ML20141L896
| ML20141L896 | |
| Person / Time | |
|---|---|
| Site: | 07002930 |
| Issue date: | 04/18/1983 |
| From: | Steven Baggett NRC OFFICE OF NUCLEAR MATERIAL SAFETY & SAFEGUARDS (NMSS) |
| To: | Buxton T WULTEX U.S., INC. |
| Shared Package | |
| ML20141J561 | List:
|
| References | |
| NUDOCS 8602270506 | |
| Download: ML20141L896 (2) | |
Text
._.
1 hDISTRIBUTION:
N**
FC Central File NMSS rf f" ~ ~ g3 FCMC rf SLBaggett C//Z ' 7Id < 3 T 70 k
Waltex U.S., Inc.
ATTN: Mr. Terry Buxton M i
125 Commonwealth Drive Thornhill Industrial Park Warrendale, PA 15086
Dear Mr. Buxton:
This refers to your request dated March 30, 1983, to registse your inline ash meter.
Our review of your submitted information yielded a need for additional infbruntion/ clarification of the following items:
l o Describe how the plexiglass shutter is.used to successfully block the radiation beam and if the shutter can be locked in plac's o In addition to the information contained in your label, the I
following should also appear on the label (s): name or logo l
of the manufacturer.1sotope and activity, date of assay of 4
the sour.ca activity, 'The' label (s) must be durable enough to l
remain legible for the useful life of the device and be l
readily visible.
o Describe the plane ~i use of the device and identify the environ-l mest and operating condit16ns expected during normal use.
Indicate the expected useful life of the source / device.
If the information provided in the brochure is the maximum climatic conditions that the device will be subjected to<, then provide the expected useful life of the source / device.
o Although the photographs of the device were very well done, they do not lend themselves to be reproduced legibly. Therefore, send us a diagram or photo that can be reproduced legibly.
o Although you provided historical data on the devices in use in l
England, we need a copy of the test results of testing that were performed on at least one device. This testing should correlate with the ANSI N538 standard enti.tled, " Classification of Industrial Ionizing Gauging Devices." Provide also the ANSI M538 classification desi0 nation. This testing should demonstrate that the device will maintain its containment integrity during the specified conditions of use.
l 8602270506 851203 l
REG 2 LIC70 i
omcE>
SURNAME >
DATE >
unc ronu ne no-son nacu o24o OFFICIAL RECORD COPY
o o
Waltex U.S., Inc.
_ APR 18 IS63 o Describe the quality control program of the device manufacturer.
The program should -identify any mechanical or material discrepancy.
that would be allowed between the units and the manufacturing specifications. An example of an acceptable quality control program can be found in Appendix 8, ANSI M538.
o Provida a paragraph which summarizes the important facts pertaining to safety and the results of the safety analysis performed by the ma9ufacturer/ distributor.
Include references to the appropriate ANSI, NBS or NRC standards used in the safety analysis.
We have enclosed a copy of " Guidelines for Applications for Registration of Devices,"
to help you formulate your response.
Once we have completed the device review, we will forward the license amendment application to our Region II licensing section. To keep Mr. Robert Stewart-infoma tion will send copies of our correspondence to him.
Please contact me if you have any questions. My phone number is (301)-427-4240.
Respectfully, Originet S!gred Rf Steven L Dccit Steven L. Baggett Material Certification and Procedures Branch Enciosure: Guidelines cc: Mr. Robert Stewart, The 5, 4, 3, Co.
"'">...E. 9.9 <
~
5 " ^"'> SLBa jll. tf. rad.
^ " >
4/.18/,83,,,
unc ronu ai. < io..o. unc" OFFICIAL RECORD COPY