample ID's affected:
match number indicated on COC?
COC form is properly signed in relinquished/received sections?.
/0 Air Bill,Tracking #'s,& &2 i-L
( L '
Licf*/
O 1 4 Additional Comments I2 c*V
[
- &L-1 4 1
c(3?
o
&u L~I~*or Z RSO RAD Receipt #
r
- If > x2 area background is observed on samples identified as 'non-Susp e a
a's regulated/non-radioactive', contact the Radiation Safety group for finther
~investigation.
A Radiological Classification?
Maximum Counts Observed*:
('7 B lPCB Regulated?
Comments-Shipped as DOT'Hazardous ciMaterial? If yes,. contact Waste Hazard Class Shipped:
[Manager or ESH Manager.
V PM,(or PMA) review of Hazard classification:
nitials 11,;
Date:
6 ~*UULJ
RADIOLOGICAL ANALYSIS Page 7 of 58
Radiochemistry Case Narrative Rancho Seco Nuclear Generating Station (SMUD)
Work Order 151732 Method/Analysis Information Product:
AIphaspec Np, Liquid Analytical Method:
DOE EML HASL 300 Analytical Batch Number:
487213 Sample ID Client ID 151732001
- 1 MW1C 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200994824 Method Blank (MB) 1200994825 151732002(#2 MW2A) Sample Duplicate (DUP) 1200994826 151732002(#2 MW2A) Matrix Spike (MS) 1200994827 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-032 REV# 6.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732002 (#2 MW2A).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Page 8 of 58
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Manual Integration No manual integrations were performed on data in this batch.
Oualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Alphaspec Am241, Cm244, Liquid Analytical Method:
DOE EML HASL-300, Am-05-RC Modified Analytical Batch Number:
487212 Sample ID Client ID 151732001
- 1 MW1C 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200994820 Method Blank (MB) 1200994821 151732001(#1 MW1C) Sample Duplicate (DUP) 1200994822 151732001(#1 MW1C) Matrix Spike (MS) 1200994823 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-01 1 REV# 14.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (OC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW1C).
Page 9 of 58
QC Information Matrix Spike 1200994822 (#1 MW1C) failed required detection limit.
Technical Information:
Holding Time All sample procedures for this sampleset were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not.generated for this SDG.
Manual Integration No manual integrations were performed on data in this batch.
Qualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Alphaspec Pu, Liquid Analytical Method:
DOE EML HASL-300, Pu-I I-RC Modified Analytical Batch Number:
487214 Sample ID Client ID 151732001
- 1 MW1C 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200994828 Method Blank (MB) 1200994829 151732001 (# 1 MW I C) Sample Duplicate (DUP) 1200994830 151732001(#1 MWIC) Matrix Spike (MS) 1200994831 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-01 1 REV# 14.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Page 10 of 58
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Ouality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW IC).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in-this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Manual Integration No manual integrations were performed on data in this batch.
Qualifier information Q
QualifierT.
Reason Analyte Sample esults rejected due to tailing from the Pu-242 tracer region.
Plutonium-244 151732004 Method/Analysis Information Product:
Liquid Scint Pu241, Liquid Analytical Method:
DOE EML HASL-300, Pu-11-RC Modified Analytical Batch Number:
487216 Sample ID Client ID 151732001
- 1 MW1C 151732003
- 3 MW3A 151732004
- 4 MW4A 1200994833 Method Blank (MB) 1200994834 151732003(#3 MW3A) Sample Duplicate (DUP) 1200994835 151732003(#3 MW3A) Matrix Spike (MS) 1200994836 Laboratory Control Sample (LCS)
Page 11 of 58
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-035 REV# 7.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (0C0 Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732003 (#3 MW3A).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Manual Integration No manual integrations were performed on data in this batch.
qualifier information Manual qualifiers were not iequired.
Method/Analysis Information Product:
Liquid Scint Pu241, Liquid Analytical Method:
DOE EML HASL-300, Pu-I 1-RC Modified Analytical Batch Number:
490003 Page 12 of 58
Sample ID Client ID 151732002
- 2 MW2A 1201001147 Method Blank (MB) 1201001148 151732002(#2 MW2A) Sample Duplicate (DUP) 1201001149 151732002(#2 MW2A) Matrix Spike (MS) 1201001150 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-035 REV# 7.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Ouality Control (0C0 Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732002 (#2 MW2A).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis Samples were reprepped once due to interference and once for low tracer yield.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that' may deviate from referenced SOP or contractual documents. The following NCR was generated for this SDG:
NCR 280211 was generated due to RDL less than MDA. 1. Sample 151732002, 1201001147, and 1201001148 did not meet the client's detection limit due to small sample aliquot. The samples were reprepped twice and the aliquots reduced due to interference and low tracer yields. I. Reporting results.
Page 13of 58
Manual Integration No manual integrations were performed on data in this batch.
Oualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Gammaspec, Gamma, Liquid (Standard List)
Analytical Method:
EPA 901.1 Analytical Batch Number:
488129 Sample ID Client ID 151732001
- 1 MWIC 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200996967 Method Blank (MB) 1200996968 151732001 (# I MW 1 C) Sample Duplicate (DUP) 1200996969 151732001(#1 MW1C) Matrix Spike (MS) 1200996970 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-013 REV# 10.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (OC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MWlC).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information Page 14 of 58
All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis Sample 1200996967 (MB) was recounted due to analyst error.
Sample 151732001 (#1 MW 1 C) was recounted due to high relative percent difference/relative error ratio.
Sample 1200996969 (#1 MW1C) was recounted due to low/high recovery.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Additional Comments The relative percent difference for K-40 did not meet the client requirements. However, precision is shown when a relative error ratio of 2.37 is calculated.
Oualifier information Qualifier Reason Analyte Sample UI Data rejected due to high peak uncertainty.
Thorium-230 151732002 UI Data rejected due to high uncertainty.
Potassium-40 1200996967 UI Data rejected due to low abundance.'
Bismuth-214 151732003 Lead-212 151732002 Lead-214 151732003 Promethium-146 1200996967 Fhorium-234 1200996968 Uranium-238 1200996968.
UI ata rejected due to no valid peak.
Bismuth-212 151732003 1
_Cerium-144 1200996967 Method/Analysis Information Product:
Analytical Method:
Analytical Batch Number:
GFPC, Gross A/B, liquid EPA 900.0 489159 Sample ID 151732001 151732002 151732003 151732004 1200999365 1200999366 1200999367 1200999368 1200999369 Client ID
- 1 MWIC
- 2 MW2A
- 3 MW3A
- 4 MW4A Method Blank (MB) 151735002(S-0000222701) Sample Duplicate (DUP) 151735002(S-0000222701) Matrix Spike (MS) 151735002(S-0000222701) Matrix Spike Duplicate (MSD)
Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-00 1 REV# 9.
Page 15 of 58
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151735002 (S-0000222701).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
-Sample Re-prep/Re-analysis Samples were reprepped due to high relative percent difference/relative error ratio.
Chemical Recoveries All chemical recoveries meet the required acceptance limits for this sample set.
Gross Alpha/Beta Preparation Information High hygroscopic salt content in evaporated samples can cause the sample mass to fluctuate due to moisture absorption. To minimize this interference, the salts are converted to oxides by heating the sample under a flame until a dull red color is obtained. The conversion to oxides stabilizes the sample weight and ensures that proper alpha/beta efficiencies are assigned for each sample. Volatile radioisotopes of carbon, hydrogen, technetium, polonium and cesium may be lost during sample heating.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Oualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
GFPC, Total Sr, Liquid Analytical Method:
EPA 905.0 Modified Page 16 of 58
Analytical Batch Number:
487482 Sample ID Client ID 151732001
- 1 MWIC 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200995396 Method Blank (MB) 1200995397 151732003(#3 MW3A) Sample Duplicate (DUP) 1200995398 151732003(#3 MW3A) Matrix Spike (MS) 1200995399 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting df analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-004 REV# 9.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (OC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732003 (#3 MW3A).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Chemical Recoveries All chemical recoveries meet the required acceptance limits for this sample set.
Miscellaneous Information:
Page 17 of 58
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Qualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Liquid Scint Tc99, Liquid Analytical Method:
DOE EML HASL-300, Tc-02-RC Modified Analytical Batch Number:
487449 Sample ID Client ID 151732001
- 1 MWIC 151732002
- 2 MW2A 151732093
- 3 MW3A 151732004
- 4 MW4A 1200995311 Method Blank (MB) 1200995312 151732001 (#1 MW I C) Sample Duplicate (DUP) 1200995313 151732001(#l MWlC) Matrix Spike (MS) 1200995314 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-005 REV# 12.
Calibration Information:'
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Ouality Control (OC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW IC).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Page 18 of 58
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Qualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Liquid Scint Fe55, Liquid Analytical Method:
DOE RESL Fe-1, Modified Analytical Batch Number:
487444 Sample ID Client ID 151732001
- 1 MW1C 151732002
.#2 MW2A 151732003
- 3 MW3A 151732004
- 4 MW4A 1200995299 Method Blank (MB) 1200995300 151732001(#l MW1C) Sample Duplicate (DUP) 1200995301 15173200 1(#1 MWI C) Matrix Spike (MS) 1200995302 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-040 REV# 3.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Ouality Control (0C) Information:
Blank Information.
The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW1C).
Page 19 of 58
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Oualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Liquid Scint Ni63, Liquid Analytical Method:
DOE RESL Ni-l, Modified Analytical Batch Number:
487445 Sample ID Client ID 151732001
- 1 MWIC 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004.
- 4 MW4A 1200995303 Method Blank (MB) 1200995304 151732001(#1 MWIC) Sample Duplicate (DUP) 1200995305 151732001(#l MW1C) Matrix Spike (MS) 1200995306 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-022 REV# 7.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Page 20 of 58
Quality Control (OC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MWIC).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Qualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Analytical Method:
Analytical Batch Number:
Sample ID 151732001 151732002 151732003 151732004 1200995315 1200995316, 1200995317 1200995318 LSC, Tritium Dist, Liquid EPA 906.0 Modified 487451 Client ID
- 1 MWIC
- 2 MW2A
- 3 MW3A
- 4 MW4A Method Blank (MB) 151732001(#l MW1C) Sample Duplicate (DUP) 151732001(#1 MWIC) Matrix Spike (MS)
Laboratory Control Sample (LCS)
SOP Reference Procedure' for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-002 REV# 10.
Calibration Information:
Calibration Information Page 21 of 58
All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (QC) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW1C).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Oualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Analytical Method:
Analytical Batch Number:
Liquid Scint C14, Liquid EPA EERF C-01 Modified 487453 Sample ID 151732001 151732002 151732003 151732004 1200995323 1200995324 1200995325 1200995326 Client ID
- 1.MW1C
- 2 MW2A
- 3 MW3A
- 4 MW4A Method Blank (MB) 151732001(#l MW1C) Sample Duplicate (DUP) 151732001(#l MWlC) Matrix Spike (MS)
Laboratory Control Sample (LCS)
Page 22 of 58
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering
- Laboratories,'LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-003 REV# 8.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Ouality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW1C).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.'
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis None of the samples in this sample set required reprep or reanalysis.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Qualifier information Manual qualifiers were not required.
Method/Analysis Information Product:
Liquid Scint Pm147, liquid Analytical Method:
EPA EERF PM-01-1 Modified Analytical Batch Number:
487442 Sample ID Client ID Page 23 of 58
151732001
- 1 MW1C 151732002
- 2 MW2A 151732003
- 3 MW3A 151732004.
- 4 MW4A 1200995295 Method Blank (MB) 1200995296 151732004(#4 MW4A) Sample Duplicate (DUP) 1200995297 151732004(#4 MW4A) Matrix Spike (MS) 1200995298 Laboratory Control Sample (LCS)
SOP Reference Procedure for preparation, analysis and reporting of analytical data are controlled by General Engineering Laboratories, LLC as Standard Operating Procedure (SOP). The data discussed in this narrative has been analyzed in accordance with GL-RAD-A-020 REV# 6.
