American Radiation Services, Inc. Preliminary Laboratory Analysis Report No. ARS1-08-01562
ML090540700
Person / Time
Site: Indian Point  
Issue date: 09/16/2008
From: Mulligan V
American Radiation Services
To: Noggle J
NRC Region 1
References
ARS1-08-01562
Download: ML090540700 (12)


Text

2609 North River Road, Port Allen, Louisiana 70767 (500) 401-4277 --

FAX (225) 381-2996 RNATIONAL American Radiation Services, Inc.

Laboratory Analysis Report ARS1-08-01562.

Prepared for:

Nuclear Regulatory Commission (NRC)

James Noggle USNRC Region 1 475 Allendale Road King of Prussia, PA 19406 3ames.Noggle@nrc.gov Phone: 610.337.5063 Fax:

Proe<rýnae eview q r1'n-W1X x

6 Management Rh)iew Notes: American Radiation Services, Inc. assumes no liability for the use or interpretation of any analytical resuhs provided other than the cost of the analysis itself.

Reproduction of this report in less than full requires the wiitten consent of the client.

Contact Person: Questions regarding this analytical report should be addressed to:

Project Manager ProjectManagers@a mrad.com Phone: 225.381.2991 Faxi 225.381.2996 LELAP Cert# 30658 NELAP Cert# E87558 LELAP Cert# 30658 NELAP Cert# E87558

2609 North River Road e Port Allen, Louisiana 70767

  • ER AE RDIATION "1 (800)401-4277 Fax (225)381-2996 SERVICES, INC.

September 16, 2008 Nuclear Regulatory Commission James Noggle 475 Allendale Road King of.Prussia, PA 19406 Client Sample: LAF-002-(009)

ARS SGD: ARSI-08-01562-001

Dear Mr. Noggle,

On August 12, 2008, American Radiation Services (ARS) received I Ground Water Sample to be analyzed for Gamma Spectroscopy, Tritium, Strontium-90, and Nickel-63.

The sample was processed and counted using the appropriate counting equipment and QA/QC for this type of analysis. Results of the analysis and QA/QC are attached in the data package.

  • The sample and QA/QC's were counted with a count time sufficient to meet a statistical sound detection limits.

Counting equipment Quality Assurance was with in acceptance criteria when the above referenced samples were processed.

If you.have any questions please do not hesitate to call at 225-381-2991.

Sincerely, Labori Ry Director American Radiation Services

'AfiRS INTERNATIONAL 2609 North River Road 9 Port Allen, Louisiana 70767 1 (800) 401-4277 9 Fax (225)381-2996 COVER PAGE Statement of Work for Analytical Laboratories PROJECT SAMPLE IDENTIFICATION CROSS-REFERENCE TO ARS SAMPLE LABORATORY IDs CLIENT SAMPLE American Radiation Services SAMPLE ID NUMBER(S)

LAF-002-(009)

ARSI-08-01562-001 SAMPLE RECEIPT The samples were received in good condition. The samples were screened for radioactive contamination as per procedure ARS-062 "Sample Receiving".

ANALYTICAL METHODS The Gamma Spectroscopy determinations were performed using American Radiation Services procedure ARS-007/EPA 901.1, "Gamma Emitting Radionuclides in Water." The tritium analyses were performed using American Radiation Services procedure ARS-054, "Tritium In Water". ". The Strontium-90 analyses were performed using American Radiation Services procedure ARS-032, "Total Strontium by Eichrom Resin Seperation." The Nickel-63 analyses were performed using American Radiation Services procedure ARS-022.

ANALYTICAL RESULTS The result data that are flagged with "U" indicates that the activity is below the MDC.

For Nickel-63 analyses, the LCSD fell outside the acceptance criteria biased low; however, the LCS did meet the acceptance criteria and the RER/DER met acceptance criteria. The data is being reported per technical review.

For Strontium analyses, the LCSD fell outside the acceptance criteria biased low; however, this was due to loss of sample during the analysis due to equipment malfunction: RER and DER didn't meet acceptance criteria due to low bias of LCSD; however the LCS is well within the acceptance criteria. The data is being reported per technical review.

AMEL CNRA DIATION

.SERVICES. INC.

