Results of Semiannual Radiation Surveys,Monitoring Station Dosimeter Readings & Environ Sample Analysis Surveys
ML20042B019
Person / Time
Site: NS Savannah
Issue date: 11/13/1981
From:
SOUTH CAROLINA, STATE OF
To:
Shared Package
ML20042B013 List:
References
NUDOCS 8203240400
Download: ML20042B019 (28)


Text

.

?

s O

RESULTS OF SEMI-ANNUAL RADIATION SURVEYS & MONITORING STATION DOSIMETER READINGS AND RESULTS OF SEMI-ANNUAL ENVIRONMENTAL SAMPLE ANALYSIS SURVEYS 8203240400 820225 DR ADOCK 05000

(

.f SEMI-ANNUAL SURVEY REPORT N. S. Savannah November 12 & 13, 1981 Contents:

TLD Monitoring Badge Locations Direct Radiation Survey Contamination Survey. Results Environmental Sample Analysis Report Conducted By:

South Carolina Department of Health and Environmental Control Bureau of Radiological Health 2600 Bull Street Coluubia, South Carolina 29201 h

[

TLD Badge Locations on the N.S. Savannah TLD No.

Location 1

Pilot House - Ceiling 30 Navigation bridge deck - Rm. NB4 Reactor Operation room on metal rack.

26 Navigation bridge deck - Officer's lounge fan 27 3rd officer room NB1 - on metal rack Navigation bridge deck.

5 Vent on main stairway to Verande and lounge on Promenade deck.

2 Outside on Promenade deck behind the swimming pool.

9 In front of the top of the reactor hatch on white column - Promenade deck.

10 Promenade deck main lounge middle of front viewing window.

11 "A" deck main desk behind counter on Orange wall.

12 "A" deck hospital entrance (near d.P. room) on wire cable above wooden wall rack, 13 "A" deck near hospital entrance sign.

14 "B" deck Cargo Hold #4 on wall near entrance in front of chalk markings on floor.

15 "B" deck inside Fire Station No. 20.

16 "B" deck handle on door leading to reactor space.

17 "B" deck door leading to fan room.

18 "B" deck main dining room - Light fixture near entrance to main galley.

19 "C" deck Fire Station No. 28.

39 "C" deck left side of ceiling in reactor viewing room.

(port side) 40 "C" deck right side of ceiling in reactor viewing room.

(starboard side)

i f'~~

Page Two

TLD locations TLD No.

Location i

33 "C" deck cold Chem lab door.

32 "C" deck door leading to reactor space on overhead pipes.

34 Pipes above stairway to "C" deck from "D" deck.

23 "D" deck Fire Station No. 31.

22-

"14 ft. flat" on right side above stabilizer door.

(starboard side) 25

" Hold" on pipe near alarm right side. (starboard side) 24 "liold" left side near alarm bell. (port side) 36 "14 ft. flat" above left side of stabilizer door.

(port side) 7 (Upper level) engine room throttle.

20 Reactor control room and alarm handle.

21 (Lower level) engine room outside pump charge hatch right side 20ft. ahead of diesel engines. '

(starboard, side) 38 (Lower level) engine room outside door of pump hatch changer left side. (portside)

Total number of TLD badges located on N.S. Savannah:

31 Date of Placement: Noveraber 13, 1981 i-

-n w-

Surveys Aboard N.S. Savannah Instruments:

November 12 &

'3, 1981 1.Ludlum Model 19 Micro R hbte Cal Date 11/81 Background 2uR/hr 2.Eberline hbdel PRS-1

,, Top of House" Deck Sodium Iodide Crystal Cal Date 10/81 Bkg 2uR/hr Location Actual uR/hr less Bkgrd. uR/hr Top of House portside 2.0 0

Top of House starboard 2.0 0

Navigation Bridge Deck Gyro & radar room 3.0 1.0 Chart room 3.0 1.0 Chief radio officer's SR-NB2 3.0 1.0 Hallway front of chief radio Officer SR 2.5 0.5-Reactor operators room 2.0 0

Radio room 3.0 1.0 Sea cabin NB1 2.5 0.5 Radio officer SR NB3 2.5 0.5 Elevator machinery room and spare parts 3.0 1.0 Fan room, portside, emergency general room 3.0 1.0 Deck toilet 3.0 1.0 In pilot house at helm 2.5 0.5 Bridge wing - Portside 2.0 0

Bridge wing - Starboard 2.0 0

Center of bridge 2.0 0

Boat Deck Officer's lounge 3.0 1.0 2nd Asst. Eng. SR BT-15 3.0 1.0 3rd Asst. Eng. Sr BT-12 3.0 1.0 Main stairway near officer lounge 3.5

1. 'a 3rd Asst. Eng. SR BT-10 4.0 2.0 ART room BT-13, 2nd Eng.

