Static MDC Worksheet for Gas Flow Proportional Detector Static Alpha MDC - Aerotest 43-37 Detector MDC From CS-FO-PR-001 Rev. 1 Rs=

7.3 Background Count Rate (cpm) tb =

1 Background Count Time (min) t =

s 1

Sample Count Time (min) c;=

0.407 Instrument Efficiency (c/d)

Es =

0.25 Surface Efficiency (c/d)

C:\Users\swcroslin\Documents\Aerotest\Survey Packages & Sampling\lnstruments\Static & Scan MDC Worksheet for Aerotest Page 1 of 1

MDC Static MDC Worksheet for Gas Flow Proportional Detector eta MDC - Aerotest 43-37 Detector 3 + 3.29 R 8

  • t, * (I + t-)

A t

s I

s l 00 cm 2 From CS-FO-PR-001 Rev. 1 Instrument S. N.

126172 Detector S. N.

Rs=

1330 Background Count Rate (cpm) tb =

1 Background Count Time (min) t =

s 1

Sample Count Time (min) e; =

0.458 Instrument Efficiency (c/d) es=

0.25 Surface Efficiency (c/d)

MDC Static MDC Worksheet for Gas Flow Proportional Detector Sta* Alpha MDC - Aerotest 43-68 Detector 3 + 3.29 A

[ 5

  • E.
  • E 2

1 s

I 00 cm From CS-FO-PR-001 Rev. 1 120588 Rs=

3.1 Background Count Rate (cpm) tb =

1 Background Count Time (min) ts=

1 Sample Count Time (min) c; =

0.428 Instrument Efficiency (c/d)

C =

s 0.25 Surface Efficiency (c/d)

Static MDC Worksheet for Gas Flow Proportional Detector Static Beta MDC - Aerotest 43-68 Detector MDC A

t

  • S. *S s

s I 00 cni 2 From CS-FO-PR-001 Rev. 1 120588 Rs=

339 Background Count Rate (cpm) tb =

1 Background Count Time (min) t =

s 1

Sample Count Time (min) t;=

0.485 Instrument Efficiency (c/d) t =

s 0.25 Surface Efficiency (c/d)

Scan MDC Worksheet for Gas Flow Proportional Detector Scan Beta MDC* Aerotest 43*68 Detector d'. 60 R. ;

i B 60 Scan MDC=

A

P

  • & i * & s
  • I O O cm 2 From CS-FO-PR-001 Rev. 1 120588

=

0.50 Scan c;=

0.485 Scanning Instrument Efficiency (c/d)

Scan £ 5 =

0.25 Scanning Surface Efficiency (c/d)

Scan S eed =

4.6 inches/sec 339 Background Count Rate (cpm)

Detector Width =

4.6 A=

i=

Scan MDC=

1,822

Scan MDC Worksheet for Gas Flow Proportional Detector Scan Beta MDC - Aerotest 43-37 Detector d' _ 60 R _ i i

B 60 Scan MDC =

A

..jp. E,- *Es* 100cm2 From CS-FO-PR-001 Rev. 1 126172 92791

=

Surve or Efficienc Scan c; =

0.458 Scanning Instrument Efficiency (c/d)

Scan c5 =

0.50 Scanning Surface Efficiency (c/d)

Scan S eed =

6.3 inches/sec 43-37 Detector Beta i=

1.00 time interval secs Scan MDC=

Alpha Scan Probability Worksheet for 43-68 Gas Flow Proportional Detectors a Scan Detetction Probability (43-37)

-l*G*&i *cs *Wdet P(n > 1)== 1-e 60*Rate Instrument S. N.

et c or.

BS=

G=

E =

s CS-FO-PR-001 Rev. 1, Section 4.8.4.1 Inputs 120588 75164 0

670 0.428 Detector Efficiency (2rc) 0.25 Surface Efficiency 1.00 source efficiency (radionuclide detectability) probability of observing a single count1 1 Must be 70% or greater

Alpha Scan Probability Worksheet for 43-37 Gas Flow Proportional Detectors a Scan Detetction Probability (43-37)

-1 *G*Ei *Es *Wdet P(n > 1)== 1-e 60*Rate CS-FO-PR-001 Rev. 1, Section 4.8.4.1 Inputs Instrument S. N.

126172 Detector S. N.

92791 s-0 710 e;=

0.407 Detector Efficiency (2n)

Es=

0.25 Surface Efficiency Sc=

1.00 source efficiency (radionuclide detectability)

Wdet width of detector in direction of scan (cm) probability of observing a single count1 1 Must be 70% or greater