A prostate brachytherapy implant using I-1A prostate brachytherapy implant using I-125 performed at Fauquier Hospital on 6/25/2015, resulted in a dose distribution having a D90 percent of 97Gy as measured on post implant CT done on 8/12/15. The written directive prescribed a D90 percent of 145Gy. The implanted dose of 97 Gy is 66.9 percent of the intended 145 Gy. The patient was notified and is being monitored to determine if additional treatment is required.</br>The final determination was not made until 10/8/15 and Fauquier notified VRMP (Virginia Radioactive Materials Program) by email on 10/8/15. A telephone conversation was held on 10/13/15 regarding the 15 day written report. VRMP will review the report and determine whether an inspection will be performed.</br>Virginia Report ID: VA-15-14</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
04:00:00, 25 June 2015  +
51,466  +
11:03:00, 13 October 2015  +
04:00:00, 25 June 2015  +
A prostate brachytherapy implant using I-1A prostate brachytherapy implant using I-125 performed at Fauquier Hospital on 6/25/2015, resulted in a dose distribution having a D90 percent of 97Gy as measured on post implant CT done on 8/12/15. The written directive prescribed a D90 percent of 145Gy. The implanted dose of 97 Gy is 66.9 percent of the intended 145 Gy. The patient was notified and is being monitored to determine if additional treatment is required.</br>The final determination was not made until 10/8/15 and Fauquier notified VRMP (Virginia Radioactive Materials Program) by email on 10/8/15. A telephone conversation was held on 10/13/15 regarding the 15 day written report. VRMP will review the report and determine whether an inspection will be performed.</br>Virginia Report ID: VA-15-14</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 13 October 2015  +
061-092-1  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
22:17:10, 24 September 2017  +
11:03:00, 13 October 2015  +
110.294 d (2,647.05 hours, 15.756 weeks, 3.626 months)  +
04:00:00, 25 June 2015  +
Agreement State Report - Medical Underdose  +