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732004 (#4 MW4A).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for. this sample set were performed within the required holding time.
Preparation Information.
All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis Samples 1200995295 (MB), 1200995296 (#4 MW4A), 151732001 (#1 MWIC), 151732002 (#2 MW2A),
151732003 (#3 MW3A) and 151732004 (#4 MW4A) were recounted due to high MDAs.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural, anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Oualifier information Manual qualifiers were not required.
Page 24 of 58
Method/Analysis Information Product:
Gamma Ni59, Liquid Analytical Method:
DOE RESL Ni-I Analytical Batch Number: 490999 Sample ID SOP Reference 151732001 Procedure for preparation, 151732002 analysis and reporting of 151732003 analytical data are 151732004 controlled by General Engineering Laboratories, 1201003355 LLC as Standard Operating 1201003356 Procedure (SOP). The data 1201003357 discussed in this narrative 1201003358 has been analyzed in accordance with GL-RAD-A-022 REV# 7.
Client ID
- 1 MWIC
- 2 MW2A
- 3 MW3A
- 4 MW4A Method Blank (MB) 151732001 (#1 MW I C) Sample Duplicate (DUP) 151732001(#1 MWIC) Matrix Spike (MS)
Laboratory Control Sample (LCS)
Calibration Information:
Calibration Information All initial and continuing calibration requirements have been met.
Standards Information Standard solution(s) for these analyses are NIST traceable and used before the expiration date(s).
Sample Geometry All counting sources were prepared in the same geometry as the calibration standards.
Quality Control (0C) Information:
Blank Information The blank volume is representative of the sample volume in this batch.
Designated QC The following sample was used for QC: 151732001 (#1 MW1C).
QC Information All of the QC samples met the required acceptance limits.
Technical Information:
Holding Time All sample procedures for this sample set were performed within the required holding time.
Preparation Information All preparation criteria have been met for these analyses.
Sample Re-prep/Re-analysis Page 25 of 58
Samples were reprepped with a higher volume to meet therequired detection limits.
Chemical Recoveries All chemical recoveries meet the required acceptance limits for this sample set.
Miscellaneous Information:
NCR Documentation Nonconformance reports are generated to document any procedural anomalies that may deviate from referenced SOP or contractual documents. An NCR was not generated for this SDG.
Qualifier information Manual qualifiers were not required.
Certification Statement Where the analytical method has-been performed under NELAP certification, the analysis has met all of the requirements of the NELAC standard unless otherwise noted in the analytical case narrative.
Review Validation:
GEL requires all analytical data to be verified by a qualified data validator. In addition, all data designated for CLP or CLP-like packaging will receive a third level validation upon completion of the data package.
The following data validator verified the information presented in this case narrative:
Reviewer:
Q o,
,,.'(
0.
Page 26 of 58
General Engineering Laboratories Form GEL-NCR Rev. 06/05 NCR Report No.:
280211 Revision No.-
COMPANY - WIDE NONCONFORMANCE REPORT Mo.Day Yr.
Division:
Quality Criteria:
Type:
29-DEC-05 Radiochemistry Specifications Process Instrument Type:
Test I Method:
Matrix Type:
Client Code:
LSC DOE EML HASL-300, Pu-11-RC Liquid SMUD Modifled Batch ID:_
Sample Numbers:
490003 See Below Potentially affected work order(s)(SDG): 151732 Application Issues:
RDL less than MDA Specification and Requirements NRG Disposition:
Nonconformance
Description:
- 1. Sample 151732002, 1201001147, and 1201001148 did not meet the
- 1. Reporting results.
client's detection limit due to small sample aliquot. The samples were reprepped twice and the aliquots reduced due to interference and low tracer yields.
Originator's Name:
Melanie Aycock 29-DEC-05 Quality Review:
Data Validator/Group Leader:
Heather Anderson 30-DEC-05 Director:
Page 1 Page 27 of 58
SAMPLE DATA
SUMMARY
Page 28 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com Certificate of Analysis Report for for SMUDOO 1 Rancho Seco Nuclear Generating Station Client SDG: 151732 GEL Work Order: 151732 Lab Sample ID 151732001 151732002 151732003 151732004 Sample(s) Contained within this report:
Client Sample ID Sample Description
- 1 MW1C
- 2 MW2A
- 3 MW3A
- 4 MW4A N/A N/A N/A N/A Collected 12/05/2005 14:00 11/29/2005 12:30 12/05/2005 14:50 12/06/2005 11:50 Where the analytical method has been performed under NELAP certification, the analysis has met all of the requirements of the NELAC standard unless qualified on the Certificate of Analysis.
This data report has been prepared and reviewed in accordance with General Engineering Laboratories, LLC standard operating procedures. Please direct any questions to your Project Manager, Martha Harrison.
Re ve.,e by Reviewed by Page 29 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732001
- 1 MW1C Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: December 05, 2005 Receive Date: December 12, 2005 Report Date: December 31, 2005 I
.Critical tyMDA RL TPU Level Units
.A n.aly te....................
ALPHA BETA H-3 C-14 Fe-55 Ni-63 Tc-99 Pm-147 Pu-241 Total Strontium Alpha Spec Np-237 Pu-238 Pu-239/240 Pu-244 Am-241 Cm-243/244 Gamma Spec Be-7 Na-22 K-40 Cr-51 Mn-54 Fe-59 Co-56 Co-57 Co-58 Co-60 Ni-59 Zn-65 Y-88 Zr-95 Nb-94 Nb-95 Ru-106 Ag-110m Sn-113 2
Run Date Qualifier Activity Uncertaini 12/23/05 12/23/05 12/14/05 12/14/05 12/15/05 12/15/05 12/19/05 12/21/05 12/16/05 12/14/05 12/15/05 12/15/05 12/15/05 12/15/05 12/15/05 12/15/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/29/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 12/28/05 5.05E-01 2.23E+00 1.13E+02 9.08E+00
-2.82E+01 1.76E+00
-5.14E+00 3.04E+00
-1.29E+00
-7.03E-01 3.85E-02 0.OOE+00
-5.96E-02 1.65E-01 2.02E-01
-4.74E-02
-6.28E-01
- 1.20E+00 3.27E+01
-2.41E+00
-1.08E+00 1.72E+00 2.13E-01 7.97E-01
-1.79E+00
-1.02E-01 1.51E+01
-7.23E-01
-3.74E-01
-2.36E-01.
6.19E-01 7.80E-01 1.15E+01
-2.34E-01
-9.27E-01 6.26E-01 1.89E+00 2.11 E+02 1.57E+01 3.71E+01 1.46E+01 1.34E+01 4.19E+00 7.06E+00 4.89E-01 1.53E-01 1.62E-01 6.74E-02 2.29E-01 2.37E-01 1.40E-01 1.88E+01 1.86E+00 2.75E+01 2.50E+01 1.57E+00 4.28E+00 1.88E+00 1.66E+00 1.94E+00 1.84E+00 1.04E+01 3.71E+00 1.65E+00 3.75E+00 1.67E+00 2.43E+00 1.64E+01 1.61E+00 2.56E+00 1.13E+00 3.71E+00 3.59E+02 2.66E+01 5.66E+01 2.51 E+0 1 2.3 1E+01 7.06E+00 1.22E+01 1.27E+00 4.39E-01 2.24E-01 5.51E-01 2.24E-01 3.28E-01 4.39E-01 3.17E+01 3.14E+00 3.05E+01 4.25E+01 2.58E+00 8.09E+00 3.34E+00 2.96E+00 3.13E+00 3.32E+00 1.66E+01 6.63E+00 2.94E+00 6.55E+00 3.01E+00 4.40E+00 3.03E+01 2.82E+00 4.25E+00 5.OOE+00 5.OOE+00 7.OOE+02 5.OOE+01 1.00E+02 5.OOE+01 5.OOE+01 1.00E+01 1.50E+01 2.OOE+00 1.00E+00 1.00E+00 1.00E+00 1.00E+00 1.00E+00 1.00E+00 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 2.OOE+01 1.00E+01.
1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 6.29E-01 1.89E+00 2.11 E+02 1.57E+0 1 3.71 E+0 I 1.46E+01 1.34E+01 4.20E+00 7.06E+00 4.90E-01 1.53E-01 1.62E-01 6.77E-02 2.30E-01 2.38E-01 1.40E-01 1.85E+01 1.82E+00 2.70E+0I 2.45E+01 1.53E+00 4.20E+00 1.84E+00 1.63E+00 1.90E+00 1.80E+00 1.02E+01 3.64E+00 1.61 E+00 3.67E+00 1.64E+00 2.38E+00 1.60E+01 1.58E+00 2.5 1E+00 4.20E-01 1.77E+00 1.73E+02 1.30E+01 2.79E+01 1.22E+01 1.13E+01 3.47E+00 5.95E+00 5.98E-01 1.1 9E-0 I 0.OOE+00 1.63E-01 0.OOE+00 7.5 1E-02 1.30E-01 1.50E+01 1.43E+00 1.39E+01 2.03E+01 1.1 9E+00 3.73E+00 1.55E+00 1.43E+00 1.45E+00 1.52E+00 7.68E+00 3.05E+00 1.28E+00 3.06E+00 1.42E+00 2.05E+00 1.43E+01 1.32E+00 2.02E+00 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/ll pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U Target analyte was analyzed for but not detected above the MDL or LOD.
UI Uncertain identification for gtamma spectroscopy.
X Lab-specific qualifier-please see case narrative, data summary package or contact your project manager for details.
Page 30 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
151732001 Client: Rancho Seco Nuclear Generating Station Client Sample ID:
- 1MW1C Collect Date: December 05, 2005 Matrix:
Ground Water Receive Date: December 12, 2005 Amount of Sample Received:
Report Date: December 31, 2005 2
Critical Analyte Run Date Qualifier Activity Uncertainty MDA R.
TPU Level Units Sb-124 12/28/05 U
-1.85E+00 3.98E+00 6.82E+00 1.00E+01 3.90E+00 2.98E+00 pCi/L Sb-125 12/28/05 U
2.44E+00 5.31E+00 8.24E+00 1.00E+01 5.20E+00 3.90E+00 pCi/L Cs-134 12/28/05 U
6.24E-01 1.82E+00 3.30E+00 1.00E+01 1.78E+00 1.54E+00 pCi/L Cs-136 12/28/05 U
5.23E+00 6.86E+00 1.30E+01 1.003E+01 6.72E+00 5.99E+00 pCi/L Cs-137 12/28/05 U
3.08E-01 2.38E+00 2'99E+00 1.00E+01 2.33E+00 1.40E+00 pCi/L Ba-133 12/28/05 U'
5.06E-01 2.60E+00 3.95E+00 1.00E+01 2.55E+00 1.88E+00 pCi/L Ba-140 12/28/05 U
1.07E+01 2.01E+01 3.69E+01 1.00E+01 1.97E+01 1.75E+01 pCi/L Ce-139 12/28/05 U
-4.91E-01.
1.86E+00 3.19E+00 1.00E+01 1.82E+00 1.54E+00 pCi/L Ce-141, 12/28/05 U
2.85E+00 6.81E+00 7.27E+00 1.00E+01 6.67E+00, 3.51E+00 pCi/L Ce-144 12/28/05 U
-6.92E+00 1.25E+01 2.14E+01 1.00E+01 1.22E+01 1.03E+01 pCi/L Nd-147 12/28/05 U
1.63E+01 4.62E+01 8.41E+01 I1.00E+01 4.53E+01 3.97E+01 pCi/L Pm-144 12/28/05 U
-5.1 IE-03 1.70E+00 2.99E+00 1.00E+01 1.67E+00 1.41E+00 pCi/L Pm-146 12/28/05 U
2.65E+00 2.38E+00 4.31E+00 1.00E+01 2.33E+00 2.05E+00 pCi/L Eu-152 12/28/05 U
4.63E+00 6.19E+00 9.76E+00 1.00E+01 6.06E+00 4.67E+00 pCi/L Eu-154 12/28/05 U
-3.39E+00 5.16E+00 8.72E+00 1.00E+01 5.05E+00.
