ENS 45621
From kanterella
Jump to navigation
Jump to search
ENS Event | |
|---|---|
05:00 Jan 11, 2010 | |
| Title | Agreement State Report - Qsa Global Report of Defect |
| Event Description | On January 8, 2010, at 1552 hrs. EST, QSA Global, Inc. reported a product defect to the Massachusetts Radiation Control Program in accordance with 105 CMR 120.142(B)(2)(a). The affected equipment that was identified was the QSA Model 87703 source assembly for a QSA Global radiography camera. A customer had received a Model 87703 source assembly and reported problems with the assembly connection on October 21, 2009. The unit was returned to their Burlington office and QSA Global, Inc. performed an evaluation of the extent of the defect. The Model 55000-3 female connector sleeve was identified as the source of the problem.
The defect causes the male connector to not be fully seated in the sleeve of the connector and it was, therefore, not locking in. This could give the user the false impression that the source is connected to the drive cable when there is only a partial connection. The operations manual requires that the user visually check the connection prior to use. If the user does not do a full visual inspection of the connection, the source assembly may not be secure and could dislodge inside guide tube. The manufacturer conducted a 100% search of a lot of 1,567 pieces in stock and found 5 to be defective. The manufacturer is sending out notice dated January 11, 2010 to all of their radiography customers describing this disconnect issue. There are 236 sources in the field with 29 different licensees that have not sent their units back to QSA Global in Burlington, MA at this time. They will continue to monitor this problem as assemblies are returned. |
| Where | |
|---|---|
| Qsa Global Burlington, Massachusetts (NRC Region 1) | |
| License number: | 12-8361 |
| Organization: | Ma Radiation Control Program |
| Reporting | |
| Agreement State | |
| Time - Person (Reporting Time:+10 h0.417 days <br />0.0595 weeks <br />0.0137 months <br />) | |
| Opened: | John Sumares 15:00 Jan 11, 2010 |
| NRC Officer: | Howie Crouch |
| Last Updated: | Jan 11, 2010 |
| 45621 - NRC Website | |
Qsa Global with Agreement State | |
WEEKMONTHYEARENS 579092025-09-08T05:00:0008 September 2025 05:00:00
[Table view]Agreement State Transported Package Exceeds Limits ENS 573642024-10-04T16:44:0004 October 2024 16:44:00 Agreement State Missing Category 2 Source ENS 559162022-05-27T11:50:00027 May 2022 11:50:00 Agreement State Lost Sealed Sources in Transit ENS 551732021-04-05T14:40:0005 April 2021 14:40:00 Agreement State Sealed Source Missing in Transit ENS 549662020-10-22T13:26:00022 October 2020 13:26:00 Agreement State Missing Category 2 Material ENS 520952016-07-14T23:35:00014 July 2016 23:35:00 Agreement State Agreement State Report - Radioactive Material Delivered to Private Residence ENS 508052015-02-09T06:00:0009 February 2015 06:00:00 Agreement State Agreement State - Radioactive Shipment Delivered to the Wrong Location ENS 490112013-05-08T14:00:0008 May 2013 14:00:00 Agreement State Agreement State Report - Potentially Lost Radioactive Source Shipment ENS 490072013-05-06T12:00:0006 May 2013 12:00:00 Agreement State Agreement State Report - Source Shipment Did Not Arrive as Expected ENS 485132012-11-15T05:00:00015 November 2012 05:00:00 Agreement State Agreement State Report - Iridium-192 Source Missing During Shipment ENS 476922012-01-31T05:00:00031 January 2012 05:00:00 Agreement State Agreement State Report - Contamination Levels Exceed Action Levels ENS 462562010-09-02T19:59:0002 September 2010 19:59:00 Agreement State Agreement State Report - Source Received with External Contamination ENS 456212010-01-11T05:00:00011 January 2010 05:00:00 Agreement State Agreement State Report - Qsa Global Report of Defect ENS 445362008-10-03T04:00:0003 October 2008 04:00:00 Agreement State Agreement State Report - Defective Component That Could Potentially Cause a Substantial Safety Hazard ENS 439952008-02-19T05:00:00019 February 2008 05:00:00 Agreement State Agreement State Report - Radioactive Material Shipment Missing ENS 438612007-12-11T05:00:00011 December 2007 05:00:00 Agreement State Massachusetts Agreement State Report - Lost Shipment of Iridium-192 Sources 2025-09-08T05:00:00 | |
Retrieved from "https://kanterella.com/w/index.php?title=ENS_45621&oldid=425455"