3.0 1.0

r^

Surveys cont'd.

Page 2 Boat Deck Location Actual uR/hr Less bkgrd. uR/hr 3rd Officer 2.5 0.5 2nd Officer 2.5 0.5 Captain's quarters 4.0 2.0 3rd officer BT-3 4.0 2.0 Hallway between Cpt. quarter and chief engineers dayroom 4.0 2.0 3rd officers OR BT-1 3.0 1.0

-1st Asst. Eng. SR BT-2 3.0 1.0 Chief Eng. dayroom 3.0 1.0 Sr. 1st Eng. SR BT-6 3.5 1.5 Fan room 3.0 1.0 2nd Asst. Eng. SR BT-8 3.0 1.0 Midale ways hallway 3.0 1.0 Outside deck 3.0 1.0 Officers laundry sink inside 11.0 9.0 Officers laundry sink underneath 11.0 9.0 Promenade Deck Top of reactor hatch 3.0 1.0 Projection booth 3.0 1.0 Fan room next to Verande 3.0 1.0 Fan roor.i portside 3.0 1.0 Floor behind bar & bar storage area 3.0 1.0 Men's room 3.0 1.0 Hallway outside exit off suite in front of water fountain 3.0 1.0 Portside Verande r som 4.0 2.0 Center Verande room 4.0 2.0 starboard side Verande room 4.0 2.0

  1. Naturally Occuring due to lavatory porcelan i

Page 3 "A" Deck location Actual uR/hr Less bk rd. uR/hr 3

Fire Station #14, port passaway 4.7 2.7 Center of main lobby 5.5 3.5 Bow of main lobby 6.0 4.0 Stern of main lobby 4.9 2.9 Fire Station #16, port pass.

3.5 1.5 Fire Station #17, starboard pass.

3.2 1.2 Fice Station #15, starboard pass.

5.0 3.0 Area outside Doctor office and H.P. lab 5.6 3.6 Room A-19 3.9 1.9 Room A-21 2.6 0.6 Room A-23 2.8 0.8 Room A-25 2.3 0.3 Room A-27 2.5 0.5 Room A-29 2.3 0.3 Area between Room 29 & ART-211 2.5 0.5 ART third FNG 2.8 0.8 "A" deck bow 2.1 0.1 Doctor's office 3.7 1.7

  • H.P. lab (floor) 8.0 6.0
  • H.P. lab (right counter top) 9.0 7.0
  • H.P. lab (left counter top) 9.0 7.0
  • H.P. lab (rear sink) 7.0 5.0 Ward A (hospital) 4.2 2.2 Ward B (hospital) 3.9 1.9 Ward C (hospital) 4.2 2.2 Ward 0 (hospital) 4.0 2.0 Operating room (hospital) 5.1 3.1 Barber shop (forward) 3.0 1.0 Barber shop (aft) 3.0 1.0 Room A-1 2.8 0.8 Room A-3 locked unable to survey Room A-5 3.4 1.4 Room A-7 4.2 2.2

p Page 4 "A" Deck cont'd Location Actual uR/hr_

Less bkgrd uR/hr Room A-9 3.5 1.5 Room A-11 3.4 1.4 Room A-13 3.6 1.6 Room A-15 3.4 1.4 Health physics office 3.0 1.0 Doctor's' stateroom 3.2 1.2 Steward's office 3.4 1.4 Room A-19 3.2 1.2 Hall behind pursar's deck' 3.5 1.5 Passenger laundry 3.8 1.8 Room A-2 2.3 0.3 Room A-4 2.8 0.8 Room A-6 2.8 0.8 Room A-8 3.2 1.2 Room A-10 2.8 0.8 Room A-12 3.0 1.0 Room A-14 locked unable to survey Room A-16 3.8 1.8 Conference room 3.6 1.6 Service pantry 4.0 2.0 Purser's office 2.9 0.9 Room A-18 3.9 1.9 Room A-20 3.2 1.2 Room A-22 3.0 1.0 Room A-24 3.0 1.0 Room A-26 2.8 0.8 Room A-28 2.5 0.5 Room A-30 2.5 0.5 Room A-32 2.9 0.9

-- Concrete pier 3ft. above surface 11.4 9, 4

-- Concrete pier contact surface 14.5 12.5 "A" deck stern 1.3 0.3 i

c-1 Page 5 "B" Deck Location _

Actual uR/hr Less bkgrd uR/hr (highest reading in area)

Room B-17 5.44 3.44 Room B-19 2.42 0.42 Room B-21 2.6 0.6 Room B-23 2.43 0.43 Room B-29 2.88 0.88 Passaway Fire Station #21 3.28

