Browse wiki
Jump to navigation
Jump to search
The following information was received by … The following information was received by the Ohio Department of Health via email:</br>A patient was scheduled to receive 200 mCi of PLUVICTO (Lu-177), however, they only received approximately 130 mCi (administered to the prostate). The injection was through a three-way stopcock; one port was for the saline solution, the other for the PLUVICTO. Near the end of the injection, the material back flowed into the syringe and leaked onto the chucks under the patient's arm.</br>Based on surveys of the syringe and chucks, approximately 70 mCi had leaked which resulted in a 35 percent underdose. The patient was surveyed and there was no indication the patient's skin was contaminated or signs of extravasation.</br>The patient's physician was notified of the event.</br>An investigation by Ohio is pending.</br>NMED Report Number: OH250003</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
04:00:00, 25 March 2025 +
57,631 +
10:02:00, 27 March 2025 +
04:00:00, 25 March 2025 +
The following information was received by … The following information was received by the Ohio Department of Health via email:</br>A patient was scheduled to receive 200 mCi of PLUVICTO (Lu-177), however, they only received approximately 130 mCi (administered to the prostate). The injection was through a three-way stopcock; one port was for the saline solution, the other for the PLUVICTO. Near the end of the injection, the material back flowed into the syringe and leaked onto the chucks under the patient's arm.</br>Based on surveys of the syringe and chucks, approximately 70 mCi had leaked which resulted in a 35 percent underdose. The patient was surveyed and there was no indication the patient's skin was contaminated or signs of extravasation.</br>The patient's physician was notified of the event.</br>An investigation by Ohio is pending.</br>NMED Report Number: OH250003</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 27 March 2025 +
02110180013 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
11:20:05, 4 April 2025 +
10:02:00, 27 March 2025 +
2.251 d (54.03 hours, 0.322 weeks, 0.074 months) +
04:00:00, 25 March 2025 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.