The reactor operated at 10 W to 25 kW steady-state power for a total of 25 minutes. At 1257
CDT the Reactor Manager noticed that the radiation monitor (AMS-4) was reading approximately 80 times the Derived Air Concentration (
DAC) for
Iodine. A second continuous air monitor and a portable thin-window ionization chamber both read background levels of radiation. A high-volume air sampler was used to sample the air near the sampling location of the
AMS, which draws air from the reactor deck. The filter from the air sampler read 3700
DPM after drawing 50 cubic feet of air, indicating elevated levels of activity. An
HPGe (High Purity
Germanium Detector) spectrograph indicated some
Cs-137 in the air filter sample, but no
I-131. Primary water activity and conductivity were both at normal levels. The indicated
DAC of
Iodine increased to a value of 149 times
DAC before decreasing to 145 times
DAC.
The Reactor Manager and Radiation Safety Officer agreed that fuel damage was unlikely, but the results of testing were inconclusive, and that it would be prudent to declare an Unusual Event.
- * * UPDATE FROM JEFFREY GUTHER TO CHARLES TEAL AT 1528 EDT ON 6/24/11 * * *
HPGe spectroscopy indicated only background levels of Cs-137 for a air sample, a primary water sample, and a swipe taken from the reactor deck. None of these samples showed I-131. A second portable air sample taken at the reactor deck, but on the opposite side of the deck from the radioactive sampling handling table, did not have elevated counts.
The Reactor Manager and Radiation Safety Officer agreed that the AMS-4 was most likely mis-calibrated, and the elevated levels in the first high-volume air sample were probably due to the proximity of the radioactive sample handling table to the sampling location.
Notified
NRR (Grobe),
NRR EO (Evans), R4DO (Deese),
IRD (Grant),
DHS (Hill),
FEMA (Blankenship)