2609 North River Road 9 Port Allen, Louisiana.70767 1 (800) 401-4277

  • Fax (225) 381-2996 American Radiation Services Project Manager/Laboratory Manager's Comments:

"I certify that this sample data package is in compliance with SOW requirements, both technically and for completeness, other than the conditions detailed above. Release of the data contained in this sample data package and the computer-readable EDD, as applicable, submitted on diskette or by modem, has been authorized by the Laboratory Manager or the Manager's designee, as verified by the following signature."

'7 certify that this electronic image, and all hardcopies produced from this image, accurately represents the data and is in.compliance with the client specific requirements, both technically and for completeness, other than the conditions detailed above or in the sample data package narrative. Release, by submission through email, the data contained in this electronic image and the computer-readable EDD (as applicable), has been authorized by the laboratory Manager/Technical Director or the Manager's designee."

Sigture-Laboratory Director, American Radiation Services Title Date

INTERNATIONAL 2609 North River Road, Port Allen, Louisiana 70767 1 (800) 401-4277 FAX (225) 381-2996 ARS Sample Delivery Group:

ARS1-08-01562 Client Sample ID:

LAF-002-(009)04-10/1048 Sample Collection Date:

04/10/08 10:48 Sample Matrix:

Aqueous Request or PO Number:

ARS Sample ID:

Date Received:

Report Date:

N/A ARS1-08-01562-001 8/12/2008 08/22/08 08:27 Analysis Analysis Analysis MDC DC Anysi Description Results E

. 2 s Units MN-54

-0:945 1.622 2.230 1.115 pCi/L AnaJysis Test Method Analysis Analysis Date/Time Technician Tracer/Chem Recovery ARS-007/EPA 901.1 8/19/08 16:11 1

JLA N/A

-1.358 1

1.338 4.470 2.3i~/

ARS-007/EPA 901.1 I8/19/08 16:11 1

ILA N/A CO-58

-0.531 2.534 2.250 1.125 pCi/L ARS-007/EPA 901.1 8/19/08 16:11 ILA N/A CO-60

-1.406 I

2.454 2.790 1.395 pCi/L ARS-007/EPA 901.1 8/19/08 16:11 ILA N/A ZN-65

-1.583 1.580 5.280 2.640

, pCi/L i

ARS-007/EPA 901.1 8/19/08 16:11 ILA N/A ZR-95 0.829 I

1.001 3.370 1.685 pCi/L ARS-007/EPA 901.1 j

8/19/08 16:11 ILA N/A NB-95

-0.093 0.526 1.800 0.900 PCi/L ARS-007/EA91 8/19/08 16:11

)LA N/A CS-134

-20.159 706.380 25.100 12.550 pCi/.

  • ARS-007/EPA 901.1 8/19/08 16:11 L

N/A 1 7 2

0.....

0....

CS-3

-0.084 I

0.698 2.370 1.185,

pCi/L AR-0/PA 901.1 8/19/08 16:11 J ILA N/A BA 140

-4.426 27.062.i 7.990 3.995 pCI/L ARS-0O7/EPA 901.1 8/19/08 16:11 1

JLA N/A S....-...---....---..---.-....-......--..--

I.......-.--..-............_

LA.140

-1.536 3.584 3.170 1.585 pCI/L ARS-007/EPA 901.1 8/19/08 16:11 IJLA N/A SR-90 0.248 0.083 0.253 0.121 pCi/L j

ARS-O32/Eichrom SRW-01 8/21/08 14:48 8.S 108.00%

16.99 8.74 CII.

ARS-034/EP I______

_____0_._0%

H-1

-8.789 96.061 i

163.989 80.754 pCiL ARS-4/EPA 906.0 9/8/08 4:31

-j BJS N/A N-63-0.583 4.788 i

8.185 4.027 pCI/L.

ARS-022 9/7/08 13:49 81 S N/A i

n R

o

/

h e

NRC-03-08-07 NOTES: Preliminary Report / Shipment# 08-152 / NRC-03-08-072 I

-v'-\\Al i 1 '4-llryl')-t-L Prq)ect Manage~V)

Notes: American Radiation Services, Inc. assumes no liability for the use or interpretation of any analytical results provided other than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent Of the client.