' 1.28 Room B-31 2.85 0.85 Room B-33 2.56 0.56 Room B-35

-2.92 0.92 Room B-37 2.43 0.43 Room B-39 2.68 0.68 Room B-43 2.70 0.70 Room B-45 2.11 0.11 Room B-47 2.69 0.69 Room B-51 2.0 0

Room B-53 2.19 0.19 Room B-10 2.02 0.02 Room B-12 1.95 0.95 Room B-14 2.60 0.60 Room B=16 2.56 0.56 Room B-18 2.04 O.04 Training room 1.68 0.68 Emergency room 1.3 0.3 Stern corner starboard side 3.06 1.06 Officer's pantry 3.46 1.46 Cabinet top center of main dining room 3.12 1.12 Room B-49 2.27 0.27 Room B-8 2.22 0.22 Room B-6 2.71 0.71 Officer's mess 2.9 0.9 Crew's lour.je 3.05 1.05 Steward's mess 3.35 1.35

-m

Page 6 "B" deck' cont'd Location Actual uR/hr Less bkgrd uR/hr

  • General cargo Hold hat.ch #4 25-23 i

Outside water door to reactor 3.6 1.6 Room B-3 9.01 7.01 Room B-1~

6.74 4.74 Room B-7 3.48 1.48 Room B-9 4 06 2.06 Room B-11 4.99 2.99 Room B-13 6.26 4.26 Room B-15 5.99 3.99 Fire Station #20, port pass.

2. %
0. %

Fire Station #22, port pass.

2.86 0.86

'Passaway outside steward's mess and lounge, portside 3.21 1.21 Center of conference room aft 1.86 0.86 Inside main galley 3.06 1.06 Paint storage area (stern) 1 0

Port winch room (stern) 1 0

Steering room portside floor 1

0 Fire Station #23-A 2

0 "C Deck Room C-23 3.0 1.0 Room C-21 3.0 1.0 l

Room C-19 2.5 0.5 Room C-17 2.5 0.5 Room C-15 2.0 0

Room C-13.

2.0 0

Co2 room 1.5 0.5 Room C-11 3.0 1.0 Room C-9 3.0 1.0 Room C-7 3.0 1.0 Room C-5 2.5 0.5

Page 7 "C" deck Location Actual uR/hr Less bkgrd uR_fh_r Hallway between Rooms C-11 to C-1 2.0 0

Room C-3 3.0 1.0 Room C-1 3.0 1.0 Hallway front of breathing apparatus & Room C-1 4.0 2.0 2nd cooks SR C-2 4.0 2.0 Butcher & chief pantry SR C-4 3.0 1.0 RoomC-6(outsidedoor) 3.0 1.0 Room C-6 3.5 1.5 Room C-8 (1st & 2nd cook) 3.0 1.0 Room C-10 3.0 1.0 Room C-12 3.0 1.0 Engine room veiwing corridor portside 3.5 1.5 Engine room center corridor portside 3.5 1.5 Mainway to engine room 3.0 1.0 Entrance to engine room 3.0 1.0 Clean linen room 3.0 1.0 Laundry room near presser and exhaust vent 6.0 4.0 Laundry room near washer & folding table 5.0 3.0 Outside door of cold water chem lab 6.0 4.0 Front stairs-to "B"

& "C" deck 60 58 Corridor on "C" & "D" decks 90 88 Outside wall near manual fire damper 50 48 Cold locker 5

3 Along pipes inside storage area 3

1 Gangway 3.0 1.0 Garbage room 3.0 1.0 Butcher shop & thaw room blocked with oil drums and locked Bottom of steps from "B" to "C" deck 110 108 Stairway to "C" deck from "D" deck 9

7 Portside in " hold" 150 148

P Page 8~

"C" Deck Location Actual uR/hr-Less bkgrd uR/hr

  • Center in Hold at bough 100 98
  • Starboard corner in hold 170 168
  • Portside corner in hold 180 178 l

Emergency H.P. lab counter

("C" deck stern) 1 0

Floor emergency H.P. lab

("C" deck stern) 1 0.'

