The following report was received via e-maThe following report was received via e-mail:</br>10-25-2013 (Friday)</br>HDR (High Dose Rate) treatment to right breast with SAVI 9 Channel HDR unit: 1st channel treated appropriately but only fractional treatments to channel 2-8 of less than 50% of prescribed dose were delivered. Problems with the applicator's curvature caused source delivery problems. The manufacturer (Varian) was contacted but was unable to arrive before the next Monday.</br> </br>10-28-2013 (Monday)</br>Licensee made another attempt, but again only fractional treatments to channel 2-8 of less than 50% of prescribed dose was delivered.</br> </br>10-29-2013 (Tuesday)</br>The patient was taken to another Mount Carmel facility (MC West), parameters were recalculated and patient treatment was successfully concluded. The licensee stated they expect that no harm was done to the patient. The licensee stated that they didn't understand that this event was reportable until they were advised by Varian it was an equipment failure so it was a 24 hr. reportable event.</br>Event ID No: OH130015</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
05:00:00, 25 October 2013  +
49,523  +
17:30:00, 8 November 2013  +
05:00:00, 25 October 2013  +
The following report was received via e-maThe following report was received via e-mail:</br>10-25-2013 (Friday)</br>HDR (High Dose Rate) treatment to right breast with SAVI 9 Channel HDR unit: 1st channel treated appropriately but only fractional treatments to channel 2-8 of less than 50% of prescribed dose were delivered. Problems with the applicator's curvature caused source delivery problems. The manufacturer (Varian) was contacted but was unable to arrive before the next Monday.</br> </br>10-28-2013 (Monday)</br>Licensee made another attempt, but again only fractional treatments to channel 2-8 of less than 50% of prescribed dose was delivered.</br> </br>10-29-2013 (Tuesday)</br>The patient was taken to another Mount Carmel facility (MC West), parameters were recalculated and patient treatment was successfully concluded. The licensee stated they expect that no harm was done to the patient. The licensee stated that they didn't understand that this event was reportable until they were advised by Varian it was an equipment failure so it was a 24 hr. reportable event.</br>Event ID No: OH130015</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 8 November 2013  +
02120250034  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
01:52:35, 2 March 2018  +
17:30:00, 8 November 2013  +
14.479 d (347.5 hours, 2.068 weeks, 0.476 months)  +
05:00:00, 25 October 2013  +
Administration of Less than the Prescribed Dose to a Patient  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.