LELAP Certificate# 30658 NELAP Certificate # E87558

RS AMERICAN RADIATION SERVICES, INC.

2609 North River Road, Port Allen, Louisiana 70767 1 (800) 401-4277 FAX (225) 381-2996 QC Results Report Sample Delivery Group:

Date Received:

ARSI-08-01562 08/12/08 Laboratory Control Sample Evaluation Anlyi Qc Alye Analysis (1)

MDC Expected Report Analysis Analy Analyois Percent nth Tye t

e ts D

Vol Ql Units Tost Method Date/Time Technican Rcovery (%)

Acceptanc.

I C Sr-90 IS1.91 21.8 1____

8/1 14:4__

750/6I___________125%___

a jARSI-B08-02213

__S-0 1.1 1

0.295 20.80 pCI/L ARS-032/EPA 905.0 8/21/08 14:48 BIS 77 75%-125%

Blank Evaluation Anelysis Qr' AnalyTe Analysis c1 U I (1 MDC Expectod r14.

Report Analysis Analysis Analysis Botch Type Results Value Units Toot Method Dote/Time Technician ARS1-8B08-02213 MBL Sr-90

-0.099 0.083 0.253 NA U

pCVL ARS-032/EPA 905.0 8/21/08 14:48 BJS RER Duplicate Evaluation Analysis QC Analysis Resuit 1 U 1 (1 0)

Resu.t 2 1SU 2 (10) QuO, Analysls Analyslis Analysis Analysis R

AcReptacR Batch Type Description UI 1 (2

)

unit.

Test Method Date/Time Technician RER ARS1-B08-02213 LCSD Sr-90 15.91 1.2 13.13 1.0 pCi/L ARS-032/EPA 905.0 8/21/08 14:48 1 BJS 1.24

< 1 DER Duplicate Evaluation Anlss o.

i; t(.

ul Analysis Analysis Analysis Analysis D R A

DEts-R Analysis Q J Anelysis Rh CSU1:(1 s)

Resuit 2 CSU 2 (1.)

QsuIaelyo*

AneloloAnolcpt nenyle Batch "ype De___riptnR eslt 1

Unit.

Teot Method Dote/Time Technician D

A 1l-B08-02213 LCSD Sr-90 1

15.91 1

1.2 13.1 1.0 pCI/L ARS-032/EPA 905.0 8/21/08 14:48 BJS 3.49 Proect Manager Notes: American Radiation Servlcem, Inc. assumes no liability for the use or intervretation of any analytical results providec other than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent of the LELAP Certificate# 30658 NELAP Certificate #E87558

A-R INTERNATIONAL 2609 North River Road, Port Allen, Louisiana 70767 1 (800) 401-4277 FAX (225) 381-2996 QC Results Report Sample Delivery Group:

Date Received:

ARSI-08-01562 08/12/08 Laboratory Control Sample Evaluation Analysis QC Anit Analysis CS. 1 (1 5)

Batch Type Results.

ARS1-B08-02204 LCS CO-60 24779.00 495,75 ARSI-808-02204 LCS CS-137 15083.00 359.87 ARS1-808-02204

,LCS At-241 34007.00 1235.00 MOC 200.80 137.80 277.10 Expected IQus.

Value 25315.22 15285.98 34794.33 Report Units pCI/9 pCI/9 PCI/g Analysis Test Method EPA 901.1 EPA 901.1 EPA 901.1 Analysis Date 08/18/08 08/18/08 08/18/08

-I.-

AnaIysi*

Percent Technician Recovery(%)

2LA 98 ILA 99 ILA 98 iLCS Acceptanlce Being.

75%-125%

75%-125%

75%-125%

i

-I.-

Blank Evaluation Analysis QC An)E dRenAnalysis valx e Repo et A etysd Oats hnalysis Batch Type Anbjyte Results U

Valo e Unit.e Test Method Date Technician ARS1-508-02204 MBL CO-60

-0.001 0.005 0.005 NA U

pCi/9 EPA 901.1 08/18/08 ILA ARS1-B08-02204 MBL CS-137 0.001 0.001 0.004 NA U

pCI/g EPA 901.1 08/18/08 ILA ARS1-808-02204 MBL AM-241

-0.001 0.014 0.006 NA U

pCI/g EPA 901.1 08/18/08 ILA RER Duplicate Evaluation

.Analysis QC Analys lyss Analyss Anlsis RER Batch Type Description Result 1 CSU 1(1.)

