ML20059A867

From kanterella
Jump to navigation Jump to search
Insp Rept 99990001/90-06 on 900401.No Violations Noted. Major Areas Inspected:Background,Teletherapy Source Exchange & Independent Measurements
ML20059A867
Person / Time
Issue date: 08/07/1990
From: Costello F, Kinneman J
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION I)
To:
Shared Package
ML20059A859 List:
References
REF-QA-99990001-900807 99990001-90-06, 99990001-90-6, NUDOCS 9008240062
Download: ML20059A867 (4)


Text

+

4 y

2 1

' )

U. S. NUCLEAR-REGULATORY COMMISSION REGION:I Report No. 99990001/90-006 Docket No. 99990001 s

.f License N4. General License (10 CFR 150.20)

(State of Maryland License No. MD-31-025-03)-

Licensee: Neutron Products, Incorporated

.22301 Mt. Ephrain Road-Dickerson, Maryland Inspection At:

Riverview Medir.41 Center Red Bank, New Jersey Inspect!on Conducted: April 1, 1990-Inspector:

/Ld4t,cle

/I I 7!7O Francis M. Costello I date Senior Health Physicist

-E!7!90 Approved by:

o n D. Kfnnemani Chief

' tlate l

Nu lear Materials Safety Section B Inspection Summary:

Special, Announced Safety Inspection Conducted-on April l',

1990 (Report No. 99990001/90-006)

'I l

l Areas Inspected:

Background; teletherapy source exchange; and independent i

l-measurements.

P Results: No violations were identified.

i

(

l l'

9008240062 900007 -

l PDR GA999 EMVfEUT 99990001 PNU.

)

1 DETAILS l

1.

Persons Contacted Marvin Turkanis - Neutron Products,- Vice President-(by telephone)

Tom Kidd - Neutron Products Service Technician Andy Koontz - Neutron Products Truck Driver _ and Assistant.

J Service Technician-Richard Tenpenny - Radiation Safety Officer, Riverview. Medical Center i

Charles Orlando - Radiology Manager, Riverview Medical Center

2..

Background

r Riverview Medical Center is authorized by NRC License No. 29-09701-02 to possess and use a cobalt-60 source in an. Advanced Medical Systems (Picker)

~

Model C teletherapy unit.: Aeutron Products, Incorporated is authorized by State of Maryland License No, MD-31-025-03'to service and install sources into these teletherapy units.

Neutron Products is authorized by the general license issued by 10 CFR 150.20 to perform these services in states where NRC maintains regulatory jurisdiction, such as New Jersey.

On April 1,1990, Neutron Products performed a source replacement at Riverview Hospital in Red Bank, New Jersey. A special NRC inspection was conducted to review this source replacement'.

3.

Teletherapy Source Exchange Procedure The replacement source, containing.approximately 6,000 curies-of cobalt-60 was delivered in a shielded Neutron ~ Products shipp.ing cask with three i

drawers for holding sources.

These drawers have a shielding plug at each end to fill the two-inch diameter cylinder which constitutes each 21.5 inch drawer.

The source exchange consists of the followingfoperations.-

1 which were described to the inspector by-the service technicians prior to the exchange:

A.

Source Removal 1.

Preparation of the teletherapy unit for exchange. This operation i

results in the norizontal alignment of.the. teletherapy head.

2.

Removal of both shielding plugs from one of the drawers of the shipping cask.

I 3.

A alignment of the cask drawer with the beam portLof the teletherapy head, l

L 4.

Movement-of the source inside the teletherapy head to-the.

exposed position.

i

  • S.

Insertion of a tool through the drawer of the shipping cask to grasp the source in the teletherapy head, k

  • 6.

Retraction of the source from the head.to the middle of the shipping cask.

t 3-

  • 7.

Disconnectin'g the tool from the source and removal-of the tool'from the shipping cask.

{

  • 8.

Insertion _ of shielding plugs at each end of the cask drawer, using a.

-tool to hold the plugs.

B.

Source Installation

  • 1.

Removal.of the shielding plug from one end of the shipping cask; drawer which contains the replacement source..

l 1

2.

Alignment of the cask drawer with the beam port of the teletherapy -head.

I 1

  • 3.

Removal of the shielding, dug from the other end of the shipping.

j cask drawer which contains-the replacement source.

-i

-l

  • 4.

Insertion of a tool through the draweraof the shipping cask to grasp j

the source in the cask.

j 5.

Insertion of the source from;the cask ir.to the teletherapy head.

i

'6.

Disconnecting the tool from the source and removal of-the tool-from.

the teletherapy head through the shipping casii.

-l 7.

Movement of the= replacement source inside the teletherapy hsad to the:

shielded position.

8.

Replacement of both shielding plugs.with the-empty drawer of the i

shipping cash.

9.

Returning -the teletherapy unit to its normal operating condition.

This includes performance of routine inspection and maintenance of the unit.

i The asterisks indicate operations where the hand of the service. technician-is near a beam of radiation coming out of the' drawer of the shipping. cask.

The technician wore vendor dosimetry on their wrists and body and also wore self reading dosimeters.

The technicians were aware of the steps-in the procedure which were performed with'the' source partially' unshielded.

The inspector did not observe any departures ~by the service-technicians from-the p'snned source exchange procedures.

4.

Independent Measurements The inspector provided Panasonic Model 801 thermoluminescent dosimeters (TLD) to the' service technicians to supplement the licensee-supplied dosimetry during the source exchange.

During.he source removal procedures, Individual B wore an NRC TLD on the-back or each hand and another TLD on the little finger of his left hand.

q i

i

e'

.c

~

4 l

During the' source installation, Individual B' wore additional NRC TLD's on the palm of each hand and another NRC TLD on his right thumb.

Furthermore, during the source installation, Individual A wore NRC TLD's at the same locations'as1 Individual B The NRC TLO's were evaluated in the NRC Region-I Dosimetry Laboratory.

The results were as follows:

i

-I Location Individual A Individual B j

t Left Palm 82 mrem 615 mrem 1

Left Back 72 mrem

  • 372 mrem i

Left Little Finger 64 mrem-

' 491 mrem; 120 mrem

}

R'ght Palm 67 mrem

  • 79 mrem l

Right Back 197 mrem

  • 152 mrem i

Right Thumb 139' mrem 75 mrem All doses except those marked with asterisk were measured during-source i n s t all a ti on ~.

1

  • - Indicates doses measured during both source removal and nstallation.

~

5.

Exit Interview 9

The inspector met with the' Neutron Products service technicians at the end I

of the inspection and discussed the scope and findings ofcthe inspection.

The inspector also discussed the inspection by telephone with Marvin Turkanis, Neutron Products Vice President.

j i

i I

-h

..