Carpenter shop floor ("C" deck stern)2 0

Passage outside emergency H.P.

lab ("C" deck stern) 1.0 0

Room C-24 2

0 Boom C-22~

2.5 0.5 Room C-20 2.5 0.5 Room C-18 2.5 0.5 Room C-16 3.0 1.0 Room C-14 2.5 0.5 Hallway in front of sink & locker 2.5 0.5 Music entertainment room 1.5 0.5 Hallway outside music enter. room 1.5 0.5 Room C-23 3.0 1.0 Room C-21 3.0 1.0 Room C-19 2.5 0.5 Room C-17 2.5 O.5 Room C-15 2.0 0

Room C-13 2.0 0

"D" Deck Storeroom starboard

~0.23 upper "D" deck 1.23 Storeroom starboard (lower) 2.00 0

Electronics work shop floor 1.1 0.1 Electronics work counter center of room 1.09 0.09 Cargo hold #4 (starboard) 1000 998

  • Middle section Cargo Hold #4 60 58

I Page 9 "D" Deck Location Actual uR/hr Less bkgrd uR/hr Electrical panels behind control 1.83 0.83 Front of control panels 2.14 0.14 Control panel 2.14 0.14 Door handle R.M. controlled area hot lab entrance 2.88 0.88 Fume hood at door to hot lab 2.78 0.78 Sink outside door to hot lab 3.13 1.13 Portside eligine area 2.64 0.64 Platform mid room next to control area 1.61 0.61 Boiler starboard side next to control 1.78 0.78 Compressor at reactor bulkhead 2.03 0.03 Stairs mid turbine area at reactor bulkhead 3.72 1.72 Heat exchange unit 2.62 0.62 (Phosphate tank) portside corner next to reactor bulkhead 1.95 0.95

  • Dry stores portside at bow 8.24 6.24 Meat dry stores aft bulk stores 2.75 0.75 Special stores #1, coffee stores 2.91 0.91 Special storcs #3, shelf 3.80 1.80 Special stores #2, refrig, floor 3.78 1.78 Meats & poultry refrigerator 5.87 3.87 Cheese room 3.08 1.08 Fish room 3.33 1.33 Sugar & flour stores 3.29 1.29
  • Fire Station #31 69.24 67.24
  • Internal wall near Fire St. #31 194 192 Machine casing & battery fuse box (Engine roomt shaft) 2.86 0.86 Lube oil gravity tank (eng. nn. shf) 2.35 0.35 Vent exhaust system from galley (engine rm. shaft) 2.22 0.22

7_

Page 10 "14ft. Flat" Area location Actual uR/hr Less bkgrd uR/hr

  • Fire Station #32 18 16
  • Refrigerator unit 11 9

Stabilizer door at floor 4

2

  • Passage outside dairy refrig.

13 11 Dairy refrigerator 4

2

  • Root veg. refrigerator 11 9

Fire Station #31 200 198

  • Starboard side stabilizer hatch door 20.31 18.31
  • Floor area outside stabilizer hatch door (starboard side) 20.31 18.31
  • Stabilizer hatch door (portside) 32.72 30.72
  • Stabilizer 32.72 30.72 Main turbine-steam effluent (starboard side) 1.07 0.07 Main turbine steam effluent (port) 1.19 0.19 Starboard deck adjacent to stabilizer room bulkhead 8.63 6.63 Starboard deck center section to stabilizer room bulkhead 1.49 O.49 Starboard deck aft to stabilizer room bulkhead

.87

.0 starboard deck stairs bottom step to stabilizer rm. bulkhead

.91 0

Main shaft access

.64 0

Main shaft turbine entrance

.90 0

H.P.

L.P. crossover center section 1.01 0.01 Starboard side access port to main steam generator 4.58 2.58'

  • Wiring entrance to stabilizer room above entrance (starboard) 7.78 5.78
  • Center stairs deck 7.31 5.31
  • Wiring above entrance to stabilizer room (portside) 7.05 5.05 Deck adjacent to bulkhead (portside) 5.47 3.47 Deck center section (portside) 2.21 0.21 Deck stairs (portside) 1.81 0.81

m Page 11 14ft. Flat Location Actual uR/hr less bkgd uR/hr Entrance to main turbine (portside) 1.26 0.26 4

Stern to main turbine (portside) 1.69 0.69 l

Portside sten control system - Main section 1.02 0.02 Portside stern deck - Stabilizer room bulkhead 2.5 0.5 i

Center deck stabilizer room bulkhead 5.94 3.94 i

" Hold Area"

  • Back of reactor at section #2 100 i
  • Floor in port compartment 100 N
  • Elevator doors -portside 45 43
  • Security check Point #2 260-258
  • Elevator doors-Starboard side 40 38
  • Linen storage floor 60 58
  • Linen dry stores at pipes 9

7

  • Drain well behind grill door inside hatch - Starboard side 500 498 Fire Station #35 4

2

  • Indicates areas of ship that may not be entered by general public.

3 Patriots Point Employees and Contractors will be limited to the annunt of time that they may occupy these areas.

r v.

