Browse wiki
Jump to navigation
Jump to search
The following was received from the State … The following was received from the State of Ohio via email:</br>On Jan. 27, 2014, the RSO of Flower Hospital provided a telephone report of medical event that occurred on Jan. 24 and 27, 2014. </br>A patient was being treated with a Varian GammaMed Plus HDR with 4.7 Ci of lr-192 in a vaginal cylinder. The written directive was for three fractions of 500 cGy for a total of 1,500 cGy. Then during the delivery of the second fraction the licensee realized that they were only delivering a planned dose of 166 cGy. The licensee completed the second fraction as planned. The treatment planning was entered as one third of the 500 cGy for three fractions instead of three fractions of 500 cGy. The license will modify the treatment plan to give the full prescribed 1500 cGy to the patient.</br>The patient was notified on Jan. 27, 2014.</br>This is a reportable medical event under OAC 3701:1-58-101 (A)(1)(c.) for a fractionated dose that differs by more than 50%.</br>Ohio Report #: OH140001</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
05:00:00, 24 January 2014 +
49,770 +
17:05:00, 27 January 2014 +
05:00:00, 24 January 2014 +
The following was received from the State … The following was received from the State of Ohio via email:</br>On Jan. 27, 2014, the RSO of Flower Hospital provided a telephone report of medical event that occurred on Jan. 24 and 27, 2014. </br>A patient was being treated with a Varian GammaMed Plus HDR with 4.7 Ci of lr-192 in a vaginal cylinder. The written directive was for three fractions of 500 cGy for a total of 1,500 cGy. Then during the delivery of the second fraction the licensee realized that they were only delivering a planned dose of 166 cGy. The licensee completed the second fraction as planned. The treatment planning was entered as one third of the 500 cGy for three fractions instead of three fractions of 500 cGy. The license will modify the treatment plan to give the full prescribed 1500 cGy to the patient.</br>The patient was notified on Jan. 27, 2014.</br>This is a reportable medical event under OAC 3701:1-58-101 (A)(1)(c.) for a fractionated dose that differs by more than 50%.</br>Ohio Report #: OH140001</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 27 January 2014 +
02120 49 0004 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
01:51:49, 2 March 2018 +
17:05:00, 27 January 2014 +
3.503 d (84.08 hours, 0.5 weeks, 0.115 months) +
05:00:00, 24 January 2014 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.