3.97E+00 pCi/L Eu-155 12/28/05 U
1.39E-01 6.87E+00 1.13E+01 1.00E+01 6.73E+00 5.46E+00 pCi/L lr-192 12/28/05 U
-1.13E+00 2.03E+00 3.36E+00 1.00E+01 1.99E+00 1.60E+00 pCi/L Hg-203 12/28/05 U
-2.83E-01 2.55E+00 4.35E+00 1.00E+01 2.50E+00 2.08E+00 pCi/L TI-208 12/28/05 U
9.45E-01 2.91E+00 3.41E+00 1.00E+01 2.85E+00 1.61E+00 pCi/L Pb-210 12/28/05 U
-1.98E+01 3.87E+02 5.80E+02 1.00E+01 379E+02 2.80E+02 pCi/L Pb-212 12/28/05 U
1.35E+00 5.89E+00 5.61E+00 1.00E+01 5.77E+00 2.70E+00 pCi/L Pb-214 12/28/05 U
4.05E+00 6.54E+00 6.16E+00 1.00E+01 6.41E+00 2.93E+00 pCi/L Bi-212 12/28/05 U
5.49E+00 1.32E+01 2.40E+01 1.00E+01 1.29E+01" 1.13E+01 pCi/L Bi-214 12/28/05 U
4.62E+00 8.74E+00 7.62E+00 1.00E+01 8.56E+00 3.64E+00 pCi/L Ra-228 12/28/05 U
-3.73E-01 6.17E+00 1.07E+01 1.00E+01 6.05E+00 4.99E+00 pCi/L Ac-228 12/28/05 U
-3.73E-01 6.17E+00 1.07E+01 1.00E+01 6.05E+00 4.99E+00 pCi/L Th-230 12/28/05 U
4.62E+00 8.73E+00 5.56E+00 1.00E+01 8.56E+00 2.61E+00 pCi/L.
Th-234 12/28/05 3,U 1.66E+02 9.50E+01 1.68E+02 1.00E+01 9.31E+01 8.18E+01 pCi/L U-235 12/28/05 U
8.78E+00 1.84E+01 2.24E+01 1.00E+01 1.81E+01 1.09E+01 pCi/L U-238 12/28/05 3,U 1.66E+02 9.50E+01 1.68E+02 1.00E+01 9.31E+01 8.18E+01 pCi/L Np-239 12/28/05 U
2.1 IE-01 1.27E+01 2.07E+01 1.00E+01 1.24E+01 1.00E+01 pCi/L Am-241 12/28/05 U
-5.11E+00 1.16E+01 1.90E+01 1.00E+01 1.13E+01 9.17E+00 pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U Target analyte was analyzed for butnot detected above the MDL or LOD.
UI Uncertain identification for gamma spectroscopy.
X Lab-specific aualifier-please see case narrative, data summary package or contact your project manager for details.
Page 31 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
151732002 Client: Rancho Seco Nuclear Generating Station Client Sample ID:
- 2 MW2A Collect Date: November 29, 2005 Matrix:
Ground Water Receive Date: December 12, 2005 Amount of Sample Received:
Report Date: December 31, 2005 2
1Critical Analyte Run Date Qualifier Activity Uncertainty MDA RL TPU Level Units ALPHA 12/22/05 3
7.07E+00 2.45E+00 2.95E+00 5.OOE+00 2.57E+00 1.34E+00 pCi/L BETA 12/22/05 3
1.71E+01 3.07E+00 4.36E+00 5.OOE+00 3.12E+00 2.12E+00 pCi/L H-3 12/14/05 U
2.62E+02 2.17E+02 3.59E+02 7,00E+02 2.17E+02 1.73E+02 pCi/L C-14 12/14/05 U
6.18E+00 1.56E+01 2.66E+01 5.00E+01 1.56E+01 1.30E+01 pCi/L Fe-55 12/15/05 U
-1.98E+01 3.26E+01 5.02E+01 1:00E+02 3.26E+01 2.47E+01 pCi/L Ni-63 12/15/05 U
3.05E+00 1.69E+01 2.89E+01' 5.00E+01 1.69E+01 1.41E+01 pCi/L Tc-99 12/19/05 U
-3.14E+00 1.17E+01 2.01E+01 5.OOE+01 1.17E+01 9.89E+00 pCi/L, Pm-147 12/21/05 U
4.64E+00 5.15E+00 8.64E+00 1.00E+01 5.16E+00 4.25E+00 pCi/L Pu-241 12/28/05 U
2.13E+01 3.44E+01 5.79E+01 1.50E+01 3.44E+01 2.85E+01 pCi/L Total Strontium 12/14/05 U
-2.20E-02 4.46E-01 1.03E+00 2.OOE+00 4.45E-01 4.77E-01 pCi/L Alpha Spec Np-237 12/15/05 U
8.02E-02 1.57E-01 2.17E-01 1.00E+00 1.57E-01 0.OOE+00 pCi/L Pu-238 12/15/05 U
-5.22E-02 2.59E-01 8.37E-01 1.00E+00 2.59E-01 3.08E-01 pCi/L Pu-239/240 12/15/05 U
-1.34E-01 2.04E-01 8.36E-01 1.00E+00 2.05E-01 3.08E-01 pCi/L Pu-244 12/15/05 U
2.05E-01 2.82E-01 4.83E-01 1.00E+00 2.83E-01 1.31E-01 pCi/L Am-241 12/15/05 U
5.45E-02 1.45E-01 2.72E-01 1.00E+00 1.45E-01 0.OOE+00 pCi/L Cm-243/244 12/15/05 U
7.64E-02 2.03E-01 5.01E-01 1.00E+00 2.03E-01 1.15E-01 pCi/L Gamma Spec Be-7 12/19/05 U
-9.72E+00 1.71E+01 2.78E+01 1.OOE+01 1.68E+01 1.30E+01 pCi/L Na-22 12/19/05 U
-2.31E-01 1.68E+00 3.01E+00 1.00E+01 1.65E+00 1.34E+00 pCi/L K-40 12/19/05 4.38E+01 3.61E+01 2.74E+01 1.00E+01 3.54E+01 1.20E+01 pCi/L Cr-51 12/19/05 U
-4.62E+00 2.46E+01 4.14E+01 1.00E+01 2.41E+01 1.97E+01 pCi/L Mn-54 12/19/05 U
-1.37E-01 1.62E+00 2.91E+00 1.00E+01 1.59E+00 1.33E+00 pCi/L Fe-59 12/19/05 U
1.59E+00 6.03E+00 8.22E+00 1.00E+01 5.91E+00 3.76E+00 pCi/L Co-56 12/19/05 U
2.72E-01 1.93E+00 3.52E+00 1.00E+01 1.89E+00 1.62E+60 pCi/L Co-57 12/19/05 U
-2.26E-01 1.76E+00 3.01E+00 1.00E+01 1.72E+00 1.46E+00 pCi/L Co-58 1.2/19/05 U
1.38E+00 1.83E+00 3.55E+00 1.00E+01 1.79E+00 1.64E+00 pCi/L Co-60 12/19/05 U
1.88E-01 1.68E+00 3.12E+00 1.00E+01
.1.65E+00 1.39E+00 pCi/L Ni-59 12/30/05 U
1.37E+01 9.82E+00 1.57E+01 2.OOE+01 9.68E+00 7.49E+00 pCi/L Zn-65 12/19/05
- U
-1.19E+00 3.98E+00 6.90E+00 1.00E+01 3.90E+00 3.14E+00 pCi/L Y-88 12/19/05 U
1.89E+00 1.99E+00 4.30E+00 1.00E+01 1.95E+00 1.91E+00 pCi/L Zr-95 12/19/05 U
3.29E+00 4.39E+00 6.18E+00 1.00E+01 4.30E+00 2.85E+00 pCi/L Nb-94 12/19/05 U
-5.65E-01 1.69E+00 2.95E+00 1.00E+01 1.66E+00 1.38E+00 pCi/L Nb-95 12/19/05 U
-4.92E-01 2.23E+00 3.95E+00 1.00E+01 2.19E+00 1.81E+00 pCi/L Ru-106 12/19/05 U
-1.15E+01 1.51E+01 2.55E+01 1.00E+01 1.48E+01 1.18E+01 pCi/L Ag-ll0m 12/19/05 U
-9.87E-01 1.42E+00 2.41E+00 1.00E+01 1.39E+00 1.1OE+00 pCi/L Sn-113 12/19/05 U
1.18E+00 2.41E+00 4.27E+00 1.00E+01 2.37E+00 2.02E+00.
pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values,
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U Target analyte was analyzed for but not detected above the MDL or LOD.
UI Uncertain identification for gamma spectroscopy.
X Lab-specific qualifier-please see case narrative, data summary package or contact your project manager for details.
Page 32 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732002
- 2 MW2A Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: November 29, 2005 Receive Date:
Report Date:
December 12, 2005 December 31, 2005 Anallyte
.A...y.
te...
Sb-124 Sb-125 Cs-134 Cs-136 Cs-137 Ba-133 Ba-140 Ce-139 Ce-141 Ce-144 Nd-147 Pm-144 Pm-146 Eu-152*
Eu-154 Eu-155 Ir-192 Hg-203 TI-208 Pb-210 Pb-212 Pb-214 Bi-212 Bi-214 Ra-228 Ac-228 Th-230 Th-234 U-235 U-238 Np-239 Am-241 2
1 Run Date Qualifier Activity Uncertainty MDA Critical TPU Level 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/1 9/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05
'12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 U
U U
U U
U U
U U
U U
U U
U U
U U
U U
UU UU U
U U
U UU U
U U
U U
3.20E+00 2.82E+00 4.83E-01
-3.68E-01 6.04E-0 1 9.28E-01 1.07E+01 9.16E-02
-5.OOE-01
-4.37E+00
-1.47E+01 2.17E-01
-7.84E-01 3.25E+00
-6.3 1E-01 3.97E+00 1.92E+00
-1.55E+00 1.55E+00 2.04E+02 0.OOE+00 5.40E+00 9.351E+00 5.80E+00 6.35E+00 6.35E+00 0.OOE+00 5.44E+01 1.26E+01 5.44E+01
-6.41 E+00
-1.38E+01 4.60E+00 5.09E+00 1.75E+00 6.08E+00 1.65E+00 2.32E+00 2.12E+01 1.85E+00 4.52E+00 1.39E+01 3.85E+01 1.59E+00 2.3 1E+00 5.27E+00 4.69E+00 7.25E+00 2.17E+00 2.47E+00 1.95E+00 5.25E+02 3.74E+00 8.77E+00 1.211E+01 7.27E+00 6.28E+00 6.28E+00 7.27E+00 1.66E+02 1.53E+01 1.66E+02 1.34E+01 1.59E+01 9.54E+00 9.01E+00 3.26E+00 1.09E+01 2.75E+00 3.64E+00 3.28E+01 3.17E+00 7.72E+00 2.35E+01 6.35E+01 2.89E+00 3.83E+00 9.34E+00 8.40E+00 1.28E+01 3.87E+00 4.07E+00 3.67E+00 8.32E+02 6.76E+00 7.77E+00 2.35E+01 8.56E+00 1.23E+01 1.23E+01 5.72E+00 1.91E+02 2.38E+01 1.91E+02 2.26E+01 2.32E+-01 1.00E+01 1.00E+01 1.001E+01 1.00E+01 1.001E+01 1.00E+01 1.60E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.001E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 4.5 1E+00 4.99E+00 1.72E+00 5.96E+00 1.61E+00 2.28E+00 2.07E+01 1.81E+00 4.43E+00 1.36E+01 3.77E+01 1.55E+00 2.26E+00 5.16E+00 4.60E+00 7.1OE+00 2.13E+00 2.42E+00 1.91E+00 5.14E+02 3.66E+00 8.60E+00 1.19E+01 7.12E+00 6.15E+00 6.15E+00 7.12E+00 1.62E1+02 1.501E+01 1.62E+02 1.31E+01 1.56E+01 4.25E+00 4.26E+00 1.50E+00 4.95E+00 I.27E+00 1.72E+00 1.54E+01 1.53E+00 3.74E+00 1.14E+01 2.95E+01 1.34E+00 1.80E+00 4.44E+00 3.73E+00 6.19E+00 1.85E+00 1.95E+00 1.73E+00 4.02E+02 3.27E+00 3.72E+00 1.08E+01 4.09E+00 5.69E+00 5.69E+00 2.67E+00 9.26E+01 1.16E+01 9.26E+01 1.1OE+01 1.12E+01
_Uni-ts.-..
pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pqCi/L pCi/L p Ci/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Note(s): 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U U'
X Target analyte was analyzed for but not detected above the MDL or LOD.