N.S. SAVANNAH SMEAR CONTAMINATION REPORT Date Collected:

November 12, 1981 Facility:

N.S. Savannah Instrument Used:

Tennelec 5100 Alpha Beta Counting System

Background:

Alpha.06 cpm / Beta-gamma 1.0 cpm Instrument Efficiency: Alpha 47.0% /. Beta-gamm 29.5%

Counting Time:

10 minutes "A" DECK Smear No.

Location Results 2

Alpha (dpm/100cm ) Beta-gamma 1

Passenger laundry 1.4 0

2 Room A-2 0

1.5 3

Room A-4 0.3 1.7 4

Room A-6 0.7 2.6 5

Room A-8 0.3 2.1 6

Room A-10 0

1.3 7

Room A-12 0

0.2 8

Room A-16 0.7 0.6 10 Conference room 0.7 3.2 11 Service pantry O.'3 2.8 12 Purser's office 0

2.8 13 Room A-18 1.0 3.8 14

' Room A-20 0.3 3.6 15 Room A-22 0.7 1.9 16 Room A-24 0.3 1.3 17 Room A-26 1.0 2.1

~

18 Room A-28 0.3 0.9 19 Room A-30 1.0 3.6 20 Room A-32 1.4 4.5 21 Doctor's office 2.0 1.9 22 H.P. lab floor 0.7 3.6 23 H.P. lab right counter top 1.4 1.9 i

f

c.-

.'..A" Deck con'd.

Results 2

Smear No.

Location Alpha (dpm/100cm ) Beta-gamma 24 H.P. lab left counter top 0

2.8 25-H.P. lab rear sink 0.7 2.1 26 Ward A 0.3 1.9 27 Ward B 1.4 2.8 28 Ward C 2.0 3.6 29 Ward D 2.7 4.3 30 Operating room 1.0 3.6 31 Main lobby (fomard) 2.0 4.5 32 Main lobby (aft) 1.4 5.3 33 Barber shop floor (fomard) 0.7 0.9 34 9arber shop floor (aft) 2.0 2.3 35 Room A-1 0.7 3.0 37 Room A-5 1.7 3.8 38 Room A-7 1.0 1.1 39 Room A-9 0.3 1.9 40 Room A-11 0.7 1.7 41 Room A-13 0.7 2.6 42 Room A-15 1.0 2.3 3.4 43 H.P. office floor 1.0 44 Doctor's state room 1.7 3.2 45 Steward's office 1.4 1.5 46 Room A-17 1.0 1.9 47 Room A-19 5.3 0.9 48 Room A-21 0

0.2 (2 smears taken,1 missing )

49 Room A-23 1.4 5.3 l

50 Room A-25 2.0 5.5 51 Room A-27 0.3 2.6 52 Room A-29 and ART-21 0

3.8 53 Floor between A-29 and ART-21 0

1.9 54 ART-21 third FHG 0

0 l

i

?

"B" D_ECK Resuus 2

Smear No.

Location

_ Alpha (dpm 100cm ) Beta-gamma 101 Dining rm. center cabinet top 0.3 4.0 102 Foyer floor to dining rm.

0 1.5 103 Conference rm. floor front of cabinet 0

0.6 104 Room B-49 floor 0-1.1 105 Passageway front of Rm. B-51 0

0.4 106 Cargo floor leading passage behind dining rm.

0.7 0.9 107 Room B-18 floor passage 0.7 0.2 108 Room B-8 floor 0.7 0

109 Room B-6 floor 0

1.9 110 Officer's mess floor 0.7 2.3 111 Officer's mess refrigerator 0.3 0.9 112 Counter bow end crew's lounge 1.0 1.1 113 Steward's mess floor 0

1.9 114 General cargo hold hatch #4 floor 1.4 1.5 115 Floor outside water down to reactor 1.7 1.9 116 Floor outside lighting load center 0.3 0.6 117 Room B-3 floor 2.0 0.9 118 Room B-1 floor 0.7 2.6 119 Floor outside fan room 0.7 1.7 120 Room B-7 floor 0.7 0

121 Room B-9 floor 0.3 0.9 122-Room B-11 floor 0

0.6 123 Room B-13 floor 0

0.6 124 Room B-15 floor 0

0.7 125 Room B-17 0

1.5 126 Room B-19 floor 0

1.1 I?7 Room B-21 floor 0

0.6 128 Room B-23 floor 0.3 1.1 129 Room B-29 0

1.7 i

_. _ _ _ ~,

r

,i "B" Deck con'd.,

)

Results 2

Smear No.

