The following information was received by the
Ohio Department of Health via email:
A patient was scheduled to receive 200 mCi of PLUVICTO [Lu-177], however, they only received approximately 130 mCi [administered to the prostate]. The injection was through a three-way stopcock; one port was for the saline solution, the other for the PLUVICTO. Near the end of the injection, the material back flowed into the syringe and leaked onto the chucks under the patient's arm.
Based on surveys of the syringe and chucks, approximately 70 mCi had leaked which resulted in a 35 percent underdose. The patient was surveyed and there was no indication the patient's skin was contaminated or signs of extravasation.
The patient's physician was notified of the event.
An investigation by Ohio is pending.
NMED Report Number: OH250003
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.