APPENDIX B NOTICE OF DEVIATION Consolidated Edison Company of New York, Inc.

New York, New York 10003 License Nos. DPR-5 and DPR-26 Accepted radiological safety practices dictate that any instrument used for the protection of individuals against radiation be rou-tinely calibrated to ensure accuracy including linearity across all scales over which the instrument may be used.

However, you cali-brate all your survey instruments, by checking only.one point on each scale with an external source.

This item is a deviation.

2 w

a

'8204070020-750705 PDR ADOCK 05000003 PDR G

.,,