Location _

Alpha (dpmfl00cm ) Beta-Sana _

130 Passageway floor front of Fire Station #21 0.3 1.9 131 Room B-31 floor 0

0.4 132 Room B-33 floor 0.7 0

133 Room B-35 floor 0.3 0.9 134 Roo.n B-37 floor 0

0.9 135 Room B-39 floor 0'

O 136 Room B-43 floor 0.7 0.4 137 Room B-45 floor 0

0.4 3

138 Room B-47 floor 1.5 0

139 Room B-51 floor 0.1 0.6 140 Room B-53 floor 0

0.9 141 Room B-10 floor 0

0 142 Room B-12 floor 0.5 0.4 143 Room B-14 floor 0.5 0.4 144 Room B-16 floor 145 Room B-18 floor 0.1 0.4 146 Galley main counter 0.5 2.1 147 Floor mid way in front of main counter in galley 1.5 2.1 148 Counter top starboard, stern corner 2.8 1.7 149

  1. 148 counter floor 6.5 3.6 150 Dishwasher counter 2.8 5.6 151 Dishwasher floor 1.8 3.4 152 Galley counter top bow, starboard 1.5 3.2 153 Galley center room counter top 0.5 1.5 154 Galley portside counter top 0.1 1.3 155,

Galley sink 1.2 1.3 156 Officer's pantry floor 0

0.9 157 Crew's pantry floor 0.1 0.9 158 Steward's laundry 0.1 0.4 159 Crew's laundry 1.5 3.6 160 Room B-5 floor 1.1 2.3

1

\\

\\

" Top of the House" Results Smear No.

Location Alpha 2

(dpm/100cm ) Beta-ganma 201 Starboard side outside deck 0.8 2.6 202 Portside outside deck 1.8 2.6 f

Navigation Br_idge Deck 203 Pilot house at helm floor 0.5 0

204 Bridge wing - portside 1.2 2.1 205 Bridge wing - starboard 0.1 0.9 206 Bridge deck center 0.8 1.9 207 Gyro and radio room floor 0.8 3.4 208 Chart room floor 0.1 0.9 209 Chief radio officer room 0.1 1.3 210 Hallway front of Chief radio officer room 0.8 0.9 211 Reactor operaton room NB 4 floor 0

2.1 212 Radio room floor 0.5 0.4 213 Sea cabin NB-1 floor 0

0.2 214 Radio officer room NB-3 floor 0

0.4 215 Reactor operator room NB-5 floor 0.1 1.7 216 Deck toilet 0.8 1.7 Boat Deck 217 Officer's lounge floor 0.5 1.3 218 3rd Asst. Eng. SR floor BT-15 0.1 1.9 219 3rd Asst. Eng. SR floor BT-12 1.2 0

220 Main stairway near officer's lounge 1.2 2.6 221 3rd Asst. Eng. SR floor BT-10 0.1 0.4 222 2nd Eng. SR floor 87-13 0.1 1.3 223 3rd Officer's room floor 0.8 0.2 224 2nd officer's room floor 0.1 0.2 225 Exec. officer's office floor 0.1 1.1 226 Exec. officer's SR floor 1.8 1.1

Boat deck con'd.

-6,

Results Smear No.

Location 2

Alpha (dpm/100cm ) Beta-gamma _

227 Cpt. quarters threshold floor 0.1 0

between lounge and SR 228 Captain's SR 0.5 0.2 229 3d officer's SR floor BR-3 0.1 0

/

230 lla11way between Cpt. quarters and Chief Eng. day room 1.2 0

231 3rd officer room floor BT-1 0.8 0.6 232 1st Asst. Eng. SR floor BT-2 0.1 2.1

~

233 Chief Eng. dayroom SR rm. floor 1.2 1.3 234 Chief Eng. day room rug in 0.5 0

lounge area 235 Sr.1st Asst. Eng. office floor BT-6 1.2 2.1 236 2nd Asst. Eng. SR floor BT-8 0.8 2.1 237 Hallway front of counter & glass bulletin board 0.1 1.7 238 Starboard side outside deck 2.5 1.3 239 Behind generator on outside deck 0.1 1.3 240 Outside deck near entrance to officer's quarters leading to navigation bridge 0.8 2.6 241 Main hallway entrance to boat deck front of staisvay 0.8 2.6 242 Officer's laundry sink 2.2 1.9 243 Officer's laurdry sink underneath OL. 8 2.8 Promenade Ceck 244 Rom deck outside Verande 1.2 1.7 245 Prom deck outside portside 1.2 1.7 246 front of reactor hatch top 0.1 1.9 247 Starboard outside deck 0.8 1.0 248 Swinming pool tile boarder 0.1 2.6 249 Verande room bar counter top 0.5 0

250 Verande room ist sink in bar 1.2 2.3 251 Verande room 2nd sink in bar 0

0.2

Promenade Deck con'd,

.7 Results 7

2 Smea_r No.

Location A_1pha (dpm/1_0_0cm ) Beta Gamma 252 Verande room middle sink in bar 0.5 2.1

!)