Result 2 CSU a(1e)

Quel Unit.

AtesMs Aeihnili-n REReAhniceag ARS1-808-02204 LCSD CO-60 24779.00 495.75 25129.00 492.32 pCI/g EPA 901.1 08/18/08

)LA 0.35

< 1 ARS1-808-02204 LCSD CS-2137 15083.00 359.87 15425.00 341.69 pCi/g EPA 901.1 08/18/08 3LA 0.49 0 1 ARS1-808-02204 LCSD AM-241 34007.00 1235.00 34050.00 1247.60 pCl/g EPA 901.1 08/18/08

)LA 0.02

< 1 DER Duplicate Evaluation Analysis QC Analysis Analyst.

Analysi.

Analysis Analysis DER Batch Type ie.o'ripU)on (is)

Qual Unit.

Test method O.ir Technician DER Acceptan.enpe ARS1-808-02204 LCSD CO-60 24779.00 495,75 25129.00 492.32 pCi/g EPA 901.1 08/18/08 3LA 1.00

< 3 ARS1-B08-02204 LCSD CS-137 15083.00 359.87 15425.00 341.69 pCi/g EPA 901.1 08/18/08 ILA 1.38

< 3 ARSI-808-02204 LCSD AM-241 34007.00 1235.00 34050.00 1247.60 pCVg EPA 901.1 08/18/08

)LA 0.05

< 3 Quality Assurance R9 Notes: American Radiation Services, Inc. assumes no flability for the use or interpretation of any analytical results provided other-than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent of the LELAP Certilficate# 30658 NELAP Certificate # E87558

~A iZA RA~iON SERVICES, INC.

2609 North River Road, Port Allen, Louisiana 70767 1 (800) 401-4277 FAX (225) 381-2996 QC Results Report Sample Delivery Group: ARS1-08-01562 Date Received: 8/12/2008 Laboratory Control Sample Evaluation Tnlyi 1S 1

(I 1

TI IL Anelyole QC Anilyte CSU 1 (1 a)

MDC Expected Qual Report Analysis Analysis Apelyseis PAept s

Retch Type Results VIlue Unit.

Test Method Date/Time Tedunicien Re-Very (%)

Rpn N/A LCS NI-63 223.41 51.9 8.5485 204.46 pCi/L ARS-022 9/5/08 19:45 BIS 109 75%-125%

Blank Evaluation Analysei QC Anslyss CSU 1 (1 5)

MDC Expected

Qua, Report Anelyeis Analysis Analysis nayss S

I (1

1)

Qul

'"Mitho Retch Type Analyte Results Value Unit.

Test Method

" Date/Time Technician N/A MBLr Ni-63 2.094 4.66 7.8386 NA U

pCi/L ARS-022 9/6/08 4:18 BJS RER Duplicate Evaluation 1:y i Result 1 CU1 (1 a)

Result 2 CSU 2 (1s) QuelatHthdc~tj~al Aniysi QC Analysis U

1

[Analysis AnalysisT Analis nsis 1

RER Retch Type Description R

[u J Unit.

Test Method Dte/Time Technician RER AcceptenceR N/A LCSD Ni-63 223.41 51.9 143.428 33.597 pCi/L ARS-022 9/6/08 0:03 BIS 0.94

< 1 DER Duplicate Evaluation Analysis QC 1Analysis t

C (1s l

f Analysis Analysis nalyuis Analysis 1

DE Retch Type Description Result 1

Units.

Tost Method Date/Time Technician DER Acceptancl N/A LCSD Ni-63 223.41 51.9 143.428 33.597 pCO/L ARS-022 9/6/08 0:03, B[ S 2.59

< 3 Projbct Manager Rr Notes: American Radiation Services, Inc. assumes no liability for the use or interpretation of any analytical results provided other than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent of the client.

LELAP Certificate# 30658 NELAP Certificate # E87558

SERVICES, INC.