Uncertain identification for gamma spectroscopy.
Lab-specific tqualifier-please see case narrative, data summary vPackage or contact your proiect manager for details.
Page 33 -of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732003
- 3 MW3A Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: December 05, 2005 Receive Date: December 12, 2005 Report Date: December 31, 2005 2
Analyte Run Date Qualifier Activity MDA I--
Critical R.
TPU Level Uncertainty Units ALPHA 12/23/05 U
-3.14E-01 3.74E-01 1.25E+00 5.OOE+00 3.74E-01 4.89E-01 pCi/L BETA 12/23/05 U
2.18E+00 2.19E+00 4.38E+00 5.OOE+00 2.19E+00 2.10E+00 pCi/L H-3 12/14/05 U
5.03E+01 2.12E+02 3.65E+02 7.OOE+02 2.12E+02 1.76E+02 pCi/L C-14 12/14/05 U.
7.OOE+00 1.57E+01 2.67E+O1 5.OOE+01 1.57E+01 1.30E+01 pCi/L Fe-55 12/15/05 U
-1.19E+01 3.88E+01 5.89E+01 1.00E+02 3.88E+01 2.90E+01 pCi/L Ni-63 12/15/05 3
2.76E+01 1.62E+01 2.64E+01 5.OOE+01 1.62E+01 1.28E+01 pCi/L Tc-99 12/19/05 U
5.45E-01 1.23E+01 2.10E+01 5.OOE+01 1.23E+01 1.03E+01 pCi/L Pm-147 12/21/05 U
6.44E-OO 4.71E+00 7.82E+00 1.00E+01 4.72E+00 3.84E+00 pCi/L Pu-241 12/16/05 U
-2.02E+00 5.63E+00 9.75E+00 1.50E+01 5.64E+00 4.77E+00 pCi/L Total Strontium 12/14/05 U
-2.79E-01 3.81E-0I 9.55E-01 2.OOE+00 3.81E-01 4.43E-01 pCi/L Alpha Spec Np-237 12/15/05 U
-1.08E-01 9.48E-02 7.03E-01 1.00E+00 9.54E-02 2.30E-01 pCi/L Pu-238 12/15/05 U
-1.33E-01 2.43E-01 8.56E-01 1.00E+00 2.43E-01 3.30E-01 pCi/L Pu-239/240 12/15/05 U
-8.40E-02 1.75E-01 6.91E-01 1.00E+00 1.75E-01 2.47E-01 pCi/L Pu-244 12/15/05 U
5.50E-02 1.46E-01 3.61E-01 1.00E+00 1.46E-01 8.25E-02 pCi/L Am-241 12/i5/05 U
6.33E-02 1.40E-0I 3.10E-01 1.00E+00 1.40E-01 7.08E-02 pCi/L Cm-243/244 12/15/05 U
4.72E-02 1.25E-01 3.10E-01 1.00E+00 1.25E-01 7.08E-02 pCi/L Gamma Spec Be-7 12/19/05 U
2.78E+00 1.90E+01 3.32E+01 1.00E+01 1.86E+61 1.57E+01 pCi/L Na-22 12/19/05
-U 7.56E-01 1.84E+00 3.49E+00.
1.00E+01 1.80E+00 1.58E+00 pCi/L K-40 12/19/05 U
1.11E+01 2.92E+01 2.90E+01 1,00E+01 2.87E+01 1.29E+01 pCi/L Cr-51 12/19/05 U
-1.17E+00 2.23E+01 3.87E+01 1;00E+01 2.18E+01 1.85E+01 pCi/L Mn-54 12/19/05 U
2.41E+00 1.93E+00 3.77E+00 1.00E+01 1.89E+00 1.77E+00 pCi/L Fe-59 12/19/05 U
4.17E-02 4.52E+00 8.08E+00 1.00E+01 4.43E+00 3.73E+00 pCi/L Co-56 12/19/05 U
-3.13E-01 1.91E+00 3.41E+00 1.00E+01 1.88E+00 1.57E+00 pCi/L Co-57 12/19/05 U
8.OOE-01 1.93E+00 2.91E+00 1.00E+01 1.89E+00 1.41E+00 pCi/L Co-58 12/19/05 U
1.54E+00
- 2. 15E+00 4.06E+00 1.00E+01 2.11 E+00 1.90E+00 pCi/L Co-60 12/19/05 U
1.32E+00 2.29E+00 3.35E+00 1.00E+01 2.25E+00 1.51E+00 pCi/L Ni-59 12/30/05 U
2.67E+00 1.07E+01 1.56E+01 2.OOE+01 1.04E+01 7.34E+00 pCi/L Zn-65 12/19/05 U
-6.19E+00 4.57E+00 7.02E+00 1.00E+01 4.48E+00 3.21E+00 pCi/L Y-88 12/19/05 U
-7.51E-02 1.95E+00 3.65E+00 1.00E+01 1.91E+00 1.60E+00 pCi/L Zr-95 12/19/05 U
-2.56E+00 4.02E+00 5.76E+00 1.00E+01 3.94E+00 2.66E+00 pCi/L Nb-94 12/19/05 U
-1.22E+00 1.86E+00 3.02E+00 1.00E+01 1.82E+00 1.41E+00 pCi/L Nb-95 12/19/05 U
1.18E+00 2.39E+00 4.44E+00 1.00E+01 2.34E+00 2.08E+00 pCi/L Ru-106 12/19/05, U
1.01E+01 1.70E+01 3.08E+01 1.00E+01 1.67E+01 1.45E+01 pCi/L Ag-110m 12/19/05 U
-1.86E+00 2.14E+00 3.42E+00 1.00E+01
- 2. i 0E+00 1.61E+00 pCi/L Sn-113 12/19/05 U
1.80E+00 2.55E+00 4.62E+00 1.00E+01 2.50E+00 2.20E+00 pCi/L Note(s): 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U UI X
Target analyte was analyzed for but not detected above the MDL or LOD.
Uncertain identification for gamma spectroscopy.
Lab-specific qualifier-please see case narrative, data summary package or contact your proiect manager for details.
Page 34 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 -:www.gel.com 10 CFR Part 50/61 Certificate.of Analysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732003
- 3 MW3A Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: December 05, 2005 Receive Date: December 12, 2005 Report Date: December 31, 2005 Sb-124 Sb-125 Cs-134 Cs-136.
Cs-137 Ba-133 Ba-140 Ce-139 Ce-141 Ce-144 Nd-147 Pm-144 Pm-146 Eu-152 Eu-154 Eu-155 Ir-192 Hg-203 TI-208 Pb-210 Pb-212 Pb-214 Bi-212 Bi-214 Ra-228 Ac-228.
Th-230 Th-234 U-235 U-238 Np-239 Am-241 2
Run Date Qualifier Activity Uncertainty MDA Critical RL TPU Level 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 12/19/05 U
U U
'U U
U U
U U
U U
U U
U U
U U
U U
U U
UUI UUI UUI U
U
.3
.U U
U U
U
-1.85E+00
-2.58E+00 7.61E-02
-3.79E-01 1.1OE+00 1.21E+00 5.81E+00
-1.16E-01 1.50E+00 8.16E+00 9.92E+00 8.05E-01
-3.64E-01
-7.22E-01 2.06E+00 2.24E-01
-1.04E-01 8.70E-01 3.1OE+00 2.6 1E+02 3.96E+00 0.OOE+00 0.OOE+00 0.00E+00 5.42E+00 5.42E+00 1.76E+01
-7.78E+01 8.68E+00
-7.78E+01 6.77E+00
-1.04E+01 4.85E+00 5.5 1E+00 2.03E+00 4.54E+00 2.25E+00 2.88E+00 1.41E+01 1.95E+00 4.3 1E+00 1.57E+01 3.03E+01 1.81E+00 2.52E+00 5.94E+00 5.14E+00 8.12E+00 2.12E+00 2.4 1E+00 2.26E+00 4.18E+02 4.04E+00 8.05E+00 5.17E+01 8.41E+00 1.48E+01 1.48E+01 8.4 1E+00 1.22E+02 1.52E+01 1.22E+02 1.48E+01 1.49E+01 8.59E+00 9.30E+00 3.67E+00 8.16E+00 3.99E+00 4.5 1E+00 2.50E+01 3.26E+00 6.79E+00 2.37E+01 5.39E+01 3.23E+00 4.34E+00 1.03E+01 9.72E+00 1.21E+01 3.70E+00 4.28E+00 4.22E+00 6.52E+02 6.92E+00 9.68E+00 2.38E+01 9.97E+00 1.34E+01 1.34E+01 6.54E+00 1.75E+02 2.29E+01 1.75E+02 2.24E+01 2.15E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+.01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 4.75E+00 5.40E+00 1.99E+00 4.45E+00 2.20E+00 2.82E+00 1.39E+01 1.9 IE+00 4.22E+00 1.54E+01 2.97E+01 1.77E+00 2.47E+00 5.82E+00 5.03E+00 7.96E+00 2.08E+00 2.36E+00 2.22E+00 4.09E+02 3.96E+00 7.89E+00 5.07E+01 8.24E+00 1.45E+01 1.45E+01 8.24E+00 1.1 9E+02 1.49E+01 1.1 9E+02 1.45E+01 1.46E+01 3.82E+00 4.4 1E+00 1.70E+00 3.7 1E+00 1.88E+00 2.15E+00 1.18E+01 1.58E+00 3.29E+00 1.15E+01 2.54E+01 1.52E+00.
2.05E+00 4.92E+00 4.4 1E+00 5.84E+00 1.77E+00 2.06E+00 2.OOE+00 3.16E+02 3.35E+00 4.68E+00
- 1. 1OE+01 4.79E+00 6.25E+00 6.25E+00 3.08E+00 8.52E+01 1.1 ]E+01 8.52E+01 I.09E+01 1.05E+01
.U -nits -----
pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U Target analyte was analyzed for but not detected above the MDL or LOD.
UI Uncertain identification for gamma spectroscopy.
X Lab-specific rlualifier-please see case narrative, data summary Packa2e or contact your project manager for details.