253 Verande room bar floor 1.2 1.3 254 Stainless steel cabinet top

[

by phone in bar 0.8 2.6 255 Large liquor rack in bar storage area 0.5 2.6 256 Bar storage area floor 0.5 2.1 257 Stainless steel liquor rack behind bar 0

1.9 258 Cash register at bar 0.8 1.9 259 Verande room center floor 0.1 1.1 260 Verande room floor portside 0

0.2 261 Verande room floor starboard 0.8 1.0 j

262 Verande room floor near back wall 0.3 0.9 263 Prom deck men's room floor 1.2 0.9 264 Hallway to main lounge and cardnn. 0.5 1.0 265 Card room floor 0

0.6 266 Main lounge center floor 1.5 1.3 267 Prom deck library and reading j

room floor 0.5 3.0 268 Prom deck powder room floor 1.8 2.8 269 Novelty shop floor 0.1 0.4 "D" DECK 301 Store room next to electronics upper deck, starboard 1.8 1.1 302 Store rm. lower deck, starboard 0.5 1.1 303 Electronics work shop floor 0.8 1.7 304 Electronic work shop room center counter 0.5 2.3 305 Electrical panel floor behind controls 0.8 1.1 306 Floor front of control panels 0.5 4.0 307 Control panel 1.2 0.6 308 Door handle R.M. control area hot lab entrance 1.2 0.4

"D" Deck con'd.

- 8'-

Results 2

Smear No.

Location Alpha (dpm/100cm )_Betajanina, t

309 Hot lab door floor 0.1 0.6 a

310 Fumehood at door to hot lab 1.5 1.3 311 Fumehood side to hot lab 0.1 0.9 312 Fume hood floor to hot lab 0.5 3.0 313 Sink outside door to hot lab 0.1 1.7 314 Enginer area floor portside 0.5 1.1 315 Platform mid room next to control area 0.9 1.5 316 Boiler next to control starboard 0.5 0.4 317' Reactor bulkhead compressor 0.5 1.5 318 Reactor bulkhead stairs middle of turbine area 0.8 1.1 319 Heat exchange unit 0.1 2.6 320 Port corner next to reactor bulkhead (phosphate tank) 0.1 0.2 321 Dry store port bow side 1.5 1.9 322 Next dry store aft bulk stores 1.2 3.2 323 Coffee & special stores #1 0.5 2.1 324 Special stores #3 shelf 0.1 1.5 325 Special stores #3 refrig. floor 0.1 0.9 8

"14ft. Flat" 326 Fire Station # 32 0.5 2.6 327 Refrigerator unit d.5 0.4 "0" Deck continued 328 Meats & poultry refrigerator 1.8 1.3 329 Cheese room 1.2 3.4 330 Fish room 0.1 3.0 331 Sugar & flour stores 1.8 2.8 332 Fire Station #31 1.2 2.8

" Hold Area" 333 Fire Station # 34 floor 1.2 1.7 l

l t

f "0" Deck con'd.

9 Results Smear No.

Location Alpha (dpm/100cm2) Beta _ gamma 334 Machine casing / bat. fuse box eng. rm, shaft 0.8 1.3 335 Lube oil gravity tank /eng. rm.

shaft 0.8 2.6 336 Vent exhaust system from galley enginer room shaft 0.1 0.2

" Hold Area" below 14ft. Flat 337 Floor front of port compartment 2.5 3.0 338 Elevator doors portside 0.1 1.1 339 Security check point #2 0

0.4 340 Elevator doors starboard side 0.8 2.3 341 Floor of line storage 0.1 3.0 342 Dry stores at pipes 0

1.7 343 Drain well behind grill door inside hatch ( starboard) 0.8 3.0 344 Fire Station # 35 1.2 2.6 14ft. Flat Area 345 Stabilizer door floor 0.1 2.3 346 Smear lost 347 Passage outside dairy refrig.

1.5 4.7 348 Dairy refrigerator l.'2 1.3 349 Root vegetable refrigerator 1.2 2.1 350 Fire Station # 31 1.8 6.0 "C" Deck Stern 351 Cargo Hold #4 0.5 1.9 352 Cargo Hold #4 starboard 1.2 1.3 353 Emergency H.P. lab counter 1.5 6.6 354 Emergency H.P. lab floor 0.8 0.9 355 Carpenter shop floor 0.5 1.5 356 Passage outside emerg. H.P. lab 0.5 0.8

"B" Deck Sten Results Smear No.

