Advises That Suppl Notice to Info Notice 80-35, Leaking & Dislodged I-125 Implant Seeds, Being Drafted & Will Be Sent for Review.Exam of Mfg & QC Process Recommended in Risk Evaluation
ML20126C309
Person / Time
Issue date: 06/22/1982
From: Cobb L
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE)
To: Cunningham R
NRC OFFICE OF NUCLEAR MATERIAL SAFETY & SAFEGUARDS (NMSS)
Shared Package
ML20126C180 List:
References
FOIA-84-720 IEIN-80-35, NUDOCS 8506140376
Download: ML20126C309 (1)


Text

'

{ $

UNIT ED ST.:.TES /

/ps *st.g%. NUCLEAR REGULATORY CO".'MisslON [p # #

[ -; gT/ WASHINGTON. D. C. 20555 ,

f,

.// JU.1 - 2 W g3, MEMORANDUM FOR: Richard E. Cunningham, Director Division of Fuel Cycle & Material Safety, NMSS FROM: L. I. Cobb, Director Division of Fuel Facilities, Materials & Safeguards. IE

SUBJECT:

AE0D REVIEW OF 10 DINE-125 SEALED SOURCE LEAKAGE INCIDENTS (C.MICHELSON'S MEMO DTD 5/26/82)

We have no objection to preparing and issuing an information notice to supplement liotice #80-35, " Leaking and Dislodged Iodine-125 Implant Seeds."

Note that Information Notice #80-35 states that licensees (1) should

" review handling procedures for the sealed sources, including methods of sterilization, to eliminate those procedures which could physically damage .

the sealed sources," and (2) establish procedures "to assure that, following I implantation, proper surveys are performed for detecting leaking or dislodged sources." We do not believe that'the issuance of a second information notice like the first one is going to eliminate the problem, but it may rekindle and heighten awareness of it. We are drafting a supplemental notice and will send it to you for review.  :

We believe the risk evaluation recommended in the AE0D memorandum is a good idea, and suggest another factor be considered. If appropriate, we should also look at the " viability"_of the procedure, examining the manufacturing  !

and quality control process to determine if they do indeed achieve an adequate degree of assurance for the structural integrity of these sources.

h '- -

L. I. Cobb, Director Division of Fuel Facilities, Materials '