Page 35 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate of Analysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732004
- 4 MW4A Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: December 06, 2005 Receive Date: December 12, 2005 Report Date: December 31, 2005 M
Critical y M A L PU Level Units Analyte 2
Run Date Qualifier Activity Uncertaint ALPHA BETA H-3 C-14 Fe-55 Ni-63 Tc-99 Pm-147 Pu-241 Total Strontium Alpha Spec Np-237 Pu-238 Pu-239/240 Pu-244 Am-241 Cm-243/244 GammaSpec Be-7 Na-22 K-40 Cr-51 Mn-54 Fe-59 Co-56 Co-57 Co-58 Co-60 Ni-59 Zn-65 Y-88 Zr-95 Nb-94 Nb-95 Ru-106 Ag-110m Sn-113 12/23/05 12/23/05 12/15/05 12/14/05 12/16/05 12/15/05 12/19/05 12/22/05 12/16/05 12/14/05 12/15/05 12/15/05 12/15/05 12/15/05 12/15/05 12/15/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/30/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 U
1.21E+00 1.80E+00 3.61E+00 5.00E+00 1.80E+00 1.65E+00 U
1.53E+00 2.13E+00 4.30E+00 5.OOE+00 2.13E+00 2.06E+00 U
-2.70E+01 2.07E+02 3.62E+02 7.OOE+02 2.07E+02 1.75E+02 U
1.25E+01 1.59E+01 2.67E+01 5.00E+01 1.59E+01 1.30E+01 U
2.07E-01 3.37E+01 5.07E+01 1.00E+02 3.37E+01 2.49E+01 U
5.74E+00 1.56E+01 2.67E+01 5.OOE+01 1.56E+01 1.30E+01 U
-3.57E+00 1.30E+01 2.23E+01 5.OOE+01 1.30E+01 1.1OE+0I U
3.20E+00 4.65E+00 7.83E+00 1.00E+01 4.65E+00 3.85E+00 U
-3.20E+00 6.15E+00
.1.07E+01 1.50E+01 6.16E+00 5.23E+00 U
-I.51E-01 4.01E-01 9.64E-01 2.OOE+00 4.01E-01 4.47E-01 U
-1.01E-01 7.50E-02 5.27E-01 1.00E+00 7.56E-02 1.82E-01 U
-2.20E-01 2.46E-01
-1.05E+00 1.00E+00
.2.47E-01 3.98E-01 U
-6.72E-02 1.98E-01 6.21E-01 1.00E+00 1.98E-01 1.84E-01 3,X 1.07E+00 6.36E-01 5.53E-01 1.00E+00 6.48E-01 1.50E-01 U
-4.60E-02 3.13E-01 4.32E-01 1.00E+00 3.14E-01 0.00E+00 U
0.00E+00 3.13E-01 4.32E-01 1.00E+00 3.13E-01 0.00E+00 U
-9.69E+00 2.06E+01 3.37E+01 1.00E+01 2.02E+01 I1.57E+01 U
-2.26E-01 2.08E+00 3.82E+00 1.00E+01 2.04E+00 1.70E+00 U
2.97E+01 2.56E+01 5.20E+0I 1.00E+01 2.51E+01 2.39E+01 U
1.77E+01 4.01E+01 4.56E+01 1.00E+01 3.93E+01 2.17E+01 U
-3.57E-01 1.87E+00 3.28E+00 1.00E+01 1.83E+00 1.49E+00 U
-1.34E+00 4.66E+00 7.16E+00 I.00E+01 4.57E+00 3.17E+00 U
-4.89E-01 2.19E+00 3.81E+00 1.00E+01 2.14E+00 1.74E+00 U
-3.93E-01 2.07E+00 3.61E+00 1.00E+01 2.03E+00 1.74E+00 U
-7.12E-01 1.99E+00 3.43E+00 1.00E+01 1.95E+00 1.55E+00 U
-1.30E+00 1.99E+00 3.38E+00 1.00E+01 1.95E+00 1.48E+00 U
8.76E+00 1.01E+01 1.56E+61 2.OOE+01 9.93E+00 7.41E+00 U
4.93E+00 4.32E+00 8.35E+00 1.00E+01 4.23E+00 3.79E+00 U
-8.22E-01 2.58E+00 4.49E-00 1.00E+01 2.53E+00 1.97E+00 U
-2.74E+00 3.83E+00 6.35E+00 1.00E+01 3.75E+00 2.88E+00 U
-5.10E-01 1.97E+00 3.44E+00 1.00E+01 1.93E+00 1.59E+00 U
-9.82E-01 2.66E+00 4.56E+00 1.00E+01 2.60E+00 2.1OE+00 U
6.36E+00 1.88E+01 3.48E+01 i.00E+01 1.85E+01 1.62E+01 U
9.81E-0I 2.08E+00 3.88E+00 1.00E+01 2.04E+00 1.80E+00 U
-2.40E+00 3.03E+00 4.87E+00 1.00E+01 2.97E+00 2.28E+00 pCi/L pCi/L pCi/L p~ifll pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L p~i/L p~i/L pCi/ll pCi/ll pCi/L pCi/L PCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U Target analyte was analyzed for but not detected above the MDL or LOD.
UI Uncertain identification for gamma spectroscopy.
X Lab-specific clualifier-please see case narrative, data summary package or contact your project manager for details.
Page 36 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com 10 CFR Part 50/61 Certificate ofAnalysis GEL Sample ID:
Client Sample ID:
Matrix:
Amount of Sample Received:
151732004
- 4 MW4A Ground Water Client: Rancho Seco Nuclear Generating Station Collect Date: December 06, 2005 Receive Date: December Report Date: December 12, 2005 31, 2005 Analyte Sb-124 Sb-125 Cs-134 Cs-136 Cs-137 Ba-133 Ba-140 Ce-139 Ce-i141 Ce-I144 Nd-147 Pm-144 Pm-146 Eu-152 Eu-154 Eu-155 Ir-192 Hg-203 TI-208 Pb-210 Pb-212 Pb-214 Bi-212 Bi-214 Ra-228 Ac-228 Th-230 Th-234 U-235 U-238 Np-239 Am-241 2
1 Run Date Qualifier Activity Uncertainty MDA Critical RL TPU Level 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05
..12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 12/20/05 3.25E-01
-2.19E+00
-6.57E-01 1.83E+00
-1.69E+00 3.61E+00 6.49E+00
-3.90E-01 1.86E+00
-6.07E-02 2.08E+01
-5.03E-01
-1.02E+00
-5.91E-01
-6.31E-01 1.34E+00 1.69E+00 2.50E+00 1.73E+00
-3.84E+01 4.07E+00 2.03E+01 8.61E+00 1.61E+01 2.79E+00 2.79E+00 1.61E+01 6.71 E+0 I 1.25E+01 6.71 E+0 1
-1.111E+01 7.88E+00 5.39E+00 6.33E+00 2.23E+00 4.90E+00 2.47E+00 3.14E+00 1.71E+01 2.23E+00 4.70E+00 1.57E+01 3.57E+01 1.87E+00 3.46E+00 7.19E+00 5.8 1E+00 8.18E+00 3.82E+00 2.91E+00 2.26E+00 4.92E+02 6.35E+00 9.05E+00 3.71E+01 7.58E+00 1.43E+01 1.43E+01 7.58E+00 1.78E+02 1.85E+01 1.78E+02 1.56E+01 2.04E+01 1.00E+01 1.05E+01 3.87E+00 9.24E+00 3.48E+00 5.29E+00 2.79E+0 1 3.87E4-00 8.38E+00 2.77E+01 5.98E+01 3.26E+00 4.97E+00 1.23E+01 1.07E+01 1.37E+01 4.36E+00 5.26E+00 4.29E+00 7.41E+02 8.07E+00 9.3 1E+00 2.50E+01 8.04E+00 1.56E+01 1.56E+01 8.04E+00 1.96E+02 2.97E+01 1.96E+02 2.48E+01 2.50E+01 1.00E+01 1.00E+01 1.001E+01 1.00E+01 1.00E+01 1,00E+01 1.00E+01 1.001E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.001E+01 1.00E+01 1.001E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.001E+01 1.00E+01 1.001E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 1.00E+01 5.28E+00 6.20E+00 2.19E+00 4.80E+00 2.42E+00 3.08E+00 1.67E+01 2.18E+00 4.6 1E+00 1.54E+01 3.49E+01 1.83E+00 3.39E+00 7.04E+00 5.70E+00 8.01E+00 3.75E+00 2.85E+00 2.2 1E+00 4.82E+02 6.22E+00 8.87E+00 3.64E+01 7.43E+00 1.40E+01 1.40E+01 7.43E+00 1.74E+02 1.82E+01 1.74E+02 1.53E+01 2.OOE+01 4.43E+00 4.94E+00 1.76E+00 4.16E+00 1.60E+00 2.5 1E+00 1.30E+01 1.86E+00 4.04E+00 1.33E+01 2.79E+01 1.50E+00 2.33E+00 5.8]E+00 4.75E+00 6.555E+00 2.07E+00 2.5 1E+00 2.01E+00 3.56E+02 3.88E+00 4.43E+00
- 1. 14E+01 3.76E+00
,7.20E+00 7.20E+00 3.76E+00 9.44E+0 I 1.44E+01 9.44E+0 I 1.19E+01 1.20E+01 Units pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Note(s):
- 1. Calculated MDAs are a-posteriori values.
- 2. Activity concentration net +/- 2 sigma overall on reference date.
- 3. Results are statistically positive at the 99.9% confidence level (activity is greater than three times the one sigma uncertainty)
U UI X
Target analyte was analyzed for but not detected above the MDL or LOD.
Uncertain identification for gamma spectroscopy.
Lab-specific oualifier-please see case narrative, data summary package or contact your Proiect manager for details.
Page 37 of 58
I.
II QUALITY CONTROL DATA I I JI Page 38 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Client:
Rancho Seco Nuclear Generating Station 14440 Twin Cities Road Revort Date: December 31, 2005 Page 1 of 20 Herald, California
Contact:
Mr. John Newey Workorder:
151732 Parmname NOM Sample Qual QC.
Units RPD%
REC%.
Range Anlst Date Time Rad Alpha Spec Batch 487212
. QC1200994821 Americium-241 Curium-243/244 QC1200994823 Americium-241 Curium-243/244 QC 1200994820 Americium-241 Curium-243/244
. QC1200994822 Americium-241 Curium-243/244 151732001 DUP U
0.202 Uncert:
+/-0.237 TPU:
+/-0.238 U
-0.0474 Uncert:
+/-0.140 TPU:
+/-0.140 U
U LCS
-0.0245
+/-0.114
+/-0.114
-0.0736
+/-0.167
+/-0.167
.26.3'
+/-2.60
+/-4.22 30.3
+/-2.79
+/-4.74 pCi/L N/A pCi/L N/A (0% - 100%) DDRI (0%- 100%)
12/15/05 16:40 MB 27.2 Uncert:
TPU:
33.0 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
54.5 U
Uncert:
TPU:
66.2.
U Uncert:
TPU:
U 0.0263
+/-0.129
+/-0.129 U
0.00
+/-0.111
+/-0.111 pCi/L.
pCi/L pCi/L pCiIL pCiIL pCi/L 92 97 (75%-125%)
12/15/05 16:40 12/15/05 16:40 151732001 MS 0.202
+/-0.237
+/-0.238
-0.0474
+/-0.140
+/-0.140 48.7
+/-4.64
+/-7.51 67.2
+/-5.45
+/-9.80 87 (75%-125%)
12/16/05 16:39 98 Batch 487213 QC1200994825 151732002 DUP Neptunium-237 QC1200994827 LCS Neptunium-237 QC1200994824 MB Neptunium-237 QC1200994826 151732002 MS Neptunium-237 U
0.0802 Uncert:
+140.157 TPU:
+/-0.157 U
-0.0939
+/-0.0752
+/-0.0757 42.7
+/-3.35
+/-5.21 pCi/L N/A (0% - 100%) DDRI 12/15/05 08:05 12/15/05 08:04 37.3 Uncert:
TPU:
Uncert:
TPU:
pCi/L pCi/L 115 (75%-125%)
U
-0.0305
+/-0.0423
+/-0.0424 12/15/05 08:05 74.5 U
0.0802 86.9 pCi/L 117 (75%-125%)
12/15/05 08:04 Page 39 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 2 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Alpha Spec Batch 487213 Batch 487214 QC1200994829 151732001 DUP Plutonium-238 Plutonium-239/240 Plutonium-244 QC1200994831 Plutonium-238 Plutonium-239/240 Plutonium-244 QC1200994828 Plutonium-238' LCS MB Uncert:
+/-0.157 TPU:
+/-0.157 U
0.00 Uncert:
+/-0.162 TPU:
+/-0.162 U
-0.0596 Uncert:
+/-0.0674 TPU:
+/-0.0677 U
0.165 Uncert:
+/-0.229 TPU:
+/-0.230 Uncert:
TPU:
25.2 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
U 0.00 Uncert:
+/-0.162 TPU:
+/-0.162 50.3 U
-0.0596 Uncert:
+/-0.0674 TPU:
+/-0.0677 U
.0.165 Uncert:
+/-0.229 TPU:
+/-0.230 U
-2.02 Uncert:
+/-5.63 TPU:
+/-5.64 U
U U
0.0724
+/-0.298
+/-0.298
.- 0.102
+/-0.156
+/-0.157.
