| title = Agreement State Report - Unable | | title = Agreement State Report - Unable to Retract Source on Nucletron Microselectron | ||
| event text = The following is an excerpt of a report received from the Oregon Department of Health via email: | | event text = The following is an excerpt of a report received from the Oregon Department of Health via email: | ||
[The Oregon Department of Health] had a teleconference with the Radiation Safety Officer (RSO). [The RSO] stated that during the morning QA check, the source would not retract. No patient was in the room and the QA check was performed with no one in the room. The RSO went inside the room just at the door and obtained an exposure [rate] reading of 2.6 mR/hr at approximately 15 feet from the HDR [high dose rate] unit, a Nucletron model microSelectron 106.990. The treatment room is secured at the current time until the Elekta FSE [Field Service Engineer] arrives. The RSO is working on treatment of another patient and will get more information as he gets time to do so. | [The Oregon Department of Health] had a teleconference with the Radiation Safety Officer (RSO). [The RSO] stated that during the morning QA check, the source would not retract. No patient was in the room and the QA check was performed with no one in the room. The RSO went inside the room just at the door and obtained an exposure [rate] reading of 2.6 mR/hr at approximately 15 feet from the HDR [high dose rate] unit, a Nucletron model microSelectron 106.990. The treatment room is secured at the current time until the Elekta FSE [Field Service Engineer] arrives. The RSO is working on treatment of another patient and will get more information as he gets time to do so. |
Latest revision as of 20:43, 1 March 2018
Where | |
---|---|
Oncology Associated Of Oregon Eugene, Oregon (NRC Region 4) | |
License number: | 90862 |
Organization: | Or Dept Of Health Rad Protection |
Reporting | |
Agreement State | |
Time - Person (Reporting Time:+78.08 h3.253 days <br />0.465 weeks <br />0.107 months <br />) | |
Opened: | Todd S Carpenter 13:05 Nov 4, 2016 |
NRC Officer: | Jeff Herrera |
Last Updated: | Nov 4, 2016 |