2609 North River Road,, Port Allen, Louisiana 70767 1 (800) 401-4277 FAX (225) 381-2996 QC Results Report Sample Delivery Group: ARS1-08-01562 Date Received: 8/12/2008 Laboratory Control Sample Evaluation Anly.

Alt Analysis su 1 (I pected u

Report Analysus Analysis AmalysIs Percent AcpLCS AnaResult coiar anptaCSan(

)

D Batch Type Results

[

va..ue Units Teat Method Data/Tim.

Technician Recoery ()

ane ARS1-B08-02293 LCS 13 1136.06 161.4 158.48416 1270.45 pCl/L ARS-DS4/EPA 906.0 9/4/08 18:01 13 S 89 75%-125%

Blank Evaluation Analysis QC j

Analyts Analysis CSU 1 (1.)

MDC Expected Qul Report Anelysis Analysis Analysil Batch Type Result.

Vaim Units Tast Method Date/Tirne Technician ARS1'-BOB-02293 MBL H3 67.923 94.439817 158.41962 NA U

pCi/L ARS-054/EPA 906.0 9/5/08 2:16 BJS RER Duplicate Evaluation Anlss Q

t2i Aayi rayi RER QC Analysis Result 1 csu (1

)

cSu 2 (is)

Qu Analysis Analysis Anlysi Analysi s RER Acceptance tch Type Descriptio Unit.

Test Method Date/Tin.

Technician I

ng ARS1-B08-02293 LCSD H3 1136.06 161.4 1143.2572 161.9636 pCi/L ARS-054/EPA 906.0 9/4/08 22:09 1

BJS 0.02 1

DER Duplicate Evaluation An alysis Qc Analysis 1asil 1

1 Analysn Analysis 1EAnely siD D ER Batch Type Description Result I C. U 1 (1ni)

Rssut 2 C.U 211.)

QuelD Anyit Analyst s Tsdlni nR I Typ De crp Uont 1Tes1 t M ethd Difte/Tinne Technician anus ARS1-B08-02293 LCSD H3 1 1136.06 161.4 1143.2572 1161.96361 1 pCi/L ARS-054/EPA 906.0 9/4/08 22:09 BS 0.06

< 3 Matrix Spike alyte Analysis CSU 1 (1s)

Unspiked SpikeCon Qua Report Analysis Analysis Analyst.

Percent

.CS Arlalyisis QC AnIs 1 )a Accapteno B3tch Type Results Activity Units Test Method Date/Time Tachnician Racoesry(%)

ange ARS1-B08-02293 MS 113 13755.623 1429.035 1862.953 13970.11 pCi/L ARS-054/EPA 906.0 9/5/08 6:24 BJS 85 60%-140%

Projedt Manager Revir-:

Notes: American Radiation Services, Inc. assumes no liability for the use or interpretation of any analytical results provided other than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent of the client.

LELAP Certificate# 30658 NELAP Certificate # E87558

2609 North River Road - Port Allen, Louisiana 70767 IA R

11 (800) 401-4277 FAX (225) 381-2996 AMEOAN RADIATION SERVICES. INC.

Notes:

Comments:

1.0)

Soil and Sludge analysis are reported on a wet basis or an as received basis unleiss otherwise indicated.

2.0)

The data in this report are within the limits of uncertainty specified in the reference method unless specified.

3.0)

Modified analysis procedures are procedures that are modified to meet the certain specifications. An example may be the use of a water method to analyze a solid matrix due to the lack of an officially recognized procedure for the analysis of the solid matrix.

(

4.0)

Derived Air Concentrations and Effluent Release Concentrations are obtained from 10 CFR 20 Appendix B.

5.0)

Total activity is actually total gamma activity and is determined utilizing the prominent gamma emitters from the naturally occurring radioactive decay chains and other prominent radioactive nucildes. Total activity may be lower than the actual total activity due to the extent of secular equilibrium achieved in the various decay chains at the time of analysis. The total activity is not representative of nuclides that emit solely alpha or beta particles.

6.0)

Ra-228 is determined via secular equilibrium with its daughter, Actinium 228. (Gamma Spectroscopy only),

7.0)

U-238 is determined via secular equilibrium with its daughter, Thorium 234. (Gamma Spectroscopy only).

8.0)

All gamma spectroscopy was performed utilizing high purity germanium detectors (HPGe).