. 0.0798,
+/-0.180
+/-0.180 pCi/L 0
pCi/L N/A pCi/L 70
+/-6.66
+/-10.5 (0% - 100%) DDRI 12/15/05 08:04 (0%- 100%)
(0%- 100%)
U 0.418 pCi/L (75%-125%)
12/15/05 08:04
+/-0.410
+/-0.413 23.5
+/-2.90
+/-3.91 0.370
+/-0.362
+/-0.365 U
-0.121
+/-0.0898
+/-0.0907 U
0.00289
+/-0.157
+/-0.157 U
0.0722
+/-0.141
+/-0.142 Plutonium-239/240 Plutonium-244 QC1200994830 151732001 MS Plutonium-238 pCi/L pCifL pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L (75%-I 25%)
12/15/05 08:04 93 (75%-125%)
Plutonium-239/240 U
U Plutonium-244
-0.152
+/-0.550
+/-0.550 51.2
+/-5.27
+/-7.39 0.636
+/-0.623
+/-0.627
-1.91
+/-6.31
+/-6.31 (75%-125%)
102 (75%-125%)
12/15/05 08:04 (75%-125%)
Batch 487216 QC1200994834
.151732003 DUP Plutonium-241 U
pCi/L N/A (0% - 100%) DDRI 12/16/05 18:46 Page 40 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 3 of 20 Parmname.
NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Alpha Spec Batch 487216 QC1200994836 LCS Plutonium-241 QC1200994833 MB Plutonium-241 QC1200994835 151732003 MS Plutonium-241 181 Uncert:
TPU:
Uncert:
TPU:
725
- U Uncert:
TPU:
175
+/-17.1
+/-24.3 U
-1.32
+/-6.13
+/-6.13 pCi/L pCifL pCiIL 97 (75%-125%)
12/16/05 20:05 12/16/05 17:43 12/16/05 19:48
-2.02
+/-5.63
+/-5.64 641
+/-64.8
+/-89.1 88 (75%-125%)
Batch 490003 QC1201001148 151732002 DUP Plutonium-241 QC1201001150 LCS Plutonium-241 QC1201001147 MB Plutonium-241 QC1201001149 151732002 MS Plutonium-241 Rad Gamma Spec Batch 488129 QC1200996968 151732601 DUP Actinium-228 U
21.3 Uncert:
+/-34.3 TPU:
+/-34.4 U
55.8
+/-38.2
+/-38.6 pCi/L 0*
(0% - 100%) DDR1 1810 Uncert:
TPU:
Uncert:
TPU:
1810 U
Uncert:
TPU:
1710
+/-56.9
+/-174 U
13.9
+/-36.9
+/-36.9 1690
+/-82.9
+/-214
,pCi/L pCi/L p~i/L 94 (75%-125%)
93 (75%-125%)
12/28/05 21:59 12/29/05 02:07 12/28/05 19:54
.12/29/05 00:03 21.3
+/-34.3
+/-34.4 Americium-241 Antimony-124 Antimony-125 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
U
-0.373
+/-6.17
+/-6.05 U
-5.11
+/-11.6
+/-11.3 U
-1.85
+/-3.98
+/-3.90 U
2.44
+/-5.31
+/-5.20 u
0.506
+/-2.60
+/-2.55 U
10.7 U
3.36
+/-9.43
+/-9.24 U
-0.209
+/-16.3
+/-16.0 U
-1.88
+/-3.01
"+/-2.95 U
0.160
+/-3.75
+/-3.67 U
1.30
+/-2.09
+/-2.05 U
-4.75 pCi/L N/A pCi/L N/A pCi/L N/A pCi/L 127 pCi/L 162 pCi/L N/A (0%- 100%) MJHI 12/20/05 18:26 (0%- 100%)
(0%- 100%)
(0% - 100%)
(0%- 100%)
(0%- 100%)
Barium-133 Barium-140 Page 41 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 4 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Tim e Rad Gamma Spec Batch 488129 Beryllium-7 Bismuth-212 Bismuth-214 Cerium-139 Cerium-141 Cerium-144 Cesium-134 Cesium-136 Cesium-137 Chromium-51 Cobalt-56 Cobalt-57 Cobalt-58 Cobalt-60 Europium-152 Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U
+/-20.1
+/-19.7
-0.628
+/-18.8
+/-18.5 5:49
+/-13.2
+/-12.9 4.62
+/-8.74
+/-8.56
-0.491
+/-1.86
+/-1.82 2.85
+/-6.81
+/-6.67
-6.92
+/-12.5
-+/-12.2
.0.624
+/-1.82
+/-1.78 5.23
+/-6.86
+/-6.72 0.308
+/-2.38
+/-2.33
-2.41
+/-25.0
+/-24.5 0.213
+/-1.88
+/-1.84 0.797
+/-1.66
+/-1.63
-1.79
+/-1.94
+/-1.90
-0.102
+/-1.84
+/-1.80 4.63 U
U U
U U
U
- U U
U U
U U
U U
U
+/-10.2
+/-10.0
-9.6
+/-13.1
+/-12.8 3.50
+/-14.4
+/-14.1
.3.05
+/-5.18
+/-5.08
-1.03
+/-1.46
+/-1.43
-0.205
+/-3.17,
+/-3.11 0.212
+/-10.5
+/-10.3 0.878
+/-1.43
+/-1.40
-1.1
+/-4.22
+/-4.13
-0.602
+/-1.22
+/-1.19
-15.9
+/-16.0
+/-15.7 0.750
+/-1.36
,+/-1.33 0.718
+/-1.34
+/-1.31
-0.526
+/-1.34
+/-1.31 0.322
+/-1.51
+/-1.48
-2.21
+/-4.08 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L N/A 82 147 N/A N/A N/A 35 N/A N/A N/A 232 846 N/A N/A N/A (0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0% - 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0% - 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
Uncert:.
+/-6.19 TPU:
+/-6.06
+/-4.00 Europium-154 U
-3.39 U
-1.44 pCi/L N/A Page 42 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel~com QC Summary Workorder:
151732 Page 5 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Gamma Spec Batch 488129 Uncert:
+/-5.16
+/-4.05 TPU:
+/-5.05
+/-3.97 Europium-155 U
0.139 U
2.75 pCi/L 1660 (0%- 100%)
Uncert:
+/-6.87
+/-5.69 TPU:
+/-6.73
+/-5.58 Iridium-192 U
-1.13 U
0.963 pCi/L N/A (0%- 100%)
Uncert:
+/-2.03
+/-1.49 TPU:
+/-1.99
+/-1.46 Iron-59 u
1.72 U
0.443 pCi/L 108 (0%- 100%)
Uncert:
+/14.28
+/-3.07.
TPU:
+/-4.20
+/-3.01 Lead-210 U
-19.8
.U 315 pCi/L N/A Uncert:
+/-387
+/-416 TPU:
+/-379
+/-407 Lead-212 U
1.35 U
2.98 pCi/L 71 (0%- 100%)
Uncert:
+/-5.89
+/-4.73 TPU:
+/-5.77
+/-4.63 Lead-214 U
4.05 U
1.79 pCi/L 128 (0%- 100%)
Uncert:
+/-6.54
+1-5.30 TPU:
+/-6.41
+/-5.20 Manganese-54 U
-1.08 U
0.740 pCi/L N/A (0%- 100%)
Uncert:
+/-1.57
+/-1.31 TPU:
+/-1.53
+/-1.28 Mercury-203 U
-0.283 U
2.19 pCi/L N/A (0% - 100%)
Uncert:
.+/-2.55
+/-1.79 TPU:
+/-2.50
+/-1.76 Neodymium-147 U
16.3 U
0.741 pCi/L 169*
(0%- 100%)
Uncert:
+/-46.2
+/-22.5 TPU:
+/-45.3
+/-22.1 Neptunium-239 U
0.211 U
4.99 pCi/L 153 (0%- 100%)
Uncert:
+/-12.7
+/-9.79 TPU:
+/-12.4
+/-9.60 Niobium-94 I U 0.619 U
-1.18 pCi/L N/A (0%- 100%)
Uncert:
+/-1.67
+/-1.42 TPU:
+/-1.64
+/-1.39 Niobium-95
. U 0.780 U
0.919 pCi/L 533 (0%- 100%)
Uncert:
+/-2.43
+/-1.81 TPU:
+/-2.38
+/-1.77 Potassium-40 32.7 43.5 pCi/L 182*
(0% - 100%)
Uncert:
+/-27.5
+/-20.2 TPU:
+/-27.0
+/-19.8 Promethium-144 U
-0.00511 U
2.23 pCi/L N/A (0%-,100%)
Uncert:
+/-1.70
+/-1.95 TPU:
+/-1.67
+/-1.91 Promethium-146 U
2.65 U
0.0729 pCi/L 262 (0%- 100%)
Uncert:
+/-2.38
+/-1.72 TPU:
+/-2.33
+/-1.69 Radium-228 U
-0.373 U
3.36 pCi/L N/A (0% - 100%)
Page 43 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com
- QC Summary Workorder:
151732 Page 6 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Gamma Spec Batch 488129 Ruthenium-106 Silver-I 10m Sodium-22 Thallium-208 Thorium-230 Thorium-234 Tin-1 13 Uranium-235 Uranium-238 Yttrium-88 Zinc-65 Zirconium-95 QC1200996970 LCS Actinium-228 Americium-241 Antimony-124 Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
. U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
U Uncert:
TPU:
Uncert:
TPU:
1220 Uncert:
TPU:
Uncert:
TPU:
+/-6.17
+/-6.05 11.5
+/-16.4
+/-16.0
-0.234
+/-1.61
+/-1.58
-1.2
+/-1.86
+/-1.82 0.945
+/-2.91
+/-2.85 4.62
+/-8.73
+/-8.56 166
+/-95.0
+/-93.1
-0.927
+/-2.56
+/-2.51 8.78
+/-18.4
+/-18.1 166
+/-95.0
+/-93.1
-0.374
+/-1.65
+/-1.61
-0.723
+/-3.71
+/-3.64
-0.236
+/-3.75
+/-3.67
+/-9.43
+/-9.24 U
-8.42
+/-11.4
+/-11.2 U
0.379
+/-1.09
+/-1.06 U
-0.511
+/-.1.45
+/-1.42 U
2.00 1-4-1.44
+/-1.41 U
3.05
+/-5.18
+/-5.08 UUI 0.00
+/-107
+/-105 U
-0.698
+/-1.86
+/-1.82 U
-0.125
+/-10.8
+/-10.6 UUI 0.00
+/-107
+/-105 U
0.147
+/-1.51
+/-1.48 U
-6.22
+/-3.75
+/-3.68 U
0.555
+/-2.61
+/-2.55 U
15.7
+/-47.7
+-/-46.7 1270
+/-111
+/-109 U
1.21
+/-14.7
+/-14.4 pCi/L pCi/L pCi/L pCifL pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L N/A N/A N/A 110 147 210*
N/A N/A 210*
N/A N/A N/A (0%
- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0%- 100%)
(0% 100%)
12/21/05 07:25 104 (75%-125%)
Page 44 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com Workorder:
151732 Parmname Rad Gamma Spec Batch 488129 Antimony-125 Barium-133 Barium-140 Beryllium-7 Bismuth-212 Bismuth-214 Cerium-139 Cerium-141 Cerium-144 Cesium-134 Cesium-136 Cesium-137 Chromium-51 Cobalt-56 Cobalt-57 Cobalt-58 QC Summary Sample Qual QC NOM Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
467 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
U 4.35
+/-27.1
+/-26.6 U
-3.36
+/-11.4
+/-11.2 U
-1.36
+/-47.7
+/-46.7 U
-87.7
+/-90.7
+/-88.9 U.
