Browse wiki

Jump to navigation Jump to search
The following is a summary of a call receiThe following is a summary of a call received from the licensee:</br>During routine maintenance, an inspector from Radiometric Services and Instruments found that the shutter for an AMC-19 gauge, s/n: 998110B, with a 100 microCi AM-241 source, would close slowly. This was not as expected. The inspector noted that the pneumatic system was over pressurized. The gauge is located on the Continuous Galvanizing Line and the shutter is open during normal operations. </br>The shutter can still be closed and a Geiger counter can be used to verify. There was no exposure to personnel.</br>* * * UPDATE ON 1/19/2021 AT 1731 EST FROM ROBIN BIRK TO JOANNA BRIDGE * * *</br>The following is a summary of a call and e-mail received from the licensee:</br>While testing the shutter today we were experiencing some continuing problems with the shutter mechanism at our Continuous Galvanizing Line. During maintenance, we discovered some error codes in the RSI software that indicated possible additional shutter problems. There was no potential for employee exposures. RSI will be back at the plant to address on 1/20/21.</br>Notified R3DO (Peterson) and NMSS Events (e-mail). R3DO (Peterson) and NMSS Events (e-mail).  +
14:15:00, 12 January 2021  +
55,065  +
13:29:00, 12 January 2021  +
14:15:00, 12 January 2021  +
The following is a summary of a call receiThe following is a summary of a call received from the licensee:</br>During routine maintenance, an inspector from Radiometric Services and Instruments found that the shutter for an AMC-19 gauge, s/n: 998110B, with a 100 microCi AM-241 source, would close slowly. This was not as expected. The inspector noted that the pneumatic system was over pressurized. The gauge is located on the Continuous Galvanizing Line and the shutter is open during normal operations. </br>The shutter can still be closed and a Geiger counter can be used to verify. There was no exposure to personnel.</br>* * * UPDATE ON 1/19/2021 AT 1731 EST FROM ROBIN BIRK TO JOANNA BRIDGE * * *</br>The following is a summary of a call and e-mail received from the licensee:</br>While testing the shutter today we were experiencing some continuing problems with the shutter mechanism at our Continuous Galvanizing Line. During maintenance, we discovered some error codes in the RSI software that indicated possible additional shutter problems. There was no potential for employee exposures. RSI will be back at the plant to address on 1/20/21.</br>Notified R3DO (Peterson) and NMSS Events (e-mail). R3DO (Peterson) and NMSS Events (e-mail).  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 19 January 2021  +
21-10459-01  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
06:47:54, 16 January 2022  +
13:29:00, 12 January 2021  +
-0.0321 d (-0.77 hours, -0.00458 weeks, -0.00105 months)  +
14:15:00, 12 January 2021  +
NON-AGREEMENT State Report - Shutter Not Functioning Correctly  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.