ENS 5245822 December 2016 17:35:00

The following information was received via fax: The ITT Conoflow models GT25CA1826 and GT25CD1826 current to pressure (l/P) transducers produce a calibrated 3.0 to 15.0 psi output pressure from a 4.0 to 20.0 mA DC (input) signal. On November 2, 2016, a discovery was made of unqualified electrical component substitutions on the internal circuit board. Supplier and internal record reviews indicate these unqualified components were used as early as May 10, 2013, as the previously qualified PCB components are no longer available. Environmental qualification of the suspect circuit boards is planned for early 2017. If any component falls to meet requirements, nuclear facilities which use this product will be advised to examine their product for unqualified printed circuit boards via notification and visual references. Qualified replacement components will be provided to all affected customers. Customer Name / Item Number / No.: Units Calvert Cliffs Nuclear / GT25CD1826 / 30 Control Components Inc. / GT2SCD 1826 / 3 Duke Energy Carolinas LLC / GT25CA1826 / 8 Duke Energy Carolinas LLC / GTI:SCD1826 / 4 Enertech Curtiss-Wright / GT25CA1816 / 19 Enertech Curtiss-Wright / GT25CD1826 / 4 Ergytech Inc. / GT25CA1826 / 3 Ergytech Inc. / GT25CD1826 / 3 Gefran Benelux NV / GT25CD1826 / 2 Ontario Power Generation / GT25CD1826 / 31 SPX Flow Technology / GT25CA1826 / 2 Weir Valves & Controls UK / GT25CA1826 / 78

  • * * UPDATE ON 2/28/17 AT 0918 EST FROM RAY SMITH TO DONG PARK VIA FAX* * *

On November 2, 2016 a discovery was made of unqualified electrical component substitutions on the internal circuit board. Supplier and internal record reviews indicate these unqualified components were used as early as May 10, 2013, as the previously qualified PCB components are no longer available. Environmental qualification of test samples containing the suspect circuit boards has been completed. The subject parts continued to operate and perform their intended safety function throughout all EQ (Equipment Qualification) tests. Notified R1DO (DeFrancisco), R2DO (Bartley), R3DO (Jeffers), R4DO(Deese), NMSS Events Resource, and Part 21 Group via email.

ENS 5230018 October 2016 10:08:00The following was received by facsimile: The ITT Conoflow models GT25CA1826 and GT25CD1826 current to pressure (l/P) transducers produce a calibrated 3.0 to 15.0 psi output pressure from a 4.0 to 20.0 mA DC (input) signal. Other dedicated l/P transducer part numbers are available with different output pressure ranges and input signal values. On August 19, 2016 a defective model GT25CA1826 (s/n 628894) was received for evaluation. This transducer would only achieve 0-9 psi output pressure, falling short of the specified 3-15 psi output range. The failure was caused by misalignment of a key internal sub-assembly component as a result of shipping damage (mechanical shock). The product has been repaired, calibrated/tested, and is being returned to the customer (Arizona Public Service Co.). The following is a list of customers, and the specific l/P transducer part numbers procured from ITT since 2006. These customers will be notified by November 1, 2016. ACSYR SL GT25CA1826 ALABAMA POWER COMPANY GT25CA1826 AMERGEN ENERGY, LLC GT25CA1826 ARIZONA PUBLIC SERVICE CO GT25CA1826 ASIAM INT'L (TAIWAN) INC. GT25CA1826 CALVERT CLIFFS NUCLEAR POWER PLANT, INC GT25CD1826 CATAWBA NUCLEAR STATION GT25CA1826 GT25CD1826 CCNPP STOREROOM GT25CD1826 CENTRAL NUCLEAR DE ALMARAZ GT25CD1826 CONTROL COMPONENTS INC GT25CD1826 DETROIT EDISON GT25CD1826 DTE ENERGY GT25CD1826 ENERGY NORTHWEST GT25CA1826 ENERTECH CURTISS-WRIGHT GT25CA1826, GT25CD1826 ENRICO FERMI POWER PLANT 2 GT25CD1826 ENTERGY OPERATIONS GT25CA1826, GT25CD1826 ERGYTECHINC GT25CD1826 EXELON BUSINESS SERVICES GT25CA1826 FLOWSERVE GT25CA1826 GEFRAN BENELUX NV GT25CA1826 GT25CD1826 KOREA HYDRO & NUCLEAR PWR GT25CA1826 NEBRASKA PUBLIC POWER GT25CA1826 NUCLEAR POWER PLANT KRSKO GT25CA1826 OMAHA PUBLIC POWER DIST GT25CA1826 GT25CD1826 ONTARIO POWER GENERATION GT25CA1826 GT25CD1826 SPX MCKEAN GT25CA1826 TVA NUCLEAR ASSURANCE AND LICENSING GT25CA1826 GT65CD1826 WATTS BAR UNIT 2 CONSTRUCTION GT25CD1826 WEIR VALVES & CONTROLS UK GT25CA1826 ZVONIMIR ZEC GT25CA1826