-18.1
+/-86.5
+/-84.7 U
0.727
+/-19.9
+/-19.5 164
+/-21.1
+/-20.6 U
9.14
+/-14.0
+/-13.7 U
-41.9
+/-57.0
+/-55.9 U
11.8
+/-12.4
+/-12.2 U
-0.399
+/-23.0
+/-22.6 486
+/-47.6
+/-46.7 U
-11.1
+/-76.7
+/-75.1 U
3.63
+/-11.8
+/-11.6 215
+/-22.7
+/-22.3 U
-0.964
+/-12.8
+/-12.6 Page 7 of 20 Units RPD%
REC%
Range AnIst Date Time pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L 104 (75%-125%)-
pCi/L pCi/L pCi/L pCi/L Page 45 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Parmname Rad Gamma Spec Batch 488129 Cobalt-60 Europium-152 Europium-154 Europium-155 Iridium-192 Iron-59 Lead-210 Lead-212 Lead-214 Manganese-54 Mercury-203 Neodymium-147 Neptunium-239 Niobium-94 Niobiumn-95 Potassium-40 Page 8 of 20 NOM Sample Qual QC Units RPD%
REC%
Range Anlst Date Tim4e 691 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
U U
U U
U U
U
- U U
U U
U U
U 708
+/-66.2
+/-64.9 7.28
+/-26.5
+/-26.0
-8.05
+/-25.6
+/-25.1
-7.14
+/-33.7
+/-33.0 1.00
+/-8.22
+/-8.06 9.36
+/-29.9
+/-29.3
-23.8
+/-244
+/-239 8.65
+/-22.2
+/-21.8 0.698
+/-18.6
+/-18.2
-7.99
+/-11.7
+/-11.5 44.9
+/-13.1
+/-12.9
-132
+/-103
+/-100
-3.93
+/-56.4
+/-55.3 4.84
+/-10.0
+/-9.82
-0.316
+/-11.8
+/-11.6
-11.9
+/-81.9
+/-80.2 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L 103 (75%-125%)
Page 46 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com Workorder:
151732 Parmname Rad Gamma Spec Batch 488129 Promethium-144 Promethium-146 Radium-228 Ruthenium-106 Silver-1 ICin Sodium-22 Thallium-208 Thorium-230 Thorium-234 Tin-I 13 Uranium-235 Uranium-238 Yttrium-88 Zinc-65 Zirconium-95 QC1200996967 MB Actinium-228 QC Summary Sample Qual 0C Page 9 of 20 Units RPD%
REC%
Ranae Anist Date Time NOM Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
U 2.27
+/-9.76
+/-9.57 U
-3.19
+/-13.2
+/-12.9 U
15.7
+/-47.7
+/-46.7 U
123
+/-98.8
+/-96.9 U
-5.37
+/-11.7
+/-11.4 U
-2.77
+/-9.14
+/-8.96 U
12.84
+/-12.1
+/-11.9 U
0.727
+/-19.9
+/-19.5 U
185
+/-203
+/-199 162
+/-32.4
+/-31.7 U
-31
+/-56.8
+/-55*6 U
185
+/-203
+/-199 325
+/-42.8
+/-41.9 U
15.7
+/-28.3
+/-27.7 U.
-4.02
+/-20.1
+/-19.7 U
6.14
+/-8.67 pCi/L pCi/L pCiIL pCiIL pCiIL pCi/L pCi/L pCiIL pCi/ll pCiIL pCifL pCiIL pCiIL pCifL pCiIL pCi/L 12/29/05 13:19 Page 47 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Parmname Rad Gamma Spec Batch 488129 Americium-241 Antimony-124 Antimony-125 Barium-133 Barium-140 Beryllium-7 Bismuth-212 Bismuth-214 Cerium-139 Cerium-141 Cerium-144 Cesium-134 Cesium-136 Cesium-137 Chromium-51 Cobalt-56.
Page 10 of 20 NOM Sample Qual QC Units RPD%
REC%
Range AnIst Date Timie TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
U U
U U
U U
U U
U U
UUI U
U U
U U
+/-8.50 1.83
+/-16.2
+/-15.9
-1.3
+/-4.40
+/-4.31
-2.94
+/-5.70
+/-5.58
-1.96
+/-2.76
+/-2.71 2.08
+1-15.1
+/-14.8 28.4
+/-20.2
+/-19.8 10.8
+/-16.9
+/-16.6 6.83
+/-5.32
- +/-5.21
-0.279
+/-1.75
+/-1.71 0.947
+/-4.22
+/-4.13 0.00
+/-15.3
+/-15.0 0.191
+/-2.08
+/-2.04
-2.64
+/-5.56
+/-5.44 1.64
+/-2.11
+/-2.07 10.7
+/-21.8
+/-21.4 2.84
+/-3.25 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Page 48 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com Workorder:
151732 Parmname Rad Gamma Spec Batch 488129 Cobalt-57 Cobalt-58 Cobalt-60 Europium-152 Europium-154 Europium-I 55 Iridium-192 Iron-59 Lead-210 Lead-212 Lead-214 Manganese-54 Mercury-203 Neodymium-147 Neptunium-239 Niobium-94 QC Summary SamDle Oual OC Page 11 of 20 Units RPD%
REC%
Ranae Anist Date Time NOM TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
+/-3.19 0.813
+/-2.05
+/-2.00 0.302
+/-2.03
+/-1.99 2.73
+/-2.37
+/-2.33 6.90
+/-5.83
+/-5.71
-0.415
+/-5.30
+/-5.19 1.88
+/-6.38
+/-6.25
-0.679
+/-2.16
+/-2.12 0.523
+/-4.37
+/-4.28 724
+/-596
+/-584
-0.886
+/-3.80
+/-3.73 5.83
+/-4.60
+/-4.51
-0.298
+/-2.14
+/-2.10 0.0663
+/-2.42
+/-2.37
-4.28
+/-34.0
+/-33.3
-5.05
+/-11.8
+/-11.5
-1.8
+/-2.17 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L Page 49 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 12 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range AnIst Date Time Rad Gamma Spec Batch 488129 TPU:
Niobium-95 Potassium-40 Promethium-144 Promethium-146 Radium-228 Ruthenium-106 Silver-110m Sodium-22 Thallium-208 Thorium-230 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
+/-2.13 U
2.71
+/-2.42
+/-2.37 UUI 0.00
+/-36.6
+/-35.8 U
-2.14
+/-2.22
+/-2.18 UUI 0.00
+/-2.79
+/-2.74 U
6.14
+/-8.67.
+/-8.50 U
-0.281
+/-20.1
+/-19.7 U
-0.852
+/-2.09
+/-2.05 U
-0.157
+/-1.90
+/-1.86 U
2.60
+/-2.39
+/-2.34 U
6.83
+/-5.32
+/-5.21 U
33.3
+/-143
+/-140 U
-1.62
+/-2.27
+/-2.23 U
3.20
+/-14.2
+/-14.0 U
33.3
+/-143
+/-140 U
1.84
+/-2.23
+/-2.19 U
2.91
+/-4.39 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCiIL pCi/L
.pCi/L pCi/L pCiIL pCi/L pCi/L pCi/L Thorium-234 Tin-] 13 Uranium-235 Uranium-238 Yttrium-88 Zinc-65 Page 50 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road' Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 13 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Gamma Spec Batch 488129 TPU:
Zirconium-95 QC1 200996969 Actinium-228 Americium-241 Antimony-124 Antimony-125 Uncert:
TPU:
151732001 MS Uncert:
. TPU 9760 Uncert:
TPU Uncert:
TPU Uncert:
TPU Barium-133 Barium-140.
Beryllium-7 Bismuth-212 Bismuth-214 Cerium-139 Cerium-141 Cerium-144 Cesium-134 Cesium-136 Cesium-137 Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU U
-0.373
+/-6.17
+/-6.05 U
-5.11
+/-11.6
+/-11.3 U
-1.85
+/-3.98
+/-3.90 U
2.44
+/-5.31
+/-5.20 U
0.506
+/-2.60
+/-2.55 U
1 10.7
+/-20.1
+/-19.7 u
-0.628
+/-18.8
+/-18.5 U
5.49
+/-13.2
+/-12.9 U
4.62
+/-8.74
+/-8.56 U
-0.491
+/-1.86
+/-1.82 U
2.85
+/-6.81
+/-6.67 U
-6.92
+/-12.5
+/-12.2 U
0.624
+/-1.82
+/-1.78 U
5.23
+/-6.86
+/-6.72 U
0.308
+/-4.31 U
-1.8
+/-3.88
+/-3.80 U
-148
+/-599
+/-590 10800
+/-2250
+/-4770 U
44.7
+/-279
+/-274 U
110
+/-297
+/-295 U
-45.2
+/-155
+/-153 U
-146
+/-1330
+/-1300 U
373
+/-1380
+/-1360 U
619
+/-1030
+/-1040 U
327
+/-251
+/-278 1540
+/-266
+/-656 U
25.1
+/-218
+/-214 U
, 603
+/-751
+/-773 U
29.0
+/-162
+/-160 U
355
.+/-647
+/-649 3850 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L 12/28/05 10:41 III Uncert:
TPU
'3740 103 Page51 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Parmname Page 14 of 20 NOM Samnle flual OC llnit*
RPDf%
~WC 0 A Rano~
Anid flat Tump Rad Gamma Spec Batch 488129 Chromium-51 Cobalt-56 Cobalt-57 Cobalt-58 Cobalt-60 Europium-152 Europium-154 Europium-155 Iridium-192 Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU:
Uncert:
TPU 5550 Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU
+/-2.38
+/-2.33 U
-2.41
+/-25.0
+/-24.5 U
0.213
+/-1.88
+/-1.84 U
0.797
+/-1.66
+/-1.63 U
-1.79
+/-1.94
+/-1.90 U
-0.102
+/-1.84
+/-1.80 U
4.63
+/-6.19
+/-6.06 U
-3.39
+/-5.16
+/-5.05 U
0.139
+/-6.87
+/-6.73 U
-1.13
+/-2.03
+/-1.99 U
1.72
+/-4.28
.+/-4.20 U
-19.8
+/-387
+/-379 U
1.35
+/-5.89
+/-5.77 U
4.05
+/-6.54
+/-6.41 U
-1.08
+/-1.57
+/-1.53 U
-0.283
+/-2.55
+/-2.50 U
16.3
+/-519
+/-1590 U
377
+/-1500
+/-1480 U
64.1
+/-165
+/-164 1460
+/-255
+/-624 U
118
+/-166
+/-169 5480
+/-729
+/-2260 U
-134
+/-313
+/-312 U
-44.1
+/-296
+/-291 U
81.8
+/-390
+/-384 U
69.4
+/-128
+/-128 U
-21
+/-415
+/-407 U
5040
+/-39000
+/-38300 U
63.6
+/-196
+/-194 U
107
+/-255
+/-253 U
-44
+/-118
+/-117 U
194
+/-260
+/-266 U
-589 Iron-59 pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L pCi/L 99 Lead&210 Lead-212 Lead-214 Manganese-54 Mercury-203 Neodymium-147 Uncert:
TPU Page 52 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 15 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range AnIst Date Time Rad Gamma Spec Batch 488129 Neptunium-239 Niobium-94 Niobium-95 Potassium-40 Promethium-144 Promethium-146 Radium-228 Ruthenium-106 Silver-I Om.
Sodium-22 Thallium-208 Thorium-230 Thoriurni-234 Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU Uncert:
TPU
+/-46.2
+/-45.3 U
0.211
+/-12.7
+/-12.4 U
0.619
+/-1.67
+/-1.64 U
0.780
+/-2.43
+/-2.38 32.7
+/-27.5
+/-27.0 U
-0.00511
+/-1.70
+/-1.67 U
2.65
+/-2.38
+/-2.33 U
-0.373
+/-6.17
+/-6.05 U
11.5
+/-16.4
+/-16.0 U
-0.234
+/-1.61
+/-1.58 U
-1.2
+/-1.86
+/-1.82 U
0.945
+/-2.91
+/-2.85 U
4.62
+/-8.73
+/-8.56 U
166
+/-95.0
+/-93.1 U
-0.927
+/-2.56
+/-2.51 U
8.78
+/-18.4
+/-18.1 U
166
+/-3090
+/-3030 U
-143
+/-707
+/-695 U
-2.06
+/-120
,+/-117 U
152
+/-201
+/-206 U
1350
+/-1210
+/-1300 U
-5.0
+/-109.
