ATTACHMENT 2 TO AEP:NRC:0856N PROPOSED CHANGES TO THE DONALD C.

COOK NUCLEAR PLANT UNIT NOS.

1 AND 2 TECHNICAL SPECIFICATIONS

'I 8511120192 851104.."

PDR ADOCK 05000315 P

PDR'

gC

'I I '

CONTAINMENT SYSTEMS 3/4.6.4 COMBUSTIBLE GAS CONTROL HYDROGEN ANALYZERS LIMITINGCONDITION,FOR OPERATION 3.6.4.1 Two containment hydrogen analyzers shall be OPERABLE.

APPLICABILITY:

MODES 1 and 2

ACTION:

a ~

b.

With one hydrogen analysis device inoperable, restore the inoperable analysis devi'ce to OPERABLE status within 30 days or be in at least HOT STANDBY within the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

With both hydrogen analysis devices inoperable, restore at least one analysis device to OPERABLE status within 72 hours8.333333e-4 days <br />0.02 hours <br />1.190476e-4 weeks <br />2.7396e-5 months <br /> or be in at least HOT STANDBY within the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

SURVEILLANCE RE UIREMENTS 4.6.4.1 Each hydrogen analysis device shall be demonstrated OPERABLE at least once per 92 days on a STAGGERED TEST BASIS by performing a

CHANNEL CALIBRATION using a four percent and fifteen percent nominal hydrogen gas, balance nitrogen.

D. C.

COOK - UNIT 1 3/4 6-'23 Amendment No.

CONTAINMENT SYSTEMS 3/4.6.4

.COMBUSTIBLE GAS CONTROL HYDROGEN ANALYZERS LIMITINGCONDITION FOR OPERATION 3.6.4.1 Two containment hydrogen analyzers shall be OPERABLE.

APPLICABILITY:

Modes 1 and 2.

ACTION:

a ~

With one hydrogen analysis device inoperable, restore the inoperable analysis device to OPERABLE status within 30 days or be in at least HOT STANDBY within the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

b.

With both hydrogen analysis devices inoperable, restore at least one analysis device to OPERABLE status within 72 hours8.333333e-4 days <br />0.02 hours <br />1.190476e-4 weeks <br />2.7396e-5 months <br /> or be in at least HOT STANDBY'ithin the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

SURVEILLANCE RE UIREMENTS 4.6.4.1 Each hydrogen analysis device shall be demonstrated OPERABLE at least once per 92 days on a STAGGERED TEST BASIS by performing a

CHANNEL CALIBRATIONusing a four percent and fifteen percent nominal hydrogen gas, balance nitrogen.

D. C.

COOK UNIT 2 3/4 6-33 Amendment No.

1'

'I

ATTACHMENT 1 to AEP:NRC:0856N REASONS AND 10 CFR 50.92 ANALYSIS FOR CHANGES TO THE DONALD C.

COOK NUCLEAR PLANT UNIT NOS.

1 AND 2 TECHNICAL SPECIFICATIONS

I

~

J'

CONTAINMENT SYSTEMS 3/4.6.4 COMBUSTIBLE GAS CONTROL HYDROGEN ANALYZERS LIMITINGCONDITION FOR OPERATION 3.6.4 '

Two containment hydrogen analyzers shall be OPERABLE-APPLICABILITY:

Modes 1 and 2.

ACTION:

a ~

With one hydrogen analysis device inoperable, restore the inoperable analysis device to OPERABLE status within 30 days or be in at least HOT STANDBY within the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

b.

With both hydrogen analysis devices inoperable, restore at least one analysis device to OPERABLE status within 72 hours8.333333e-4 days <br />0.02 hours <br />1.190476e-4 weeks <br />2.7396e-5 months <br /> or be in at least HOT STANDBY within the next 6 hours6.944444e-5 days <br />0.00167 hours <br />9.920635e-6 weeks <br />2.283e-6 months <br />.

SURVEILLANCE RE UIREMENTS 4.6.4.1 Each hydrogen analysis device shall be demonstrated OPERABLE at least once per 92 days on a STAGGERED TEST BASIS by performing a CHANNEL CALIBRATIONusing a four percent and fifteen percent nominal hydrogen gas, balance nitrogen.

D. C.

COOK - UNIT 2 3/4 6-33 Amendment No.