The following information was received by … The following information was received by the State of Ohio via email:</br>On July 14, 2015, the licensee reported that the intended delivery of Y-90 SirSpheres went to the small bowel instead of the right lobe of the liver during a procedure that morning. The intervention physician felt that the dose delivery was not going where it should be going and discontinued the treatment. Scanning the patient identified that the Y-90 microspheres were delivered to the small bowel. The original prescribed dose to the right lobe of the liver was 78 Gy with 20.5 mCi. The delivered dose of 36 Gy with 7.79 mCi went to the small bowel instead of the liver right lobe. The patient was notified at the time of the event. The interventional physician was the referring physician and AU (Authorized User).</br>Ohio Report: OH150007</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
14:30:00, 14 July 2015 +
51,233 +
10:50:00, 15 July 2015 +
14:30:00, 14 July 2015 +
The following information was received by … The following information was received by the State of Ohio via email:</br>On July 14, 2015, the licensee reported that the intended delivery of Y-90 SirSpheres went to the small bowel instead of the right lobe of the liver during a procedure that morning. The intervention physician felt that the dose delivery was not going where it should be going and discontinued the treatment. Scanning the patient identified that the Y-90 microspheres were delivered to the small bowel. The original prescribed dose to the right lobe of the liver was 78 Gy with 20.5 mCi. The delivered dose of 36 Gy with 7.79 mCi went to the small bowel instead of the liver right lobe. The patient was notified at the time of the event. The interventional physician was the referring physician and AU (Authorized User).</br>Ohio Report: OH150007</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 15 July 2015 +
OH-0211018007 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:05:01, 24 November 2018 +
10:50:00, 15 July 2015 +
0.847 d (20.33 hours, 0.121 weeks, 0.0278 months) +
14:30:00, 14 July 2015 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.