+/-107 U
-60.5
+/-174
+/-172 U
-148
+/-599
+/-590 U
517
+/-1270
+/-1260 U
-10
+/-139
+/-136 U
-16.4
+/-106
+/-104 U
79.9
+/-142
+/-143 U
327
+/-251
+/-278 U
3030
+/-7400
+/-7350 1700
+/-324
+/-740 U
49.0
+/-639
+/-627 U
3030 pCi/L pCi/L pCi/L pCi/L pCi/L pCiIL pCiIL pCiIL pCi/L pCiJL pCiIL pCiIL pCiIL pCi/L pCiIL pCiIL Tin-I 13 Uranium-235 Uranium-238 Page 53 'of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 -www.gel.com QC Summary Workorder:
151732 Page 16 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range AnIst Date Time Rad Gamma Spec Batch 488129 Yttrium-88 Zinc-65 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
+/-95.0
+/-93.1 U
-0.374
+/-1.65
+/-1.61 U
-0.723
+/-3.71
+/-3.64 U
-0.236
+/-3.75
+/-3.67 U
U
+/-7400
+/-7350
- 2470
+/-537
+/-1100
-19.9
+/-299
+/-294
-39.5
+/-297
+/-291 pCi/L pCi/L
,pCi/L Zirconium-95 Batch 490999 QC1201003356 151732001 DUP Nickel-59 QC1201003358 LCS Nickel-59 QC1201003355
- MB Nickel-59 QC1201003357 151732001 MS Nickel-59 SU 15.1 Uncert:
+/-10.4 TPU:
+/-10.2 U
.0.141
+/-11.6
+/-11.4 pCi/L 196 (0% - 100%) BXFI 380 Uncert:
TPU:
Uncert:
TPU:
380 U
Uncert:
TPU:
358
+/-45.5
+/-52.7 U
7.92
+/-10.2.
+/-10.1 pCi/L pCi/L pCi/L 94 (75%-125%)
12/30/05 16:51 12/29/05 17:59 12/30/05 14:49.
12/29/05 19:06 15.1
+/-10.4
+/-10.2 355
+/-50.4
+/-56.7 93 (75%-125%)
Rad Gas Flow Batch 487482 QC1200995397 151732003 DUP Total Strontium QC1200995399 LCS Total Strontium U
-0.279 Uncert:
+/-0.381 TPU:
+/-0.381 51.8 Uncert:
TPU:
U
-0.434
+/-0.357
+/-0.357 pCi/L N/A (0% - 100%) 31XW I QC1200995396 MB Total Strontium QC1200995398 151732003 MS Total Strontium Uncert:
TPU:
104 U
Uncert:
TPU:
46.6
+/-3.22
+/-3.51 U
0.212
+/-0.428
+/-0.428 92.0
+/-6.57
+/-7.12 pCi/L pCi/L pCi/L 90 (75%-125%)
12/14/05 17:28 12/14/05 17:33 12/14/05 17:28 12/14/05 17:28
-0.279
+/-0.381
+/-0.381 89 (75%-125%)
Batch 489159 QC1200999366 151735002 DUP Page 54 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 17 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anlst Date Time Batch Alpha 489159 Beta QC1200999369 LCS Alpha Beta QC1200999365 MB Alpha Beta QC1200999367 151735002 MS Alpha Beta QC1200999368.151735002 MSD Alpha Beta Rad Liquid Scintillation Batch 487442 QC1200995296 151732004 DUP Promethium-147 Uncert:
TPU Uncert:
TPU U
0.831
+/-1.27
+/-1.27 U
. -1.94
+/-2.06
+/-2.07 U
1.01
+/-1.36
+/-1.37 U
1.07
+/-2.11
+/-2.11 pCi/L
.19 pCi/L N/A (0%
-100%) CXo I (0%- 100%)
12/23/05 15:37 108" Uncert:
TPU:
311 Uncert:
TPU:
Uncert:
TPU:
Uncert:
TPU:
108 U
Uncert:
TPU:
311 U
Uncert:
TPU:
108 U
Uncert:
TPU:
311 U
Uncert:
TPU:
109
+/-10.7
+/-13.4 311
+/-14.8
+/-19.3 U
0.421
+/-1.16
+/-1.17 U
-1.54
+/-1.86
+/-1.86 pCiIl pCi/ll pCiIL.
pCiIL pCi/L pCiIL 101 (75%-125%)
100 (75%-125%)
12/23/05 15:48 12/23/05 15:48 0.831
+/-1.27
+/-1.27
-1.94
+/-2.06
+/-2.07 0.831
+/-1.27
+/-1.27
. -1.94
+/-2.06
+/-2.07 89.6
+/-12.1
+/-13.7 328
+/-15.6
+/-18.6 96.5
+/-12.5
+/-13.7 364
+/-16.8
+/-22.9 83 (75%-125%)
106 (75%-125%)
12/23/05 15:48 pCi/L 7
89 (0%-20%)
pCi/L 11 117 (0%-20%)
U 3.20 Uncert:
+/-4.65 TPU:
+/-4.65 QC1200995298 Promethium-147 LCS QC1200995295 MB Promethium-147 QC1200995297 151732004 MS 138 Uncert:
TPU:
Uncert:
TPU:
U 4.90
+/-4.84
+/-4.84
-137
+/-12.3
+/-13.5' U
4.32
+/-4.52
+/-4.52 pCi/L 0
pCi/L pCi/L 100 (75%-125%)
(0% - 100%)
AFI 12/22/05 04:23 12/21/05 05:25 12/22/05 02:51 Page 55 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 18 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Liquid Scintillation Batch 487442 Promethium-147 278 U
Uncert:
TPU:
3.20
+/-4.65
+/-4.65 Batch 487444 QC1200995300 151732001 DUP Iron-55 U
-28.2 Uncert:
+/-37.1 TPU:
+/-37.1 U
229
+/-22.1
+/-24.0
-1.51
+/-28.2
+/-28.2 2030
+/-179
+/-251 pCi/L 82 (75%-125%)
pCi/L N/A (0%- 100%) BXFI
-12/16/05 06:35 QC1200995302 lron-55 LCS 2000 Uncert:
TPU:
QC1200995299 MB Iron-55 QC1200995301 151732001 lron-55 Uncert:
TPU:
U
-16.4
+/-36.0
+/-36.0 pCi/L pCifL pCi/L 101 (75%-125%)
12/16/05 08:55 12/16/05 04:29 12/16/05 08:38 M.S Batch 487445
. QC1200995304 151732001 DUP Nickel-63 QC1200995306 LCS Nickel-63 QC1200995303 MB Nickel-63 QC1200995305 151732001 MS Nickel-63*
2010 U
Uncert:
TPU:
U Uncert:
TPU:
1570 Uncert:
TPU:
Uncert:
-TPU:
1570 U
Uncert:
TPU:.
-28.2
+/-37.1
+/-37.1 1940
+/-129
+/-210 8.46
+/-13.9
+/-13.9 97 (75%-125%)
1.76 t/-14.6
+/-14.6 U
pCi/L 0
(0% - 100%) BXFI 1370
+/-39.0
+/-51.0 U
-3.38
+/-13.4
+/-13.4 pCi/L pCi/L pCi/L 87 (75%-125%)
12/15/05 05:00 12/15/05 06:34 12/15/05 04:13 12/15/05 05:47 1.76
+/-14.6
+/-14.6 Batch 487449 QC1200995312 151132001 DUP Technetium-99 QC1200995314 LCS Technetium-99 QC1200995311 MB Technetium-99 U
-5.14 Uncert:
+/-13.4 TPU:
+/-13.4 U
1390
+/-38.2
+/-50.7 0.989
+/-10.4
+/-10.4 1270
+/-50.6
+/-59.6 89 (75%-125%)
pCi/L N/A (0% - 100%) SLN1 1180 Uncert:
TPU:
pCi/L 107 (75%-125%)
12/19/05 23:54 12/20/05 01:13 12/19/05 22:51 U
5.13 pCi/L Page 56 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 19 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time Rad Liquid Scintillation Batch 487449 Uncert:
TPU:
QC1200995313 151732001 MS Technetium-99
+/-12.2
+/-12.2 989
+/-42.5
+/-49.2 1180 U
Uncert:
TPU:
-5.14
+/-13.4
+/-13.4 pCi/L 84 (75%-125%)
12/20/05 00:56 Batch 487451
ý QC1200995316 151732001 DUP Tritium IU 113 Uncert:
+/-211 TPU:
+/-211 U
129
+/-215
+/-215 pCi/L 0
(0% - 100%) MXP1 QC1200995318 Tritium LCS QC1200995315 MB Tritium QC1200995317 151732001 MS Tritium 2990 Uncert:
TPU:
Uncert:
TPU:
3000 U
Uncert:
TPU:
2900
+/-488
+/-490 U
23.1
+/-212
+/-212 pCi/L pCi/L pCi/L 97 (75%-125%)
107 (75%-125%)
12/15/05 01:35 12/15/05 02:39' 12/15/05 00:48 12/15/05 02:22 113
+/-211
+/-211 3200
+/-504
+/-567 Batch 487453 QC1200995324 151732001 DUP Carbon-14 QC1200995326 LCS Carbon-14 QC1200995323 MB Carbon-14 QC1200995325 151732001 MS Carbon-14 U
9.08 Uncert:
+/-15.7 TPU:
+/-15.7 1080 Uncert:
TPU:
Uncert:
TPU:
U 9.32
+/-15.7
+/-15.7 1090
+/-36.4
+/-40.2 U
8.73
+/-15.7
+/-15.7 1050
+/-35.8
+/-39.4 pCi/L 0
(0% - 100%) SLNI pCi/L pCi/L pCi/L 101 (75%-125%)
12/14/05 22:44 12/15/05 00:19 12/14/05 21:56 12/14/05 23:32 1080 U
Uncert:
TPU:
9.08
+/-15.7
+/-15.7 97 (75%-125%)
Notes:
The Qualifiers in this report are defined as follows:
B Target analyte was detected in the sample as well as the associated blank.
BD Results below the MDC or low tracer recovery.
E Concentration of the target analyte exceeds the instrument calibration range.
Analytical holding time exceeded.
Page 57 of 58
GENERAL ENGINEERING LABORATORIES, LLC 2040 Savage Road Charleston SC 29407 - (843) 556-8171 - www.gel.com QC Summary Workorder:
151732 Page 20 of 20 Parmname NOM Sample Qual QC Units RPD%
REC%
Range Anist Date Time H
J Indicates an estimated value.
U Target analyte was analyzed for but not detected above the MDL or LOD.
UI Uncertain identification for gamma spectroscopy.
X Lab-specific qualifier-please see case narrative, data summary package or contact your project manager for details.
d The 2:1 depletion requirement was not met for this sample h
Sample preparation or preservation holding time exceeded.
N/A indicates that spike recovery limits do not apply when sample concentration exceeds spike conc. by a factor of 4 or more.
- Indicates analyte is a surrogate compound.
^ The Relative Percent Difference (RPD) obtained from the sample duplicate (DUP) is evaluated against the acceptence criteria when the sample is greater than five times (5X) the contract required detection limit (RL). In cases where either the sample or duplicate value is less than 5X the RL, a control limit of +/- the RL is used to evaluate the DUP result.
For PS, PSD, and SDILT results, the values listed are the measured amounts, not final concentrations.
Where the analytical method has been performed under NELAP certification, the analysis has met all ofthe requirements of the NELAC standard unless qualified on the QC Summary..
Page 58 of 58