Location Alpha (dpm/100cm ) Beta Gamma 2

357 Paint storage area floor 0.5 1.7 358 Port winch room 0.8 2.1 359 Steering room floor portside 0.8 3.2-360 Fire Station # 23-A 2.8 5.5 "C" Deck 406 Room C-24 floor 1.5 3.8 401 Room C-22 floor 0.8 3.6 402 Room C-20 floor 1.2 3.8 403 Room C-18 floor 0.1 5.3 404 Room C-16 floor 1.2 3.4 405 Room C-14 floor 1.8 3.8 406 Itallway front of sink & locker 1.2 1.3 407 Water fountain outside music entertainment room 0.8 3.0 408 Room C-23 floor 2.5 6.6 409 Room C-21 floor 1.5 4.3 410 Room C-19 floor 2.5 1.9 411 Room C-12 floor 0.5 3.0 412 Room C-15 "

1.2 2.6 413 Room C-13 "

3.5 4.0 414 Room C-11 "

1.8 3.2 415 Room C-9 "

1.8 3.2 416 Room C-7 "

1.8 4.5 417 Room C-5 "

2.8 3.4 418 Hallway outside Rms. C-11 through C-1 0.1 0.2 419 Room C-3 floor 1.2 2.6 420 Room C-1 floor 2.5 2.6 421 2nd Cook room C-2 floor 0.8 1.3 422 Butcher & chief pantry floor C-4 0.1 1.3 423 Porters Room C-6 floor 1.5 1.5 424 Room C-8, 1st & 2nd cook 1.5 1.3

y

' 11-

"C" deck con'd.

Results Smear No.

Location Alpha (dpm/100cm2) Beta gamma 425 Ste<ard & Store keeper Room C-10 floor 0.1 2.6 426 Room C-12 floor cook & baker 0.8

1. 9 427 Entrance to engine room viewing corridor portside 0

1.1 428 Center of engine room viewing corridor 0.5 3.6 429 Engine room viewing corridor starboard side 0.5 1.9 430 Laundry rm. floor near folding table 0.8 2.1 431 Laundry rm. floor near steam press and exhaust vent 1.2 2.8 432 Cold water chem lab outside door 0.8 2.3 433 Corridor from C-deck to D-deck 0.5 0.9 434 Cold locker floor 0.8 1.1 435 Garbage room floor 0.8 1.9 14ft. Flat Deck 501 Outside stabilizer hatch (closed) starboard side 1.2 4.4 502 Stabilizer room outside floor starboard side 1.2 0.4 503 Stabilizer hatch room outside portside 1.8 4.0 504 Stabilizer hatch room floor outside - portside 0.8 4.5 505 Main turbine steam effluent starboard side 2.5 1.9 506 Main turbine steam effluent portside 1.2 2.6 507 Starboard deck adjacent to star-board side room bulkhead 1.2 0.6 508 Starboard deck center section 0.1 0.4 509 Starboard deck aft section 0.1 3.4 510 Starboard - 14ft. flat deck stairs bottom step 1.2 1.1 511 Main shaft access 0.1 2.6 512 Main shaft turbine entrance 0.8 6.7 l

l

14ft. Flat con'd.

Smear lio.

Location Alpha (dpm/100cm ) Beta gamma 2

513 H.P. L.P. crr"'over central station 0.8 1.7 514 Access port to main star genera-tor starboard side 0.5 1.3 515 Wiring entrance to stabilizer room starboard side 0.8 3.0 516 Center stairs to D-deck

,2. 2 3.4 517 Wiring entrance to stabilizer room - portside 0.1 4.7 518 Deck adjacent to bulkhead-portside 0.5 1.5 519 Portside deck center section 2.8 2.6 520 Portside stairs to D-deck 1.2 1.7 521 Entrance to main turbine-portside 0.5 1.5 522 Stern of deck - portside 0.8 3.8 523 Stern near electric control system main section - portside 0.8 1.7 524 Stabilizer room bulkhead-portside 0.8 2.8 524 Center stablizer room bulkhead 0

0 i

i 6

e

--.,n-r.

I O

Semi-Annual Environmental Sample Analysis N. S. Savannah 1.

Date Collected Sample No.

Type Location November 13, 1981 71-0950 Silt N.S. Savannah Berth Resul ts:

Cc-144 0.76 0.20

p. /gn Ra-226 2.3 1 0.6 pci/gm Be-7 2.3 10.3 pci/gm Cs-137 0.20

.06 pci/gm Nb-95 0.24 1 0.07 pci/gm 2.

Date Collected Sample No.

Type location November 13, 1981 10-0950 Surface water

,N.S. Savannah berth Results:

Tritiun 865 29 pci/l Ra-226 55 30 pci/l Bi-214 20 5

pci/l Pb-214 20 t 5 pci/l

)

4