e APPENDIX-B NOTICE OF DEVIATION L
Texas Engineering Experiment Station Docket: _ 50-128/90-02 Texas A&M University Operating License:
R-83 iy Based onLthe results of an NRC-inspection conducted on: February 21-23, 1990, a;
deviation of your safety analysis report was identified. _The deviation q
' involved the failure to follow instrument calibration procedures In accordance with the " General Statement of Policy and Procedure for.NRC~
j Enforcement Actions," 10 CFR Part 2, Appendix C-(1989) (Enforcement-Policy),
l the deviation is listed below:
!g Calibration of Neutron Survey Instrument-Section X of the Safety Analysis Report requires written procedures and states, "The procedures are followed by all personnel using the facilities _
of the Nuclear Science Center."
L i
Standard Operating Procedure VIIB-13 requires portable radiation' survey-instruments be calibrated on an annual basis.
In deviation from the above, the neutron survey instrument used to perform radiation surveys of the beam ports was not calibrated between July 15, l
1988, and January 9,1990..(128/9002-02) t 1
1 Texas A&M University is hereby requested to submit a written statement or explanation to this Off. ice within 30 days of the date of the letter-transmitting this Notice. This reply, should include for each deviation:
(1) the reason for the deviation if admitted, (2) the corrective steps that have been taken and the results achieved, (3) the corrective steps-that will be taken to avoid further deviations, and (4) the date when full-complianceEwill i
be achieved. Where good cause is shown, consideration will be given to extending the response time.
q; Dated at Arlington, Texas, this j g Zb dayof(23q1)[