9.0)

ARS makes every attempt to match sample density to calibrated density; however, in some cases, it is not practical or possible to do so and data results may be affected.

Method

References:

1.0)

EPA 600/4-80-032, Prescribed Procedures for the Measurements of Radioactivity in Drinking Water, August 1980.

2.0)

Standard Methods for Examination of Water and Waste Water, 18 h, 1992.

23.0)

EPA SW-846, Test Methods for Evaluating Solid Waste, Third Edition, (9/86). (Updated through 1995).

.4.0)

EPA 600/4/79-020, Methods for Chemical Analysis of Water and Waste, March 1983.

5.0)

HASL 300 Definitions:

1.0)

BDL Analyte not detected because the value was below the detection limit.

2.0)

ND Not detected above the detection limit.

3.0)

Detection Limit The minimum amount of the analyte that ARS can detect utilizing the specific analysis.

4.0)

B Method Blank 5.0)

D Method Duplicate 6.0)

MS Matrix Spike 7.0)

S Spike 8.0)

RS Reference Spike 9.0)

SC Subcontracted out to another qualified laboratory 10.0)

NR Not Referenced 11.0)

NIA Not Applicable 12.0)

Reported as a calculated value 13.0)

False Positive due to interference from Bi-214 Notes:

American Radiation Services, Inc. assumes no liability for the use or interpretation of any analytical results provided other than the cost of the analysis itself. Reproduction of this report in less than full requires the written consent of the client.

LELAP Cert# 30658 NELAP Cert# E87558

PAGE 1 of 2 NRC FORM 303 U.S. NUCLEAR REGULATORY COMMISSION LABORATORY USE ONLY (4-2004)

CONTROL NUMBER REQUEST FOR ANALYSIS AND CHAIN OF CUSTODY LABORATORY:

SAMPLE LOCATION (LICENSEE)

LICENSEE NUMBER DOCKET NO.

INDIAN POINT ENERGY CENTER,

SAMPLE SUBMITTED

  1. TOTAL TYPE VOLUME WEIGHT DATE SAMPLES SUBMITTED PRIORITY

]

ROUTINE 1

GROUND WATER 2000 ML

-2 Kg URGENT SAMPLE COLLECTION INTERVAL MONTH DAY YEAR TIME INSPECTOR RESPONSIELi.

TELEPHONE NUMBER START Jim Noggle (USNRC)

(610) 337-5063 STOP LIST DESIRED LIST DESIRED ANALYSIS TO BE PERFORMED LLD (Opbional)

OTHER TYPE OF ANALYSIS (Specify)

LLD (Optional)

D GROSS ALPHA (GA)

STRONTIUM-90 (Sr90)

D GROSS BETA (GB)

NICKEL-63 (Ni63)

[] GAMMA SPEC (GS)

I TRITIUM (H3)

D CARBON-14 (C14)

I IODINE-125 (1125)

DF.

RELENOUWSHED BY RECEIVED BY DATE TIME REASON FOR CHANGE OF CUSTODY

-c

/

£t

-4 t ep. -

7,

('e,

S//

.5;;ieA(- f 8-7-2oi Zbe 10 6,

FEE RECOVERABLE D NO 7'

YES TAC NUMBER REMARKS:

-Inspection Report: 1007005 NOTE: SAMPLES WILL BE DISCARDED AFTER ANALYSIS UNLESS REASON ARE NOTED IN REMARKS ABOVE.

NRC FORM303 (4-2004)

PRINTED ON RECYCLED PAPER

PAGE 2 of 2 NRC FORM 303 U.S. NUCLEAR REGULATORY COMMISSION LABORATORY USE ONLY (4-2004)

CONTROL NUMBER SAMPLE RECORD - Continued LABORATORY:

SAMPLE SAMPLE NAME COLLECTION REMARKS, PRESERVATIVE NUMBER AND DESCRIPTION DATEITIME ANALYSIS REQUESTED, ETC.

LAF-002-(009)

Ground Water Split Sample of LAF-0024{ 009)

Sample is unfiltered & unpreserved. Analyze fo

  • S/ V//

Gamma Spec, H3, Sr9O, Ni63 NRC FORM303 (4-2004)

PRINTED ON RECYCLED PAPER