ENS 51621: Difference between revisions

From kanterella
Jump to navigation Jump to search
(Created page by program invented by Mark Hawes)
 
(Created page by program invented by StriderTol)
 
Line 16: Line 16:
| event date = 12/21/2015 EST
| event date = 12/21/2015 EST
| last update date = 12/23/2015
| last update date = 12/23/2015
| title = Part 21 Report - Error In Reporting The Variability In The Dose Delivered On Certificate Of Processing
| title = Part 21 Report - Error in Reporting the Variability in the Dose Delivered on Certificate of Processing
| event text = The following information was received from Homewood Products Corporation via fax:
| event text = The following information was received from Homewood Products Corporation via fax:
Steris Isomedix Services measuring and test equipment used by NTS (Formerly Wyle Laboratories, Inc.) to determine the applied radiation dose did not account for all of the uncertainties involved, and therefore the actual radiation dose applied to components and reported to Homewood Products Corporation by Wyle Laboratories, could be less than the requested service condition dose.
Steris Isomedix Services measuring and test equipment used by NTS (Formerly Wyle Laboratories, Inc.) to determine the applied radiation dose did not account for all of the uncertainties involved, and therefore the actual radiation dose applied to components and reported to Homewood Products Corporation by Wyle Laboratories, could be less than the requested service condition dose.

Latest revision as of 21:46, 1 March 2018

ENS 51621 +/-
Where
National Technical Systems, Inc. Formerly Wyle Laboratories
Pittsburgh, Pennsylvania (NRC Region 1)
Organization: Homewood Products Corporation
Reporting
10 CFR 21.21(d)(3)(i), Failure to Comply or Defect
Time - Person (Reporting Time:+54.25 h2.26 days <br />0.323 weeks <br />0.0743 months <br />)
Opened: Richard Martin
11:15 Dec 23, 2015
NRC Officer: Howie Crouch
Last Updated: Dec 23, 2015
51621 - NRC Website